TRANSTEK Medel Soft TMB-1581-S User Manual

USER MANUAL
Wrist blood pressure monitor
MEDEL SOFT
Model: TMB-1581-S
EN
1
A
LCD DISPLAY
GRADE
D
MEM BUTTON
C
SET BUTTON
E
DIASTOLIC
F
PULSE RATE
G
START/STOP BUTTON
B
2
+
-
-
+
4
A B
1
3
EN
C
D
E
G
F
H
2
EN
I
5
A B
C D
3
6
A B
7
A B
C
EN
4
EN
(DISPLAY SYMBOL (PIC. 3
SYMBOL DESCRIPTION EXPLANATION
SYS
DIA
INTRODUCTIONS
Dear customer, ank you for choosing one of our products. Our name stands for hi­gh-quality, thoroughly tested products. Please read these instructions for use carefully and keep them for later use, be sure to make them accessible to other users and observe the information they contain. With kind regards, Your Medel team
INDICATIONS FOR USE
Medel So is a digital monitor intended for use in measuring blood pres­sure and heartbeat rate with a wrist circumference ranging from 13.5cm to 21.5 cm ( about 5½˝-8½˝ ). It is intended for adult indoor use only.
Systolic blood pressure High blood pressure
Dystolic blood pressure Low blood pressure
Pulse display Pulse in bests per minutes
Excessive Body Motion Detector
Low battery Batteries are running low and
mmHg (1mmHg=0.133kPa)Measurement
Irregular heartbeat Detects an irregular heartbeat
Current time Month/Day/Year,Hour : Minute
Blood pressure level indicator
Heartbeat Month/Day/Year,Hour : Minute
Motion may result in an inaccura­te measurement
need to be replaced
unit the blood pressure
during measurement.
Indicate the blood pressure level
5
CONTRAINDICATION
1.e device should not be used by any person who may be suspected of,or is pregnant .
2.e device is not suitable for use on patients with implanted,electrical devices, such as cardiac pacemakers, debrillators.
MEASUREMENT PRINCIPLE
is product uses the Oscillometric Measuring Method to detect blood pressure. Before every measurement, the unit establishes a “zero point” equivalent to the atmospheric pressure. en it starts inating the cu. Meanwhile, the unit detects pressure oscillation generated by beat-to-beat pulsatile, which is used to determine the systolic pressure and diastolic pressure as well as pulse rate.
SAFETY INFORMATION
e signs below might be in the user manual, labeling or other compo­nent.ey are the requirement of standard and using.
Symbol for “THE OPERATION GUIDE MUST BE READ”
Symbol for “COMPLIES WITH MDD 93/42/EEC REQUIREMENTS”
Symbol for “MANUFACTURER”
Symbol for “SERIAL NUMBER”
SN
Symbol for “RECYCLE”
Symbol for “MANUFACTURE DATE”
The Green Dot is the license symbol of a European network of industry-funded systems for recycling the packaging materials of consumer goods.
Symbol for “TYPE BF APPLIED PARTS”
Symbol for “ENVIRONMENT PROTECTION - Electrical waste products should not be disposed of with household waste. Please recycle where facilities exist. Check with your local authority or retailer for recycling advice”
Symbol for “DIRECT CURRENT”
Symbol for “Authorised Representative
EC REP
in the European Community
Caution: These notes must be observed to prevent any damage to the device.
ATTENTION!
is device is intended for adult use in homes only. * e device is not suitable for use on neonatal patients, pregnant women,patients with implanted, electronical devices, patients with pre-eclampsia, premature ventricular beats, atrial brillation, periph ral, arterial disease and patients undergoing intravascular therapy or arterio-venous shunt or people who received a mastectomy. Please consult your doctor prior to using the unit if you suer from illnesses. * e device is not suitable for measuring the blood pressure of children. Ask your doctor before using it on older children. * e device is not intended for patient transport outside a healthcare facility. * e device is not intended for public use. * is device is intended for no-invasive measuring and monitoring of arterial blood pressure. It is not intended for use on extremities other than the wrist or for functions other than obtaining a blood pressure measurement. * Do not confuse self-monitoring with self-diagnosis. is unit allows you to monitor your blood pressure.Do not begin or end medical treatment without asking a physician for treatment advice. * If you are taking medication,consult your physician to determine the most appropriate time to measure your blood pressure. Never change a prescribed medication without consulting your physician.
