TRANSTEK LS808-BS User Manual

User Manual
Blood Pressure Monitor
LS808-BS
Please do read the user manual carefully and thoroughly so as to ensure the safe usage of this product, and keep the manual well for your further reference in case you have problems.
Thank you very much for selecting TRANSTEK Blood Pressure Monitor LS808-BS.
Version:1.0
Guangdong Transtek Medical Electronics Co., Ltd.
Zone B, No.105 ,Dongli Road, Torch Development District, Zhongshan,528437,Guangdong,China
FCC ID: OU9LS808-B-S
INTRODUCTION ................................................................
2
Indications for Use
LCD Display Signal Monitor Components List
BEFORE YOU START ........................................................... 9
MEASUREMENT ...............................................................
16
INFORMATION FOR USER...................................................... 22
Tips for Measurement Maintenance
ABOUT BLOOD PRESSURE..................................................... 24
What are systolic pressure and diastolic pressure? What is the standard blood pressure classification?
Irregular heartbeat detector
Why does my blood pressure fluctuate throughout the day?
TROUBLESHOOTING .......................................................... 26
SPECIFICATIONS .............................................................. 27
ATHORIZED COMPONENT ................................................ 28
CONTACT INFORMATION....................................................... 28
Safety Information
Power Supply and Charge Power
Setting the Time,Date and Unit
Start Measurement
DATA MANAGEMENT .....................................
18
Recall the Records Delete the Records
......................
COMPLIED EUROPEAN STANDARDS LIST ........................ 29
....................................................... 31......
Table of Contents
................
Measurement Principle
General Description
EMC GUDIANCE
Tie the Cuff
Pair up with Your Device
Why do I get a different blood pressure at home compared to the hospital? Is the result the same if measuring on the right arm?
Contraindications
FCC STATEMENT
.............................................................. 30
1
TABLE OF CONTENTS
FEATURES:
86.1mm×24mm Blue LCD Display with White Backlight Measure-during-inflating Technology
Up to 60 pieces of record stored per each user
This product uses the Oscillometric Measuring Method to detect blood pressure. Before every measurement, the unit establishes a “zero point” equivalent to the atmospheric pressure. Then it starts inflating the cuff. Meanwhile, the unit detects pressure oscillation generated by beat-to-beat pulsatile, which is used to deter­mine the systolic pressure and diastolic pressure as well as pulse rate.
Measurement Principle
General Description
The signs below might be in the user manual, labeling or other components. They are the requirement of standard and using.
Safety Information
T1A/250V Φ3.6*10CCC
Indications for Use
* Thank you for selecting TRANSTEK blood pressure monitor (LS808-BS).
The monitor features blood pressure measurement, pulse rate measurement and the result storage. The design provides you with two years of lifetime.
* This manual contains important safety information and caution, and provides step
by step instructions for using the product.
* Please do read this user manual carefully and thoroughly before use.
The Transtek Blood Pressure Monitor is digital monitors intended for use in measuring blood pressure and heartbeat rate with arm circumference ranging
from 22 cm to 32 cm ( about 8¾˝-12½˝ ) or 22cm to 42cm(about 8¾˝-16½˝).
It is intended for adult indoor use only.
Symbol for “ENVIRONMENT PROTECTION - Electrical waste products should not be disposed of with household waste. Please recycle where facilities exist. Check with your local authority or retailer for recycling advice”
SN
Symbol for “TYPE BF APPLIED PARTS”
Symbol for “THE OPERATION GUIDE MUST BE READ”
Symbol for “MANUFACTURER”
Symbol for “SERIAL NUMBER”
Symbol for “DIRECT CURRENT”
Symbol for “MANUFACTURE DATE”
Caution: These notes must be observed to prevent any damage to the device.
Symbol for “Recycle”
The Green Dot is the license symbol of a European network of industry-funded systems for recycling the packaging materials of consumer goods.
Symbol for “Class II Equipment”
For indoor use only
F1
1.The device is not suitable for use on may be pregnant women or pregnant women.
2.The device is not suitable for use on patients with implanted,electrocical devices, such as cardiac pacemakers, defibrillators.
Contraindications
INTRODUCTION INTRODUCTION
2 3
CAUTION
* This device is intended for adult use in homes only. * The device is not suitable for use on neonatal patients, pregnant women,patients with
implanted, electronical devices, patients with pre-eclampsia, premature ventricular beats, atrial fibrillation, peripheral, arterial disease and patients undergoing intravascular therapy or arterio-venous shunt or people who received a mastectomy. Please consult your doctor prior to using the unit if you suffer from illnesses.
* The device is not suitable for measuring the blood pressure of children. Ask your doctor
before using it on older children.
* The device is not intended for patient transport outside a healthcare facility. * The device is not intended for public use. * This device is intended for no-invasive measuring and monitoring of arterial blood pressure.
It is not intended for use on extremities other than the arm or for functions other than obtaining a blood pressure measurement.
* Do not confuse self-monitoring with self-diagnosis. This unit allows you to monitor your blood
pressure.Do not begin or end medical treatment without asking a physician for treatment advice.
