CATALOGUE CATALOGUE
Table of Contents
INTRODUCTION...................................................................................................................2
General Description
Indications for Use
Contraindications
Measurement Principle
Safety Information
LCD Display Signal
Monitor Components
List
BEFORE YOU START...........................................................................................................9
The Choice of Power Supply
Installing and Replacing the Batteries
Setting Date, Time and Measurement Unit
Select the User
MEASUREMENT...................................................................................................................14
Tie the Cuff
Start the Measurement
DATA MANAGEMENT...........................................................................................................17
Recall the Records
Delete the Records
INFORMATION FOR USER..................................................................................................20
Tips for measurement
Maintenances
ABOUT BLOOD PRESSURE................................................................................................22
What are systolic pressure and diastolic pressure?
What is the standard blood pressure classification?
Irregular heartbeat detector
Why does my blood pressure fluctuate throughout the day?
Why do I get a different blood pressure at home compared to the hospital?
Is the result the same if measuring on the right arm?
TROUBLESHOOTING..........................................................................................................24
SPECIFICATIONS................................................................................................................25
AUTHORIZED COMPONENT .............................................................................................26
CONTACT INFORMATION...................................................................................................26
COMPLIED STANDARDS LIST............................................................................................27
FCC STATEMENT.................................................................................................................28
EMC GUIDANCE..................................................................................................................28
1

INTRODUCTION
INTRODUCTION
General Description
Thank you for selecting TRANSTEK arm type blood pressure monitor
(TMB-1585-BS). The monitor features blood pressure measurement, pulse
rate measurement and the result storage. The design provides you with
two years of reliable service.
Readings taken by the TMB-1585-BS are equivalent to those obtained by
a trained observer using the cuff and stethoscope auscultation method.
This manual contains important safety and care information, and
provides step by step instructions for using the product.
Read the manual thoroughly before using the product.
Features:
100mm×68 mm Digital LCD display
Maximum 60 records per each user
3rd technonoly: Measuring during inflation
(The updated technology in the world)
Indications for Use
The Transtek Blood Pressure Monitor is digital monitors intended for use in
measuring blood pressure and heartbeat rate with arm circumference ranging
from 22 cm to 32 cm ( about 8¾˝-12½˝ ) or 22cm to 42cm(about 8¾˝-16½˝).
It is intended for adult indoor use only.
Contraindications
1.The device is not suitable for use on may be pregnant women or pregnant women.
2.The device is not suitable for use on patients with implanted,electrocical devices,
such as cardiac pacemakers, defibrillators.
Measurement Principle
This product uses the Oscillometric Measuring method to detect blood pressure.
Before every measurement, the unit establishes a “zero pressure” equivalent to the air
pressure. Then it starts inflating the arm cuff, meanwhile, the unit detects pressure
oscillations generated by beat-to-beat pulsatile, which is used to determine the systolic
and diastolic pressure, and also pulse rate.
2
Safety Information
The signs below might be in the user manual, labeling or other component.
They are the requirement of standard and using.
Symbol for “THE OPERATION
GUIDE MUST BE READ”
Caution: These notes must be
observed to prevent any damage
to the device.
Symbol for “MANUFACTURER”
SN
Symbol for “SERIAL NUMBER”
Symbol for “DIRECT CURRENT”
Symbol for “MANUFACTURE
DATE”
F1
T1A/250V Φ3.6*10CCC
The Green Dot is the license
symbol of a European network
of industry-funded systems for
recycling the packaging
materials of consumer goods.
Symbol for “TYPE BF APPLIED
PARTS”
Symbol for “ENVIRONMENT
PROTECTION - Electrical waste
products should not be disposed of
with household waste. Please recycle
where facilities exist. Check with your
local authority or retailer for recycling
advice”
Symbol for “Recycle”
For indoor use only
Symbol for “Class II Equipment”
3

INTRODUCTION
INTRODUCTION
CAUTION
* This device is intended for adult use in homes only.
* The device is not suitable for use on neonatal patients, pregnant women,patients with implanted,
electronical devices, patients with pre-eclampsia, premature ventricular beats, atrial fibrillation, peripheral,
arterial disease and patients undergoing intravascular therapy or arterio-venous shunt or people who
received a mastectomy. Please consult your doctor prior to using the unit if you suffer from illnesses.
* The device is not suitable for measuring the blood pressure of children. Ask your doctor before using it on
older children.
* The device is not intended for patient transport outside a healthcare facility.
* The device is not intended for public use.
* This device is intended for no-invasive measuring and monitoring of arterial blood pressure.
It is not intended for use on extremities other than the arm or for functions other than obtaining a blood
pressure measurement.
* Do not confuse self-monitoring with self-diagnosis. This unit allows you to monitor your blood pressure.Do
not begin or end medical treatment without asking a physician for treatment advice.
* If you are taking medication,consult your physician to determine the most appropriate time to measure your
blood pressure. Never change a prescribed medication without consulting your physician.
