Transducer Techniques DPM-3 Operator's Manual

DPM-3
DIGITAL PANEL MOUNT METER
PLUG AND PLAY IEEE 1451.4 COMPLIANT
OPERATOR MANUAL
4006497
Transducer
Techniques
R
TABLE OF CONTENTS
REVISED 07/2017
1. TEDS IEEE 1451.4 INTRODUCTION .......................................................................... 3
2. GENERAL INTRODUCTION ....................................................................................... 3
3. RECEIVING & UNPACKING ....................................................................................... 4
4. SAFETY CONSIDERATIONS ...................................................................................... 4
5. CONNECTOR WIRING INFORMATION ..................................................................... 6
6. MECHANICAL ASSEMBLY ......................................................................................... 8
7. FRONT PANEL SETUP KEYS .................................................................................... 10
8. ENABLING AND LOCKING OUT MENU ITEMS ......................................................... 12
9. TEDS SIGNAL CONDITIONER IN TEDS MODE ........................................................ 13
10. SETUP OF NON-TEDS TRANSDUCERS ................................................................... 17
11. METER SCALING BY APPLYING KNOWN LOADS .................................................. 18
12. TEDS SIGNAL CONDITIONER WITH NON-TEDS TRANSDUCERS ......................... 19
13. LOAD CELL & MICROVOLT SIGNAL CONDITIONER ............................................... 23
14. DUAL RELAY OUTPUT OPTION ............................................................................... 25
15. ANALOG OUTPUT OPTION ....................................................................................... 28
16. SERIAL COMMUNICATION OPTIONS ....................................................................... 29
17. EXCITATION OUTPUTS & POWER SUPPLY ............................................................ 34
18. DIGITAL CONTROL INPUTS ...................................................................................... 35
19. INSTRUMENT SETUP VIA PC .................................................................................... 36
20. CUSTOM CURVE LINEARIZATION ............................................................................ 39
21. METER CALIBRATION ............................................................................................... 41
22. SPECIFICATIONS ....................................................................................................... 41
23. GLOSSARY OF TERMS .............................................................................................. 44
24. ACCESSORIES ........................................................................................................... 50
25. WARRANTY & REPAIR POLICY ................................................................................. 52
1. TEDS IEEE 1451.4 INTRODUCTION
The DPM-3 is a TEDS IEEE 1451.4 Plug and Play Smart Load Cell Meter. TEDS, or Transducer Electronic Data Sheet, is a set of electronic data in a standardized format defined within the IEEE 1451.4 standard. Stored in an EEPROM, this data specifies what type of sensor is present, describes its interface, and gives technical information such as sensitivity, bridge type, excitation, etc.
The DPM-3 automatically detects when a TEDS IEEE 1451.4 compliant Load Cell / Torque Sensor has been connected to it by using a built-in EEPROM detector. Once such a sensor has been detected, the DPM-3 displays a front panel TEDS indicator light, reads the EEPROM, stores the information in memory, and performs an automatic configuration. The built-in, sensor-related EEPROM may be of any of the following types: DS1973/DS2433, DS2431 or DS1971/DS2430A. The automatic system configuration function performs all steps needed to calibrate the TEDS IEEE 1451.4 compliant Load Cell / Torque Sensor and DPM-3 as a system. This includes the configured precision of 32 bits, 19 bits or 11 bits and the configured excitation voltage. Using the DPM-3 with a TEDS IEEE 1451.4 compliant Load Cell / Torque Sensor is as easy as plugging a mouse into a computer, making it a true plug and play experience.
The DPM-3 is safety certified to UL 61010-1 and to CSA C22.2#61010-1. It carries the ETL certification mark for the USA and Canada.
2. GENERAL INTRODUCTION
DPM-3 digital panel meters are a versatile, cost-effective solution to a wide variety of monitoring and control applications. They are easily set up to produce an accurate display of weight, load or direct reading of microvolts. Setup can be via front panel pushbuttons or the meter’s serial interface. Digital scaling of zero and span provides direct readout in engi­neering units. Digital calibration of all ranges eliminates drift associated with potentiometers found in non-microcomputer-based meters. Selective security lockout of the front panel keys protects against accidental changes to the meter setup.