* If you are taking medication,consult your physician to determine the most appropriate time to
measure your blood pressure. Never change a prescribed medication without consulting your physician.
* Do not take any therapeutic measures on the basis of a self measurement. Never alter the
dose of a medicine prescribed by a doctor. Consult your doctor if you have any question about your blood pressure.
* When the device was used to measure patients who have common arrhythmias such as
atrial or ventricular premature beats or atrial fibrillation, the best result may occur with deviation. Please consult your physician about the result.
* Don't kink the connection tube during use, otherwise, the cuff pressure may continuously
increase which can prevent blood flow and result in harmful injury to the PATIENT.
* When using this device, please pay attention to the following situation which may interrupt
blood flow and influence blood circulation of the patient, thus cause harmful injury to the patient: connection tubing kinking too frequent and consecutive multiple measurements; the application of the cuff and its pressurization on any arm where intravascular access or
therapy, or an arterio-venous (A-V) shunt, is present; inflating the cuff on the side of a
mastectomy.
* Warning: Do not apply the cuff over a wound;otherwise it can cause further injury. *Do not inflate the cuff on the same limb which other monitoring ME equipment is applied
around simultaneously, because this could cause temporary loss of function of those simultaneously-used monitoring ME equipment.
*On the rare occasion of a fault causing the cuff to remain fully inflated during measurement, open the cuff immediately. Prolonged high pressure (cuff pressure 300mmHg or constant pressure > 15mmHg for more than 3 minutes) applied to the arm may lead to an ecchymosis. *Please check that operation of the device does not result in prolonged impairment of patient
blood circulation.
* When measurement, please avoid compression or restriction of the connection tubing. * The device cannot be used with HF surgical equipment at the same time.
CAUTION
* The ACCOMPANYING DOCUMENT shall disclose that the SPHYGMOMANOMETER was
clinically investigated according to the requirements of ISO 81060-2:2013.
* To verify the calibration of the AUTOMATED SPHYGMOMANOMETER, please contact the
manufacturer.
* This device is contraindicated for any female who may be suspected of, or is pregnant.
Besides providing inaccurate readings, the effects of this device on the fetus are unknown.
* Too frequent and consecutive measurements could cause disturbances in blood circulation
and injuries.
* This unit is not suitable for continuous monitoring during medical emergencies or
operations.Otherwise, the patient’s arm and fingers will become anaesthetic, swollen and even purple due to a lack of blood.
* When not in use, store the device with the adapter in a dry room and protect it against
extreme moisture, heat, lint, dust and direct sunlight. Never place any heavy objects on the storage case.
* This device may be used only for the purpose described in this booklet. The manufacturer
cannot be held liable for damage caused by incorrect application.
*This device comprises sensitive components and must be treated with caution. Observe the
storage and operating conditions described in this booklet. * The maximum temperature that the applied part can be achieved is 42.5℃ while the environmental temperature is 40.
* The equipment is not AP/APG equipment and not suitable for use in the presence of a
flammable anesthetic mixture with air of with oxygen or nitrous oxide.
* Warning: No servicing/maintenance while the ME equipment is in use. * The patient is an intended operator. * The patient can measure,transmit data and charge power under normal circumstances and
maintain the device and its accessories according to the user manual.
* To avoid measurement errors, please avoid the condition of strong electromagnetic field
radiated interference signal or electrical fast transient/burst signal.
* The blood pressure monitor, its adaptor, and the cuff are suitable for use within the patient environment. If you are allergic to polyester, nylon or plastic, please don't use this device. * During use, the patient will be in contact with the cuff. The materials of the cuff have been
tested and found to comply with requirements of ISO 10993-5:2009 and ISO 10993-10:2010. It will not cause any potential sensization or irritation reaction.
* Adaptor is specified as a part of ME EQUIPMENT. * If you experience discomfort during a measurement, such as pain in the arm or other
complaints, press any button to release the air immediately from the cuff. Loosen the cuff and remove it from your arm.
* If the cuff pressure reaches 40 kPa (300 mmHg), the unit will automatically deflate. Should the cuff not deflate when pressures reaches 40 kPa (300 mmHg), detach the cuff from the
arm and press any button to stop inflation.
* Before use, make sure the device functions safely and is in proper working condition.
Check the device, do not use the device if it is damaged in any way. The continuous use of a damaged unit may cause injury, improper results, or serious danger.
* Do not wash the cuff in a washing machine or dishwasher!
INTRODUCTION
INTRODUCTION
4 5
CAUTION
LCD Display Signal
Systolic Blood Pressure
High blood pressure
Diastolic Blood Pressure
Low blood pressure
Unit Measurement unit of blood pressure
Memory Query
Low Battery
Low battery and please charge the power.
Current time
Data pending to transmit
Year/Month/Day(Hour:Minute)
Measurement data stored in the device
Data transmitting Data transmission succeeds.
User ID
Start measurement for selected user, and transmit the measuring result
Blood pressure monitor is detecting an irregular heartbeat during measurement.
Irregular heartbeat
Heartbeat
Blood pressure monitor is detecting a heartbeat during measurement.
Motion indicator
Motion may result in an inaccurate measurement.
