Contact:
HONGKONG XUHUA INDUSTRIAL CO., LTD.
RM 605, 6/F DAVID HSE 8-20 NANKING ST JORDAN KL, HONGKONG
E-mail: service.usuk@lifehoodglobal.com
Guangdong Transtek Medical Electronics Co., Ltd.
Zone B, No.105 ,Dongli Road, Torch Development District,
Zhongshan,528437,Guangdong,China
If any quality questions or confusion, feel free to contact us via
E-mail: service.usuk@lifehoodglobal.com.
Please do read the user manual carefully and thoroughtly so as to ensure the
safe usage of this product, and keep the manual well for further reference in
case you have problems.
DATA MANAGEMENT.....................................................................17
Recall the Records
Delete the Records
INFORMATION FOR USER............................................................19
Tips for measurement
Maintenances
1
Page 3
CATALOGUE
INTRODUCTION
ABOUT BLOOD PRESSURE........................................................21
What are systolic pressure and diastolic pressure?
What is the standard blood pressure classification?
Irregular heartbeat detector
Why does my blood pressure fluctuate throughout the day?
Why do I get a different blood pressure at home compared to
the hospital?
Is the result the same if measuring on the right arm?
Thank you for selecting LIFEHOOD arm type blood pressure
monitor (TMB-1775). The monitor features blood pressure
measurement, pulse rate measurement and the result storage.
The design provides you with two years of reliable service.
Readings taken by the TMB-1775 are equivalent to those
obtained by a trained observer using the cuff and stethoscope
auscultation method.
This manual contains important safety and care information,
and provides step by step instructions for using the product.
Read the manual thoroughly before using the product.
Features:
65×50mm LCD Display
Maximum 60 records
3rd technonoly: Measuring during inflation
Indications for Use
The LIFEHOOD Blood Pressure Monitor is digital monitors intended
for use in measuring blood pressure and heartbeat rate with arm
circumference ranging from 22cm to 42cm(about 83⁄4 -161⁄2 ).
It is intended for adult indoor use only.
Contraindications
1.The device should not be used by any person who may be
suspected of,or is pregnant .
2.The device is not suitable for use on patients with implanted,
electrical devices, such as cardiac pacemakers, defibrillators.
Contraindications
23
Page 4
INTRODUCTION
Measurement Principle
This product uses the Oscillometric Measuring method to detect
blood pressure. Before every measurement, the unit establishes
a “zero pressure” equivalent to the air pressure.Then it starts
inflating the arm cuff, meanwhile, the unit detects pressure
oscillations generated by beat-to-beat pulsatile, which is used to
determine the systolic and diastolic pressure, and also pulse rate.
Screen Elements
mmHg
SYMBOLDESCRIPTION
Systolic pressure
Diastolic pressureLow blood pressure
mmHg
Blood pressure
level indicator
INTRODUCTION
EXPLANATION
High blood pressure
Pulse in beats per minutePulse display
Measurement Unit of the blood
pressure
(1mmHg=0.133kPa)
Indicate the blood pressure
level
Low battery
mmHg
Irregular heartbeat
Current TimeYear/Month/Day, Hour : Minute
Heartbeat
/min
Average value
45
Batteries are low and need to
be replaced
Blood pressure monitor is
detecting an irregular heartbeat
during measurement.
Blood pressure monitor is
detecting a heartbeat during
measurement.
The average value of lastest
tree grops blood pressure
Page 5
INTRODUCTION
INTRODUCTION
Monitor Components
CUFF
AIR TUBE
AIR CONNECTOR
PLUG
MEMORY BUTTON
LCD DISPLAY
BATTERY
COMPARTMENT
mmHg
M
D
Y
mmHg
/min
START/STOP
BUTTON
Component list of
pressure measuring system
1 Cuff
2 Air pipe
3 PCBA
4 Pump
5 Valve
List
1.Blood Pressure Monitor
mmH
g
mmH
M
D
Y
2.Cuff (Type BF applied part)
3. 4×AAA batteries
g
/mi
n
: 8.75"- 16.5"(22cm~42cm)
(TMB-1775)
(Please use LIFEHOOD authorized
cuff. The size of the actual cuff
please refer to the label on the
attached cuff.)
4.User manual
67
Page 6
BEFORE YOU START
BEFORE YOU START
The Choice of Power Supply
1.Battery powered mode:
6VDC 4×AAA batteries
CAUTION
In order to get the optimal performance and protect your
monitor, please use the correct batteries.
Installing and Replacing the Batteries
• Slide off the battery cover.
• Install the batteries by matching the correct polarity, as shown.