EN
6
EN
* Do not take any therapeutic measures on the basis of a self measurement. Never alter the dose of a medicine prescribed by a doctor. Consult your doctor if you have any question about your blood pressure. * When the device was used to measure patients who have common arrhythmias such as atrial or ventricular premature beats or atrial brillation, the best result may occur with deviation. Please consult your physician about the result. * Warning: Do not apply the cu over a wound;otherwise it can cause further injury. *Do not inate the cu on the same limb which other monitoring ME equipment is applied around simultaneously, because this could cause temporary loss of function of those simultaneously-used monitoring ME equipment. *On the rare occasion of a fault causing the cu to remain fully inated during measurement, open the cu immediately. Prolonged high pressure (cu pressure > 300mmHg or constant pressure >15mmHg for more than 3 minutes) applied to the wrist may lead to an ecchymosis. *Please check that operation of the device does not result in prolonged impairment of patient blood circulation. * When measurement, please avoid compression or restriction of the connection tubing. * e device cannot be used with HF surgical equipment at the same time. * e ACCOMPANYING DOCUMENT shall disclose that the SPHYGMOMANOMETER was clinically investigated according to the requirements of ISO 81060-2:2013. * To verify the calibration of the AUTOMATED SPHYGMOMANOMETER, please contact the manufacturer. * is device is contraindicated for any female who may be suspected of, or is pregnant. Besides providing inaccurate readings, the eects of this device on the fetus are unknown. * Too frequent and consecutive measurements could cause disturbances in blood circulation and injuries. * is unit is not suitable for continuous monitoring during medical emergencies or operations.Otherwise, the patient’s wrist and ngers will become anaesthetic, swollen and even purple due to a lack of blood. * When not in use, store the device in a dry room and protect it against extreme moisture, heat, lint, dust and direct sunlight. Never place any heavy objects on the storage case. * is device may be used only for the purpose described in this booklet. e manufacturer cannot be held liable for damage caused by incorrect application. *is device comprises sensitive components and must be treated with caution. Observe the storage and operating conditions described in this booklet. * e maximum temperature that the applied part can be achieved is 42.5°C while the environmental temperature is 40°C. * e equipment is not AP/APG equipment and not suitable for use in the presence of a ammable anesthetic mixture with air of with oxygen or nitrous oxide. * Warning: No servicing/maintenance while the ME equipment is in use. * e patient is an intended operator. * e patient can measure data and change batteries under normal circumstances and maintain the device and its accessories according to the user manual. * To avoid measurement errors, please avoid the condition of strong electromagnetic eld radiated interference signal or electrical fast transient/burst signal. * e blood pressure monitor and the cu are suitable for use within the patient environment. If you are allergic to polyester, nylon or plastic, please don’t use this device. * During use, the patient will be in contact with the cu. e materials of the cu have been tested and found to comply with requirements of ISO 10993-5:2009 and ISO 10993-10:2010. It will not cause any potential sensization or irritation reaction. * If you experience discomfort during a measurement, such as pain in the wrist or other complaints, press the START/STOP button to release the air immediately from the cu. Loosen the cu and remove it from your wrist. * If the cu pressure reaches 40 kPa (300 mmHg), the unit will automatically deate. Should the cu not deate when pressures reaches 40 kPa (300 mmHg), detach the cu from the wrist and press the START/STOP button to stop ination. * Before use, make sure the device functions safely and is in proper working condition. Check the device, do not use the device if it is damaged in any way. e continuous use of a damaged unit may cause injury, improper results, or serious danger. * Do not wash the cu in a washing machine or dishwasher! * e service life of the cu may vary by the frequency of washing, skin condition, and storage state. e typical service life is 10000 times. * It is recommended that the performance should be checked every 2 years and aer maintenance and repair, by retesting at least the requirements in limits of the error of the cu pressure indication and air leakage (testing at least at 50mmHg and 200mmHg). * Manufacturer will make available on request circuit diagrams, component part lists, descriptions, calibration instructions,etc., to assist to service personnel in parts repair. * e operator shall not touch output of batteries and the patient simultaneously. Cleaning :Dust environment may aect the performance of the unit. Please use the so cloth to clean the whole unit before and aer use. Don’t use any abrasive or volatile cleaners. * e device doesn’t need to be calibrated within two years of reliable service. * If you have any problems with this device, such as setting up, maintaining or using, please contact the SERVICE PERSONNEL of Transtek. Don’t open or repair the device by yourself in the event of
.