* Do not take any therapeutic measures on the basis of a self measurement. Never alter the dose of a
medicine prescribed by a doctor. Consult your doctor if you have any question about your blood pressure.
* When the device was used to measure patients who have common arrhythmias such as atrial or ventricular
premature beats or atrial fibrillation, the best result may occur with deviation. Please consult your physician
about the result.
* Don't kink the connection tube during use, otherwise, the cuff pressure may continuously increase which
can prevent blood flow and result in harmful injury to the PATIENT.
* When using this device, please pay attention to the following situation which may interrupt blood flow and
influence blood circulation of the patient, thus cause harmful injury to the patient: connection tubing kinking
too frequent and consecutive multiple measurements; the application of the cuff and its pressurization on any
arm where intravascular access or therapy, or an arterio-venous (A-V) shunt, is present; inflating the cuff on
the side of a mastectomy.
* Warning: Do not apply the cuff over a wound;otherwise it can cause further injury.
*Do not inflate the cuff on the same limb which other monitoring ME equipment is applied around
simultaneously, because this could cause temporary loss of function of those simultaneously-used monitoring
ME equipment.
*On the rare occasion of a fault causing the cuff to remain fully inflated during measurement, open the cuff
immediately. Prolonged high pressure (cuff pressure > 300mmHg or constant pressure > 15mmHg for more
than 3 minutes) applied to the arm may lead to an ecchymosis.
*Please check that operation of the device does not result in prolonged impairment of patient blood
circulation.
* When measurement, please avoid compression or restriction of the connection tubing.
* The device cannot be used with HF surgical equipment at the same time.
* The ACCOMPANYING DOCUMENT shall disclose that the SPHYGMOMANOMETER was clinically
investigated according to the requirements of ISO 81060-2:2013.
* To verify the calibration of the AUTOMATED SPHYGMOMANOMETER, please contact the manufacturer.
* This device is contraindicated for any female who may be suspected of, or is pregnant. Besides providing
inaccurate readings, the effects of this device on the fetus are unknown.
* Too frequent and consecutive measurements could cause disturbances in blood circulation and injuries.
* This unit is not suitable for continuous monitoring during medical emergencies or operations.Otherwise, the
patient’s arm and fingers will become anaesthetic, swollen and even purple due to a lack of blood.
* When not in use, store the device with the adapter in a dry room and protect it against extreme moisture,
heat, lint, dust and direct sunlight. Never place any heavy objects on the storage case.
* This device may be used only for the purpose described in this booklet. The manufacturer cannot be held
liable for damage caused by incorrect application.
*This device comprises sensitive components and must be treated with caution. Observe the storage and
operating conditions described in this booklet.
4
CAUTION
* The equipment is not AP/APG equipment and not suitable for use in the presence of a flammable
anesthetic mixture with air of with oxygen or nitrous oxide.
* Warning: No servicing/maintenance while the ME equipment is in use.
* The patient is an intended operator.
* The patient can measure ,transmit data and change batteries under normal circumstances and
maintain the device and its accessories according to the user manual.
* To avoid measurement errors, please avoid the condition of strong electromagnetic field radiated
interference signal or electrical fast transient/burst signal.
*The blood pressure monitor, its adaptor, and the cuff are suitable for use within the patient environment.
If you are allergic to polyester, nylon or plastic, please don't use this device.
* During use, the patient will be in contact with the cuff. The materials of the cuff have been tested and
found to comply with requirements of ISO 10993-5:2009 and ISO 10993-10:2010. It will not cause any
potential sensization or irritation reaction.
* Adaptor is specified as a part of ME EQUIPMENT.
* If you experience discomfort during a measurement, such as pain in the arm or other complaints, press
the START/STOP button to release the air immediately from the cuff. Loosen the cuff and remove it from
your arm.
* If the cuff pressure reaches 40 kPa (300 mmHg), the unit will automatically deflate. Should the cuff not
deflate when pressures reaches 40 kPa (300 mmHg), detach the cuff from the arm and press the
START/STOP button to stop inflation.
* Before use, make sure the device functions safely and is in proper working condition. Check the device,
do not use the device if it is damaged in any way. The continuous use of a damaged unit may cause
injury, improper results, or serious danger.
* Do not wash the cuff in a washing machine or dishwasher!
* The service life of the cuff may vary by the frequency of washing, skin condition, and storage state. The
typical service life is 10000 times.
* It is recommended that the performance should be checked every 2 years and after maintenance and
repair, by retesting at least the requirements in limits of the error of the cuff pressure indication and air
leakage (testing at least at 50mmHg and 200mmHg).
* Please dispose of ACCESSORIES, detachable parts, and the ME EQUIPMENT according to the local
guidelines.
* Manufacturer will make available on request circuit diagrams, component part lists, descriptions,
calibration instructions,etc., to assist to service personnel in parts repair.
* The plug/adapter plug pins insulates the device from the main supply. Do not position the device in a
position where it is difficult to disconnect from the supply mains to safely terminate operation of ME
equipment.
* The operator shall not touch output of batteries /adapter and the patient simultaneously.
