This equipment has been tested and found to comply with the limits for a Class A digital device,
pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment.
This equipment generates, uses and can radiate radio frequency energy and, if not installed and
used in accordance with the instructions manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his/her own expense.
Page 3
TABLE OF CONTENTS NEPTUNE 4500
Chapter 1: Introduction To The Neptune 4500 Series Indicator..................................................... 1-1
1-1 Neptune 4500 Series Product Matrix ...................................................................................... 1-1
4-1 Invalid Setup Selections for commercial applications............................................................. 4-3
6-1 Calibration Value Table........................................................................................................... 6-3
7-1 Neptune 4500 Series Annunciator Definitions........................................................................ 7-2
C-1 Minimum Recommended Span Gain Table............................................................................ C-2
Neptune 4500
iii
Page 6
CHAPTER 1: INTRODUCTION TO THE Neptune 4500 SERIES DIGITAL INDICATORS
The Neptune 4500 Series Digital Indicator is a general purpose, industrial grade weight indicator.
All models operate identically, can readout up to 50,000 display divisions and can supply enough
current for up to 4-350Ω load cells. All setup parameters may be entered via the front panel keys,
including calibration.
If your Neptune Model 4500 Series Digital Indicator is part of a complete floor scale or has been
installed for you, you may skip to Chapter 7 for operating instructions. Prior to using the indicator,
please read this chapter carefully and completely. Store the manual in a safe and convenient
place so it will be available if you have questions concerning the operation of the scale.
If you are an installer, the indicator's installation and wiring instructions are found in Chapter 2.
The indicator contains two main setup menus: The Setup (“F”) menu, which configures the
indicator to your weigh platform and the User (“A”) menu, which configures the serial
communication port and enables some user options. Chapter 3 gives an overview and explains
how to use the two front panel keys to maneuver and save settings in both menus. Chapters 4
and 5 explain the Setup and User Menu options, respectively. Chapter 6 covers system
calibration. Prior to installing the indicator, please read this manual carefully and completely.
Store the manual in a safe and convenient place so it will be available if you have questions
concerning the setup and operation of the scale.
MODEL DISPLAY
TYPE
ENCLOSURE
TYPE
POWER SOURCE
Neptune 4500 LED Stainless Steel 110/220 VAC, 50/60 Hz
TABLE 1-1: Neptune Series Product Matrix
FIGURE 1-1: Neptune 4500 Front Panel
Neptune 4500 Page 1-1
Page 7
CHAPTER 2: INSTALLATION
2.1 STAINLESS STEEL ENCLOSURE
For indicators contained in a stainless steel enclosure, the rear cover must first be removed to
make the appropriate connections to the weigh platform, printer, remote display and power supply.
o remove the rear cover, simply remove the screws that secure it to the enclosure and set aside. T
NOTE: On earlier units, the rear cover must remain off to access the Setup Menu and calibration
procedures.
J4
J6
J3
J8
S- SHE+ S+ E-
J1
P+ P-
TXD RXD GND
JP2
+5
CPU
EPROM
Figure 2-1: Neptune 4500 Main Circuit Board Overview
2.1.1 CONNECTING THE WEIGH PLATFORM
1. Connect your shielded load cell cable (not included) to the appropriate terminal on the
main board. Connection assignments for the Load Cell Terminals are shown in Figure
2-6.
Figure 2-2: Connection assignments for the Load Cell Terminal
Neptune 4500 Page 2-1
Page 8
2.1.2 CONNECTING THE SERIAL PRINTER, REMOTE DISPLAY OR COMPUTER
The Neptune 4500 Series indicator comes standard with one full duplex RS-232 serial port,
designed for connection to either a PC or a serial printer. The same port may be also used
as a simplex, RS-232 port designed for connection to a remote display.
Connection assignments for all serial communication terminals are shown in Figure 2-7.
NOTE: Do not connect any RS-232 equipment to the “+5V” terminal (not shown).
1. Connect your serial printer, remote display or computer communication cable (not
included) to the appropriate terminal on the main board.
