Trek 341B User Manual

Trek Model 341B
High-Speed, High-Voltage Electrostatic Voltmeter
The Trek Model 341B is a DC-stable, precision electrostatic voltmeter for making noncontacting surface voltage measurements. The 341B employs a field-nulling technique that achieves DC stability and high accuracy even if the probe-to­surface spacing changes. This permits measurements of either stationary or moving surfaces without the need to establish fixed spacing to maintain accuracy. The instrument also utilizes a patented probe design that eliminates the need for close tolerance components which significantly improves noise and drift under conditions of high humidity and wide temperature ranges.
Key Specifications
! Measurement Range: 0 to ±20 kV DC or peak AC
! Measurement Accuracy: Better than ±0.1% of full scale ! Speed of Response: Less than 200 µs for a 1 kV step
Typical Applications Include
! Charge accumultion monitoring of LCD production processes ! Monitoring surface potentials in the electrostatic painting process ! Measuring of electrostatic potentials on polymers, rubber, fabrics and paper
Features and Benefits
Superb noise and drift performance
!
Precision voltage monitor output
!
Monitor provides a low voltage replica of the measured electrostatic potential for
!
monitoring purposes or for use as a feedback signal in a closed loop system Easy-to-read LED display
!
Optional probes offer versatility (order separately)
!
Can be operated on a bench top, or with optional hardware, in a standard 19-inch rack
!
NIST-traceable Certificate of Calibration provided with each unit
!
compliant
!
Available Probes Available Configurations
Standard Resolution Model 341B
PN 17157: Model 3450 Side-viewing PN 341B-L, 341B Electrostatic Voltmeter (90-127 V AC)
PN 341B-H, 341B Electrostatic Voltmeter (180-250 V AC)
High Temperature (up to 100°C) Model 341B-1 (for use with 3460-1 Line Driver) PN 17284 Model 3455ET End-viewing PN 341B-1-L, 341B-1 Electrostatic Voltmeter (90 to 127 V AC) PN 17285 Model 3453ST Side viewing PN 341B-1-H, 341B-1 Electrostatic Voltmeter (180 to 250 V AC)
PN 17181, Model 3460 Line Driver (used with 341B-1) (Model 341B-1 utilizes a separate line driver for extended probe cable lengths)
TREK, INC. • 190 Walnut Street • Lockport, NY 14094 • USA • 800-FOR TREK
716-438-7555 • 716-201-1804 (fax) • www.trekinc.com • sales@trekinc.com
Model 347 Specifications
Performance
Measurement Range
0 to ±20 kV DC or peak AC
Measurement Accuracy
Better than ±0.1% of full scale, referred to the voltage monitor
Speed of Response (10% to 90%)
Less than 200 µs for 1 kV step. Less than 5 ms for 20 kV step change
Full Signal Bandwidth
DC to better than 25 Hz
Stability Drift with Time
Less than 100 ppm/hour, noncumulative Drift with Temperature
Less than 100 ppm/˚C
Voltage Monitor
Output
A buffered output provides a low-voltage replica of the measured voltage
Ratio
1/100th of the measured voltage
Output Noise
Less than 20 mV rms*
Output Impedance
Less than 0.1 W
Voltage Display
Voltage Display
4 ½ digit LED display
Range
0 to V±19.99 k
Resolution
1 V
Zero Offset
±2 counts, referred to the voltage monitor
Sampling Rate
3 readings per second
Features
High Voltage Ready LED
LED indicator illuminates when the Model 341B is ready to make high-voltage measurements
High Voltage ON-OFF
Two-position toggle switch that turns on and off the high-voltage power supply inside the instrument
Zero Control
A 10-turn control to null offsets or other zero errors that occur within the system
Mechanical
Dimensions
230 mm H x 441 mm W x 432 mm D (9.06” H x 17.36” W x 17” D)
Weight
17 kg (37 lb)
Voltage Monitor Output Connector
BNC connector Ground Receptacle
Green binding post
Operating Conditions
Temperature
0°C to 40°C (32°F to 104°F)
Relative Humidity
To 90%, noncondensing
Altitude
To 2000 m (6561.68 ft.)
Probe-to-Surface Separation
3 mm ±1mm (recommended)
Electrical
AC Line Cord Receptacle
Standard 3-prong with integral fuse holder
Line Voltage
Factory set for one of two ranges: 90 to 127 V AC or 180 to 250 V AC, at 48 to 63 Hz
Power ON/OFF
Two-position rocker switch that turns ON and OFF the main power to the instrument
Supplied Accessories
Operator’s Manual
PN: 23306
Line Cord
PN: N5002 (for 90 to 127 V AC) PN: Determined by the geographical destination (for 180 to 250 V AC)
Optional Accessories
Probes
Please refer to Page 1
Probe Line Driver (required when used with the 341B-1 and an extended cable length)
Model: 3460-1
Probe Extension Cable (from the 341B to the probe)
PN: 17218, Model 3450EC Probe Extension Cable Full-Rack Mount Kit
Model 603RA (19-inch)
Certification
TREK, INC. certifies that each Model 341B is tested and calibrated to specifications using measurement equipment traceable to the National Institute of Standards and Technology or traceable to consensus standards.
*Measured using the true rms feature of the Hewlett Packard Model 34401A digital multimeter
Measurement and Power Solutions
Copyright ã 2012 TREK, INC. All specifications are subject to change. 1234/DEC