• Replace the battery cover.
Replace the batteries whenever the below happens
The shows
The display is dim
The display does not light up.
CAUTION
Do not use new and used batteries together.
Do not use different types of batteries together.
Do not dispose the batteries in fire. Batteries may explode or
leak.
Remove batteries if the device is not likely to be used for some
time.
Worn batteries are harmful to the environment. Do not dispose
with daily garbage.
Remove the old batteries from the device following your local
recycling guidelines.
89
Page 7
BEFORE YOU START
BEFORE YOU START
Setting Date and Time.
It is important to set the date and time before using your blood
pressure monitor, so that a time stamp can be assigned to
each record that is stored in the memory. (The setting range
of the year :2019—2059; Time format:24H)
1.
When the monitor is off, hold
pressing “ START/STOP ”
for 3 seconds to enter the
mode for year setting.
2.
Press “ M ” button to change
the [ ]. Each press will
YEAR
increase the number by one
in a cycling manner.
mmHg
mmHg
/mi n
mmHg
mmHg
/mi n
3.
Press “ START/STOP ” button
to confirm [YEAR].
Then the monitor diverts to
[MONTH] and [DAY] setting.
4.
Repeat step 2 and 3 to set the [MONTH] and [DAY].
mmHg
mmHg
/mi n
mmHg
mmHg
/mi n
mmHg
mmHg
/mi n
1011
Page 8
BEFORE YOU START
MEASUREMENT
Repeat step 2 and 3 to set the [HOUR] and [MINUTE]. 5.
mmHg
mmHg
/mi n
6.
After hour and minute is set,the LCD will display “donE” first,
then display all the settings you have done and then turn off.
mmHg
mmHg
/mi n
mmHg
mmHg
/mi n
Applying the Arm Cuff
To achieve an accurate blood pressure reading, follow these steps to
position the blood pressure cuff properly.
1. Remove tight-fitting clothing or tight rolled up sleeve from your left
upper arm. Do not place the arm cuff over thick clothes.
2. Insert the air plug into the air jack securely.
3. Wrap the arm cuff firmly in place around your left upper arm.
The bottom edge of the cuff should be 2 to 3 centimeters (0.8 to
1.2inches) above the bend in your arm, and the tube and artery
marker should appear on your arm as shown.
Note: If your doctor has diagnosed you with poor circulation in your left
arm, please use your right arm.
4. Keeping the artery marker and the tube in place, wrap the cuff snugly
around your arm, but not too tight. If you can insert just two fingers
between the cuff and your arm, the tightness is acceptable.
1213
Page 9
MEASUREMENT
MEASUREMENT
How to Sit Correctly
To take a measurement, you need to be relaxed and comfortably
seated,under comfortable room temperature. Avoid bathing,
drinking alcohol or caffeine, smoking, exercising or eating 30
minutes before taking a measurement.
Sit comfortably in a chair that supports your back.
Keep legs uncrossed.
Place feet flat on the floor.
Support your arm on the flat surface.
The arm cuff should be placed on your arm at the same level
as your heart.
Start the Measurement
1.
Press the “START/STOP” button to power up the monitor and
begin the blood pressure measurement.
LCD display
mmHg
mmHg
/min
Inflating and measuring.
mmHg
mmHg
/min
START
mmHg
mmHg
/min
Display and save the
measurement
mmHg
mmHg
/min
1415
Page 10
MEASUREMENT
DATA MANAGEMENT
2. Press the “START/STOP” button to turn the monitor off.
It will automatically turn off within 1 minute.
Recall the Records
1.
When the monitor is off, please press “ M ”
button to show the average value of the
latest three records.If the records are less
than 3 groups, it will display the latest
record instead.
Press the “M” to get the record
2.
you want.
The date and time of the record will be shown alternately.
mmHg
mmHg
/min
mmHg
mmHg
/min
The current No. is No 1.
Five records in total.
The corresponding
date is January 1st
The corresponding
time is A.M. 6:00.
2017.
1617
Page 11
DATA MANAGEMENT
INFORMATION FOR USER
Delete the Records
If you did not get the correct measurement, you can delete
all results for the selected user by following steps below .
1. Hold pressing “ M ” button for 3 seconds
when the monitor is in the memory recall
mode ,the flash display “dEL ALL” will
show.
2. Hold Press “ ” button to
confirm deleting and the monitor
will turn off.
Note: To exit out of delete mode
without deleting any records,
press “START/STOP” button.
3. If there is no record, press “M” button,
the right display will be show.