Figure 2-3: Connection assignments for the serial communication terminal
2.1.3 CONNECTING THE POWER SUPPLY
The Neptune 4500 indicator ships with an AC line cord attached to the indicator. Simply plug
the unit into a standard wall outlet.
1. Simply plug the AC adapter into the indicator’s DC Power Jack first, and then plug into a
standard wall outlet. Make sure that the AC voltage appearing at the wall outlet matches the input voltage marked on the AC adapter.
Neptune 4500 Page 2-2
Page 9
CHAPTER 3: CONFIGURATION
3.1 CONFIGURATION OVERVIEW
The indicator contains two main setup menus: The Setup (“F”) menu, which configures the indicator
to your weigh platform and the User (“A”) menu, which configures the serial communication port
and enables some user options. The Setup and User menus consist of several menu selections, each with
its own sub-menu of choices.
To set up the indicator, you must first enter the appropriate menu mode. Once there, the front panel keys
become directional navigators to move around in the menus, and one key is used to save or SET the
selections.
3.2 SETUP (“F”) MENU
3.2.1 ENTERING THE SETUP MENU – STAINLESS STEEL ENCLOSURE
1. Power off the indicator.
2. Locate the slide switch on the rear cover and move it to the right.
NOTE: A metal plate held on by two drilled-head screws may conceal the slide switch.
3. Power on the indicator. The indicator shows ” F 1” to indicate that you are in Setup
Menu mode.
Note 1: Access to the back cover is inhibited if the indicator has been sealed for commercial
use. For more information, please refer to Chapter 8.
3.2.2 NAVIGATING IN THE SETUP MENU
Use the directional keys shown in Figure 3-1 to move around in the Setup Menu Chart
shown in Figure 3-2 on the following page.
1. To move to a new “F” heading, use the ZERO (left) key to move in the Setup Menu
Chart.
2. To move to the selection level, press the UNIT/SET (down) key once. The current
saved selection is shown.
3. To view the available selections for the current “F” heading, use the ZERO (left) key to
move through the selection field.
4. To save/exit a new selection, press the UNIT/SET key .To exit without saving, press
the UNITS (up) key to return to the current “F” heading.
5. Repeat Steps 1 through 4 until the Setup Menu is programmed.
Neptune 4500 Page 3-1
Page 10
Figure 3-1: Setup Menu Key Assignments
Neptune 4500 Page 3-2
Page 11
Figure 3-2: Setup Menu Chart
3.2.3 NOTES ON THE SETUP MENU
1. There is an F21 sub-menu present that is for FACTORY USE ONLY!
2. Detailed descriptions of the setup menu parameters can be found in Chapter 4 of this
manual.
3. The User (“A”) menu sub-menus appear when scrolling left or right from the “F” menu.
4. Some selections shown are not available on all models in the series. Likewise,
additional selections may be available on other models in the series. See Chapter 4
for more information.
3.2.4 EXITING THE SETUP MENU – STAINLESS STEEL ENCLOSURE
1. Power off the indicator.
2. Move the slide switch on the rear cover back to the left.
3. Power on the indicator. The display will go through a digit check, then settle into Normal
Operating mode. All front panel keys will now return to their normal mode of operation.
3.3 USER (“A”) MENU
3.3.1 ENTERING THE USER MENU
1. Enter the Setup (“F”) menu by following the directions in Section 3.2.1 or 3.2.2.
2. Use the ZERO directional keys shown in Figure 3-3 to move right or left in the Setup
(“F”) menu until the indicator shows ” A 1”.
3.3.2 NAVIGATING IN THE USER MENU
Use the directional keys shown in Figure 3-3 to move around in the User Menu Chart shown
in Figure 3-4 on the following page.
1. To move to a new “A” heading, use the ZERO (left) or PRINT (right) key to move right
or left in the User Menu Chart.
2. To move to the selection level, press the ZERO (down) key once. The current saved
selection is shown.
3. To view the available selections for the current “A” heading, use the ZERO (left) or
PRINT (right) key to move through the selection field.
4 . To save/exit a new selection, press the UNIT/SET key .
