Trek 541A User Manual

Trek Model 541A Non-Contacting
Electrostatic Voltmeter for EOS/ESD
The Trek Model 541A Electrostatic Voltmeter (pictured at left with
the Model 541PR-S probe) provides accurate non-contacting measurements of the electrostatic surface voltage associated with EOS/ESD processes. The instrument is configured with a miniature electrostatic field chopper probe that can be remotely located and easily positioned within process equipment to provide highly accurate, non-contacting, DC-stable, spacing-independent voltage measurements in either ionized or non-ionized environments. A 20x4 alphanumeric LCD screen displays the present measured voltage, the positive peak voltage, the negative peak voltage and additional menu information.
Key Specifications
!
Measurement Range
Model 541A-1: ±1 kV DC or peak AC
Model 541A-2: ±100 V DC or peak AC
Measurement Accuracy: Better than ±1% of full scale over a
!
probe-to-surface separation of
2.5 mm ± 1mm
Alphanumeric LCD Display: 20 x 4 characters
!
Walking Test Adapter
Typical Applications Include
! Semiconductor ! LCD ! Electronic assembly ! ESD-sensitive processes
Optional Walking Test Adapter kits for the Model 541A allow analys i s o f c h a rge leve l s accumulated on the human body per compliance with ANSI/ESD STM97.2. and IEC Standard 61340-4-5. See page 2 for details.
Features and Benefits
! Chopper probe is DC-stable with or without incident air ion flow ! Drift-free measurements ! LCD screen displays present voltage and holds the most positive and negative values ! Visual and audible alarms activate when the preset voltage threshold levels are reached ! Voltage output monitor for remote monitoring or control ! USB and RS-232 serial ports ! NIST-traceable Certificate of Calibration provided with each unit ! compliant ! Optional Walking Test Adapter kits available
TREK, INC. • 190 Walnut Street • Lockport, NY 14094 • USA • 800-FOR TREK
716-438-7555 • 716-201-1804 (fax) • www.trekinc.com • sales@trekinc.com
Model 541A Specifications
Performance
Measurement Ranges
Model 541A-1
0 to ±1 kV DC or peak AC
Model 541A-2
0 to ±100 V DC or peak AC
Speed of Response (10% to 90%)
Less than 50 ms for a ±1 kV step (541A-1)
Less than 50 ms for a ±100 V step (541A-2) Accuracy
1% of full scale
Resolution
0.1%
Monitor Output
Model 541A-1
1/100th of the measured voltage
Model 541A-2
1/10th of the measured voltage
Output Noise
Less than 30 mV rms*
Output Impedance
47 W
Features
Alarms
Activated if measured voltage exceeds preset threshold limits; positive/negative limits may be programmed separately
Visual
Front-panel LEDs illuminate when thresholds are reached
Audible
ON/OFF programmable pulsating or continuous tone. (+) and (-) alarms have different tone rates for the pulsating tone selection
Alarm Relay Output
Form C relay contact with barrier strip terminals; changes state when voltage thresholds are met or exceeded
Alarm Digital Output
TTL output with a TTL low (0 to 0.8 V) as the alarm “ON” status; TTL high (2.5 to 5.0 V) indicates a normal condition
Reset Button
Resets alarms and peak hold to zero
Zero Control
Adjustable to produce zero volts when probe coupled to a known zero voltage source
Alphanumeric LCD Display
20 character by four line (20x4) LCD displays the present voltage and holds the most positive and most negative values
Ground Receptacle
Banana jack
Serial Port and USB Port
Provides control of specific functions and acquires sensor data utilizing Trek software and a PC connected to the RS-232 serial port or the USB Type B port (connectors are on back panel)
Current Output
Provides a current of 4 mA to 20 mA representing -1 kV to +1 kV (541A-1) and
-100 V to +100 V (541A-2)
Features (cont.)
Menu Buttons
Select and program menu options - the set the alarm threshold voltages, alarm conditions and alarm reset type
Mechanical
Dimensions
97 mm H x 152 mm W x 204 mm D (3.8” H x 6” W x 8” D)
Weight
0.77 kg (1.7 lb.)
Operating Conditions
Temperature
15°C to 35°C (59°F to 95°F)
Relative Humidity
5% to 85% RH, non-condensing
Altitude
To 2000 m (6561.68 ft.)
Electrical
Power
15 V DC ±20%, 800 mA, power bus or AC/DC adapter with a 2.1 mm DC plug.
Power ON/OFF
Rear panel switch
Supplied Accessories
PN: 24005 Operator Manual (with software CD), PN: N9056 6P/4C Plug, PN: BA108 Serial Cable, PN: N9044 Ground Cord
Optional Accessories
AC/DC Adapter PN: F5054R
Walking Test Adapter - Round Body Probe CN: 1K037
Walking Test Adapter - Square Body Probe CN: 1K038
Probes
Probe
Miniature electrostatic field chopper probe
End-view orientation (round body)
Model No: 541PR-E. Aperture 1.6 mm (0.06”) diameter. Dimensions 9.7 mm dia x 50 mm L (0.38” dia. x 2” L)
Side-view orientation (round body)
Model No: 541PR-S. Aperture 2.3 mm (0.09”) diameter. Dimensions 9.7 mm dia x 53 mm L (0.38” dia x 2.1” L)
Side-view orientation (square body)
Model 541P-S. Aperture 2.5 mm (0.1”) diameter. Dimensions 5.6 mm sq x 50 mm L (0.22” sq x 2” L)
Probe Cable Length
3 meters nominal(9.8 ft.)
Probe-to-Surface Separation Distance
2.5 mm ±1 mm (recommended)
End View Probe Aperture
1.6 mm
50 mm
Side View Probe Aperture
2.3 mm
50 mm
50 mm
*Measured using the true rms feature of the HP Model 34401A digital multimeter
Measurement and Power Solutions
Side View Probe Aperture
2.5 mm
Copyright ã 2013 TREK, INC. All specifications are subject to change. 1340/JRB
TM
5.6 mm
Loading...