1-1PC Series Front Panel........................................................................................................ 1-1
2-1Location of Calibration Switch............................................................................................ 2-1
2-2Detail of Calibration Switch ................................................................................................ 2-1
5-1Opening up the PC-30 and PC-70 Base Unit...................................................................... 5-1
5-2Opening up the PC-150 Base Unit...................................................................................... 5-3
6-1PC Series Chassis Wiring Diagram.................................................................................... 6-1
B-1Consolidated Controls “Demand” Weigh Data Record........................................................ B-1
B-2Consolidated Controls “Continuous” Weigh Data / Status Record....................................... B-2
B-3Toledo 8213 “Demand” Weigh Data Record....................................................................... B-3
B-4Toledo 8213 “Demand” Status Record ............................................................................... B-3
B-5NCI 3800 / Triner “Demand” Weigh Data Record ............................................................... B-4
B-6NCI 3800 / Triner “Demand” Status Record........................................................................ B-4
B-7NCI 3800 / Triner “Demand” Invalid Command Record ...................................................... B-4
B-8Transcell Technology “Demand” Weigh Data Record......................................................... B-5
B-9Transcell Technology “Demand” Status Record.................................................................. B-5
B-10 Transcell Technology “Continuous” Weigh Data / Status Record........................................ B-6
B-11 Detecto Weigh Data / Status Record.................................................................................. B-7
B-12 Fairbanks “Demand” Weigh Data / Status Record.............................................................. B-8
ii
Page 4
CHAPTER 1: INTRODUCTION
Service and maintenance personnel who desire more technical information than is presented
in the PC Series User’s Guide should use this manual. Any component level troubleshooting
or board replacement should be done by factory authorized personnel only.
The PC Series has three ranges of weight capacities. The PC-30 is a 30 lb (15 kg) scale, the
PC-70 is a 70 lb (30 kg) and the PC-150 is a 150 lb (70 kg) scale.
GROSS
NET
TARE
ZERO
CLASS
lb
kg
Transcell Technology inc.
FIGURE 1-1: PC Series Front Panel
CAPACITY: 70 x 0.05 lb, 30 x 0,02 kg
III
Temp. Range: 0 to 40° C
ZERO
NET
GROSS
n
= 1500
max
TAREPRINT
lb
kg
STABLE
MODEL
PC-70
1-1
Page 5
CHAPTER 2: LEGAL FOR TRADE SEALING
2.1OVERVIEW
The scale can be prepared for commercial use by sealing access to the calibration switch, to
meet the requirements for a legal-for-trade scale.
2.1.1 SEALING THE SCALE
Once calibration has been successfully completed, then the scale can be sealed as
follows:
1. Referring to Figure 2-1, locate the calibration switch on the back of the scale.
2. Replace the calibration switch cover by securing with the two drilled head
screws.
3. Thread a wire security seal through the openings in both drilled head screws.
This prevents access to the calibration switch. Refer to Figure 2-2 for more
detail.
Calibration Switch
Figure 2-1: Location of Calibration Switch
Drilled Head Screws
Figure 2-2: Detail of Calibration Switch Cover Plate with Drilled Head Screws
2-1
Page 6
3-1
CHAPTER 3: GENERAL TROUBLSHOOTING FLOWCHART
START
Does
Scale Turn On
?
NO
YES
Try The Following In Order Of Appearance:
1. Replace AC Adapter.
2. Check the Display Cable.
3. Troubleshoot or Replace Main PCB.
NO
Try The Following In Order Of Appearance:
3. Replace Load Cell.
4. Replace Main PCB.
2. Check Load Cell Wiring.
Does
Scale Weigh
?
YES
NO
Try The Following In Order Of Appearance:
Do the
Keys Work
?
YES
1. Check Display Cable.
NO
Try The Following In Order Of Appearance:
1. Check for proper AC Adapter Voltage .Is the
Scale Stable
?
YES
2. Check environment.
NO
Try The Following In Order Of Appearance:
1. Check for platform obstruction (i.e. ovld stop).Is the
Scale Linear
?