Pulse in beats per minute
Pulse display
Indicate it is in the memory mode and which group of memory it is.
* The service life of the cuff may vary by the frequency of washing, skin condition, and
storage state. The typical service life is 10000 times.
* It is recommended that the performance should be checked every 2 years and after
maintenance and repair, by retesting at least the requirements in limits of the error of the cuff
pressure indication and air leakage (testing at least at 50mmHg and 200mmHg). * Please dispose of ACCESSORIES, detachable parts, and the ME EQUIPMENT according
to the local guidelines.
* Manufacturer will make available on request circuit diagrams, component part lists,
descriptions, calibration instructions,etc., to assist to service personnel in parts repair.
* The plug/adapter plug pins insulates the device from the main supply. Do not position the
device in a position where it is difficult to disconnect from the supply mains to safely terminate operation of ME equipment.
* The operator shall not touch output of batteries /adapter and the patient simultaneously. * Cleaning :Dust environment may affect the performance of the unit. Please use the soft
cloth to clean the whole unit before and after use. Don’t use any abrasive or volatile cleaners.
* The device doesn’t need to be calibrated within two years of reliable service. * If you have any problems with this device, such as setting up, maintaining or using, please
contact the SERVICE PERSONNEL of Transtek. Don’t open or repair the device by yourself in the event of malfunctions. The device must only be serviced, repaired and opened by individuals at authorized sales/service centers.
* Please report to Transtek if any unexpected operation or events occur. * Keep the unit out of reach of infants, young children or pets to avoid inhalation or
swallowing of small parts. It is dangerous or even fatal.
* Be careful to strangulation due to cables and hoses, particularly due to excessive length. * At least 30 min required for ME equipment to warm from the minimum storage temperature
between uses until it is ready for intended use. At least 30 min required for ME equipment to cool from the maximum storage temperature between uses until it is ready for intended use. * This equipment needs to be installed and put into service in accordance with the information provided in the ACCOMPANYING DOCUMENTS;
* Wireless communications equipment such as wireless home network devices, mobile
phones, cordless telephones and their base stations, walkie-talkies can affect this equipment and should be kept at least a distance d away from the equipment. The distance d is calculated by the MANUFACTURER from the 80MHz to 5.8 GHz column of Table 4 and Table 9 of IEC 60601-1-2:2014, as appropriate.
* Please use ACCESSORIES and detachable partes specified/ authorised by
MANUFACTURE. Otherwise, it may cause damage to the unit or danger to the user/patients.
* There is no luer lock connectors are used in the construction of tubing, there is a possibility
that they might be inadvertently connected to intravascular fluid systems, allowing air to be pumped into a blood vessel.
* Please use the device under the environment which was provided in the user manual.
Otherwise, the performance and lifetime of the device will be impacted and reduced.
INTRODUCTION
INTRODUCTION
6 7
Monitor Components
1. The battery of LS808-BS is built-in rechargeable li-polymer battery, the battery current is 1000 mAh.
2. Please use the AC adaptor to charge the battery, just like the following picture:
Power Supply and Charge Power
Charging the power under following circumstances:
displays on the LCD
The LCD display is dim.
When powering on the monitor, the LCD doesn’t light up.
AC adaptor
CAUTION
* The battery of LS808-BS is built-in rechargeable lithium-ion battery,
please do not disassemble it by the unauthorized maintenance personel.
* Under the normal using, it can charge power about 300 times, if the
battery cannot charge the power normally or the blood pressure monitor cannot use normally, please connect with the authorized maintenance personel.If measured three times per day, and the battery is fully charged, it can be used for about 20 days. * Storge and use the blood pressure monitor at the cool, dry and ventilated environment. Avoid to approach to the fire and the heat source, or it will cause the battery explode.
* Only can use the Transtek’s authorized AC Adaptor (Model: BLJ06L060100P-U) to charge the power. You cannot use the blood
pressure monitor during the process of charging.
* During the process of charging, the blood pressure monitor display
When the charging is finished, please pull the plug in time.
* When charging, shall not touch charging connector and the patient
simultaneously.
List
1.Blood Pressure Monitor
(LS808-BS)
2.AC Adaptor
( Model: BLJ06L060100P-U)
3.User Manual
4. Cuff (22cm-32cm or 22cm-42cm) (Type BF Applied Part)
Component list of pressure measuring system 1 PCBA 2 Air pipe 3 Pump 4 Valve
5 Cuff
SYSTOLIC PRESSURE
DIASTOLIC PRESSURE
QUERY/UP
BUTTON
USER 2/ACTIVATION BUTTON
USER 1/SET BUTTON
CUFF
(Type BF applied part)
PULSE RATE
ON/OFF BUTTON
DC POWER SOCKET
(Please use TRANSTEK
authorized cuff. The size of the actual cuff please refer to the label on the
attached cuff.)
BEFORE YOU START
8 9
INTRODUCTION
Setting the Time, Date and Unit
(1) When the monitor is off, press and hold User 1 button for 3s
to enter Time Setting Mode.
To ensure the stored measurement result has correct time record, please set time and unit before device is used.
