Transtek Medical Electronics BS 1507 User Manual

Welch Allyn Remote Monitoring Scale RPM-Scale100
Directions for use
901077 Weight scale, Software Version 1.0
© 2015 Welch Allyn. All rights are reserved. To support the intended use of the product described in this publication, the purchaser of the product is permitted to copy this publication, for internal distribution only, from the media provided byWelch Allyn. No other use, reproduction, or distribution of this publication, or any part of it, is permitted without written permission from Welch Allyn.
product, that may result from failure to use this product in accordance with the instructions, cautions, warnings, or statement of intended use published in this manual.
The
Bluetooth
®
word mark and logos are registered trademarks owned by
Bluetooth SIG, Inc. and any
use of such marks by Welch Allyn is under license. Software in this product is Copyright 2015 Welch Allyn or its vendors. All rights are reserved. The
software is protected by United States of America copyright laws and international treaty provisions applicable worldwide. Under such laws, the licensee is entitled to use the copy of the software incorporated with this instrument as intended in the operation of the product in which it is embedded. The software may not be copied, decompiled, reverse-engineered, disassembled, or otherwise reduced to human-perceivable form. This is not a sale of the software or any copy of the software; all right, title, and ownership of the software remain with Welch Allyn or its vendors.
For information about any Welch Allyn product, contact your local Welch Allyn representative:
www.welchallyn.com/about/company/locations.htm.
723809
DIR 80020550 Ver. A Revised 2016–01
This manual applies to
901077 Weight scale.
Manufactured for Welch Allyn, Inc. 4341 State Street Road Skaneateles Falls, NY 13153-0220 USA
Guangdong Transtek Medical Electronics Co., Ltd. Zone A, No. 105, Dongli Road Torch Development District Zhongshan, 528437, Guangdong, China
www.welchallyn.com

Contents

Introductory information ........................................................................ 1
Use the scale ............................................................................................ 7
iii
Introduction .......................................................................................................... 1
Symbols ............................................................................................................... 1
About warnings and cautions ............................................................................... 3
Ensure accurate measurements .......................................................................... 4
Content list ........................................................................................................... 4
Controls and indicators ........................................................................................ 4
Screen elements .................................................................................................. 5
Startup ................................................................................................................. 7
Insert or replace the batteries .............................................................................. 7
Pair a connection hub with the scale ................................................................... 7
Select the measurement unit .............................................................................. 8
Obtain measurement ........................................................................................... 8
Stop a measurement ........................................................................................... 8
Maintain the scale ................................................................................................ 8
Troubleshooting .................................................................................................. 9
Appendices ............................................................................................ 11
Specifications ..................................................................................................... 11
General radio compliance ................................................................................... 12
Complied standards list ...................................................................................... 13
Emissions and immunity information ................................................................. 14
Warranty ............................................................................................................ 17
Approved accessories ........................................................................................ 18
iv Contents Welch Allyn Remote Monitoring Scale RPM-Scale100

Introductory information

Introduction

The Welch Allyn Remote Monitoring Scale is intended for adults and children (10-17 years old) to measure weight in a home environment.
Directions for use contains important safety and care information and provides step
This by step instructions for using the scale. Read the manual thoroughly before using the scale.
1
For information about any Welch Allyn product, contact your local Welch Allyn representative at www.welchallyn.com/about/company/locations.htm.

Indications for use

The Welch Allyn Remote Monitoring Scale is intended for adults and children (10–17 years old) to measure weight in home.

Contraindications

This device is contraindicated for any person who is connected to a wearable or implantable electronic device or instrument, such as a pacemaker or defibrillator.

Symbols

Documentation symbols

Warning: The warning statements in this manual identify conditions or practices that could lead to illness, injury, or death. Warning symbols will appear with a grey background in a black and white document.
Caution: The caution statements in this manual identify conditions or practices that could result in damage to the equipment or other property, or loss of data.
Mandatory - Consult Directions for Use
2 Introductory information Welch Allyn Remote Monitoring Scale RPM-Scale100

Power symbols

Direct current (DC)

Shipping, storing, and environment symbols

Humidity limitation Separate collection of
Electrical and Electronic Equipment. Do not dispose as unsorted municipal waste.
Temperature limit Atmospheric pressure
limitation
Recycle Stacking limit by number