YES
2. Replace load cell if zero offset count too high.
5.1 PC-30 and PC-70
Tools Required:5 mm Allen Wrench
Phillips Head Screwdriver
1. Remove the scale platter (1) and set aside.
2. Using the 5 mm Allen wrench, remove the [2] load cell mounting bolts (2).
3. Lift off the sub-platform assembly (3), and set aside.
4. Remove the [4] mounting screws (5) for the splash plate (4) and set aside.
5. Lift off the splash plate and set aside.
1
2
3
4
5
Figure 5-1: Opening up the PC-30 and PC-70 Base Unit
5-1
Page 12
5.2 PC-150
Tools Required:5 mm Allen Wrench
Phillips Head Screwdriver
1. Remove the scale platter (1) and set aside.
2. Using the 5 mm Allen wrench, remove the [2] load cell mounting bolts (2).
3. Lift off the sub-platform assembly and aluminum spacer (3), and set aside.
4. Remove the [4] mounting screws (5) for the splash plate (4) and set aside.
5. Lift off the splash plate and set aside.
5-2
Page 13
1
2
3
4
5
Figure 5-2: Opening up the PC-150 Base Unit
5-3
Page 14
CHAPTER 6: INTERNAL WIRING DIAGRAM
6-1
RED
GRN
BLK
WHT
3
5
2
Load Cell
DC Power Jack
Female DSUB9
JP2
J1
–
+
Main / ADC Board Assy
GND
SCALE RXD
SCALE TXD
RED
BLK
RED
Setup Switch
+
On/Off Switch
1 2
+
LED
RED
GRN
BLK
WHT
1
2
3
4
E+
S+
E–
S–
J8
2
1
Ferrite Bead Core
Ferrite Bead Core
J3
1 2 3 4
J2
Female 14-PIN CENTRONICS
TO REMOTE
DISPLAY/ KEY
UNIT (PC-30 &
1 2 3 4 5 6 7 8 9 10
10
9
8
7
6
5
4
3
2
1
PC-70 ONLY)
Display/Key Board Assy (PC-150 only)
Figure 6-1: PC Series Chassis Wiring Diagram
Page 15
CHAPTER 7: THEORY OF OPERATION
7.1THEORY OF OPERATION OVERVIEW
Refer to Appendix C for reference designators.
7.1.1 MAIN / ADC BOARD ASSEMBLY
Power to the scale is supplied by a 12 VDC, 500 mA adapter. Regulator U1 supplies
+5 VDC to the digital components. Regulator U8 supplies + 10 VDC to the analog
components including load cell excitation. Switching regulator V5 inverts the + 10
VDC to – 10 VDC to act as the negative supply rail. Regulators V4 and V6 further
regulate to ± 8 VDC for the A/D converter IC, U9.
Op-amp U11 forms a high impedance bridge amplifier for the load cell with a fixed
gain of 100. Op-amp U10:B forms a low pass filter to prevent unwanted noise from
entering the circuit. The other half of U10 (U10:A) provides a buffered reference
voltage for op-amp U11. The A/D converter, U1, is a dual slope type converter.
Timing is controlled by the microcontroller.
Microcontroller U2 executes several functions including A/D fetch, digital signal
conditioning, key scan, serial communication, and display output. EEPROM U5
contains calibration data as well as the setup parameters. IC U7 is the EPROM,
which contains program memory. The network around U4 boosts the MCU’s UART
signal to RS-232 level for serial communication with a host device.
7.1.2 DISPLAY/KEY BOARD ASSEMBLY
This circuit contains footprints for 5 keys, display driver (U4) and the LED Display
(U1-U3).
7-1
Page 16
APPENDIX A: TECHNICAL SPECIFICATIONS
ANALOG SPECIFICATIONS
Full Scale Input Signal16 mV, including dead load
Internal ResolutionApproximately 150,000 counts
Displayed ResolutionUp to 3,500 graduations max
Measurement Rate10 Meas/sec, nominal
System LinearityWithin 0.02% of FS
Calibration MethodSoftware Calibration, with long term storage in EEPROM
Excitation Voltage+10 VDC
RFI ProtectionSignal lines protected by LC Low Pass Filters
Excitation & Power lines protected by EMI Chokes
DIGITAL SPECIFICATIONS
MicrocontrollerIntel 80C32
Program Memory:
EEPROM:
32K x 8, external to µC
64 x 16, external to µC
SERIAL COMMUNICATIONS
RS-232Full Duplex, Selectable Baud, Data Bits and Parity
OPERATOR INTERFACE
Display6-Digit, 0.56” 7-segment LED Display
Keyboard5-key Tactile Keypad
POWER
AC Adapter12 VDC, 500mA
ENVIRONMENTAL
Operating Temperature
Storage Temperature
41°F (0°C) to 95°F (40° C)
-4°F (-20°C) to 167°F (+75°C)
Operating RH20% to 95%, non-condensing
Storage RH20% to 98%
PHYSICAL DIMENSIONS:
DC POWER
ON/OFF
SWITCH
JACK
CALIBRATION
SWITCH COVER
13.0 in [330.2 mm]
RS-232
PORT
ID TAG
REMOTE
DISPLAY
CONNECTOR
(WP-30 & 70 ONLY)
13.0 in [330.2 mm]
4.4 in [112 mm]
PC-150 Only
10.2 in [258.0 mm]
PC-30/70
A-1
Page 17
APPENDIX B: SERIAL DATA FORMATS INFORMATION
B.1SERIAL DATA FORMATS
B.1.1 SERIAL DATA FORMATS OVERVIEW
The COM1 serial port is a full duplex RS-232 port designed for connection to a serial
printer, computer, or remote display. Consult the PC Series Operation Manual for
information on pinouts and cable diagrams.