5. Repeat Steps 2 through 5 until the User Menu is programmed.
Neptune 4500 Page 3-3
Page 12
Figure 3-3: User Menu Key Assignments
3.3.3 NOTES ON THE USER MENU
1 . Detailed descriptions of the user menu parameters can be found in Chapter 5 of this manual.
2. Some selections shown are not available on all models in the series. Likewise, additional
selections may be available on other models in the series. See Chapter 5 for more information.
3.3.4 EXITING THE USER MENU
1. Exit the User (“A”) menu by sliding the setup switch to its original position. Replace the metal
cover. The display will go through a digit check, then settle into Normal Operating mode. All front
panel keys will now return to their normal mode of operation.
Figure 3-4: User Menu Chart
Neptune 4500 Page 3-4
Page 13
CHAPTER 4: SETUP MENU DESCRIPTIONS AND PROCEDURES
4.1 SETUP MENU DESCRIPTIONS
This section provides more detailed descriptions of the selections found in the Setup Menu Chart.
Factory-set defaults are shown in bold with a checkmark (√).
Table 4-1 shows the selections that are not
allowed for “Legal-for-Trade” applications:
NAME/CODE DESCRIPTION CODE/VALUE
F1
Graduations
F2
Span Gain
F3
Zero Track
Band
F4
Zero Range
F5
Motion Band
Specifies number of full-scale graduations. Value should be consistent with legal requirements and environmental limits on the useful
system resolution.
Span Gain is related to A/D integration time. The larger the span
gain, the higher the internal resolution, but the slower the update
speed. Note that the scale must be re-calibrated whenever this parameter is altered. See Appendix C for more information.
Selects the range within which the scale will automatically zero. Note
that the scale must be in standstill to automatically zero. Selections
are in Display Divisions.
Selects the range within which the scale may be zeroed. Note that the
indicator must be in standstill to zero the scale.
Sets the level at which motion is detected by comparing the present
display update with the previous one. If motion is not detected for two
seconds or more, scale is in standstill and can process a Print or Zero
command. Maximum value varies depending on local regulations.
Averages weight readings to produce higher stability. The higher the
filter setting, the greater the stability but the slower the indicator’s
response time. Choose 8 unless a very fast response is needed.
Neptune 4500
1 2
4 8√
Page 4-1
Page 14
F7
Overload Limit
F8
Calib. Unit
F9
Display
Divisions
F10
Decimal Pt.
F16
Zero Calibration
F17
Span Calibration
F18
View Calibration
Selects the desired formula which determines the point at which the
indicator shows overload. All selections are based on the primary unit
selected in F8.
"FS" = Full scale in primary units.
Selects the primary base unit to be used in the calibration process.
Also the default unit for normal operation.
"1" = primary unit is lb. "2" = primary unit is in kg.
Determines the desired weight increments. Value should be consistent with legal requirements.
Determines location of the decimal point.
Places indicator into the zero calibration routine. Scrolling down with
the UNIT/SET key one level begins the procedure.
Places indicator into the span calibration routine. Scrolling down with
the UNIT/SET key one level begins the procedure. NOTE – press and
hold the UNIT/SET for three seconds for effect.
Actuates the function that allows you to view both the zero and span
calibration value. The values displayed in this function are valid only
after Calibration (F16 & F17) has been successfully completed.
Scrolling down with the UNIT/SET key one level begins the procedure.
FS
FS + 2%√
FS + 1d
FS + 9d
1√
2
1√
2
5
0√ 0.0
0.00 0.000
0.0000 00
Press UNIT/SET
key to begin sequence
Press UNIT/SET
key to begin sequence
Press UNIT/SET
key to begin sequence
F19
Key-in Zero
F20
Key-in Span
F21
Factory Reset
F24
Piece Count
Enable
Allows you to key-in known zero calibration value in case of memory
loss in the field. Scrolling down with the UNIT/SET key one level begins the procedure.
Allows you to key-in a known span calibration value in case of memory loss in the field. Scrolling down with the UNIT/SET key one level
begins the procedure.
This sub-menu will reset all parameters in the “F” and “A” menu to the
default settings. USE WITH CAUTION!
Used to enable or disable the piece count feature. This feature must
be disabled for commercial applications.