Note: You cannot partially delete the
values stored in the memory. All
values for the user will be deleted.
mmHg
mmHg
/min
mmHg
mmHg
/min
Tips for Measurement
Measurements may be inaccurate if taken in the following
circumstances.
Within 1 hour
after dinner or drinking
Within 20 minutes
after taking a bath
In a very cold environment
Immediate measurement
after tea, coffee, smoking
When talking or moving
your fingers
When you want to discharge urine
1819
Page 12
INFORMATION FOR USER
ABOUT BLOOD PRESSURE
Maintenance
In order to get the best performance, please follow the
instructions below.
water.
mmHg
mmHg
/m
i
n
mmHg
mmHg
/m
in
mmHg
m
mHg
/mi
n
Put in a dry place and avoid
the sunshine
mmHg
mmH
g
/
min
Avoid intense shaking
and collisions
m
mH
g
mmHg
/min
Using wet cloths to remove dirt
Avoid touching water,
clean it with a dry cloth in case.
Avoid dusty and unstable
temperature environment
Do not attempt to clean the
reusable cuff with water and
never immerse the cuff in
What are systolic pressure and diastolic pressure?
When ventricles contract and pump blood out of
the heart, the blood pressure reaches its maximum
value in the cycle, which is called systolic pressure.
When the ventricles relax, the blood pressure
reaches its minimum value in the cycle, which is
called diastolic pressure.
Systolic
blood discharging
press
Diastolic
blood entering
vein
relax
What is the standard blood pressure classification?
The chart on the right is the standard blood pressure
classification published by American Heart Association (AHA).
AHA Home Guideline for Upper Limit of Normal BP
SYS135 mm Hg
DIA85 mm Hg
This chart reflects blood pressure categories defined by American
Heart Association.
Blood Pressure
Category
Normal
Elevated
High Blood Pressure
(Hypertension) Stage 1
High Blood Pressure
(Hypertension) Stage 2
Hypertensive Crisis
(Emergency care needed)
Systolic
mmHg (upper#)
less than 120
120-129
130-139
140 or higher
Higher than 180
Diastolic
mmHg (lower#)
less than 80
and
and
less than 80
or
80-89
or
90 or higher
and
Higher than
/or
120
artery
2021
Page 13
ABOUT BLOOD PRESSURE
ABOUT BLOOD PRESSURE
CAUTION
Only a physician can tell your normal BP range. Please contact a
physician if your measuring result falls out of the range. Please
note that only a physician can tell whether your blood pressure
value has reached a dangerous point.
Irregular Heartbeat Detector
An irregular heartbeat is detected when a heartbeat rhythm varies
while the device is measuring systolic pressure and diastolic
pressure. During each measurement, blood pressure monitor will
keep a record of all the pulse intervals and calculate the average
value of them. If there are two or more pulse intervals , the
difference between each interval and the average is more than
the average value of ±25% , or there are four or more pulse
intervals ,the difference between each interval and the average is
more than the average value of ±15%, then the irregular heartbeat
symbol will appear on the display with the measurement result.
CAUTION
The appearance of the IHB icon indicates that a pulse irregularity
consistent with an irregular heart-beat was detected during
measurement. Usually this is NOT a cause for concern. However,
if the symbol appears often, we recommend you seek medical
advice. Please note that the device does not replace a cardiac
examination, but serves to detect pulse irregularities at an early
stage.
Why does my blood pressure fluctuate through
out the day?
1. Individual blood pressure varies multiple times everyday.
It is also affected by the way you tie yourcuff and your
measurement position, so please takethe
measurement under the same conditions.
2. If the person takes medicine, the pressure
will vary more.
3. Wait at least 3 minutes for another
measurement.
Is the result the same if measuring on the right
arm?
The blood pressure is differenteven throughout the day due to
weather, emotion, exercise etc, Also, there is the “white coat”
effect, which means blood pressure usually increases in clinical
settings.
2223
Page 14
ABOUT BLOOD PRESSURE
TROUBLE SHOOTING
What you need to pay attention to when you measure your
blood pressure at home:
If the cuff is tied properly.
If the cuff is too tight or too loose.
If the cuff is tied on the upper arm.
If you feel anxious.
Taking 2-3 deep breaths before beginning will be better for
measuring. Advice: Relax yourself for 4-5 minutes until you calm
down.
Is the result the same
if measuring on the
right arm?
It is ok for both arms, but there
will be some different results for
different people. We suggest
you measure the same arm
every time.
This section includes a list of error messages and frequently asked
questions for problems you may encounter with your blood pressure
monitor. If the products not operating as you think it should, check
here before arranging for servicing.