* Cleaning :Dust environment may affect the performance of the unit. Please use the soft cloth to clean
the whole unit before and after use. Don’t use any abrasive or volatile cleaners.
* The device doesn’t need to be calibrated within two years of reliable service.
* If you have any problems with this device, such as setting up, maintaining or using, please contact the
SERVICE PERSONNEL of Transtek. Don’t open or repair the device by yourself in the event of
malfunctions. The device must only be serviced, repaired and opened by individuals at authorized
sales/service centers.
* Please report to Transtek if any unexpected operation or events occur.
* Keep the unit out of reach of infants, young children or pets to avoid inhalation or swallowing of small
parts. It is dangerous or even fatal.
5

INTRODUCTIONINTRODUCTION
CAUTION
* Be careful to strangulation due to cables and hoses, particularly due to excessive length.
* At least 30 min required for ME equipment to warm from the minimum storage temperature between
uses until it is ready for intended use. At least 30 min required for ME equipment to cool from the
maximum storage temperature between uses until it is ready for intended use.
* This equipment needs to be installed and put into service in accordance with the information provided in
the ACCOMPANYING DOCUMENTS;
* Wireless communications equipment such as wireless home network devices, mobile phones, cordless
telephones and their base stations, walkie-talkies can affect this equipment and should be kept at least a
distance d away from the equipment. The distance d is calculated by the MANUFACTURER from the 80
MHz to 5.8 GHz column of Table 4 and Table 9 of IEC 60601-1-2:2014, as appropriate.
* Please use ACCESSORIES and detachable partes specified/ authorised by MANUFACTURE.
Otherwise, it may cause damage to the unit or danger to the user/patients.
* There is no luer lock connectors are used in the construction of tubing, there is a possibility that they
might be inadvertently connected to intravascular fluid systems, allowing air to be pumped into a blood
vessel.
* Please use the device under the environment which was provided in the user manual. Otherwise, the
performance and lifetime of the device will be impacted and reduced.
6
LCD display signal
SYMBOL DESCRIPTION
Systolic blood pressure High blood pressure
Diastolic blood pressure Low blood pressure
Memory
Motion indicator
kPa
mmHg
Low battery Batteries are low and need to be replaced
Irregular heartbeat
Blood pressure
level indicator
Current Time
Heartbeat
User 1
User 2
Bluetooth icon
Average value
Bluetooth connection
EXPLANATION
Pulse in beats per minutePulse display
Indicate it is in the memory mode and
which group of memory it is.
Motion may result in an inaccurate
measurement
Measurement Unit of the blood pressure
(1kPa=7.5mmHg)
Measurement Unit of the blood pressure
(1mmHg=0.133kPa)
Blood pressure monitor is detecting an
irregular heartbeat during measurement.
Indicate the blood pressure level
Year/Month/Day, Hour : Minute
Blood pressure monitor is detecting a
heartbeat during measurement.
Start measurement,save and transmit the
measuring results for User A
Start measurement,save and transmit the
measuring results for User B
The bluetooth icon blinks when the
bluetooth is working
The average value of blood pressure
It blinks when the bluetooth connection
fails or the data is not sent.
7

INTRODUCTION BEFORE YOU START
Monitor Components
CUFF
AIR HOSE
AIR CONNECTOR PLUG
SET BUTTON
START/STOP
MEM BUTTON
List
1.Blood Pressure Monitor
(TMB-1585-BS)
3. 4×AAA alkaline batteries
8
Component list of
pressure measuring system
1 Cuff
2 Air pipe
3 PCBA
4 Pump
5 Valve
LCD DISPLAY
DC POWER SOCKET
2.Cuff (Type BF applied part)
(22cm~32cm or 22cm~42cm)
4.User manual
(Please use TRANSTEK
authorized cuff. The size
of the actual cuff please
refer to the label on the
attached cuff.)
5. AC Adaptor
BLJ06L060100P-U)
(
BATTERY COMPARTMENT
The Choice of Power Supply
1.Battery powered mode:
6VDC 4×AAA alkaline batteries
2.AC adaptor powered mode:
6V 1A
(Please only use the recommended AC
adaptor model).
Please unplug the adaptor to depart from
the using utility power.
AC adaptor
CAUTION
In order to get the best effect and protect your monitor,please use the right
battery and special power adaptor which complies with CE safety standard.
Installing and Replacing the Batteries
• Open the battery cover.
• Install the batteries by matching
the correct polarity, as shown.
• Replace the battery cover.
Replace the batteries whenever the below happen
The shows
The display is dim.
The display does not light up.
CAUTION
Do not use new and used batteries together.
Do not use different types of batteries together.
Do not dispose the batteries in fire. Batteries may explode or leak.
Remove batteries if the device is not likely to be used for some time.
Worn batteries are harmful to the environment. Do not dispose with daily garbage.
Remove the old batteries from the device following your local recycling guidelines.
The typical service life of the new and unused batteries is 300 measurements for the
operation time is 60s
9