"0" = Disable piece count "1" = Enable piece count
Press UNIT/SET
key to begin sequence
Press UNIT/SET
key to begin sequence
Press the
UNIT/SET key
twice to execute.
0√
1
Neptune 4500
Page 4-2
Page 15
SUB-MENU TITLE SELECTIONS
F1 Graduations 6,000 8,000
10,000 12,000
20,000 30,000
40,000 50,000
F3 Zero Tracking
Band (SAZSM)
0d
5d
F5 Motion Band 3d 5d
10d 15 d
20d 30d
40d 50d
F6 Digital Filter 1
2
4
Table 4-1: Invalid Setup Menu selections for commercial applications
Neptune 4500
Page 4-3
Page 16
CHAPTER 5: USER MENU DESCRIPTIONS AND PROCEDURES
5.1 USER MENU DESCRIPTIONS
This section provides more detailed descriptions of the selections found in the User Menu Chart.
Factory-set defaults are shown in bold with a checkmark (√).
NAME/CODE DESCRIPTION CODE/VALUE
A1
Baud Rate
A2
Data Bits and
Parity
A3
Mode of Serial
Transmission
A4
Display Check
A5
Disable the
lb/kg Key
A6
Serial Port
Mode
Selects the baud rate for data transmission through the serial port. 1200 2400
4800 9600√
Selects the number of data bits and parity of serial transmission.
"8n" = 8 data bits with no parity bit and one stop bit
"7O" = 7 data bits with odd parity bit and one stop bit
"7E" = 7 data bits with even parity bit and one stop bit
"7n" = 7 data bits with no parity bit and two stop bits
Selects when data will be sent out of the serial port to a printer or
computer:
"C" = Continuous mode; send data continuously
"d" = Demand mode; send data when a PRINT command is issued
from the printer, computer, or indicator.
Actuates the function that illuminates all digit segments, decimal
points, and LCD annunciators in a test sequence. Pressing the
UNIT/SET key to scroll down one level begins the test sequence.
Allows the lb/kg key to be disabled so that an operator cannot
accidentally press the key and change the displayed units.
"0" = Disable the lb/kg key "1" = Enable the lb/kg key
Selects the mode of the RS-232 serial port: Refer to Appendix B for
more information.
"0" = Full Duplex Mode
"1" = Print Ticket Mode
8n√
7O
7E
7n
C
d√
Press UNIT/SET
key to begin
sequence
0
1√
0√
1
A7
ID No. Enable
A8
ID No. Entry
A9
No. of Line
Feeds
Allows the ID number to be disabled in the Print Ticket mode. Valid
only when A6 is set to “1”.
"0" = Disable the ID No. "1" = Enable the ID No.
Actuates the function that allows entry of a new ID No. Valid only
when A6 is set to “1”. Pressing the UNIT/SET key to scroll down one
level begins the sequence.
Actuates the function that allows entry of the desired number of line
feeds to be printed in Print Ticket Mode. Valid only when A6 is set to
“1”. Pressing the UNIT/SET key to scroll down one level begins the
sequence.
0√
1
123456√
0 - 99
5√
Neptune 4500 Page 5-1
Page 17
5.2 USER MENU PROCEDURES
This section provides instructions for all of the User Menu procedures.
5.2.1 ID Number Entry (A8)
1. While in the User Menu mode, scroll to "A 8", then scroll down once using the
UNIT/SET key to enter the ID Number menu.
2. The display will momentarily show "ID NO", followed by a value with one flashing
digit. This value will be the current ID number value.
3. Use the directional keys to adjust the displayed value to the actual ID Number
value. Increase the flashing digit by pressing the UNITS key. Pressing the ZERO to
change the position of the flashing digit.
4. After setting the exact value, hold the UNIT/SET key to save the ID Number value.
The display will show "END" momentarily, then revert back up to A8.
Figure 5-1: User Menu Key Assignments
Neptune 4500 Page 5-2
Page 18
5.2.2 LF (Line Feeds) Number Entry (A9)
1. While in the User Menu mode, scroll to "A 9", then scroll down once using the
UNIT/SET key to enter the Line Feeds menu.