PRO
-BLEM
No
power
Low
batteries
Error
message
Warning
message
SYMPTOM
Display will
not light up
Display is
dim or show
E 01 shows
E 02 shows
E 03 shows
E 04 shows
EExx,shows
on the
display
“out”shows
CHECK THIS
Batteries are
exhausted.
Batteries are
inserted incorrectly.
Batteries are low.
The cuff is too tight
or too loose.
The monitor
detected motion
while measuring.
The measurement
process does not
detect the pulse
signal.
The treatment of the
measurement failed.
A calibration error
occurred.(XX can be
some digital symbol,
such as 01, 02,etc.,
if this similar situation
appear, all belong to
calibration error.)
Out of measurement
range
REMEDY
Replace with new
batteries
Insert the batteries
correctly
Replace with new
batteries
Refasten the cuff and
then measure again.
Movement can affect
the measurement.
Relax for a moment
and then measure again.
Loosen the clothing on
the arm and then
measure again.
Relax for a moment and
then measure again.
Retake the measurement.
If the problem persists,
contact the retailer or our
customer service
department for further
assistance.Refer to the
warranty for contact
information and return
instructions.
Relax for a moment.
Refasten the cuff and
then measure again.
If the problem persists,
contact your physician.
A temperature range of :+5°C to +40°C
A relative humidity range of 15% to 90%,
non-condensing, but not requiring a
water vapour partial pressure greater
than 50 hPa
An atmospheric pressure range of :
700 hPa to 1060 hPa
Temperature:-20°C to +60°C
A relative humidity range of ≤ 93%,
non-condensing, at a water vapour
pressure up to 50hPa
22cm~42cm
External dimensions
Attachment
Mode of operation
Degree of protection
Protection against
ingress of water
Device Classification
Approx.120.2mm×108.2mm×68.5mm
4×AAA batteries,user manual
Continuous operation
Type BF applied part
IP21 It means the device could protected
against solid foreign objects of 12.5mm
and greater, and protect against vertically
falling water drops.
Battery Powered Mode:
Internally Powered ME Equipment
Software VersionA01
WARNING: No modification of this equipment is allowed.
Weight
Approx.225g(Excluding the dry cells
and cuff)
2627
Page 16
SAFETY INFORMATIONSAFETY INFORMATION
Safety Information
The signs below might be in the user manual,labeling or other
component.
They are the requirement of standard and using.
Symbol for “THE OPERATION GUIDE MUST BE READ”
Symbol for “MANUFACTURER”
Symbol for “SERIAL NUMBER”
SN
Symbol for “DIRECT CURRENT”
Symbol for “MANUFACTURE DATE”
Caution: These notes must be observed to prevent any
damage to the device.
Symbol for “TYPE BF APPLIED PARTS”
Symbol for “ENVIRONMENT PROTECTION - Electrical
waste products should not be disposed of with
household waste. Please recycle where facilities exist.
Check with your local authority or retailer for recycling
advice”
Symbol for “RECYCLE”
The Green Dot is the license symbol of a European
network of industry-funded systems for recycling the
packaging materials of consumer goods.
2829
CAUTION
* Be careful to strangulation due to cables and hoses, particularly
due to excessive length.
* At least 30 min required for ME equipment to warm from the
minimum storage temperature between uses until it is ready for
intended use. At least 30 min required for ME equipment to cool
from the maximum storage temperature between uses until it is
ready for intended use.
* This equipment needs to be installed and put into service in
accordance with the information provided in the
ACCOMPANYING DOCUMENTS;
* Wireless communications equipment such as wireless home
network devices, mobile phones, cordless telephones and their
base stations, walkie-talkies can affect this equipment and
should be kept at least a distance d away from the equipment.
The distanced is caculated by the MANUFACTURER from
the 80 MHz to 5.8 GHz column of Table 4 and Table 9 of
IEC 60601-1-2:2014,as appropriate.
* Please use ACCESSORIES and detachable partes specified/
authorised by MANUFACTURE.
Otherwise, it may cause damage to the unit or danger to the
user/patients.
* There is no luer lock connectors are used in the construction of
tubing, there is a possibility that they might be inadvertently
connected to intravascular fluid systems, allowing air to be
pumped into a blood vessel.
* Please use the device under the environment which was provided in
the user manual. Otherwise, the performance and lifetime of the
device will be impacted and reduced.
* This device is intended for adult use in homes only.