Connectivity symbols

Bluetooth® wireless technology enabled

Miscellaneous symbols

Serial number Manufacturer
Product identifier, product family catalogue number Non-ionizing electromagnetic
Reorder number, Model number Global Trade Item Number
IP22
The ingress protection: the device could protected against solid foreign objects of 12.5mm and greater, and against vertically falling water drops when ENCLOSURE tilted up to 15°
radiation
Type BF applied parts
Directions for use Introductory information

About warnings and cautions

Caution statements can appear on the Welch Allyn weight scale device, the packaging, the shipping container, or in this
The Welch Allyn weight scale is safe for patients and clinicians when used in accordance with the instructions and caution statements presented in this Directions for use.
Before using the device, you must familiarize yourself with all cautions, with the steps to power up the device, and with the sections of this Directions for use that pertain to your use of the device. In addition to reviewing the general cautions presented in the next section, you must also review the more specific cautions that appear throughout the manual in conjunction with setup/startup, operation, and maintenance tasks.
Failure to understand and observe any warning statement in this manual could lead to patient injury or illness.
Failure to understand and observe any caution statement in this manual could lead to damage to the equipment or other property, or loss of patient data.

General warnings and cautions

Directions for use.
3
Warning Patient injury risk. Do not use the scale on a slippery floor, such as tile.
Warning Patient injury risk. Do not jump onto the platform immediately after bathing or with wet hands.
Warning Patient injury risk. Keep the device clean. Patient contact with contaminated equipment can spread infection. Clean the device after each patient use.
Warning Patient injury risk. Dispose of accessories, detachable parts, and the ME equipment according to the local guidelines.
Warning Patient injury risk. The device is not suitable for use in the presence of flammable anesthetic mixtures with air or with oxygen or nitrous oxide.
Warning Patient injury risk. No modification of this equipment is allowed. Modifying the equipment could cause patient injury.
Warning Patient injury risk. Do not touch output of the batteries and the patient simultaneously.
Warning Patient injury risk. Before use, make sure the device functions safely and is in proper working condition.
Warning Patient injury risk. When the scale is in use, there should not be any electrical appliances nearby, such as high voltage cables, X-ray machines, or ultrasound equipment.
Warning Possible injury risk. The device contains small parts. Keep the equipment away from children and pets.
CAUTION Avoid storing the device where the device may be exposed to extreme temperatures, humidity, moisture, direct sunlight, dust, or salt air.
4 Introductory information Welch Allyn Remote Monitoring Scale RPM-Scale100
CAUTION Avoid storing the device where there is risk of shock or dropping.
CAUTION Avoid storing the device where you store chemicals.
CAUTION Avoid storing the device where there are corrosive gases.
CAUTION Avoid storing the device near infants or children.
CAUTION The device does not need to be calibrated in two years of
reliable service. CAUTION Do not attempt to repair the unit yourself in the event of
malfunctions. Only have repairs carried out by authorized service centers. CAUTION Manufacturer will make available on request circuit diagrams,
component parts lists, and the like. CAUTION Wireless communications equipment, such as wireless home
network devices, mobile phones, cordless telephones and their base stations, walkietalkies can affect this equipment and should be kept at least a distance d=3,3m away from the equipment. Note: As indicated in Table 6 of IEC 60601-1-2:2007 for ME Equipment, a typical cell phone with a maximum output power of 2 W yields d=3,3m at an IMMUNITY LEVEL of 3V/m).
CAUTION Do not use rechargeable batteries.

Ensure accurate measurements

To ensure the accuracy of measurements, follow the recommendations below when you start a measurement.
Place the scale on a flat, hard surface. If you put the scale on carpet, fit the extension feet before taking the measurement. Placing the scale sensors directly on carpet may affect the accuracy of the measurement.
Always start a measurement at the same time on the same scale located on the same flat surface.