(2) As pictured in the right, the blinking numeral representing [HOUR]. Press “Query” button to change the numeral. Each press will increase
the numeral by one in a cycling manner.
(3) Press “User 1” button again to confirm [HOUR]. Then the numeral representing [MINUTE] blinks.
Before use,switch the button to the “ON” side to turn on the monitor. Note: If the button is on the “OFF” side, there is no reaction when you press any button.
CAUTION
• Do not attempt to replace your blood pressure monitor’s battery. It is built-in
and not changeable.
• Only charge the battery in accordance with the user instructions supplied
with the blood pressure monitor.
• Avoid charging your blood pressure monitor in extremely high or low
temperatures.
• Do not use your blood pressure monitor while you are charging it.
• Do not attempt to disassemble the blood pressure monitor or force open the
built-in battery.
• Do not clean the blood pressure monitor when it is being charged. Always
unplug the charger first before cleaning the blood pressure monitor.
• Do not dispose of your blood pressure monitor in a fire. The battery could
explode causing injury or death.
• Batteries (battery pack or batteries installed) shall not be exposed to
excessive heat such as sunshine, fire or the like.
BEFORE YOU START
10 11
BEFORE YOU START
(4) Repeat steps 2 and 3 to confirm [MINUTE].
(5) Repeat steps 2 and 3 to confirm[MONTH], [DAY] and [YEAR].
(6) Repeat steps 2 and 3 to confirm the measurement unit.
(7) After confirming the measurement unit, the LCD will
display “dOnE” and the monitor will shut off.
12 13
BEFORE YOU START
BEFORE YOU START
Tie the Cuff
1.
2.
3.
4.
5.
Remove all accessories (watch, bracelet,etc) from your arm. If your physician has
diagnosed you with poor circulation in your arm, use the other one.
Roll or push up your sleeve to expose the skin. Apply the cuff to your arm with your palm facing up.
Position the edge of the cuff about 2cm~3cm from elbow. Fasten the cuff around your arm, leaving no extra room between the cuff and your
skin. If the cuff is too loose, the measurement will not be accurate.
Sit comfortably with your tested arm resting on a flat surface. Place your elbow on a table so that the cuff is at the same level as your heart. Turn your palm upwards. Sit upright in a chair, and take 5-6 deep breaths.
Rest for 5 minutes before first measuring. Wait at least 3 minutes between measurements. This allows your blood circulation to recover.
The patient must relax as much as possible and do not move and talk during the measurement procedure.
For a meaningful comparison, try to measure under similar conditions. For example, take daily measurements at approximately the same time, on the same arm, or as directed by a physician.
6.
7.
Take the measurement in a silent room.
The cuff should maintain at the same level as the right atrium of the heart.
Please sit comfortably. Do not cross your legs and keep your feet flat on the ground. Keep your back against the backrest of the chair.
Helpful tips for Patients, especially for Patients with Hypertension:
Pair up the Blood Pressure Monitor with Your Device
(2)Press “My devices”,the app will display My Devices tap.
Bluetooth Module No. : LS51802
Frequency Range
Output Power Range
Supply Voltage2.402 - 2.480 GHz
0 dBm
1.8-3.6 V
Transmitting Distance
10 meters
(5)After selecting the device,select the “account” and return;then enter device to “Set device” .When the app display“Receive data automatically”please choose “Yes”.
(1) Turn on Bluetooth and the app. Enter to “MedM Health” then
press “My Setting” ,the app will display My Setting tap.
(4) Press in “MedM Health to reseach,then select “808A0” to connect the monitor.
(3)When the moniter is off,press press“User 1” to measure,
the will show and will blink after measuring,then the datas are transmitting.
(6)When the monitor is connected,the datas will transmit to “MedM”,
if transmit successfully,the symbel and will disappear,then the monitor turn off;or not,it will turn off within one minute.
List of compatible devices: For iOS devices: The operating system must be iOS 8 or more, such as iPhone 4S, iPhone 5/5C/5S, iPhone 6/6 Plus and so on. For Android devices: The operating system must be 4.3 or more.
14 15
BEFORE YOU START
BEFORE YOU START
Start Measurement
(1) When the monitor is off, press the User 1 button
to turn on the monitor.
LCD display Adjust to zero.
Inflating and measuring.
Display and save the results. The data transmission will proceed.
(2) Press User 1 button to power off,
otherwise it will turn off within one minute.
Tips: A. When finish the whole measurement, press another user button, the blood monitor will begin measuring again. B. Maximum 60 records are both for user 1 and user 2.
1.With LS808-BS successfully pair-up with your mobile device, the measurement data will be automatically transmitted to your mobile device via Bluetooth.
2.The symbol will disappear after successful data transmission, and you may check your personal health data stored in your mobile device.
3.If the data transmission fails, the symbol will remain. The pending measurement data will be transmitted to your mobile device when next measurement is complete.
CAUTION
When the monitor is off, press User 1 button to turn on the monitor and it will finish the whole measurement, and then save the measure data for User 1.
The same to the User 2. (Take User 1 for example.)
(Note: Select the same user on your app and BPM to take
the measurement, or the measurement data won’t be
transmitted to the app.)