7
malfunctions. e device must only be serviced, repaired and opened by individuals at authorized sales/service centers. * Please report to Transtek if any unexpected operation or events occur. * Keep the unit out of reach of infants, young children or pets to avoid inhalation or swallowing of small parts. It is dangerous or even fatal. * Be careful to strangulation due to cables and hoses, particularly due to excessive length. * At least 30 min required for ME equipment to warm from the minimum storage temperature between uses until it is ready for intended use. At least 30 min required for ME equipment to cool from the maximum storage temperature between uses until it is ready for intended use. * is equipment needs to be installed and put into service in accordance with the information provided in the ACCOMPANYING DOCUMENTS; * Wireless communications equipment such as wireless home network devices, mobile phones, cordless telephones and their base stations, walkie-talkies can aect this equipment and should be kept at least a distance d away from the equipment. e distance d is caculated by the MANUFACTURER from the 80MHz to 5.8 GHz column of Table 4 and Table 9 of IEC 60601-1-2:2014, as appropriate. * Please use ACCESSORIES and detachable partes specied/ authorised by MANUFACTURE. Otherwise, it may cause damage to the unit or danger to the user/patients. * ere is no luer lock connectors are used in the construction of tubing, there is a possibility that they might be inadvertently connected to intravascular uid systems, allowing air to be pumped into a blood vessel. * Please use the device under the environment which was provided in the user manual. Otherwise, the performance and lifetime of the device will be impacted and reduced.
2. BLOOD PRESSURE MONITOR COMPONENT (PIC. 1)
A.
LCD Display
B.
Start / Stop button
C.
Memory button / Set button (MEM)
D.
Hypertension grade indicator
E.
Systolic blood pressure
F.
Diastolic blood pressure
G.
Pulse rate
COMPONENT LIST OF PRESSURE MEASURING SYSTYEM:
1. PCBA;
2. Air Pipe;
3. Pump;
4. Valve;
5. Cu.
LIST OF ACCESSORIES:
1. Wrist blood pressure monitor
2. Storage case
3. 2 x AAA batteries
4. User manual
EN
8
EN
3. BEFORE YOU START
3.1 INSTALLING AND REPLACING THE BATTERIES (PIC. 2)
Slide o the battery cover. Install the batteries by matching the correct polarity, as show pic. 2. Always use the correct battery type (2 x AAa batteries). Replace the battery cover.
3.2 SETTING DATE, TIME AND MEASUREMENT UNIT (PIC. 4)
It is important to set the clock before using your blood pressur moni­tor, so that a time stamp can be assigned to each record that is stored in the memory. (year: 2018 - 2058, time format: 12H/24H)
1. When the monitor is o, hold pressing “MEM” button for about 3 seconds to enter into setting mode. e blinking numeral represen­ts [YEAR] (pic. 4.A)
2. Press the “MEM” button to change the [YEAR]. Each press will increase the numeral by one in a cycling manner (pic. 4.B).