A unique method of analog-to-digital conversion provides 60 conversions per second (50 for 50 Hz operation), while integrating the signal over a full line cycle for maximum noise rejection. Self-calibration cycles reduce the average reading rate to 56 per second (47 for 50 Hz). This fast read rate provides an accurate display of peak signal input and quick response in control applications.
An adaptive auto-filter automatically selects a time constant appropriate for the encountered signal noise level. This ensures stable displayed readings and outputs while responding rapidly to changes of the input signal that exceed a selected threshold value. Input signal polarity may be selected as normal or reverse.
The DPM-3 uses a lightweight, high-efficiency switching power supply that operates from either AC or DC voltages and complies with safety regulations. The meter can be powered worldwide without changes to the supply. An optional low voltage supply operates on 10 to 48 Vdc from batteries or 12 to 32 Vac from sources such as 400 Hz aircraft power. Both supplies have an isolated 10 Vdc excitation supply to power transducers.
The NEMA-4 (IP65) 1/8 DIN case is made of high impact 94V-0 UL-rated plastic. Mounting is from the front of the panel and requires less than 110 mm behind the panel. All wiring is by removable plugs conforming to UL61010C safety standards. All output options are isolated from meter and power ground by 250 Vac minimum.
Alarm or setpoint control is provided by an optional relay board with two or four Form C 8A mechanical relays or two or four Form A 120 mA solid state relays. The setpoints may be latching or non-latching, be energized above or below the setpoint, or operate in a fail-safe mode. The relays can operate from the filtered signal to reduce relay chatter or from the unfiltered signal for fastest response. Snubber circuits and a programmable relay switching time delay extend relay contact life.
An isolated analog output of 4-20 mA, 0-20 mA, 0-10V or -10 to +10V can be provided by an optional analog output board. The output is linearized to the display and can operate from the filtered or unfiltered signal input. It can be scaled via front panel pushbuttons or the meter’s serial interface.
Optional RS232, RS485, or USB serial interfaces allow the DPM-3 to communicate bidirec­tionally with computers, PLCs or other digital devices. An optional USB-to-RS485 converter board allows a primary DPM-3 to be interfaced to a computer and to be the device server for a network of up to 31 other DPM-3’s on an RS485 bus, while itself retaining all capabilities of a meter.
Windows-based software is available from Transducer Techniques to program our meter and transmitters via a PC using the serial interface. This software can be downloaded at no charge from www.transducertechniques.com/online-manuals.aspx. Please see Section 19 for details.
3. RECEIVING & UNPACKING
Your DPM-3 meter was carefully tested and inspected prior to shipment. Should the meter be damaged in shipment, notify the freight carrier immediately. In the event the meter is not configured as ordered or the unit is inoperable, return the unit to Transducer Techniques for repair or replacement. Please include a detailed description of the problem.
4. SAFETY CONSIDERATIONS
Visually inspect the instrument for signs of damage. If damaged, do not attempt to operate.
Warning: Use of this equipment in a manner other than specified may impair the pro­tection of the device and subject the user to a hazard. Visually inspect the unit for signs of damage. If the unit is damaged, do not attempt to operate.
Caution:
This unit must be powered with AC (mains) from 85-264 Vac with the high voltage power
supply option, or 12-32 Vac (10-48 Vdc) with the low voltage power supply option. Verify
that the proper power option is installed for the power to be used. This meter has no AC
4006497
ETL Mark. Indicates that product conforms to UL Std. 61010-1 and is certified to CAN/USA Std. C22.2 No. 61010-1
CE Mark. Indicates that product meets EU safety, health and environmental requirements.
Caution (refer to accompanying documents)
Caution, risk of electric shock.
Earth (ground) terminal.
Equipment protected throughout by double insulation or reinforced insulation.
Both direct and alternating current.
RoHS Symbol. Indicates that product is free from hazardous substances defined in EC directive 2002/95/EC.