16 17
MEASUREMENT
MEASUREMENT
Recall the Records
(1) When the monitor is off, press “Query” button
to access the memory.
(2) The LCD will display the latest measuring
result of the user ID which completes the last measurement.
The record number, measuring date and measuring time will be displayed alternatively.
(3) Press “Query” button to rotate the history records.
(4) When in the memory mode, press the User 1 button
to recall the measurement history of User 1, or press the User 2 button to recall the measurement history.
(5) When no history stored for the specific user in the monitor, press “Query” button and the LCD will display
as pictured to the right.
The most recent record (1) is shown first. Each new measurement is assigned to the first (1) record. All other records are pushed back one digit (e.g., 2 becomes 3, and so on), and the last record (60) is dropped from the list.
CAUTION
18 19
DATA MANAGEMENT
DATA MANAGEMENT
Delete the Records
(1) When under the query mode, press and hold “Query” button for 3 seconds to clear the memory.
(2) When the LCD display “dEL ALL”, press “Query”
button to confirm.
(3) The LCD will display “dEL dONE” and then shut off.
(4) If you wish to stop clearing the memory, you may press the other button, rather than “Query” button to
turn off the monitor, or wait until the monitor shuts off.
CAUTION
.
• Interference may occur in the vicinity of equipment marked with the following
symbol .And LS808-BS may interfering vicinity electrical equipment.
• Sensitive people, including pregnant women pre-eclamptic and those who implanted medical electronic instruments, should avoid using the unit whenever possible.
• Keep the monitor at least 20 centimeters away from the human body (especially the head) when the data transmission is proceeding after measurement.
• To enable the data transmission function, this product should be paired to
Bluetooth end at 2.4 GHz.
How to mitigate possible interference?
1. The range between the device and BT end should be reasonably close, from 1 meter to 10 meters. Please ensure no obstacles between the device and BT end so as to obtain quality connection and to lower the RF output range.
2. To avoid interference, other electronic devices (particularly those with wireless transmission / Transmitter) should be kept at least 1 meter away from the monitor.
2120
DATA MANAGEMENT
DATA MANAGEMENT
Tips for Measurement
Within 1 hour
after dinner or drinking
Within 20 minutes after taking a bath
In a very cold environment
Immediate measurement after tea, coffee, smoking
When talking or moving your fingers
When you want to discharge urine
Maintenance
To obtain the best performance, please follow instructions below.
Measurements may be inaccurate if taken in the following circumstances.
Put in a dry place and
avoid the sunshine
Avoid intense shaking and collisions.
Use the slightly damp cloth
to remove the dirt.
Avoid immersing it in the water. Clean it with a dry cloth in case.
Avoid dusty environment and
unstable temperature surrounding
Avoid washing the cuff
22 23
INFORMATION FOR USER
INFORMATION FOR USER
What are systolic pressure and diastolic pressure?
press
artery
vein
blood discharging
Systolic
relax
blood entering
Diastolic
When ventricles contract and pump blood out of the heart, the blood pressure reaches its maximum value in the cycle, which is called systolic pressure. When the ventricles relax, the blood pressure reaches its minimum value in the cycle, which is called diastolic pressure.
Irregular Heartbeat Detector
CAUTION
Why does my blood pressure fluctuate throughout the day?
1. Individual blood pressure varies multiple
times everyday. It is also affected by the way
you tie your cuff and your measurement
position, so please take the measurement
under the same conditions.
2.If the person takes medicine, the
pressure will vary more.
3.Wait at least 3 minutes for another
measurement.
Why do I get a different blood pressure at home compared to the hospital?
The blood pressure is different even
throughout the day due to weather,
emotion, exercise etc, Also, there is
the “white coat” effect, which means
blood pressure usually increases in
clinical settings.
What you need to pay attention to when you measure your blood pressure at home:
If the cuff is tied properly. If the cuff is too tight or too loose. If the cuff is tied on the upper arm. If you feel anxious. Taking 2-3 deep breaths before beginning will be better for measuring. Advice: Relax yourself for 4-5 minutes until you calm down.
Is the result the same if measuring on the right arm?
It is ok for both arms, but there will be some different results for different people. We suggest you measure the same arm every time.
An irregular heartbeat is detected when a heartbeat rhythm varies while the unit is measuring the systolic and diastolic blood pressure.During each measurement, the monitor records all the pulse intervals and calculate the average ; if there are two or more pulse intervals ,the difference between each interval and the average is more than the average value of ±25% , or there are four or more pulse intervals ,the difference between each interval and the average is more than the average value of ±15%,the irregular heartbeat symbol appears on the display when the measurement results are appear.
The appearance of the IHB icon indicates that a pulse irregularity consistent with an irregular heartbeat was detected during measurement. Usually this is NOT a cause for concern. However, if the symbol appears often, we recommend you seek medical advice. Please note that the device does not replace a cardiac examination, but serves to detect pulse irregularities at an early stage.
What is the standard blood pressure classification?
The chart on the right is the standard blood pressure classifi­cation published by American
Heart Association (AHA).