3. When you get the right year, press “START/STOP” button to conrm your selection and it will turn to the next step (pic. 4.C) (pic.4.D).
4. Repeat step 2 and 3 to conrm [MONTH] and [DAY]
5. (pic. 4.E) (pic. 4.F)
6. Repeat step 2 and 3 to conrm the time formate [12H] and [24H] (pic. 4.G) (pic. 4.H)
7. Repeat step 2 and 3 to conrm [HOUR] and [MINUTE] (pic. 4.I)
Replace the batteries whenever the below happen
The shows
The display is dim.
The display does not light up
CAUTION
Do not use new and used batteries together. Do not use different types of batteries together. Do not dispose the batteries in fire. Batteries may explode or leak. Remove batteries if the device is not likely to be used for some time. Worn batteries are harmful to the environment. Do not dispose with daily garbage. Remove the old batteries from the device following your local recycling guidelines.
9
3.3 TIE THE CUFF
1. Remove all accessories (watch, bracelet,etc) from your wrist. If your physician has diagnosed you with poor circulation in your wrist, use the other one.
2. Roll or push up your sleeve to expose the skin.
3. Apply the cu to your wrist with your palm facing up.
4. Position the edge of the cu about 1cm~1.5cm from wrist joints.
5. Fasten the wrist cu around your wrist, leaving no extra room between the cu and your skin. If the cu is too loose, the measurement will not be accurate.
6. Sit comfortably with your tested wrist resting on a at surface. Place your el­bow on a table so that the cu is at the same level as your heart. Turn your palm upwards. Sit upright in a chair, and take 5-6 deep breaths.
7. Patients with Hypertension: e middle of the cu should be at the level of the right atrium of the heart; Before starting measurement, please sit comfortably with legs uncrossed, feet at on the oor, back and wrist supported. Rest for 5 minutes before measuring.
Wait at least 3 minutes between measurements. is allows your blood circulation to recover. Take the measurement in a silent room. e patient must relax as much as possible and do not move and talk during the measurement procedure.
e cu should maintain at the same level as the right atrium of the heart. Do not cross your legs and keep your feet on the ground. Keep your back against the backrest of the chair. For a meaningful comparison, try to measure under similar conditions. For example, take daily mea­surements at approximately the same time, on the same wrist, or as directed by a physician.
3.4. START THE MEASUREMENT
1. When the monitor is o, press “START/STOP” button to turn on the monitor, and it will nish the whole measurement (pic. 5.A) (pic. 5.B).
2. Press “START/STOP” button to power o, otherwise it will turn o within 1 minute (pic. 5.C) (pic. 5.D).
DATA MANAGEMENT
4.1 RECALL THE RECORDS
1. When the monitor is o, press “MEM” button to show the average value of the latest three measurement records.If the records are less than 3 groups,it will display the latest record instead (pic. 6. A).
2. Press “MEM” button again, it will display the latest measurement result, date and time. Press “MEM” button again, it will display the
EN
10
EN
next record,and so on. During the process of displaying the results, if there is no operation, the blood pressure monitor will turn o in one minute. Or you can press “START/STOP” button to turn it o. (pic. 6.B).
It means the total records is 5, the current No. is No. 1.
CAUTION
The most recent record )1( is shown first. Each new m easurement is assigned t o the first )1( record. All other records a re pushed back one digit ) e.g., 2 becomes 3, and so on(, and the last record ) 60( dropped from the list.
4.2 DELETE THE RECORDS
If you did not get the correct measurement, you can delete all results by followings steps below.
1. Hold pressing “MEM” button for 3 seconds when the monitor is in the memory recall mode, the “dEL All” will ash on the display (pic. 7.A).
2. Press “MEM” to conrm deleting , the LCD displays “ dEL donE” and the monitor will turn o (pic. 7.B).
Note: To exit out of delete mode without deleting any records, press START/ STOP button before pressing “MEM” to conrm any delete commands.
The corresponding date is May 14 .
th
The corresponding time is 8:16.