WEEE Symbol. Indicates that product should be recycled and not disposed of as general waste.
(mains) switch. It will be in operation as soon as power is connected.
The 85-264 Vac mains connector (P1 Pins 1-3) is colored Green to differentiate it from
other input and output connectors. The 12-32 Vac (10-48 Vdc) mains connector is colored Black.
Do not make signal wiring changes or connections when power is applied to the instru-
ment. Make signal connections before power is applied. If reconnection is required,
disconnect the AC (mains) power before such wiring is attempted. To prevent electrical or fire hazard, do not expose the instrument to excessive moisture. Do not operate the instrument in the presence of flammable gases or fumes; such an
environment constitutes a definite safety hazard. This meter is designed to be mounted in a metal panel. Verify the panel cutout dimen-
sions, and mount according to instructions.
Symbols applicable to this product:
Operating environment:
The meter is Class II (double insulated) equipment designed for use in Pollution degree 2.
5. CONNECTOR WIRING INFORMATION
5.1 CONNECTOR LOCATION
Connectors for signal and power are UL-rated screw-clamp terminal blocks that plug into mating jacks on the printed circuit board. Communication connectors can be a USB jack, a single RJ11 jack for RS232, dual RJ11 jacks for RS485, or dual RJ45 jacks for RS485.
Note: For wiring color, refer to Load Cell Calibration Certificate or to
www.transducertechniques.com/wiring-color-code.aspx
P3 - SERIAL COMMUNICATIONS P4 - ANALOG OUTPUT
6 5 4 3 2 1
N/C ISO GND RX TX RTS N/C
GND TX RX RTS
RS232 INTERFACE Computer
6 5 4 3 2 1
ISO GND BRX ARX ATX BTX ISO GND
GND BTX ATX ARX ARX GND
RS485 INTERFACE - FULL DUPLEX
6 5 4 3 2 1
ISO GND
ATX / ARX BTX / BRX ISO GND
GND
ATX / ARX BTX / BRX GND
RS485 INTERFACE - HALF DUPLEX
1 2 3 4 5 6 7 8
(A') RXD0 (B') RXD1
(B) TXD1 (A) TXD0
ISO GND
TXD0 TXD1
RXD1 RXD0
GND
RS485-MODBUS - FULL DUPLEX RS485-MODBUS - HALF DUPLEX
1 2 3 4 5 6 7 8
(B) TX/RXD1 (A) TX/RXD0
ISO GND
D1 D0
GND
6. MECHANICAL ASSEMBLY
Rear Panel
Option Board
Main Board Jack
Rear Panel Jack
Power supply Relay board Serial interface board Analog output board Signal conditioner board
P11 P12 P13 P14 P15
J1 J2 J3 J4 J5
Retaining tab with tab release
Retaining tab
with tab release
Retaining tab
Retaining tab
REMOVING THE REAR PANEL
First remove any connectors. Use one hand to press in the two sides of the rear of the case, and the other hand to press down the two protruding tab releases at the top of the rear panel (see figure below). This will unhook the rear panel from the case.
REMOVING THE ELECTRONICS
With the rear panel removed, grasp the power supply board to the left and signal condi­tioner board to the right, and carefully slide the electronic assembly out through the rear of the case (see figure below).
INSTALLING NEW OPTION BOARDS
Options boards plug into the main board at the front of the meter. These are plug-and-play and may installed in the field. They will be recognized by the software, which will provide access to the menu items associated with that board. If necessary, remove rear panel knockouts for new boards. Boards plug into connectors as follows:
Note: Corresponding main board and option board connectors have the same number of
The DPM-3 uses UL / VDE rated detachable screw terminal connectors for signal and power.
electrical lines. When an option board is correctly installed, the top and bottom edges of the main board and option board are aligned.
REASSEMBLING YOUR METER
Slide the electronics assembly into the case until the display board is seated flush against the front overlay. Insert the bottom tabs of the rear panel into the case, then carefully align the board connectors with the openings in the rear panel. If necessary, remove any rear panel knockouts for new option boards that may have been installed. Ensure that all option boards are properly aligned with the molded board retaining pins on the inside of the rear panel. With the rear panel in place, reinstall the input/output screw clamp terminal plugs.