AHA Home Guideline for Upper Limit of Normal BP
Please consult a physician if your measuring result falls outside the range. Please note that only a physician can tell whether your blood pressure value has reached a dangerous point.
CAUTION
SYS 135 mm Hg
DIA 85 mm Hg
Blood Pressure Category
Normal
Prehypertension
High Blood Pressure
(Hypertension) Stage 1
High Blood Pressure
(Hypertension) Stage 2
Hypertensive Crisis
(Emergency care needed)
Systolic
mmHg (upper#)
Diastolic
mmHg (lower#)
less than 120
120-139
140-159
160 or higher
Higher than 180
and
or
or
or
or
less than 80
80-89
90-99
100 or higher
Higher than 110
This chart reflects blood pressure categories defined by American Heart Association.
ABOUT BLOOD PRESSURE
ABOUT BLOOD PRESSURE
24 25
This section includes a list of error messages and frequently asked questions for problems you may encounter with your blood pressure monitor. If the products not operating as you think it should, check here before arranging for servicing.
3.7V 1000mAH Built-in rechargeable li-polymer battery, 6V 1A AC Adaptor
Blue LCD with White Backlight
V.A.= 86.1mm(L) x24mm(W)
Approx.285g
Approx.130.9mm×73mm×29.4mm AC Adaptor,user manual
A01
Power supply
Display mode
Measurement mode
Oscillographic testing mode
Measurement range
Accuracy
Normal working condition
Storage & transportation
condition
Measurement perimeter
of the arm
Weight
External dimensions
Attachment
Mode of operation Continuous operation
Degree of protection Type BF applied part
Protection against
ingress of water
Software version
Device classification
Battery Powered Mode: Internally Powered ME Equipment AC Adaptor charged Mode: Class II ME Equipment
WARNING: No modification of this equipment is allowed.
Pressure:
5°C-40°C within±3mmHg(0.4kPa)
Pulse value:±5%
Rated cuff pressure:
0mmHg~299mmHg(0kPa ~ 39.9kPa)
Measurement pressure:
SYS: 60mmHg~230mmHg (8.0kPa~30.7kPa) DIA: 40mmHg~130mmHg (5.3kPa~17.3kPa) Pulse value: (40-199)beat/minute
IP22: The first number 2: Protected against solid
foreign objects of 12,5mm Ф and greater. The second
number: Protected against vertically falling water drops when enclosure titled up to 15º. Vertically falling drops shall have no harmful effects when the enclosure is titled at any angle up to 15º on either side of the vertical.
About 22cm-32cm or 22cm-42cm
A temperature range of :+5°C to +40°C A relative humidity range of 15% to 90%, non-condensing, but not requiring a water vapour partial pressure greater than 50 hPa An atmospheric pressure range of : 700 hPa to 1060 hPa
Temperature:-20°C to +60°C
A relative humidity range of ≤ 93%, non-condensing,
at a water vapour pressure up to 50hPa
PROBLEM SYMPTOM CHECK THIS REMEDY
No power
Low
batteries
Error
message
Display will not
light up.
E1 shows
The cuff is not secure.
E 3 shows
Relax for a moment and then measure again.
E 10 or E 11 shows
The monitor detected motion,talking or the pulse is too poor while measuring.
E 20 shows
E 21 shows Relax for a moment and
then measure again.
EExx,shows on
the display.
A calibration error occurred.
Retake the measurement. If the problem persists, contact the retailer or our customer service department for further assistance.Refer to the warranty for contact information and return instructions.
The measurement process does not detect the pulse signal.
Loosen the clothing on the arm and then measure again.
The treatment of the measurement failed.
Power is exhausted. Charge the power
Power is low. Charge the power
Display is dim or shows
AC adaptor is inserted incorrectly.
Insert the AC adaptor tightly
Refasten the cuff and then measure again.
Warning message
“out ” shows
Out of measurement range
the measurement result is out of
the measurement range (SYS:
60mmHg to 230mmHg; or DIA: 40mmHg to 130mmHg;or
Pulse: 40-199 pulse/minute)
Relax for a moment and then measure again.
The pressure of the cuff is excess.
TROUBLESHOOTING
SPECIFICATIONS
26 27
For more information about our products, please visit www.transtek.cn.you can get customer service, usual problems and customer download, transtek will serve you anytime.
Contact Information
Athorized Component
Please use the TRANSTEK authorized adaptor
Manufactured by:
Company:
Address:
Guangdong Transtek Medical Electronics Co., Ltd.
Zone B, No.105 ,Dongli Road, Torch Development District, Zhongshan,528437,Guangdong,China
Guangdong Transtek Medical Electronics Co., Ltd.