* The device is not suitable for use on neonatal patients, pregnant
women,patients with implanted, electronical devices, patients with
pre-eclampsia, premature ventricular beats, atrial fibrillation,
peripheral, arterial disease and patients undergoing intravascular
therapy or arterio-venous shunt or people who received a
mastectomy.
Please consult your doctor prior to using the unit if you suffer from
illnesses.
* The device is not suitable for measuring the blood pressure of
children. Ask your doctor before using it on older children.
Page 17
SAFETY INFORMATION
CAUTIONCAUTION
* The device is not intended for patient transport outside a
healthcare facility.
* The device is not intended for public use.
* This device is intended for no-invasive measuring and monitoring
of arterial blood pressure. It is not intended for use on extremities
other than the arm or for functions other than obtaining a blood
pressure measurement.
* Do not confuse self-monitoring with self-diagnosis. This unit allows
you to monitor your blood pressure.Do not begin or end medical
treatment without asking a physician for treatment advice.
* If you are taking medication,consult your physician to determine
the most appropriate time to measure your blood pressure.
Never change a prescribed medication without consulting your
physician.
* Do not take any therapeutic measures on the basis of a self
measurement. Never alter the dose of a medicine prescribed by
a doctor. Consult your doctor if you have any question about your
blood pressure.
* When the device was used to measure patients who have
common arrhythmias such as atrial or ventricular premature beats
or atrial fibrillation, the best result may occur with deviation.
Please consult your physician about the result.
* Don't kink the connection tube during use, otherwise, the cuff
pressure may continuously increase which can prevent blood flow
and result in harmful injury to the PATIENT.
* When using this device, please pay attention to the following
situation which may interrupt blood flow and influence blood
circulation of the patient, thus cause harmful injury to the patient:
connection tubing kinking too frequent and consecutive multiple
measurements; the application of the cuff and its pressurization
on any arm where intravascular access or therapy, or an arterio
-venous (A-V) shunt, is present; inflating the cuff on the side of a
mastectomy.
* Warning: Do not apply the cuff over a wound;otherwise it can
cause further injury.
* Do not inflate the cuff on the same limb which other monitoring ME
equipment is applied around simultaneously, because this could
cause temporary loss of function of those simultaneously-used
monitoring ME equipment.
SAFETY INFORMATION
* On the rare occasion of a fault causing the cuff to remain fully
inflated during measurement, open the cuff immediately.
Prolonged high pressure (cuff pressure > 300mmHg or constant
pressure > 15mmHg for more than 3 minutes) applied to the arm
may lead to an ecchymosis.
* Please check that operation of the device does not result in
prolonged impairment of patient blood circulation.
* When measurement, please avoid compression or restriction of the
connection tubing.
* The device cannot be used with HF surgical equipment at the
same time.
* The ACCOMPANYING DOCUMENT shall disclose that the
SPHYGMOMANOMETER was clinically investigated according to
the requirements of ISO 81060-2:2013.
* To verify the calibration of the AUTOMATED
SPHYGMOMANOMETER, please contact the manufacturer.
* This device is contraindicated for any female who may be
suspected of, or is pregnant. Besides providing inaccurate
readings, the effects of this device on the fetus are unknown.
* Too frequent and consecutive measurements could cause
disturbances in blood circulation and injuries.
* This unit is not suitable for continuous monitoring during medical
emergencies or operations. Otherwise, the patient’s arm and
fingers will become anaesthetic, swollen and even purple due to a
lack of blood.
* When not in use, store the device in a dry room and protect it
against extreme moisture, heat, lint, dust and direct sunlight.
Never place any heavy objects on the storage case.
* This device may be used only for the purpose described in this
booklet. The manufacturer cannot be held liable for damage
caused by incorrect application.
* This device comprises sensitive components and must be treated
with caution. Observe the storage and operating conditions
described in this booklet.
* The equipment is not AP/APG equipment and not suitable for use
in the presence of a flammable anesthetic mixture with air of with
oxygen or nitrous oxide.
3031
Page 18
SAFETY INFORMATION
CAUTIONCAUTION
SAFETY INFORMATION
* Warning: No servicing/maintenance while the ME equipment is
in use.
* The patient can measure data and change batteries under normal
circumstances and maintain the device and its accessories
according to the user manual.
* To avoid measurement errors, please avoid the condition of strong
electromagnetic field radiated interference signal or electrical fast
transient/burst signal.
* The blood pressure monitor and the cuff are suitable for use within
the patient environment. If you are allergic to polyester, nylon or
plastic, please don't use this device.
* During use, the patient will be in contact with the cuff. The materials
of the cuff have been tested and found to comply with requirements
of ISO 10993-5:2009 and ISO 10993-10:2010. It will not cause any
potential sensization or irritation reaction.