3. If there is no record, the right display will show (pic. 7.C).
5. INFORMATION FOR USER
5.1 MAINTENANCE
To obtain the best performance, please follow the instructions below.
Put in a dry place and avoid the sunshine
Avoid immersing it in the water. Clean it with a dry cloth in case.
Avoid shaking and collision.
Avoid dusty environment and unstable temperature surrounding.
Use the slightly damp cloth to remove the dirt.
Avoid washing the cu.
11
6. ABOUT BLOOD PRESSURE
6.1 WHAT ARE SYSTOLIC PRESSURE AND DIASTOLIC PRESSURE?
When ventricles contract and pump blood out of the heart, blood pressure reaches its maximum value, the highest pressure in the cycle is known as systolic pressure.
When the heart relaxes between heartbeats, the lowest blood pressure is diastolic pressure.
6.2 WHAT IS THE STANDARD BLOOD PRESSURE CLASSIFICATION?
e blood pressure classication publi­shed by World He­alth Organization (WHO) and Inter­national Society of Hypertension (ISH) in 1999 is as follows:
Level
Blood pressure
(mmHg)
Optimal Normal Mild Moderate SevereNormal
<120
SYS
<80
DIA
120~129
80~84
Hight-
130~139
85~89
140~159
90~99
Systolic
blood discharging
artery
press
160~179
100~109
Diastolic
blood entering vein
relax
>
180
110
>
EN
12
EN
6.3 IRREGULAR HEARTBEAT DETECTOR
An irregular heartbeat is detected when a heartbeat rhythm varies while the device is measuring systolic pressure and diastolic pressure. During each measurement, blood pressure monitor will keep a record of all the pulse intervals and calculate the average value of them. If there are two or more pulse intervals , the dierence between each interval and the average is more than the average value of ±25% , or there are four or more pulse intervals ,the dierence between each interval and the average is more than the average value of ±15%, then the irregular heartbeat symbol will appear on the display with the measurement result.
6.4 WHY DOES MY BLOOD PRESSURE FLUCTUATE THROUGHOUT THE DAY?
1. Individual blood pressure varies multiple times everyday. It is also aected by the way you tie your cu and your measurement posi­tion, so please take the measurement under the same conditions.
2. If the person takes medicine, the pressure will vary more.
3. Wait at least 3 minutes for another measurement.
7. ERROR MESSAGE MALFUNCTION
PROBLEM SYMPTOM CHECK THIS REMEDY
No power
Low batteries
Error message
Warning message
13
Batteries are exhausted.
Display is dim or
Batteries are inserted
will not light up.
incorrectly.
Show on
Batteries are low.
the display
E1 shows
E 2 shows The cuff is very tight
E 3 shows
E 10 or E 11 shows
E 20 shows
E 21 shows
EExx,shows on the display.
“out ” shows
The cuff is not secure.
The pressure of the cuff is excess.
The monitor detected motion, talking,or the pulse is too poor while measuring.
The measurement process does not detect the pulse signal.
Measure incorrectly.
A calibration error occurred.
Out of measurement range
Replace with new batteries
Insert the batteries correctly
Replace with new batteries
Refasten the cuff and then
Refasten the cuff and then measure again.
Relax for a moment and then measure again.
movement can affect the measurement. Relax for a moment and then measure again.
Loosen the clothing on the arm and then measure again.
Relax for a moment and then measure again.
Retake the measurement.If the problem persists,contact the retailer or our customer service department for further assistance. Refer to the warranty for contact information and return instructions.
Relax for a moment. Refasten the cuff and thenmeasure again. If the problem persists, contact your physician.
measure again.