BLJ06L060100P-U
100-240V
50-60Hz,0.2Amax
6V 1000mA
Adaptor
Type:
Input: Output:
Complied Standards List
EN ISO 14971:2012 / ISO 14971:2007 Medical devices -
Application of risk management to medical devices
EN 1041:2008 Information supplied by the manufacturer of medical
devices
EN 60601-1:2006+A1:2013/ IEC 60601-1:2005+A1:2012 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
EN 60601-1-11:2015/ IEC 60601-1-11:2015 Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
EN 60601-1-2:2015/ IEC 60601-1-2:2014 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic disturbances - Requirements and tests
EN ISO 81060-1:2012 Non-invasive sphygmomanometers - Part 1: Requirements and test methods for non-automated measurement type
EN 1060-3:1997+A2:2009 Non-invasive sphygmomanometers ­Part 3: Supplementary requirements for electro-mechanical blood pressure measuring systems
EN 1060-4:2004 Non-invasive sphygmomanometers - Part 4: Test procedures to determine the overall system accuracy of automated non-invasive sphygmomanometers
EN 60601-1-6:2010+A1:2015/IEC 60601-1-6:2010+A1:2013 Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability
IEC 62366-1:2015 Medical devices - Part 1: Application of
usability engineering to medical devices
EN 62304:2006/AC: 2008 / IEC 62304: 2006+A1:2015 Medical
device software - Software life-cycle processes
Risk management
Labeling
User manual
General Requirements for Safety
Electromagnetic compatibility
Performance requirements
Clinical investigation
Usability
Software life-cycle processes
Bio-compatibility
ISO 10993-1:2009 Biological evaluation of medical devices- Part
1: Evaluation and testing within a risk management process
ISO 10993-5:2009 Biological evaluation of medical devices -
Part 5: Tests for in vitro cytotoxicity
ISO 10993-10:2010 Biological evaluation of medical devices -
Part 10: Tests for irritation and skin sensitization
EN ISO 15223-1:2016 / ISO 15223-1:2016 Medical devices. Symbols to be used with medical device labels, labelling and information to be supplied. Part 1 : General requirements
IEC 80601-2-30:2009+A1:2013 Medical electrical equipment- Part 2-30: Particular requirements for the basic safety and essential performance of automated non-invasive sphygmomanometers
ISO 81060-2:2013 Non-invasive sphygmomanometers - Part 2: Clinical validation of automated measurement type
COMPLIED STANDARDS LIST
28 29
ATHORIZED COMPONENT
EMC Guidance
1)This product needs special precautions regarding EMC and needs to be
installed and put into service according to the EMC information provided, and this unit can be affected by portable and mobile RF communications equipment.
2)* Do not use a mobile phone or other devices that emit electromagnetic
fields, near the unit. This may result in incorrect operation of the unit.
3)Caution: This unit has been thoroughly tested and inspected to assure proper performance and operation!
4)* Caution: This machine should not be used adjacent to or stacked with
other equipment and that if adjacent or stacked use is necessary, this machine should be observed to verify normal operation in the configuration in which it will be used.
Table 1
Guidance and manufacturer’s declaration – electromagnetic emissions
RF emissions
CISPR 11
Group 1
Class B
Class A
Complies
Compliance
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
RF emissions
CISPR 11
Emissions test Electromagnetic environment - guidance
The device is intended for use in the electromagnetic environment specified below. The customer or the user of the device should assure that it is used in such an environment.
The device uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
The device is suitable for use in all establishments, other than domestic and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.
This device complies with Part 15 of the FCC Rules. Operation is subject
to the following two conditions: (1) this device may not cause harmful interference, and (2) this device must accept any interference received,
including interference that may cause undesired operation. Caution: The user is cautioned that changes or modifications not expressly approved by the party responsible for compliance could void
the user's authority to operate the equipment.
NOTE: This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:
-- Reorient or relocate the receiving antenna.
-- Increase the separation between the equipment and receiver.
-- Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
-- Consult the dealer or an experienced radio/TV technician for help. FCC Radiation Exposure Statement: This equipment complies with FCC radiation exposure limits set forth for an uncontrolled environment. This transmitter must not be co-located or operating in conjunction with any other antenna or transmitter.
FCC Statement
FCC ID: OU9LS808-B-S
FCC STATEMENT
30 31
EMC GUIDANCE
Guidance and manufacturer’s declaration – electromagnetic immunity
Immunity test
±8 kV contact ±15 kV air
±8 kV contact ±15 kV air
±2 kV
power supply lines:
line(s) to line(s): ±1 kV line(s) to earth: ±2 kV
0%UT; 0.5 cycle At 0°, 45°, 90°, 135°, 180°,225°,270° and 315° 0%U
T
; 1 cycle and 70%U
T
; 25/30 cycles Single phase: at 0° 0% U
T
; 300 cycle
30 A/m 50Hz/60Hz
NOTE UT is the a.c. mains voltage prior to application of the test level.
Table 2
Compliance level
Electrostatic discharge (ESD) IEC 61000-4-2
Electromagnetic environment - guidance
30 A/m 50Hz/60Hz
Power frequency
(50Hz/60Hz) magnetic field
IEC 61000-4-8
Voltage dips, short interruptions and voltage variations on power supply input lines
IEC 61000-4-11
Electrical fast transient/burst IEC 61000-4-4
Surge IEC61000-4-5
input/output lines: ±1 kV
The device is intended for use in the electromagnetic environment specified below. The customer or the user of the device should assure that it is used in such an environment.
IEC 60601 test level
Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%.