2. The display will momentarily show "LF", followed by the current line feeds value.
3. Use the directional keys (shown in Figure 5-1 above) to adjust the displayed value
to the actual line feeds value. Increase the flashing digit by pressing the UNITS key.
Pressing the ZERO key will change the position of the flashing digit.
4. After setting the exact value, hold the UNIT/SET key to save the line feeds value.
The display will show "END" momentarily, then revert back up to A9.
Neptune 4500 Page 5-3
Page 19
CHAPTER 6: CALIBRATION
6.1 CALIBRATION OVERVIEW
The indicator is calibrated by following the procedures embedded in F16 (Zero) and F17 (Span) of
the Setup Menu. Each procedure enters a value into the indicator's non-volatile memory - F16 the
zero value (deadweight) and F17 the span value (test weight). The minimum test weight that can be
used is 1% of full-scale capacity. After the two calibration procedures are executed successfully,
you should record both calibration values in Table 6-1 using the F18 View procedure.
In the unlikely event that either value is lost while in the field, the setup menu makes provisions for
re-entering these values via F19 and F20, thus eliminating the need for re-calibration with test
weights.
NOTE: This chapter assumes that the indicator is in Setup (“F”) Menu mode. If the indicator is not
in Setup Menu mode, refer to Chapter 3 for instructions.
6.2 ZERO CALIBRATION (F16)
1. While in the Setup mode, scroll to "F 16", then scroll down once using the UNIT/SET key to
enter zero calibration menu. The display will momentarily show "C 0" followed by a value. This
value is the internal A/D count and can prove useful when trying to troubleshoot setup
problems.
2. After making sure that there are no test weights on the platform, press the ZERO key again to
zero out the displayed value.
3. Press the UNIT/SET key to save the zero point value. The display will show "EndC0"
momentarily, then revert back up to F16. At this time, proceed to the F17 span calibration to
complete indicator calibration.
6.3 SPAN CALIBRATION (F17)
1. While in the Setup mode, scroll to "F 17", then scroll down once using the UNIT/SET key to
enter span calibration menu.
2. The display will momentarily show "C 1" for the span calibration, followed by a value with one
flashing digit. This value will be zero with the Decimal Point parameter selected in F10. Place
the test weight on the weighing mechanism.
3. Use the directional keys (shown in Figure 6-1 below) to adjust the displayed value to the actual
test weight value. Change the flashing digit by pressing the UNIT/SET key. Pressing the ZERO
key will change the position of the flashing digit.
Neptune 4500 Page 6-1
Page 20
Figure 6-1: Setup Menu Key Assignments
4. After setting the exact value, press and hold the UNIT/SET key for three seconds to save the
value.
5. If the calibration was successful, the display will show "EndC1" momentarily, then revert back
up to F17. At this time it is suggested that the calibration values be recorded for future use (see
Section 6.4).
6. If the calibration was not successful, one of the error messages below will appear. Take the
indicated action to correct the problem, then perform a new calibration.
"Err0" - The calibration test weight or the adjusted keyed-in weight is larger than the full
capacity of the scale. Change the calibration test weight or check the input data.
"Err1" - The calibration test weight or the adjusted keyed-in weight is smaller than 1% of the full
capacity of the scale. Change the calibration test weight or check the input data.
"Err2" - The internal resolution of the scale is not high enough to accept the calibration value.
Select a larger parameter for the Span Gain (F2). SEE APPENDIX C FOR MORE
INFORMATION.
Neptune 4500 Page 6-2
Page 21
6.4 VIEW CALIBRATION VALUES (F18)
Note: The values displayed in this procedure are valid only after a successful calibration has been
performed using F16 and F17.
1. While in the Setup mode, scroll to "F 18", then scroll down once using the UNIT/SET key to
enter View calibration menu.
2. The display will momentarily show "CAL 0" followed by a value. This value is the zero calibration value and should be recorded in the table below. Press any key to continue.
3. The display will momentarily show "CAL 1" followed by another value. This value is the span calibration value and should also be recorded in the table below. Press any key to return to
upper level (F18).