PANEL MOUNTING
Ensure that the panel mounted gasket is in place against the back of the bezel. Turn the two mounting screws counterclockwise until the space between the mounting pawl and the rear of the gasket is greater than the panel thickness. Insert the meter in the panel cutout. Turn the mounting screws clockwise until the meter is securely mounted in the panel. Do not overtighten.
7. FRONT PANEL INDICATORS AND SETUP KEYS
Alarms key (Reverse menu)
Alarms status
indicators
1 through 4
Menu key
Peak display key
(Digit select)
Reset/Tare key (Value select)
Sensor Calibration indicator
Instrument Calibration
indicator
TEDS indicator
TEDS INDICATOR
Meter Front Panel
On steady when DPM-3 has detected a TEDS sensor, has read the EEPROM, and has performed an automatic meter configuration.
CALIBRATION INDICATORS
The following two indicators are active when the Calibration indicators are enabled by the
tArE menu item:
1. INSTRUMENT CALIBRATION INDICATOR
Flashing indicates either that the instrument is due for calibration or will be due in less than 30 days. On steady when instrument is past due for calibration. Stops flashing when acknowledged from the front panel by selecting the CALrt menu item and changing the first digit from a 1 to a 0.
2. SENSOR CALIBRATION INDICATOR
Active when a TEDS Plug-and-Play Sensor is detected in Plug-and-Play mode, other­wise stays dark. Flashing indicates either that the sensor is due for calibration or will be due in less than 30 days. Stops flashing when acknowledged from the front panel by selecting the CALrt menu item and changing the second digit from a 1 to a 0. Once acknowledged, it is on steady when the TEDS Plug-and-Play sensor is past due for calibration.
FRONT PANEL KEYS
There are four front panel keys, which change function for the Run Mode and Menu Mode, effectively becoming eight keys. The keys are labeled with alphanumeric captions (MENU,
PEAK, RESET, ALARMS) for the Run Mode and with symbols ( Menu, Digit select,
Value select, Reverse menu) for the Menu Mode.
- 10 -
KEYS IN MENU MODE
In the Menu Mode, pressing a key momentarily advances to the next menu item. Holding down a key automatically advances through multiple menu items for fast menu navigation.
(Menu). Pressing MENU steps the meter through all menu items that have been enabled and then back to the Run Mode. With the DC signal conditioner board and no option boards, available menu items are _InPut, SEtuP, ConFG, _FiLtr, dEc.Pt, Lo in,
Lo rd, Hi in, Hi rd, tArE, Loc 1, Loc 2, Loc 3, Loc 4 . If a change has been made to a
menu item, that change is saved to non-volatile memory when the MENU key is pressed next, and StorE is displayed briefly.
(Digit Select). Pressing digit select from the InPut menu brings up all meter functions available with the meter's signal conditioner. For the Load cell signal conditioner, these are, Strn and _dC u.
Pressing digit select from the SEtuP, ConfFG, FiLtr, Lo in, Lo rd, Hi in, Hi rd, tArE,
Loc 1, Loc 2 or Loc 3 menus items sequentially selects digit positions 1 - 5, as indicated by a flashing digit: 00000, 00000, 00000, 00000, 00000.
(Value Select). Pressing value select for a flashing item (digit position or decimal point position) will increment that item. Pressing MENU will save any changes.
(Reverse Menu). Pressing has the same effect as the MENU key, except that menu items are brought up in reverse order.
KEYS IN RUN MODE
MENU Key. Pressing MENU from the Run Mode enters the Menu Mode. Pressing
MENU repeatedly will step the meter through the various menu items (if these have not been locked out) and then back to the Run Mode.
PEAK Key. Pressing PEAK normally causes the peak value of the input signal to be displayed. The peak display then blinks to differentiate it from the normal present value display. Pressing PEAK again returns the display to the present value. The PEAK key can also be programmed to display Valley, alternating Peak or Valley, or to Tare the reading to zero. When Peak or Valley is selected, periodic horizontal bars at the top of the display indicate Peak, and periodic horizontal bars at the bottom indicate Valley.