Content list

The following items are in the device box:
Remote Monitoring Scale
(4) AAA alkaline batteries (1.5V each)
(4) extension feet for use on carpet
Directions for use

Controls and indicators

The Welch Allyn weight scale is simple to use.
Directions for use Introductory information

Scale top

5
No. Feature Description
1 LCD display Displays measurement
2 UNIT and activation button The button enables you to select kg or lb unit measure. Press and
3 Battery compartment (behind cover) Houses (4) AAA alkaline batteries (1.5V each)
4 Sensors (4) 4 load cell sensors which conduct weight measurement

Screen elements

The liquid crystal display (LCD) display registers the measurement. The display also may show other icons that provide information about the functionality of the scale.
hold for 3 seconds to enter pairing mode.
6 Introductory information Welch Allyn Remote Monitoring Scale RPM-Scale100
Icon Description Icon Description
kg
lb
Kilogram Data transmitting
Pound Data pending to transmit to
wireless wellness system
Low battery charge Full battery charge
Bluetooth wireless technology is powered on
Data being transferred using Bluetooth wireless technology (dotted area flashes)
Bluetooth wireless technology connected (dotted area is not flashing)
Scale not paired with device

Use the scale

Startup

The Welch Allyn weight scale uses SENSE ON™ technology to turn on the scale. Step on the scale with bare feet to begin a measurement.

Insert or replace the batteries

7
CAUTION When the symbol appears, the device will power off in four seconds. Replace the old batteries with a new set of four batteries. Always replace all four batteries with new batteries. Do NOT mix old batteries with new batteries
CAUTION Remove the batteries if the device is not to be used for some time.
CAUTION Dispose old batteries by following your local recycling guidelines.
CAUTION Do not burn batteries. Batteries may leak or explode.
1. Open the battery door on the under side of the scale. The digits “8888” appear on the LCD screen.
2. Insert the four AAA alkaline batteries (1.5V each) into the battery compartment according to the polarity indications marked inside the battery compartment.
3. Close the battery door.
4. Wait until the digits “0.0lb”appear on the LCD screen.

Pair a connection hub with the scale

With advanced Bluetooth 4.0 Technology applied, a communication hub is capable of receiving your personal health data. Contact your Remote Patient Monitoring Solution provider for connectivity instructions and troubleshooting.
1. Confirm that both the connection hub and the weight scale are powered on.
Note
Bluetooth wireless connectivity is only available for devices which support
Bluetooth 4.0 technology.
8 Use the scale Welch Allyn Remote Monitoring Scale RPM-Scale100
2. Press and hold the Unit button on the bottom of the scale for 2 seconds. The Bluetooth pairing sequence is activated. The dotted line surrounding the
Bluetooth symbol flashes while discovering the weight scale.
If successful, a steady
LCD screen. If unsuccessful, the
symbol surrounded by a dotted line appears on the scale's

Select the measurement unit

The default measurement unit is pound (lb). Press the Unit button on the bottom of the scale to change from pound (lb) to
kilogram (kg).

Obtain measurement

With the original SENSE ON patent technology, the scale automatically powers on when you step on the platform in your bare feet.
1. Step on the platform in your bare feet. The scale automatically powers on and begins the measurement.
2. Stand still and keep full contact with the scale until the LCD stops blinking and displays the weight.

Stop a measurement

appears.
You can stop a measurement before it is complete. Step off the platform.
The scale returns to a ready / zero state. After approximately 2 seconds, the scale powers down automatically.

Maintain the scale

Warning Patient injury risk. Keep the device clean. Patient contact with contaminated equipment can spread infection. Clean the device after each patient use.
Clean on a routine basis according to facility protocols and standards or local regulations. Disinfect according to facility protocols and standards or local regulations. The following agents are compatible with the device:
CaviWipes or Sani-Cloth Plus
70 percent isopropyl alcohol
10 percent chlorine bleach/90 percent water solution
Directions for use Use the scale 9

Troubleshooting

This section includes a list of error messages and frequently asked questions for problems you may encounter with your weight scale. If the device is not operating as you think it should, check here before arranging for servicing.