INDICATOR ZERO CALIBRATION VALUE SPAN CALIBRATION VALUE
S/N:
Table 6-1: Calibration Value Table
6.5 KEY-IN ZERO CALIBRATION VALUE (F19)
Note: This procedure is intended for emergency use only in the case of non-volatile memory loss.
A valid zero calibration value, obtained from a successful F16 calibration procedure, must
be used.
1. While in the Setup mode, scroll to "F 19", then scroll down once using the UNIT/SET key.
2. The display will momentarily show "CAL 0", followed by a flashing zero. Use the directional
keys (shown in Figure 6-1) to adjust the displayed value to the zero calibration value.
3. After setting the exact value, press and hold the UNIT/SET key to save the value.
4. The display will show "E CAL 0" momentarily, then revert back up to F19.
6.6 KEY-IN SPAN CALIBRATION VALUE (F20)
Note: This procedure is intended for emergency use only in the case of non-volatile memory loss.
A valid span calibration value, obtained from a successful F17 calibration procedure, must
be used.
1. While in the Setup mode, scroll to "F 20", then scroll down once using the UNIT/SET key.
2. The display will momentarily show "CAL 1", followed by a flashing zero. Use the directional
keys (shown in Figure 6-1) to adjust the displayed value to the span calibration value.
3. After setting the exact value, press and hold the UNIT/SET key to save the value.
4. If the entered value is greater than zero, the display will show " E CAL 1" momentarily, then
revert back up to F20. If a value of zero is entered, the indicator will briefly show "Err 5", then
revert back to the screen described above in Step # 2.
Neptune 4500 Page 6-3
Page 22
CHAPTER 7: OPERATION
7 .1 DISPLAY
As mentioned in Chapter 1, some models utilize a 6 digit LCD (Liquid Crystal Display) while others
utilize a 6-digit LED (Light Emitting Diode) display. Typically, LCD’s are used for outdoor applications
while LED’s are used indoors where brightness is needed. Table 7-1 summarizes both types of display annunciators.
7.1.1 LIGHT EMITTING DIODE (LED) DISPLAY
Figure 7-1 shows the display detail of the LED display indicators.
FIGURE 7-1: Neptune 4500 Series LED Display Detail
Neptune 4500 Page 7-1
Page 23
LCD Annun-
ciator
!0" ZERO
N NET
G GROSS
T TARE
Lb, Kg Lb, Kg, Lb-Oz,
LED Annun-
ciator
MEANING
Better known as the “Center of Zero” annunciator, this light is active whenever the displayed weight is within ± 0.25 divisions of true zero.
Indicates that the indicator is displaying net weight.
Indicates that the indicator is displaying gross weight.
Indicates that a tare weight has been established in the system.
Indicates the unit of the displayed weight. PCS stands for “pieces”.
Oz, PCS
STABLE
This light is on whenever the scale is stable.
TABLE 7-1: NEPTUNE 4500 Series Annunciator Definitions
7 .2 KEYBOARD
The keyboard is composed of TWO function keys. Refer to Figures 7-2 for the overall layout and key
locations.
7.2.1 FUNCTION KEYS
Unit/Set – This key toggles the indicator among the available weight units if enabled in the
User (“A”) menu. Available weight units include lb, kg and pieces
nformation. Hold the key for three seconds for value to be accepted. i
FIGURE 7-2: Function Keys Layout
1
. See Chapter 5 for more
1
Newer units only.
Neptune 4500 Page 7-2
Page 24
Zero - This key sets the indicator to display zero provided the following conditions are met:
1. The indicator is displaying Gross weight.
2. The displayed weight is within the zero reset range that is programmed in F4 of the Setup
(“F”) Menu.
3. The scale is not in motion.
4 . The scale is not in overload (see Appendix D for error codes).
1 . The scale is not in overload (see Appendix D for error codes).
7.3 GENERAL SCALE OPERATION
7.3.1 WEIGHING AN ITEM
1. Select the desired weighing unit by pressing the UNIT key until that unit is indicated on
the display.
2. If necessary, press the ZERO key to obtain a weight reading of zero.
3. Place the object to be weighed on the scale’s platter and allow the weight indication to
stabilize. If the item weight exceeds the scale’s weight capacity, it displays “!!!!!!”.