Through the User Menu, you may choose one of the available SDF’s (Serial Data Formats).
These serial data formats emulate certain makes of scales and make up the Application
Layer. From a Data Link Layer perspective, there are two types: Demand and Continuous.
Some SDF’s can be used with both types but most can be used with only one type.
For Demand types, the scale acts as the client and requires a host (usually a PC). The host
issues a command (typically a single ASCII character) and the scale responds with a weigh
data record, a status record or no record depending on the specific SDF chosen.
NOTE: Use Demand types for interfacing to a serial printer. In this case, data output can be
initiated by pressing the PRINT key on the front panel of the scale. (Host is not required).
For Continuous types, host control is not required – but not restricted either. In this type, the
scale automatically transmits a weigh data / status record out of the serial port at the end of
each display update. Continuous types are used to interface to computers, scoreboards,
and other remote devices requiring constant data updating. Host commands, although not
required, are accepted and acted upon.
B.1.2 CONSOLIDATED CONTROLS DEMAND FORMAT
Figure B-1 shows the weigh data record sent to the host when the scale receives a Print
(“P”) command. The same weigh data record is sent out of the serial port if the PRINT key
is pressed on the scale’s front panel. If the weigh data is invalid, the scale does not
respond. See section below for more information.
<STX> <POL> xxxxx.xx<LB/KG><GR/NT> <CR> <LF>
Start
Transmission
Polarity:
<SP> = Positive
"–" = Negative
Weight Data
<SP><SP>
SpaceSpace
Units:
lb = pound
kg = kilogram
FIGURE B-1. Consolidated Controls “Demand” Weigh Data Record
B.1.2.1 RECOGNIZED HOST COMMANDS
These commands are issued from the host to the scale and are single ASCII
characters.
“P” - This command is sent to the scale to print the indicated display. The scale will
not respond if the scale is in motion, positive overload or negative overload.
Gross/Net:
GR = Gross
NT = Net
Carriage
Return
Line
Feed
Page B-1
Page 18
“Z” - This command is sent to the scale to zero the scale. The scale will disregard
this command if the scale is in motion, positive overload or negative
overload. The scale will also disregard this command if it is not in gross mode
or within the zero reset range. Even if successful, the host receives no
response from the scale. The host must issue a “P” command to determine if
the command was processed by the scale.
“T” - This command is sent to the scale to tare the scale. The scale will disregard
this command if the scale is in motion, positive overload or negative
overload. The scale will also disregard this command if it displaying a
negative gross value. Even if successful, the host receives no response from
the scale. The host must issue a “P” command to determine if the command
was processed by the scale.
“G” - This command is sent to the scale to revert to gross mode. The scale will
disregard this command if the scale is in motion, positive overload or
negative overload. The scale will also disregard this command if it is not in
net mode. Even if successful, the host receives no response from the scale.
The host must issue a “P” command to determine if the command was
processed by the scale.
“N” - This command is sent to the scale to revert to net. The scale will disregard
this command if the scale is in motion, positive overload or negative
overload. The scale will also disregard this command if it is not in gross mode
or a tare has yet to be established. Even if successful, the host receives no
response from the scale. The host must issue a “P” command to determine if
the command was processed by the scale.
“C” - This command is sent to the scale to toggle among the configured units. Even
if successful, the host receives no response from the scale. The host must
issue a “P” command to determine if the command was processed by the
scale.