Error prompt

Error Description Solution
Dotted lines Overload. The device will power off Stop using this scale for measurement
Error condition At any point, when a situation occurs

When measuring

Problem Root cause Solution
No display on the LCD screen with the scale powers on
Low Battery. The device will power off in four seconds.
that prevents a successful test acquisition, an error code will be displayed in the LCD area. The message will persist for several seconds until the device times out and powers down.
Batteries are not installed or not installed properly
Worn batteries Replace all four batteries at the same
Replace all four batteries at the same time. Purchase the authorized batteries for replacement.
Check the following items:
Bluetooth wireless technology is
• on.
The application is on.
Both devices are within the transmission distance of Bluetooth devices.
Install the batteries. Go to "Insert or replace the batteries"
time. Purchase the authorized batteries for replacement.
The scale powers off automatically
Low Battery.
Replace all four batteries at the same time. Purchase the authorized batteries for replacement.
10 Use the scale Welch Allyn Remote Monitoring Scale RPM-Scale100

When transmitting data

Problem Root cause Solution
Data transmission failed
Bluetooth wireless technology is off Turn on the Bluetooth wireless
technology via the Settings.
Application is off
Out of range of Bluetooth transmission. Place your mobile or portable device
Press the icon to run your application.
closer to the scale.

Appendices

Specifications

Warning: No modification of this equipment is allowed.
Item Specification
11
Power supply
Display mode Blue LCD with White Backlight V.A.: 74x53mm
Measurement unit Kilogram / Pound
Measurement range
Resolution
Normal working condition Temperature: 5 to 40 Humidity: 85%RH
Division 0.1kg / 1.2lb
Storage & transportation condition
6V 4*AAA alkaline batteries
5kg to 200kg / 11lb to 441lb
The measurement resolution shall be:
0–100kg: 0.1kg 100–200kg: 0.2kg
100-150kg: ±0.5kg; 150-200kg: ±0.7kg
Atmospheric pressure: 86kPa to 106kPa
Temperature:-20to 60 Relative Humidity: 10% RH–93% RH
Atmospheric Pressure: 50kPa to 106 kPa
0–220 lb: 0.2lb 220–440lb: 0.5lb
Net Weight Approximately 1.98kg ( Excluding the dry cells)
Dimensions 360mm x 408mm x 32.5mm with carpet feet
Accessories
(4) AAA alkaline batteries
(4) extension feet
Directions for use
12 Appendices Welch Allyn Remote Monitoring Scale RPM-Scale100
Item Specification
Mode of operation Continuous operation
Auto-ON SENSE ON technology
Auto-OFF About 10 seconds when showing 0.0
About 15 seconds after the weight data is locked
Degree of protection Type BF applied part
Protection against ingress of water
Software Version Version 1.0
IP22

General radio compliance

Item Specification
Bluetooth Module No. AW8001
RF Frequency Range 2402 MHz to 2480 MHz
Output Power Range -2.27dBm to -5.88 dBm
Supply Voltage 1.8-3.6 V
Antenna Gain 1.0dBi
Transmitting Distance 10 meters, (30 feet)

Federal Communication Commission (FCC) Interference Statement

This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions.
•This device may not cause harmful interference.
•This device must accept any interference received, including interference that may cause undesired operation.
This equipment (FCC ID: OU9BS-1507) has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one of the following measures.
1.Reorient or relocate the receiving antenna.
Directions for use Appendices 13
2.Increase the separation between the equipment and receiver.
3.Consult the dealer or an experienced radio/TV technician for help. Any changes or modifications not expressly approved by the party responsible for
compliance could void the user's authority to operate this equipment.

FCC Radiation Exposure Statement

This equipment complies with FCC radiation exposure limits set forth for an uncontrolled environment. This transmitter must not be co-located or operating in conjunction with any other antenna or transmitter.

Industry Canada (IC) compliance

To ensure compliance with FCC and Industry Canada RF exposure requirements, this device must be installed in a location where the antennas of the device will have a minimum distance of at least 20 cm from all persons. Using higher gain antennas and types of antennas not certified for use with this product is not allowed. The device shall not be co-located with another transmitter.
Installez l'appareil en veillant à conserver une distance d'au moins 20 cm entre les éléments rayonnants et les personnes. Cet avertissement de sécurité est conforme aux limites d'exposition définies par la norme CNR-102 at relative aux fréquences radio.
This device complies with Industry Canada licence-exempt RSS standard(s). Operation is subject to the following two conditions: (1) this device may not cause interference, and (2) this device must accept any interference, including interference that may cause undesired operation of the device.
Le présent appareil est conforme aux CNR d'Industrie Canada applicables aux appareils radio exempts de licence. L'exploitation est autorisée aux deux conditions suivantes : (1) l'appareil ne doit pas produire de brouillage, et (2) l'utilisateur de l'appareil doit accepter tout brouillage radioélectrique subi, même si le brouillage est susceptible d'en compromettre le fonctionnement.
This radio transmitter (IC ID: 12725A-BS1507) has been approved by Industry Canada to operate with the antenna listed in the specification table.
This Class B digital apparatus complies with Canadian ICES-003. Cet appareil numérique de la classe B est conform à la norme NMB-003 du Canada