* If you experience discomfort during a measurement, such as pain
in the arm or other complaints, press the START/STOP button to
release the air immediately from the cuff. Loosen the cuff and
remove it from your arm.
* If the cuff pressure reaches 40 kPa (300 mmHg), the unit will
automatically deflate. Should the cuff not deflate when pressures
reaches 40 kPa (300 mmHg), detach the cuff from the arm and
press the START/STOP button to stop inflation.
* Before use, make sure the device functions safely and is in proper
working condition. Check the device, do not use the device if it is
damaged in any way. The continuous use of a damaged unit may
cause injury, improper results, or serious danger.
* Do not wash the cuff in a washing machine or dishwasher!
* The service life of the cuff may vary by the frequency of washing,
skin condition, and storage state. The typical service life is 10000
times.
* It is recommended that the performance should be checked every
2 years and after maintenance and repair, by retesting at least the
requirements in limits of the error of the cuff pressure indication
and air leakage (testing at least at 50mmHg and 200mmHg).
* Please dispose of ACCESSORIES, detachable parts, and the ME
EQUIPMENT according to the local guidelines.
* Manufacturer will make available on request circuit diagrams,
component part lists, descriptions, calibration instructions,etc.,
to assist to service personnel in parts repair.
* The operator shall not touch output of batteries and the patient
simultaneously.
* Cleaning :Dust environment may affect the performance of the unit.
Please use the soft cloth to clean the whole unit before and after
use. Don’t use any abrasive or volatile cleaners.
* The device doesn’t need to be calibrated within two years of
reliable service.
* If you have any problems with this device, such as setting up,
maintaining or using, please contact the SERVICE PERSONNEL
of Transtek. Don’t open or repair the device by yourself in the event
of malfunctions. The device must only be serviced, repaired and
opened by individuals at authorized sales/service centers.
* Please report to Transtek if any unexpected operation or events
occur.
* Keep the unit out of reach of infants, young children or pets to
avoid inhalation or swallowing of small parts. It is dangerous or
even fatal.
3233
Page 19
CONTACT INFORMATION
COMPLIED STANDARDS LIST
Contact Information
For more information about our products, please contact us by
mail,you can get customer service, usual problems and customer
download, Lifehood will serve you anytime.
Manufactured by:
Guangdong Transtek Medical Electronics Co., Ltd.
Company:
Guangdong Transtek Medical Electronics Co., Ltd.
Address:
Zone B, No.105 ,Dongli Road, Torch Development District,
Zhongshan,528437,Guangdong,China
If any quality questions or confusion, feel free to contact us :
HONGKONG XUHUA INDUSTRIAL CO., LTD.
RM 605, 6/F DAVID HSE 8-20 NANKING ST JORDAN KL, HONGKONG
E-mail: service.usuk@lifehoodglobal.com
FCC Statement
This device complies with Part 15 of the FCC Rules. Operation is
subject to the following two conditions: (1) this device may not
cause harmful interference, and (2) this device must accept any
interference received, including interference that may cause
undesired operation.
Complied Standards List
Risk
management
Labeling
User manual
General
Requirements
for Safety
Electromagnetic
compatibility
Performance
requirements
EN ISO 14971:2012 / ISO 14971:2007 Medical
devices - Application of risk management to
medical devices
EN ISO 15223-1:2016/ ISO 15223-1:2016 Medical devices.
Symbols to be used with medical device labels, labelling
and information to be supplied.
