Trek 10-40A-hs User Manual

Trek Model 10/40A-HS
High-Speed High-Voltage Power Amplifier
The Model 10/40A-HS is a DC-stable, high-speed, high­voltage power amplifier used in industrial and research applications. It features an all-solid-state design for high slew
rate, wide bandwidth and low-noise operation. The four­quadrant, active output stage sinks or sources current into reactive or resistive loads throughout the output voltage range.
This type of output is essential to achieve an accurate output response and high slew rate demanded by a variety of loads such as highly capacitive or reactive loads. It is configured as a non-inverting amplifier.
Key Specifications
Output Voltage Range:
!
Output Current Range: peak AC for 1 ms (must not exceed 40 mA rms max)
!
Slew Rate: Greater than 900 V/
!
Large Signal Bandwidth (-3 dB): DC to greater than 23 kHz, typical
!
DC Voltage Gain: 1000 V/V
!
0 to ±10 kV DC or peak AC 0 to ±40 mA DC or 120 mA
µs
Typical Applications Include
! Electrostatic deflection ! Electrophoresis ! Electrorheological fluids ! Electro-optic modulation ! Material poling ! AC or DC biasing ! Ion beam steering ! Particle accelerators ! Mass spectrometers ! Material characterization ! Ferroelectrics ! Atmospheric plasma ! Dielectric barrier discharge
Features and Benefits
!
Four-quadrant output for driving capacitive loads
!
Closed loop system for high accuracy
!
Short-circuit protected for equipment protection All solid-state design for maintenance free operation
!
DC-stable for programmable supply applications
!
Low output noise for ultra-accurate outputs
!
NIST-traceable Certificate of Calibration provided with each unit
!
TREK, INC. • 190 Walnut Street • Lockport, NY 14094 • USA • 800-FOR TREK
716-438-7555 • 716-201-1804 (fax) • www.trekinc.com • sales@trekinc.com
Features (cont.)
Dynamic Adjustment
Graduated one-turn panel potentiometer is used to optimize the AC response for various load parameters.
Limit/Trip Mode
Switch selectable for either limit or trip. Graduated 1-turn panel potentiometer is used to adjust limit or trip level from 0 to 100% peak current. There is one LED indicator and one BNC connector
Current Trip Limit Status Indicator/ Connector
An indicator will Illuminate and a BNC will provide a TTL low when the Model 10/40A-HS fails to produce the required high-voltage output such as during current limit.
Out of Regulation Status
Illuminates and a TTL low is provided when unit fails to produce required HV output such as during a current limit
Mechanical
Dimensions
279 mm H x 482 mm W 654 mm D (11” H x 19” W x 25.75” D)
Weight
24.9 kg (55 lb)
HV Connector
Alden High Voltage Connector
BNC Connectors
Amplifier Input, Voltage Monitor, Current Monitor, Remote High Voltage ON/OFF, Out of Regulation Status, Limit/Trip Status
Operating Conditions
Temperature
0°C to 40°C (32°F to 104°F)
Relative Humidity
To 85%, noncondensing
Altitude
To 2000 meters (6561.68 ft.)
Electrical
Line Voltage
Factory Set for one of two ranges: 104 to 127 V AC or 180 to 250 V AC, either at 48 to 63 Hz
AC Line Receptacle
Standard IEC 320 three-prong AC line connector
Power Consumption
1000 VA, maximum
Supplied Accessories
Operators’ Manual
PN: 23462
HV Output Cable
PN: 43463
Line Cord, Spare Fuses
PN: N5011. destination
Selected per geographic
Model 10/40A-HS Specifications
Performance
Output Voltage Range
0 to ±10 kV DC or peak AC
Output Current Range
0 to ±40 mA DC or ±120 mA peak for 1 ms (must not exceed 40 mA rms, max)
Input Voltage Range
0 to ±10 V DC or peak AC
Input Impedance
25 kW, nominal
DC Voltage Gain
1000 V/V
DC Voltage Gain Accuracy
Better than 0.1% of full scale DC Offset Voltage
Less than ±2 V
Output Noise
Less than 0.5 V rms*
Slew Rate (10% to 90%, typical)
Greater than 900 V/µs
Small Signal Bandwidth (-3dB)
DC to greater than 25 kHz
Large Signal Bandwidth (-3 dB)
DC to greater than 23 kHz, typical
Large Signal Bandwidth (1% distortion)
DC to greater than 9 kHz, typical (The unit will trip when the maximum bandwidth is reached)
Stability
Drift with Time
Less than 50 ppm/hr, noncumulative
Drift with Temp
Less than 100 ppm/°C
Voltage Monitor
Ratio
1/1000th of the high-voltage output signal
DC Accuracy
Better than 0.1% of full scale
DC Offset Voltage
Less than ±2 mV
Output Noise
Less than 10 mV rms*
Output Impedance
47 W
Current Monitor
Ratio
1 V/12 mA
DC Accuracy
Better than 1% of full scale
Offset Voltage
Better than ±10 mV
Output Noise
Less than 30 mV rms*
Bandwidth (-3dB)
DC to greater than 20 kHz
Output Impedance
47 W
Features
High-Voltage On/Off
Local
Individual push-button switch
Remote
TTL compatible input. TTL high (or open) turns off high-voltage output. TTL low turns on high­voltage output.
*Measured using the true rms feature of the HP Model 34401A digital multimeter
Measurement and Power Solutions
Copyright ã 2013 TREK, INC. All specifications are subject to change. 1322/JRB
www.trekinc.com
TM
Loading...