Calibration
We recommend the scale to be calibrated at periodic intervals with the frequency based on usage.
The minimum test weight that can be used is 70% of full-scale capacity.
To adjust the zero point using the F16 zero adjust procedure:
1. Turn the power OFF
2. Put Calibration Switch to the Calibration position.
3. Press the ON key.
4. Scroll to "F 16". Use TARE or ON to move up or down.
5. Press ZERO. The display will momentarily show "C 0" followed by a value. (If you want, press Dot
to go back to parameter number.)
6.
After making sure that there are no weights on the platform, press ZERO to zero out the displayed
value.
7.
Press NET to save the zero point value. The display will show "EndC0"momentarily, and then revert
back up to F16. At this time, proceed to the F17 span calibration to complete indicator calibration.
To adjust the max point using the F17 span adjusts procedure:
1. Scrollto"F 17",thenpress ZERO to enterspancalibration menu.
2. The display will momentarily show "C 1" for the span calibration, followed by a value with one flashing digit. This
value will be zero. Place the test weight on the Scale.
3. Pressing TARE or ON will change the position of the flashing digit.
4. Increase the flashing digit by pressing Dot. Decrease the flashing digit by pressing ZERO.
5. Aftersettingtheexactvalue,press NETto savethevalue.
6. Iftheadjustmentwassuccessful, the displaywillshow"EndC1" momentarily,and thenrevertbackup to F17.
"
Err0
" – The calibrationtestweight or theadjustedkeyed-inweight is largerthanthefullcapacityof
thescale.Change thecalibration testweightorcheck theinput data.
"Err1"– Changethecalibration test weightorchecktheinputdata.
"Err2"– Theinternalresolutionof thescaleis nothighenough toacceptthecalibrationvalue.
Selecta largerparameterfortheSpanGain(F2).
7. Put the Calibration Switch to the normal mode.
8. Check the Scale.