Part 1 : General requirements
EN 1041:2008 Information supplied by the
manufacturer of medical devices
EN 60601-1:2006/ IEC 60601-1:2005+A1:2012
Medical electrical equipment - Part 1: General
requirements for basic safety and essential
performance
EN 60601-1-11:2010/ IEC 60601-1-11:2015
Medical electrical equipment - Part 1-11: General
requirements for basic safety and essential
performance - Collateral standard: Requirements
for medical electrical equipment and medical
electrical systems used in the home healthcare
environment
EN 60601-1-2:2015/ IEC 60601-1-2:2014 Medical
electrical equipment - Part 1-2: General
requirements for basic safety and essential
performance - Collateral standard: Electromagnetic
compatibility - Requirements and tests
EN ISO 81060-1:2012 Non-invasive sphygmomanometers
-Part 1: Requirements and test methods for non
-automated measurement type
EN 1060-3:1997+A2:2009 Non-invasive
sphygmomanometers-Part 3: Supplementary
requirements for electro-mechanical blood pressure
measuring systems
IEC 80601-2-30:2013 Medical electrical equipment
-Part 2-30: Particular requirements for the basic
safety and essential performance of automated
non-invasive sphygmomanometers
3435
Page 20
COMPLIED STANDARDS LIST
EMC GUIDANCE
Complied Standards List
EN 1060-4:2004 Non-invasive sphygmomanometers
Clinical
investigation
Usability
processes
Bio
-compatibility
-Part 4: Test procedures to determine the overall
system accuracy of automated non-invasive
sphygmomanometers
ISO 81060-2:2013 Non-invasive sphygmomanometers
-Part 2: Clinical validation of automated measurement type
EN 60601-1-6:2010/IEC 60601-1-6:2010+A1:2013
Medical electrical equipment - Part 1-6: General
requirements for basic safety and essential
performance - Collateral standard: Usability
EN 62366:2008/ IEC 62366-1:2015 Medical devices
-Application of usability engineering to medical devices
EN 62304:2006/AC: 2008 / IEC 62304:2006 Medical
device software - Software life-cycle processes
ISO 10993-1:2009 Biological evaluation of medical devices
-Part 1:Evaluation and testing within a risk
management process
ISO 10993-5:2009 Biological evaluation of medical devices
- Part 5: Tests for in vitro cytotoxicity
ISO 10993-10:2010 Biological evaluation of medical devices
- Part 10: Tests for irritation and skin sensitization
EMC Guidance
1)This product needs special precautions regarding EMC and needs
to be installed and put into service according to the EMC information
provided, and this unit can be affected by portable and mobile RF
communications equipment.
2)* Do not use a mobile phone or other devices that emit electromagnetic
fields, near the unit. This may result in incorrect operation of the unit.
3)Caution: This unit has been thoroughly tested and inspected to
assure proper performance and operation!
4)* Caution: This machine should not be used adjacent to or stacked
with other equipment and that if adjacent or stacked use is necessary,
this machine should be observed to verify normal operation in the
configuration in which it will be used.
Table 1
Guidance and manufacturer’s declaration
– electromagnetic emissions
The device is intended for use in the electromagnetic environment
specified below. The customer or the user of the device should
assure that it is used in such an environment.
Compliance
Group 1
Class B
Class A
Complies
Electromagnetic environment
- guidance
The device uses RF energy only
for its internal function. Therefore,
its RF emissions are very low and
are not likely to cause any
interference in nearby electronic
equipment.
The device is suitable for use in all
establishments, other than domestic
and those directly connected to the
public low-voltage power supply
network that supplies buildings used
for domestic purposes.
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic
emissions
IEC 61000-3-2
Voltage
fluctuations/
flicker emissions
IEC 61000-3-3
3637
Page 21
EMC GUIDANCE
Table 2
Guidance and manufacturer’s declaration
– electromagnetic immunity
The device is intended for use in the electromagnetic environment
specified below. The customer or the user of the device should
assure that it is used in such an environment.
Immunity
test
Electrostatic
discharge
(ESD)
IEC 61000
-4-2
Electrical
fast
transient/
burst
IEC 61000
-4-4
Surge
IEC61000
-4-5
Voltage dips,
short
interruptions
and voltage
variations on
power supply
input lines
IEC 61000
-4-11
Power
frequency
(50Hz/60Hz)
magnetic
field
IEC 61000
-4-8
IEC 60601 test
level
±8 kV contact
±15 kV air
power supply
lines:±2 kV
input/output
lines:±1 kV
line(s) to line(s):
±1 kV
line(s) to earth:
±2 kV
100 kHz repetition
frequency
0% 0.5 cycle
At 0°,45°,90°,135°
180°,225°,270°and
315° 0% 1 cycle
and 70% 25/30
cycles
Single phase:
at 0 0% 300 cycle
30 A/m
50Hz/60Hz
Compliance
level
±8 kV contact
±15 kV air
power supply
lines:±2 kV
line(s) to line(s):
±1 kV
100 kHz repetition
frequency
0% 0.5 cycle
At 0°,45°,90°,135°
180°,225°,270°and
315° 0% 1 cycle
and 70% 25/30
cycles
Single phase:
at 0 0% 300 cycle
30 A/m
50Hz/60Hz
Electromagnetic
environment
- guidance
Floors should be wood,
concrete or ceramic tile.
If floors are covered
with synthetic material,
the relative humidity
should be at least 30%
Mains power quality
should be that of a
typical commercial or
hospital environment.
Mains power quality
should be that of a
typical commercial or
hospital environment.
Mains power quality
should be that of a
typical commercial
or hospital
environment.