B.1.3 CONSOLIDATED CONTROLS CONTINUOUS FORMAT
Figure B-2 shows the weigh data / status record automatically transmitted by the scale at
the end of each display update. The same host commands listed in Section B.1.2.1 are
accepted and acted upon by the scale.
<STX> <POL> xxxxx.xx<L/K> <G/N><STAT> <CR> <LF>
Start
Transmission
Polarity:
<SP> = Positive
"–" = Negative
Weight Data
Units:
L = pound
K = kilogram
Gross/Net:
G = Gross
N = Net
Status:
<SP> = Valid
M = Motion
O = Over/under range
Carriage
Return
Line
Feed
FIGURE B-2. Consolidated Controls “Continuous” Weigh Data / Status Record
Page B-2
Page 19
B.1.4TOLEDO 8213 DEMAND FORMAT
Figure B-3 shows the weigh data record sent to the host when the scale receives a Print
(“W”) command. The same weigh data record is sent out of the serial port if the PRINT key
is pressed on the scale’s front panel. If the weigh data is invalid, the scale responds with a
status record instead – as shown in Figure B-4. See section below for more information.
NOTE: Since this SDF (Serial Data Format) does not support units of measure. It is
recommended that lb/kg conversion be disabled.
<STX> XXX.XX<CR>
Start
Transmission
Weight Data
FIGURE B-3. Toledo 8213 “Demand” Weigh Data Record
<STX> <?> <SS> <CR>
Start
Transmission
Invalid Weight
Status Byte: ( "1" = "True")
See Table
Carriage
Return
FIGURE B-4. Toledo 8213 “Demand” Status Record
B.1.4.1 RECOGNIZED HOST COMMANDS
These commands are issued from the host to the scale and are single ASCII
characters.
Carriage
Return
STATUS BYTE TABLE
BitFunction
B7Parity bit
B6
Always "1"
Always "1"B5
Center of zero
B4
Outside zero capture range
B3
Under zero
B2
B1B0Out of range
Motion
“W” - This command is sent to the scale to print the indicated display. The scale
will respond with the status record if the weight is less than zero or the scale
is in motion, positive overload or negative overload. Otherwise, the scale
responds with the weigh data record.
“H” – Same as the “W” command except that the scale outputs an extra character
at the beginning and end of the weigh data record.
“Z” - This command is sent to the scale to zero the scale. The scale will disregard
this command if the scale is in motion, positive overload or negative
overload. The scale will also disregard this command if it is not in gross mode
or within the zero reset range. Whether the command was successful or not,
the scale sends a status record to the host.
Page B-3
Page 20
“A” - This command is sent to the scale to request an acknowledgement. This
acknowledgement consists of two characters: <STX> and <CR>.
B.1.5 NCI 3800 / TRINER DEMAND FORMAT
Figure B-5 shows the weigh data record sent to the host when the scale receives a Print
(“W”<CR>) command. The same weigh data record is sent out of the serial port if the
PRINT key is pressed on the scale’s front panel. The status record is shown in Figure B-6. If
an invalid host command is received, the scale responds with the invalid command record
shown in Figure B-7. See section below for more information.
<LF> (1) <xx.xxx> (2)<HSB>
Line
Feed
Weight Data
(1) Insert <SP> for Triner.
(2) Insert <SP> for NCI.
(3) Use lower case for NCI.
(4) Insert <LF> for NCI.
<LB/KG>
Units (3):
LB = pound
KG = kilogram
<CR> (4)
Carriage
Return
High Status Byte:
See Table
Low Status Byte:
See Table
FIGURE B-5. NCI 3800 / Triner “Demand” Weigh Data Record
<LF><HSB>
S
Line
Feed
High Status Byte:
Same as "W"
<LSB>
<CR>
Carriage
Return
Low Status Byte:
See as "W"
<ETX>
End of
Transmission
<LSB>
<ETX>
End of
Transmission
Line
Feed
High Status Byte Table
("1" = "True")
BitFunction
B3
B2
B1B0Not at zero
Low Status Byte Table
("1" = "True")
BitFunction
Not Used
B3
B2
Not Used
B1B0Over capacity
Under capacity
<LF>?<CR>
Carriage
Return
Not Used
Not Used
Motion
FIGURE B-6. NCI 3800 / Triner “Demand” Status
Record
B.1.5.1 RECOGNIZED HOST COMMANDS
These commands are issued from the host to the scale and consist of two ASCII
characters.