8. SPECIFICATION
EN
Power supply
Display mode
Measurement mode
Measurement range
Accuracy
Working condition
Storage & transportation
condition
Measurement perimeter
of the upper arm
Weight
External dimensions
Attachment
Battery powered mode:
Digital LCD V.A.32mmx45mm
Oscillographic testing mode
Rated cuff pressure: 0mmHg~299mmHg)0kPa ~ 39.9kPa( Measurement pressure: SYS: 60mmHg~230mmHg )8.0kPa~30.7kPa( DIA: 40mmHg~130mmHg )5.3kPa~17.3kPa( Pulse value: )40-199(beat/minute
Pressure:5°C-40°C within±3mmHg)0.4kPa( Pulse value:±5%
A temperature range of :+5°C to +40°C A relative humidity range of 15% to 90%, non-condensing, but not requiring a water vapour partial pressure greater than 50 hPa An atmospheric pressure range of : 700 hPa to 1060 hPa
About 13.5cm-21.5cm
Approx.104g)Excluding the batteries(
Approx.85mm×67mm×23mm)Excluding the cuff(
2*AAA batter ies,user manual,PP case
Mode of operation
Degree of protection Type BF applied part
Device Classification
Continuous operation
Internally Powered ME Equipment
2*AAA batteries
IP Classification
Software Version
A01
WARNING: No modication of this equipment is allowed.
14
EN
9.WARRANTY CONDITIONS
The device is guaranteed for 5 years from the date of original purchase against any defect in materials or workmanship.
The warranty consists in the replacement and/or repair, free of charge, of originally defective components.
The warranty does not cover the accessories supplied and the parts subject to normal wear and tear. The device may only be repaired by authorised technical service centres.
The appliance must be sent to MEDEL CUSTOMER SERVICE for repairs.
The transport costs shall be borne by the user.
Any repair out of warranty shall be borne by the user. The warranty lapses if the device has been tampered with, if the defect was caused by improper use or in case the damage is not due to the manufacturer (accidental fall, incorrect transport etc.).
The warranty does not involve any direct or indirect damages of any kind to people or property during the period of ineffi­ciency of the product.
The warranty is valid from the date of purchase certified by the receipt or invoice.
15
10. REFERENCE TO STANDARD
10.1 COMPLIED LIST
Risk management
Labeling
User manual
General Requirements for Safety
Electromagnetic compatibility
Performance requirements
Clinical investigation
Usability
Software life-cycle processes
Bio-compatibility
EN ISO 14971:2012 / ISO 14971:2007 Medical devices -
Application of risk management to medical devices
EN ISO 15223-1:2016 / ISO 15223-1:2016 Medical devices. Symbols to be used with medical device labels, labelling and information to be supplied. Part 1 : General requirements
EN 1041:2008 Information supplied by the manufacturer of medical devices
EN 60601-1:2006+A1:2013/ IEC 60601-1:2005+A1:2012 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
EN 60601-1-11:2015/ IEC 60601-1-11:2015 Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
EN 60601-1-2:2015/ IEC 60601-1-2:2014 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic disturbances - Requirements and tests
EN ISO 81060-1:2012 Non-invasive sphygmomanometers - Part 1: Requirements and test methods for non-automated measurement type
EN 1060-3:1997+A2:2009 Non-invasive sphygmomanometers ­Part 3: Supplementary requirements for electro-mechanical blood pressure measuring systems
IEC 80601-2-30:2009+A1:2013 Medical electrical equipment­Part 2-30: Particular requirements for the basic safety and essential performance of automated non-invasive sphygmomanometers
EN 1060-4:2004 Non-invasive sphygmomanometers - Part 4: Test procedures to determine the overall system accuracy of automated non-invasive sphygmomanometers
ISO 81060-2:2013 Non-invasive sphygmomanometers - Part 2: Clinical validation of automated measurement type
EN 60601-1-6:2010+A1:2015/IEC 60601-1-6:2010+A1:2013 Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability
IEC 62366-1:2015 Medical devices - Part 1: Application of usability engineering to medical devices
EN 62304:2006/AC: 2008 / IEC 62304: 2006+A1:2015 Medical device software - Software life-cycle processes
ISO 10993-1:2009 Biological evaluation of medical devices­Part 1: Evaluation and testing within a risk management process
ISO 10993-5:2009 Biological evaluation of medical devices -
Part 5: Tests for in vitro cytotoxicity
ISO 10993-10:2010 Biological evaluation of medical devices -
Part 10: Tests for irritation and skin sensitization
EN
16
EN
12. EMC Guidance
1)is product needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information pro­vided, and this unit can be aected by portable and mobile RF communications equipment.