RESET/TARE Key. Pressing RESET with PEAK resets peak and valley values. Pressing RESET with ALARMS resets latched alarms. Pressing RESET with MENU performs a meter reset (same as power on). Meter reset can also be applied via a rear panel connector. Pressing and releasing RESET/TARE tares the meter value to zero.
ALARMS Key. Pressing ALARMS once displays the setpoint for Alarm 1. Pressing it again displays the setpoint for Alarm 2. Pressing it again returns to the present value.
- 11 -
8. ENABLING & LOCKING OUT MENU ITEMS
Check lockout status. DPM-3 meters may have lockouts in place. This causes
menu items described in this manual not to appear.
Enabled or Disabled Menu Items
Loc 1
1 - Input type selection. 2 - Meter setup, configuration &
decimal point
3 - Filter selection. 4 - Scale or Lo, Hi input. 5 - Offset or Lo, Hi reading
Loc 2
2 - Alarm setup. 3 - Alarm setpoint value
programming.
4 - Analog output scaling. 5 - Serial interface setup.
Loc 3
2 - View peak value 3 - View alarm setpoints 4 - Reset (peak & latched alarms) 5 - Reset (meter reset)
Loc 4 (TEDS only)
1 - dAtE 2 – SerNo of sensor & Units 3 - Calibration dates and CALrt 4 - Calibrator’s initials & bridge ID 5 - Calibration Period
For security reasons and ease of meter operation, any or all menu items may be disabled or "locked out" so that they are no longer accessible from the front panel. Each function to be disabled can be set to "1" under menu headers Loc 1-4, while each function to be enabled can be set to "0." Access to the menu headers Loc 1-4 can in turn be locked out by installing a hardware jumper on the power supply board. With the jumper installed, the operator only has access to previously enabled menu items, not to the menu headers Loc 1-4 and hence not to the menu items below. With the jumper removed, the operator has access to menu headers Loc 1-4 and hence to the menu items below.
SETTING HARDWARE LOCKOUT JUMPER
To access the lockout jumper, remove the rear panel per Section 6 and locate jumper “a” in the lower portion of the power supply board next to the input connectors (see figure above).
SETTING SOFTWARE LOCKOUTS
When setting up the meter, it may be necessary to enable specific menu items by setting the corres­ponding lockout digit to 0. Be sure to reset the lockout digit to "1" if you do not want the menu item to be changed by an operator.
Loc 1 Loc 2 Loc 3 Loc 4
Press the MENU key until Loc 1, Loc 2 Loc 3 or Loc 4 is displayed. Note: Hardware lockout jumper a
must be removed (see above).
11111
Press the PEAK to display the lockout status, con-
sisting of 1’s and 0’s. The left digit will flash. Press
again to step to the next digit, which will flash.
00000
12345 Press RESET to set the flashing digit to "0" to
enable the menu item or to "1" to disable. Press MENU to enter. See the table to the right for list of menu items that can be enabled or disabled.
- 12 -
9. TEDS SIGNAL CONDITIONER IN TEDS MODE
9.1 INTRODUCTION
The DPM-3 is a TEDS IEEE 1451.4 Plug and Play Smart Load Cell Meter. TEDS, or Transducer Electronic Data Sheet, is a set of electronic data in a standardized format defined within the IEEE 1451.4 standard that is stored in an EEPROM. This data specifies what type of sensor is present, describes its interface, and gives technical information such as sensitivity, bridge type, excitation, etc. The DPM-3 will automatically detect when a TEDS IEEE 1451.4 compliant Load Cell / Torque Sensor is connected to it, each and every time, by a built in EEPROM detector. Once a TEDS Sensor has been detected, the DPM-3 displays a front panel TEDS indicator light, reads the EEPROM and stores the information in memory and performs an automatic configuration. The automatic system configuration function performs all steps needed to calibrate the TEDS IEEE 1451.4 compliant Load Cell / Torque Sensor and DPM-3 as a system, including selecting the correct sensor excitation. Using the DPM-3 with a TEDS IEEE 1451.4 compliant Load Cell / Torque Sensor is as easy as plugging a mouse into a computer, making it a true plug and play experience.