Power frequency
magnetic fields
should be at levels
characteristic of a
typical location in a
typical commercial or
hospital environment.
NOTE UT is the a.c. mains voltage prior to application of the test level.
38
Table 3
Guidance and manufacturer’s declaration
– electromagnetic immunity
The device is intended for use in the electromagnetic environment
specified below. The customer or the user of the device should
assure that it is used in such an environment.
Immunity
test
IEC 61000
-4-6
Radiated RF
IEC 61000
-4-3
IEC 60601
Test level
150 kHz to
80 MHz:
3 Vrms
6Vrms
(in ISM and
amateur
radio bands)
80% Am at
1kHz
10V/m, 80%
Am at 1kHz
Compliance
level
150 kHz to
80 MHz:
3 Vrms
6Vrms
(in ISM and
amateur
radio bands)
80% Am at
1kHz
10V/m, 80%
Am at 1kHz
Electromagnetic environment
- guidance
Portable and mobile RF
communications equipment
should be used no closer to
any part of the device, including
cables, than the recommended
separation distance calculated
from the equation appropriate
for the frequency of the
transmitter.
Recommended separation
distances:d=0.35;d=1.2
80 MHz to
800 MHz:
d=1.2
800 MHz to
2.7 GHz:
d=2.3
EMC GUIDANCE
where,P is the
maximum output
power rating of the
transmitter in watts
(W)according to the
transmitter
manufacturer, d is
the recommended
separation distance
in meters (m).
Field strengths from
fixed RF transmitters,
as determined by an
electromagnetic site
survey,a should be
less than the
compliance level in
each frequency range.
b
Interference may
occur in the vicinity
of equipment marked
with the following
symbol:
39
Page 22
EMC GUIDANCE
Table 3
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations.
Electromagnetic propagation is affected by absorption and reflection
from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for
radio (cellular / cordless) telephones and land mobile radios,
amateur radio, AM and FM radio broadcast and TV broadcast
cannot be predicted theoretically with accuracy. To assess the
electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the
measured field strength in the location in which the device is
used exceeds the applicable RF compliance level above, the
device should be observed to verify normal operation.
If abnormal performance is observed, additional measures may
be necessary, such as re-orienting or relocating the device.
b Over the frequency range 150 kHz to 80 MHz, field strengths
should be less than 3V/m.
EMC GUIDANCE
Table 4
Recommended separation distances between portable and
mobile RF communications equipment and the device.
The device is intended for use in an electromagnetic environment
in which radiated RF disturbances are controlled. The customer
or the user of the device can help prevent electromagnetic
interference by maintaining a minimum distance between portable
and mobile RF communications equipment (transmittters) and the
device as recommended below, according to the maximum output
power of the communications equipment.
Rated maximum
output power of
transmitter
(W)
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above,
the recommended separation distance d in metres (m) can be
estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the
higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations.
Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
Separation distance according to frequency
of transmitter (m)
800 MHz to
2.7 GHz
d = 2.3
0.23
0.73
2.3
7.3
23
150 kHz to
80 MHz
d = 3.5
0.120.12
0.38
1.2
3.8
12
80 MHz to
800 MHz
d = 1.2
0.38
1.2
3.8
12
4041
Page 23
EMC GUIDANCE
Guidance and manufacturer’s declaration
- electromagnetic immunity
The device is intended for use in the electromagnetic environment
specified below. The customer or the user of the device should
assure that it is used in such an environment.
Radiated RF
IEC61000
-4-3
(Test
specifications
for
ENCLOSURE
PORT
IMMUNITY
to RF wireless
communic
NOTE If necessary to achieve the IMMUNITY TEST LEVEL,
the distance between the transmitting antenna and the ME
EQUIPMENT or ME SYSTEM may be reduced to 1 m.
The 1 m test distance is permitted by IEC 61000-4-3.
a) For some services, only the uplink frequencies are included.
b) The carrier shall be modulated using a 50% duty cycle square
wave signal.
c) As an alternative to FM modulation, 50% pulse modulation at
18 Hz may be used because while it does not represent actual
modulation, it would be worst case.
The MANUFACTURER should consider reducing the minimum
separation distance, based on RISK MANAGEMENT, and using
higher IMMUNITY TEST LEVELS that are appropriate for the
reduced minimum separation distance. Minimum separation
distances for higher IMMUNITY TEST LEVELS shall be calculated
using the following equation:
Where P is the maximum power in W, d is the minimum separation
distance in m, and E is the IMMUNITY
TEST LEVEL in V/m.
E=
6
P
d
44
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.