2)* Do not use a mobile phone or other devices that emit electromagnetic elds, near the unit. is may result in incorrect operation of the unit.
3)Caution: is unit has been thoroughly tested and inspected to assure proper performance and operation!
4)* Caution: is machine should not be used adjacent to or stacked with other equipment and that if adjacent or stacked use is necessary, this ma­chine should be observed to verify normal operation in the conguration in which it will be used.
Table 1
Guidance and manufacturer’s declaration – electromagnetic emissions
The device is intended for use in the electromagne tic environment specified below. The customer or the user of the device should assure that it is used in such an environment.
Emissions test Electromagnetic environment - guidance
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Compliance
Group 1
Class B
Class A
Complies
The device uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
The device is suitable for use in all establishments, other than domestic and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.
17
Table 2
Guidance and manufacturer’s declaration – electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below. The customer or the user of the device should assure that it is used in such an environment.
Immunity test
Electrostatic discharge )ESD( IEC 61000-4-2
Electrical fast transient/burst IEC 61000-4-4
Surge IEC61000-4-5
Voltage dips, short interruptions and voltage variations on power supply input lines
IEC 61000-4-11
IEC 60601 test level
±8 kV contact ±15 kV air
power supply lines: ±2 kV
input/output lines: ±1 kV
line)s( to line)s(: ±1 kV line)s( to earth: ±2 kV
100 kHz repetition frequency
0%UT; 0.5 cycle At 0°, 45°, 90°, 135°, 180°,225°,270° and 315° 0%U and 70%U Single phase: at 0° 0% U
T ; 1 cycle
T ; 25/30 cycles
T ; 300 cycle
Compliance level
±8 kV contact ±15 kV air
power supply lines: ±2 kV
line)s( to line)s(: ±1 kV
100 kHz repetition frequency
0% U
T ; 0.5 cycle
At 0°, 45°, 90°, 135°, 180°,225°,270° and 315° 0% U
T ; 1 cycle
and 70% U
T ; 25/30 cycles
Single phase: at 0° 0% U
T ;300 cycle
Electromagnetic environment - guidance
Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%.
Mains power quality should be that of a typical commercial or hospital environment.
Mains power quality should be that of a typical commercial or hospital environment.
Mains power quality should be that of a typical commercial or hospital environment.
EN
Power frequency
)50Hz/60Hz( magnetic field
IEC 61000-4-8
NOTE UT is the a.c. mains voltage prior to application of the test level.
30 A/m 50Hz/60Hz
30 A/m 50Hz/60Hz
Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment.
18
EN
Table 3
Guidance and manufacturer’s declaration – electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below. The customer or the user of the device should assure that it is used in such an environment.
Immunity test
Conducted RF IEC 61000-4-6
Radiated RF IEC 61000-4-3
IEC 60601 Test level
150 kHz to 80 MHz: 3 Vrms 6Vrms )in ISM and amateur radio bands( 80% Am at 1kHz
10V/m, 80% Am at 1kHz
Compliance level
150 kHz to 80 MHz: 3 Vrms 6Vrms )in ISM and amateur radio bands( 80% Am at 1kHz
10V/m, 80% Am at 1kHz
Electromagnetic environment - guidance
Portable and mobile RF communications equipment should be used no closer to any part of the device, including cables, than the recommended separation distance calculated from the equation appropriate for the frequency of the transmitter. Recommended separation distances: d=0.35 ;
P
d=1.2
P
P
P
where,
output power rating of the transmitter in watts )W( according to the transmitter manufacturer, d is the recommended separation distance in meters )m(. Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey, should be less than the compliance level in each frequency range. Interference may occur in the vicinity of equipment marked with the following symbol:
80 MHz to 800 MHz: d=1.2 800 MHz to 2.7 GHz: d=2.3
P is the maximum
a
b
NOTE 1 A t 80 MHz and 800 MHz, the higher frequency range applies. NOTE 2 T hese guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio )cellular / cordless( telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the device is used exceeds the applicable RF compliance level above, the device should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the device.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3V/m .