9.2 SCALING When a DPM-3 connected to a TEDS transducer is powered on, the meter reads the data
stored in the transducer. The excitation supply is automatically set to the correct value and the DPM-3 calculates the correct scaling to calibrate the meter and the transducer as a system. The jumper on the signal conditioner board must be set for the 50 mV range. See Section 13.1. If the Maximum Physical Value of the TEDS sensor exceeds 100,000, the calculated scale and offset are divided by 10 and the reading is in 10's of units.
If the full scale of the transducer is 50 pounds, the meter will display 50 at full scale output. If more resolution is desired, the meter decimal point can be set so that the meter will read
50.0, 50.00, or 50.000. In this example, if another unit of measure, such as kilograms, is desired, there are two alternatives.
If the meter has a serial communication board installed, TEDS Reader_Editor software can be used (see Section 24 "Accessories" for information). The electronic data sheet for your transducer can then be read electronically and be displayed on a computer screen. The values can be modified (e.g., to display kilograms instead of pounds), and then written back to the transducer.
The meter Menu item "Tare," digits 4 and 5, can be set to 1 to configure the meter as a non-TEDS transducer type.
9.3 Error Messages If there is a problem with the TEDS connection between the meter and the TEDS trans-
ducer, one of the following error messages will be displayed at power on or when plugging in a TEDS transducer while the power is on:
Err 1 - TEDS data line shorted Err 2 - No Presence pulse from TEDS Err 3 - Improper Presence pulse width Err 4 - TEDS family code in ROM in error Err 5 - TEDS checksum error in the first or second 32 bytes
- 13 -
9.4 Software
MENU KEY
DIGIT SELECT KEY
VALUE SELECT KEY
SEtuP
Meter Setup
_0__0
Power line frequency
0 Noise minimized for 60 Hz 1 Noise minimized for 50 Hz
_0__0
Rear control inputs A & B
True = logic 1 (0V or tied to digital ground)
False = logic 0 (5V or open)
0 A = Reset, B = Meter Hold 1 A = Function Reset B = Peak or Valley Displ. 2 A = Meter Hold B = Peak or Valley Displ. 3 A = Meter Hold B = Tare 4 A = Peak or Valley B = Tare 5 A = Tare B = Reset 6 A = 0, B = 0, decimal point 1= XXXXX
A = 1, B = 0, decimal point 1 = XXXX.X A = 0, B = 1, decimal point 1 = XXX.XX A = 1, B = 1, decimal point 1 = XX.XXX
7 A = 0, B = 0, decimal point 2 = XXXX.X
A = 1, B = 0, decimal point 2 = XXX.XX A = 0, B = 1, decimal point 2 = XX.XXX A = 1, B = 1, decimal point 2 = X.XXX.X
8 A = Function Reset B = Display Blank 9 A = Hold B = Display Blank A A = Peak or Valley B = Display Blank B A = Tare B = Display Blank C A = Valley Display B = Peak Display D A = Tare B = Tare Reset
Both control inputs A and B set to logic 1 for selections 2, 4, A, C = Function Reset. Both control inputs A and B set to logic 1 for selections 0, 1, 3, 5, 8, 9, B, D = Meter Reset.
ConFG
Meter Configuration
_00___
Operation of front panel PEAK button
0 Peak Display (max reading) 1 Valley Display (min reading) 2 Peak (1st push), Valley (2nd push)
_00___
Signal polarity
0 Normal Input Signal Polarity 1 Reverse Signal Polarity
TEDS Reader-Editor Software, P/N DPM-3-TRES, is available from Transducer Tech­niques. This software allows the user to read and edit information stored in the TEDS transducer. See Section 24 "Accessories" for ordering information.
9.5 Meter Setup
When setting up the meter, it may be necessary to enable some of the menu items. See Section 8 “Enabling and Locking Out Menu Items” for the procedure.
