Trek Model 30/20A
High-Voltage Power Amplifier
The Model 30/20A is a DC-stable, high-voltage power amplifier featuring an all solid-state design for high slew
rate, wide bandwidth, and low-noise operation. t is configured as noninverting with a fixed gain of 3000 V/V and
is protected against overvoltage and overcurrent conditions that may be generated by active loads or by
output short circuits to ground. Precision voltage and current monitors provide low-voltage
representations of the high-voltage output and load current for monitoring purposes or for use as
feedback signals in a closed-loop system.
The 4-quadrant, active output stage sinks or sources current to reactive or resistive loads throughout the
output voltage range. This is essential to achieve the accurate output response and high slew rates demanded
by reactive loads.
Key Specifications
! Output Voltage Range:
! Output Current Range: peak AC
! Slew Rate: Greater than 550 V/
Large Signal Bandwidth DC to greater than 2.5 kHz
!
(2% distortion):
DC Voltage Gain: 3000 V/V
!
0 to ±30 kV DC or peak AC
0 to ±20 mA DC or
µs
Typical Applications Include
! Dielectric studies
! Electron beam ion traps and ion sourcing
! Electrospinning
! Electrostatic deflection (including ion beam steering)
! Electrostatic flame control
! Electrostatic levitation
! Electrostatic precipitation
! High-voltage cable testing
! High-voltage component testing
! Plasma studies (including dielectric barrier discharge)
Features and Benefits
! Four-quadrant output for driving capacitive loads
Closed loop system for high accuracy
!
Short-circuit protected for equipment protection
!
All solid-state design for maintenance free operation
!
DC-stable for programmable supply applications
!
Low output noise for ultra-accurate outputs
!
NIST-traceable Certificate of Calibration provided with each unit
!
compliant
!
TREK, INC. • 190 Walnut Street • Lockport, NY 14094 • USA • 800-FOR TREK
(fax) • www.trekinc.com • sales@trekinc.com 716-438-7555 • 716-201-1804
Individual push-button switches
TTL compatible input. TTL high (or open) turns
off high-voltage output. TTL low turns on highvoltage output.
Switch selectable for limit or trip. Graduated 1turn panel potentiometer is used to adjust limit
or trip level from 0 to ±20 mA.
Out of Regulation
Status Indicator and
Connnector
Illuminates and TTL low is provided when unit
fails to produce required HV output such as
during current limit.
Limit/Trip Status
Indicator and
Connector
An indicator will illuminate and a BNC will
provide a TTL low when the high-voltage output
is disabled due to the output current exceeding
the current trip level, the detection of a highvoltage supply fault, the removal of one of the
panels, or if the Model 30/20A is out of
regulation for greater than 500 ms.
103.9 cm H x 43 cm W x 87 cm D
(40.9” H x 17” W x 34” D) Depth dimension
includes wheels, handles and spacing for air
flow.
73 kg (160 lb) approximate
Caton high-voltage Connector
Amplifier Input, Voltage Monitor, Current Monitor,
Remote High Voltage ON/OFF, Out of Regulation
Status, Fault/Trip Status
0°C to 40°C (32°F to 104°F)
To 1524 meters (5000 ft.)
Factory set for one of two ranges:
104 to 127 V AC or 180 to 250 V AC at 48 to
63 Hz (specify when ordering)
Standard 3-prong with integral fuse holder
Selected per geographic destination
Model 30/20A Specifications
0 to ±30 kV DC or peak AC
0 to ±20 mA DC or peak AC
Input Voltage Range
25 kW nominal (inverting/differential option
50 kW, nominal)
Better than 0.1% of full scale
Offset Voltage
Slew Rate
(10% to 90%, typical)
Small Signal
Bandwidth (-3dB)
DC to greater than 30 kHz
Large Signal
Bandwidth
(2% distortion)
DC to greater than 2.5 kHz
Less than 50 ppm/hr, noncumulative
Better than 0.1% of full scale
Better than 2% of full scale
Less than 200 µs for a 0-30 kV step
Graduated 1-turn panel potentiometer is used
to optimize the AC response for various load
parameters.
*Measured using the true rms feature of the Hewlett Packard Model 34401A digital multimeter
Measurement and Power Solutions
Copyright ã 2013 TREK, INC. All specifications are subject to change. 1330/JRB
TM