Complied standards list

Item Standard
Risk management ISO/EN 14971 Medical devices — Application of risk
management to medical devices
Labeling ISO/EN 15223-1 Medical devices. Symbols to be used with
medical device labels, labeling and information to be supplied. General requirements
Tested in accordance with ISO 10993 Biocompatibility
14 Appendices Welch Allyn Remote Monitoring Scale RPM-Scale100
Item Standard
User manual EN 1041 Medical equipment manufacturers to provide
information
General Requirements for Safety
Electromagnetic compatibility IEC/EN 60601-1-2 Medical electrical equipment - Part 1-2:
Software life-cycle processes
Usability
IEC 60601-1+A1 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
IEC 60601-1-11 Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests
IEC/EN 62304+AC Medical device software - Software life cycle processes
IEC 62366 Medical devices - Application of usability engineering to medical devices (IEC 62366)
IEC 60601-1-6 Medical electrical equipment - Part 1 -6 : General requirements for basic safety and essential performance - collateral standard : Usability

Emissions and immunity information

The Welch Allyn Remote Monitoring Scale RPM-SCALE100 is intended for use in the electromagnetic environment specified below. The customer or user of the Welch Allyn Remote Monitoring Scale RPM-SCALE100 should assure that it is used in such an environment.
This equipment needs to be installed and put into service in accordance with the information provided in the user manual.
CAUTION Wireless communications equipment, such as wireless home network devices, mobile phones, cordless telephones and their base stations, walkie-talkie can affect this equipment and should be kept at least a distance d=3,3m away from the equipment.
Note As indicated in Table 6 of IEC 60601-1-2:2007 for ME Equipment, a typical
cell phone with a maximum output power of 2 W yields d=3,3m at an IMMUNITY LEVEL of 3V/m).
Electromagnetic emissions
The Welch Allyn Remote Monitoring Scale RPM-Scale100 is intended for use in the electromagnetic environment specified below. The customer or user of the Welch Allyn Remote Monitoring Scale RPM-Scale100 should assure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment - guidance
Directions for use Appendices 15
Electromagnetic emissions
RF emissions CISPR 11
Group 1 The Welch Allyn Remote Monitoring Scale RPM-Scale100 uses RF
energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
RF emissions CISPR 11
Class B
The Welch Allyn Remote Monitoring Scale RPM-Scale100 is suitable for use in all establishments other than domestic establishments and those directly connected to the public low-voltage power supply
Harmonic emissions
Not applicable
network that supplies buildings used for domestic purposes.
IEC 61000-3-2
Voltage fluctuations/
Not applicable
flicker emissions IEC 61000-3-3
Electromagnetic immunity
The Welch Allyn Remote Monitoring Scale RPM-Scale100 is intended for use in the electromagnetic environment specified below. The customer or the user of the Welch Allyn Remote Monitoring Scale RPM-Scale100 should assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment - guidance
Electrostatic discharge (ESD)
IEC 61000-4-2
±6 kV contact ± 8 kV air
± 6 kV contact ± 8 kV air
Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%.
Electrical fast transient/burst
IEC 61000-4-4
Surge IEC 61000-4-5
Voltage dips, short interruptions and voltage variations on power supply input lines
IEC 61000-4-11
±2 kV for power supply lines
±2 kV for power supply lines
± 1 kV line(s) to line(s) ± 1 kV differential
mode
<5% U
T
(>95% dip in UT) for 0.5 cycle
40% U
T
(60% dip in UT) for 5 cycles
70% U
T
(30% dip in UT) for 25 cycles
<5% U
T
(>95% dip in UT) for 5 seconds
<5% U
T
(>95% dip in UT) for 0.5 cycle
40% U
T
(60% dip in UT)for 5 cycles
70% U
T
(30% dip in UT) for 25 cycles
<5% U
T
(>95% dip in UT) for 5 seconds
Mains power quality should be that of a typical commercial or hospital environment.
Mains power quality should be that of a typical commercial or hospital environment.
Mains power quality should be that of a typical commercial or hospital environment. If the user of the Welch Allyn Remote Monitoring ScaleRPM-Scale100 requires continued operation during power mains interruptions, it is recommended that the Welch Allyn Remote Monitoring Scale RPM-Scale100 be powered from an uninterruptible power supply or a battery.
16 Appendices Welch Allyn Remote Monitoring Scale RPM-Scale100
Electromagnetic immunity
Power frequency (50Hz) magnetic field IEC 61000-4-8
3A/m 3A/m Power frequency magnetic fields should be at
levels characteristic of a typical location in a typical commercial or hospital environment.
Note: UT is the AC mains voltage prior to application of the test level.
Electromagnetic immunity
The Welch Allyn Remote Monitoring Scale RPM-Scale100 is intended for use in the electromagnetic environment specified below. The customer or the user of the Welch Allyn Remote Monitoring Scale RPM-Scale100 should assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment - guidance
Portable and mobile RF communications equipment should be used no closer to any part of the Welch Allyn Remote Monitoring Scale RPM-Scale100, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
Recommended separation distance
Conducted RF IEC 61000-4-6
3 Vrms 150 kHz to 80 MHz
3 Vrms
d= (1.167)
Radiated RF
3 V/m 80 MHz to 1 GHz 3 V/m
d= (1.167) 80 to 800 MHz
IEC 61000-4-3
d= (2.333) 800 MHz to 2,5 GHz
P is the maximum output power rating of
where the transmitter in watts (W) andd is the recommended separation distance in meters (m). Field strengths from fixed RF transmitters, as determined by an electromagnetic site
surveya, should be less than the compliance level in each frequency rangeb. Interference
may occur in the vicinity of equipment marked with the following symbol:
Note1: At 80 MHz and 800 MHz, the higher frequency range applies. Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects, and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast, and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the Welch Allyn Remote Monitoring Scale RPM­Scale100 is used exceeds the applicable RF compliance level above, the Welch Allyn Remote Monitoring Scale RPM-
Directions for use Appendices
Electromagnetic immunity
Scale100 should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the Welch Allyn Remote Monitoring Scale RPM-Scale100.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Recommended separation distances between portable and mobile RF communications equipment and the Welch Allyn Remote Monitoring scale RPM-SCALE100
The Welch Allyn Remote Monitoring Scale RPM-Scale100 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or user of the Welch Allyn Remote Monitoring Scale RPM­Scale100 can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the Welch Allyn Remote Monitoring Scale RPM-Scale100 as recommended below, according to the maximum output power of the communications equipment.
Separation distance according to frequency of transmitter (m)
17