19
Table 4
Recommended separation distances between portable and mobile RF communications equipment and the device.
The device is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the device can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment )transmittters( and the device as recommended below, according to the maximum output power of the communications equipment.
EN
Rated maximum output
power of transmitter
)W(
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance
d in metres )m( can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts )W( according to the transmitter manufacturer. NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
Separation distance according to frequency of transmitter )m(
150 kHz to 80 MHz 8 0 MHz to 800 MHz 800 MHz to 2.7 GHz
=d
3.5
P
0.12 0.12
0.38
1.2
3.8
12
=d 1.2
0.38
1.2
3.8
12
P
0.23
0.73
2.3
7.3
23
2.3=d
P
20
EN
Table 5
Guidance and manufacturer’s declaration - electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below. The customer or the user of the device, should assure that it is used in such an environment.
Radiated RF IEC61000-4-3 )Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless communica­tions equipment(
NOTE If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the transmitting antenna and the ME EQUIPMENT or ME SYSTEM may be reduced to 1 m. The 1 m test distance is permitted by IEC 61000-4-3.
a( For some services, only the uplink frequencies are included. b( The carrier shall be modulated using a 50% duty cycle square wave signal. c( As an alternative to FM modulation, 50% pulse modulation at 18 Hz may be used because while it does not represent actual modulation, it would be worst case.
The MANUFACTURER should consider reducing the minimum separation distance, based on RISK MANAGEMENT, and using higher IMMUNITY TEST LEVELS that are appropriate for the reduced minimum separation distance. Minimum separation distances for higher IMMUNITY TEST LEVELS shall be calculated using the following equation: E=
Where P is the maximum power in W, d is the minimum separation distance in m, and E is the IMMUNITY TEST LEVEL in V/m.
Test
Band a(
Service a( Modulation b( Modulation b(
)MHz(
Frequency )MHz(
385
450 430-470
710 704-787
745
780
810
870
930
1720
1845
1970
5240
5500
5785
380-390 TETRA
800-960
1700­1990
2570
5100­5800
400
GMRS 460 FRS 460
LTE Band 13, 17
GSM 800/900, TETRA 800, iDEN 820, CDMA 850, LTE Band 5
GSM 1800; CDMA 1900; GSM 1900; DECT; LTE Band 1, 3, 4,25; UMTS
Bluetooth, WLAN,
802.11 b/g/n, RFID 2450, LTE Band 7
WLAN
802.11 a/n
Pulse modulation b( 18Hz
FM c( ± 5kHz deviation 1kHz sine
Pulse modulation b( 217Hz
Pulse modulation b( 18Hz
Pulse modulation b( 217Hz
Pulse modulation b( 217 Hz
Pulse modulation b( 217 Hz
Distance )m(
)W(
1.8
20 .3 28
0.20 .3 9
2
2
2
0.2
21
IMMUNITY TEST LEVEL )V/m(
0.32 7
0.32 8
0.32 8
0.32 82450 2400-
0.39
Manufacturer: Guangdong Transtek Medical Electronics Co., Ltd. Zone A, No.105 ,Dongli Road, Torch Deve­lopment District, Zhongshan,528437 Guang­dong,China
MDSS - Medical Device Safety Service GmbH
EC REP
Schigraben 41, 30175 Hannover- Germany
Distributed by:
Medel International Srl
Via Villapizzone 26 - 20156 Milano / Italy
Loading...