“W”<CR> - This command is sent to the scale to print the indicated display. The
FIGURE B-7. NCI 3800 / Triner “Demand”
Invalid Command Record
host should check the status bytes to determine if the scale is in
motion, positive overload or negative overload.
Page B-4
Page 21
“Z”<CR> - This command is sent to the scale to zero the scale. The scale will
disregard this command if the scale is in motion, positive overload or
negative overload. The scale will also disregard this command if it is not
in gross mode or within the zero reset range. Even if successful, the host
receives no response from the scale. The host must issue an “S”<CR>
command to determine if the command was processed by the scale.
“S”<CR> - This command is sent to the scale to request the status record.
B.1.6 TRANSCELL TECHNOLOGY DEMAND FORMAT
Figure B-8 shows the weigh data record sent to the host when the scale receives a Print
(“P”) command. The same weigh data record is sent out of the serial port if the PRINT key
is pressed on the scale’s front panel. If the weigh data is invalid, the scale sends the status
record shown in Figure B-9. See section below for more information.
<STX> <+/–>XXXX.XXX<UNITS><GR/NT> <CR> <LF>
Start
Transmission
Polarity:
+ = Positive
– = Negative
Weight digits*
<SP><SP>
SpaceSpace
Units:
lb = pound
kg = kilogram
Carriage
Return
Gross/Net:
GR = Gross
NT = Net
* with leading zero suppression
FIGURE B-8. Transcell Technology “Demand” Weigh Data Record
<STX>
Question
Mark
Start
Transmission
?
Status Character 1:
See Continuous Mode
<S2>
End
Transmission
Status Character 2:
See Continuous Mode
<S1><ETX>
FIGURE B-9. Transcell Technology “Demand” Status Record
<ETX>
End
Transmission
Line
Feed
B.1.6.1 RECOGNIZED HOST COMMANDS
These commands are issued from the host to the scale and are single ASCII
characters.
“P” - This command is sent to the scale to print the indicated display. The scale will
respond with the status record instead if the scale is in motion, positive
overload or negative overload.
Page B-5
Page 22
“Z” - This command is sent to the scale to zero the scale. The scale will disregard
this command if the scale is in motion, positive overload or negative
overload. The scale will also disregard this command if it is not in gross mode
or within the zero reset range. Even if successful, the host receives no
response from the scale. The host must issue a “P” command to determine if
the command was processed by the scale.
“T” - This command is sent to the scale to tare the scale. The scale will disregard
this command if the scale is in motion, positive overload or negative
overload. The scale will also disregard this command if it displaying a
negative gross value. Even if successful, the host receives no response from
the scale. The host must issue a “P” command to determine if the command
was processed by the scale.
“G” - This command is sent to the scale to revert to gross mode. The scale will
disregard this command if the scale is in motion, positive overload or
negative overload. The scale will also disregard this command if it is not in
net mode. Even if successful, the host receives no response from the scale.
The host must issue a “P” command to determine if the command was
processed by the scale.
“N” - This command is sent to the scale to revert to net. The scale will disregard
this command if the scale is in motion, positive overload or negative
overload. The scale will also disregard this command if it is not in gross mode
or a tare has yet to be established. Even if successful, the host receives no
response from the scale. The host must issue a “P” command to determine if
the command was processed by the scale.
“C” - This command is sent to the scale to toggle among the configured units. Even
if successful, the host receives no response from the scale. The host must
issue a “P” command to determine if the command was processed by the
scale.
B.1.7 TRANSCELL TECHNOLOGY CONTINUOUS FORMAT
Figure B-10 shows the weigh data / status record automatically transmitted by the scale at
the end of each display update. The same host commands listed in Section B.1.6.1 are
accepted and acted upon by the scale.
<+/–> XXXX.XXX<UNITS><S1><ETX>
<STX>
Line
Feed
Start
Transmission
<LF>
Polarity:
+ = Positive
– = Negative
Weight Data
<G/N>
Gross/Net:
G = Gross
N = Net
Units:
LB = pounds
KG = kilograms
<LF>
Line
Feed
<S2>
End
Transmission
Status Character 2:
0 = valid
1 = Motion
2 = Zero
3 = Motion and Zero
4 = Not Ready
Status Character 1:
0 = valid
1 = Under
2 = Over
FIGURE B-10. Transcell Technology “Continuous” Weigh Data / Status Record
Page B-6
Page 23
B.1.8 DETECTO DEMAND / CONTINUOUS FORMAT
The Detecto SDF (Serial Data Format) can work both as a Demand type and a Continuous
type, depending on the commands received from the host. Upon power up, the scale
defaults to Demand type. The scale disregards the type setting programmed for Mode of
Serial Transmission.