±2 kV
power supply lines:
100 kHz repetition frequency
line(s) to line(s): ±1 kV
100 kHz repetition frequency
Mains power quality should be that of a typical commercial or hospital environment.
Mains power quality should be that of a typical commercial or hospital environment.
Mains power quality should be that of a typical commercial or hospital environment.
Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment.
0% U
T
; 0.5 cycle At 0°, 45°, 90°, 135°, 180°,225°,270° and 315° 0% U
T
; 1 cycle and 70% U
T
; 25/30 cycles Single phase: at 0° 0% U
T
;300 cycle
Table 3
Guidance and manufacturer’s declaration – electromagnetic immunity
Immunity test
Compliance level
IEC 60601 Test level
Conducted RF IEC 61000-4-6
Radiated RF IEC 61000-4-3
10V/m, 80% Am at 1kHz
150 kHz to 80 MHz: 3 Vrms
6Vrms (in ISM
and amateur
radio bands)
80% Am at 1kHz
Electromagnetic environment - guidance
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies. NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular / cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the device is used exceeds the applicable RF compliance level above, the device should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the device. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3V/m.
b
The device is intended for use in the electromagnetic environment specified below. The customer or the user of the device should assure that it is used in such an environment.
150 kHz to 80 MHz: 3 Vrms
6Vrms (in ISM
and amateur
radio bands)
80% Am at 1kHz
Portable and mobile RF communications equipment should be used no closer to any part of the device, including cables, than the recommended separation distance calculated from the equation appropriate for the frequency of the transmitter. Recommended separation distances: d=0.35 ; d=1.2
10V/m, 80% Am at 1kHz
80 MHz to 800 MHz: d=1.2 800 MHz to 2.7 GHz: d=2.3
where, P is the maximum output power rating of the
transmitter in watts (W)
according to the transmitter manufacturer, d is the recommended separation distance in
meters (m). Field
strengths from fixed RF transmitters, as determined by an electromagnetic site survey, should be less than the compliance level in each frequency range. Interference may occur in the vicinity of equipment marked with the following symbol:
P
P
P
P
EMC GUIDANCE
32 33
EMC GUIDANCE
150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2.7 GHz
Table 4
Recommended separation distances between portable and mobile RF communications equipment and the device.
Rated maximum output
power of transmitter
(W)
Separation distance according to frequency of transmitter (m)
0.01
0.1
1
10
100
0.12 0.12
0.38
1.2
3.8
12
0.23
0.73
2.3
7.3
23
=d
=d 1.2
2.3=d
0.37
1.2
3.8
12
3.5
The device is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the device can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment
(transmittters) and the device as recommended below, according to the maximum output power of the
communications equipment.
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer. NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
P
P
P
EMC GUIDANCE
Table 5
Guidance and manufacturer’s declaration - electromagnetic immunity
Test Frequency
(MHz)
385
0.3 27
The device is intended for use in the electromagnetic environment specified below. The customer or the user of the device, should assure that it is used in such an environment.
NOTE If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the transmitting antenna
and the ME EQUIPMENT or ME SYSTEM may be reduced to 1 m. The 1 m test distance is permitted by
IEC 61000-4-3.
Radiated RF IEC61000-4-3
(Test
specifications for ENCLOSURE PORT IMMUNITY to RF wireless communicatio
ns equipment)
Band a) (MHz)
Service a) Modulation b) Modulation b)
(W)
Distance (m)
IMMUNITY TEST LEVEL
(V/m)
380-390 TETRA
400
Pulse
modulation b)
18Hz
1.8
450 380-390
GMRS 460 FRS 460
FM c) ± 5kHz
deviation 1kHz sine
2 0.3 28
710 704-787
745
780
LTE Band 13, 17
Pulse
modulation b)
217Hz
0.2 0.3 9
810
870
930
800-960
GSM 800/900, TETRA 800, iDEN 820, CDMA 850, LTE Band 5
Pulse
modulation b)
18Hz
2
0.3 28
1720
1845
1970
1700­1990
GSM 1800; CDMA 1900; GSM 1900; DECT; LTE Band 1, 3, 4,25; UMTS
Pulse
modulation b)
217Hz
2
0.3 28
2
0.3 282450 2400-
2570
Bluetooth, WLAN,
802.11 b/g/n, RFID 2450, LTE Band 7
Pulse
modulation b)
217 Hz
5240
5240
5785
5100­5800
WLAN
802.11 a/n
Pulse
modulation b)
217 Hz
0.2
0.3 9
a) For some services, only the uplink frequencies are included. b) The carrier shall be modulated using a 50% duty cycle square wave signal. c) As an alternative to FM modulation, 50% pulse modulation at 18 Hz may be used because while it does
not represent actual modulation, it would be worst case.
The MANUFACTURER should consider reducing the minimum separation distance, based on RISK MANAGEMENT, and using higher IMMUNITY TEST LEVELS that are appropriate for the reduced minimum separation distance. Minimum separation distances for higher IMMUNITY TEST LEVELS shall be calculated using the following equation: E=
Where P is the maximum power in W, d is the minimum separation distance in m, and E is the IMMUNITY TEST LEVEL in V/m.
34 35
EMC GUIDANCE