Warranty

Rated max. output power of transmitter (W)
0.01 0.167 0.167 0.233
0.1 0.369 0.369 0.738
1 1.167 1.167 2.333
10 3.690 3.690 7.338
100 11.67 11.67 23.33
For transmitters rated at a maximum output power not listed above, the recommended separation distanced in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects, and people.
150 kHz to 80 MHz
d= (1.167)
80 MHz to 800 MHz
d= (1.167)
800 MHz to 2.5 GHz
d= (2.333)
Welch Allyn will warranty the Transtek weight scale to be free of defects in material and workmanship and to perform in accordance with manufacturer specifications for the period of two years from the date of retail purchase.
The warranty period shall start on the date of purchase. The date of purchase is: 1) the invoiced ship date if the device was purchased directly from Welch Allyn, 2) the date specified during product registration, 3) the date of purchase of the product from a Welch Allyn authorized distributor as documented from a receipt from said distributor.
This warranty does NOT cover damages caused by misuse or abuse, including but not limited to:
Failure caused by unauthorized repairs or modifications
Damage caused by shock or dropping during transportation
18 Appendices Welch Allyn Remote Monitoring Scale RPM-Scale100
Damage caused by improper use of the power supply
Failure caused by improper operation not consistent with the instructions stated in
this Directions for use
Should this device require maintenance (or replacement at our option) under warranty, please deliver the original package to Welch Allyn prepaid. Please return the store receipt (with the retail purchase date) and a note with the reasons for the return.

Approved accessories

Item Description
RPM-SCALEACC-01 Carpet Feet 40EA