- 14 -
MENU KEY
DIGIT SELECT KEY
VALUE SELECT KEY
FiLtr
Filtering
00000
Alarm filtering
0 Alarm unfiltered 1 Alarm filtered
00000
Peak & Valley filtering
0 Peak & Valley unfiltered 1 Peak & Valley filtered
00000
Display filtering
0 Display batch average every 16 readings 1 Display filtered signal
00000
Adaptive filter threshold
0 Low adaptive filter threshold level 1 High adaptive filter threshold level
00000
Input signal filtering.
Can be applied to dis­play, setpoint, analog output, data output.
0 Autofilter 1 Batch average, 16 readings.
Moving averages with equivalent RC time constants:
2 0.08 sec. 3 0.15 sec. 4 0.3 sec. 5 0.6 sec. 6 1.2 sec. 7 2.4 sec. 8 4.8 sec. 9 9.6 sec. A Unfiltered
dEc.Pt
Decimal point selection.
d.dddd
Determines decimal point position & reading resolution.
d.dddd dd.ddd ddd.dd dddd.d ddddd.
.ddddd Press ^ key to shift decimal point.
Decimal point is stored in TEDS memory. Display shows TEds instead of StorE.
Option board dependent menu items
ALSEt. ALS34 dEU1H dEU2H dEU1b dEU2b dEU3H DEU4H DEU3b DEU4b
Menu items related to alarm setup These will only appear if a relay board is detected. If so, please see Section14.
AnSEt. An Lo. An Hi..
Menu items related to analog output setup. These will only appear if an analog output board is detected. If so, see Section 15.
tArE_
Tare & TEDS selections.
00000 Calibration due
indicator.
0 Calibration due indicators enabled 1 Calibration due indicators disabled
00000 Front panel
pushbutton Tare.
0 Front panel tare enabled 1 Front panel tare disabled
00000
Auto-Tare.
0 Tare after meter reset 1 No tare after meter reset
00000 TEDS Plug and
Play indicator.
0 TEDS Plug and Play indicator enabled 1 TEDS Plug and Play Indicator disabled
00000
TEDS Plug and Play.
0 TEDS Plug and Play enabled 1 TEDS Plug and Play disabled
- 15 -
MENU KEY
DIGIT SELECT KEY
VALUE SELECT KEY
Option board dependent menu items
SEr 1. SEr 2. SEr 3. SEr 4. _Addr
Menu items related to serial communications. These will only appear if an RS232, RS485, or USB I/O board is detected. If so, see Section 16.
CAL rt
Cal Reset
The Cal Instrument indicator starts flashing 30 days before instrument calibration is due. To acknowledge and turn off flashing, change the first digit of this menu item from 1 to 0. The Cal Sensor indicator starts flashing 30 days before sensor calibration is due. To acknowledge and turn off flashing, change the second digit of this menu item from 1 to 0.
dAtE
Current date
Use keypad to enter the current date in MM.DD.YY format. This menu item can also appear as an error message when the battery has just been replaced, or in other instances where the instrument date is not correct. The meter remains non-functional until the current date has been entered.
CALd II
Instr cal date
00.00.00 Meter calibration date in MM.DD.YY format. Read-only.
CALdS
Sens cal date
00.00.00 Sensor calibration date in MM.DD.YY format. Read-only.
SEr no
Transducer Serial No.
If the serial number is 6 digits or less, press the Peak key once to display all digits 123456. If the serial number is greater than 6 digits, press Peak once to display the first digits 12–___, and press the Peak key again to
display the remaining digits 345678. Read only.
UnitS_
Units of mea­sure. Read only.
CALInL
Cal initials
_A_ b_C Initials of the person who performed the transducer calibration. Read only. Characters used are:
CALPEr
Cal period
___365 Required transducer calibration interval in days. Read only.
-M – Id-
Bridge ID
brdg-A or brdg-b or brdg-C Identifies whether meter is monitoring
bridge A, bridge B, or bridge C.
- 16 -
Loading...
+ 37 hidden pages