Figure B-11 shows the weigh data / status record sent to the host when the scale receives a
Print (“~”) command. The same weigh data record is sent out of the serial port if the PRINT
key is pressed on the scale’s front panel. The output records are identical when operating as
a Demand type or a Continuous type.
NOTE: Since this SDF (Serial Data Format) does not support units of measure. It is
recommended that lb/kg conversion be disabled.
<STX> <POL> XXXLB<SB>
Start
Transmission
Polarity:
<SP> = Positive
– = Negative
<SP><SP>
Weight digits*
SpaceSpace
XX.X
OZ <SP>
Weight digits*
SpaceChecksum Byte 1
Status Byte:
M = Motion
C = Over
FIGURE B-11. Detecto Weigh Data / Status Record
B.1.8.1 RECOGNIZED HOST COMMANDS
These commands are issued from the host to the scale and are single ASCII
characters.
“~” - This command is sent to the scale to print the indicated display. The host
should check the status bytes to determine if the scale is in motion, positive
overload or negative overload.
“<SO>” - This command is sent to the scale to change the scale to Continuous
type. Even if successful, the host receives no response from the scale.
“<SI>” - This command is sent to the scale to change the scale to Demand type.
Even if successful, the host receives no response from the scale.
<CHK1> <CHK2>
Checksum Byte 2
* with leading zero suppression
<ETX>
End
Transmission
NOTE: You can execute the <SO> and <SI> host commands by hitting Ctrl-N for
<SO> and Ctrl-O for <SI> on your computer’s keyboard.
Page B-7
Page 24
B.1.9 FAIRBANKS 70-2453-4 DEMAND FORMAT
Figure B-12 shows the weigh data / status record sent to the host when the scale receives a
Print (<CR>) command. The same weigh data record is sent out of the serial port if the
PRINT key is pressed on the scale’s front panel.
XXXX.XXX<UNITS><GR/NT>
Weight digits*
<SP><SP>
SpaceSpace
Units:
lb = pounds
kg = kilograms
<SP>
SpaceCarriage
Two Status Bytes:
GR = gross/stable
gr = gross/unstable
NT = net/stable
nt = net/unstable
<CR> <LF>
<SP>
Return
Space
* with leading zero suppression
FIGURE B-12. Fairbanks “Demand” Weigh Data / Status Record
B.1.9.1 RECOGNIZED HOST COMMANDS
These commands are issued from the host to the scale and are single ASCII
characters.
“<CR>” - This command is sent to the scale to print the indicated display. The
scale will not respond if the scale is in positive overload. The host should
check the status bytes to determine if the scale is in motion.
“Z” - This command is sent to the scale to zero the scale. The scale will disregard
this command if the scale is in motion, positive overload or negative
overload. The scale will also disregard this command if it is not in gross mode
or within the zero reset range. Even if successful, the host receives no
response from the scale. The host must issue a <CR> command to determine
if the command was processed by the scale.
<EOT>
End
Transmission
Carriage
Return
“T” - This command is sent to the scale to tare the scale. The scale will disregard
this command if the scale is in motion, positive overload or negative
overload. The scale will also disregard this command if it displaying a
negative gross value. Even if successful, the host receives no response from
the scale. The host must issue a <CR> command to determine if the
command was processed by the scale.
“G” - This command is sent to the scale to revert to gross mode. The scale will
disregard this command if the scale is in motion, positive overload or
negative overload. The scale will also disregard this command if it is not in
net mode. Even if successful, the host receives no response from the scale.
The host must issue a <CR> command to determine if the command was
processed by the scale.
“N” - This command is sent to the scale to revert to net. The scale will disregard
this command if the scale is in motion, positive overload or negative
overload. The scale will also disregard this command if it is not in gross mode
or a tare has yet to be established. Even if successful, the host receives no
response from the scale. The host must issue a <CR> command to determine
if the command was processed by the scale.
Page B-8
Page 25
“C” - This command is sent to the scale to toggle among the configured units. Even
if successful, the host receives no response from the scale. The host must
issue a <CR> command to determine if the command was processed by the
scale.
Page B-9
Page 26
APPENDIX C: CIRCUIT DIAGRAMS
Please see attached schematic diagram sheets.
C-1
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.