4. Read the weight shown on the display.
7.3.2 ZEROING THE WEIGHT OF AN ITEM
To weigh an item in a container, the weight of that container must first be subtracted from the
overall weight to obtain an accurate weight reading.
1. Select the desired weighing unit by pressing the UNIT key until that unit is indicated on
the display.
2. Place the empty container on the scale’s platter and allow the weight indication to stabilize.
3. Press the ZERO key. The display shows zero weight.
4. Place the material to be weighed in the container and allow the weight indication to stabilize.
5. Read the weight shown on the display.
7.3.3 PIECE COUNTING
IMPORTANT NOTE: The piece counting function cannot be used in commercial (NTEP) ap-
plications.
This mode is used to indicate the number of pieces of an item you have placed on the scale’s
platform and is accessed by pressing the UNITS key. To ensure accuracy, the parts you are
counting must be consistent in weight.
The indicator uses the sampling method to determine the average piece weight (APW) of the
items you wish to count. When sampling items, always count the parts in your hand and
place them on the platform all at once. If the APW of the items is too light or the total weight
of the sample is too light, accuracy cannot be guaranteed. You will get an error message, but
piece counting will still be allowed.
1. If the items you will be counting require a container, you must first tare the container off by pressing
the ZERO key.
Neptune 4500 Page 7-3
Page 25
2. Press the UNITS key until “5 0” is indicated on the display. If the screen does not show “5 0”, press
the ZERO key once. The indicator is prompting you to place five identical items on the platform.
NOTE: If you wish to change the sample number, simply press the UNITS key repeatedly until the
desired sample number appears. Available choices are 5, 10, 20, 50 and 100. If you continue to push
the UNITS key, the indicator will resort back to weighing mode and you must start again from Step 2.
3. Place the sample items on the platform all at once and allow the weight indication to stabilize. Once
this is done, press and hold the UNIT/SET for three seconds. If the sample size is large enough, the
indicator now displays the number of pieces on the platform and the “PCS” annunciator is lit. If not,
the indicator briefly displays “Lo” and automatically increments the sample size. Repeat step 3.
NOTE: If the indicator continues to display “Lo” even after sampling 100 pieces, the unit weight of the
items you wish to count is too light for your scale to process accurately.
4. To exit the piece count mode, press the UNIT/SET key.
NOTE: The APW will NOT remain in scale memory when you exit piece counting mode.
Neptune 4500 Page 7-4
Page 26
CHAPTER 8: LEGAL FOR TRADE SEALING
8.1 STAINLESS STEEL ENCLOSURE
Indicators in the stainless steel enclosure can be sealed for commercial (Legal for Trade)
pplications as follows. a
1. Power off the indicator.
2. On the rear cover of the indicator, locate the calibration switch cover.
3. Thread a wire security seal through both drilled head screws securing the calibration
switch cover as well as the two drilled head screws holding on the rear panel.
Neptune 4500 Page 8-1
Page 27
APPENDIX A: SPECIFICATIONS
ANALOG SPECIFICATIONS
Full Scale Input Signal 30mV, including dead load
Minimum Sensitivity - Non H-44
Minimum Sensitivity - H-44
Input Impedance
Internal Resolution – Neptune 4500
Display Resolution 50,000 display division max
Measurement Rate 10 Meas/sec, nominal
System Linearity Within 0.02% of FS
Calibration Method Software Calibration, with long term storage in EEPROM
Excitation Voltage – Neptune 4500
0.4 µV / grad
1.0 µV / grad
30MΩ, typical
Approximately 260,000 counts
+10 VDC, 4 x 350Ω load cells
SERIAL COMMUNICATIONS
Serial Port Full Duplex, 1200, 2400, 4800, 9600 Baud
8 data bits, no parity, 1 stop bit
7 data bits, odd parity, 1 stop bit
7 data bits, even parity, 1 stop bit
7 data bits, no parity, 2 stop bits
APPENDIX B: SERIAL PORT INFORMATION
B.1 SERIAL PORT MODES
B.1.1 FULL DUPLEX MODE
The Full Duplex Mode provides a Demand serial transmission mode and is selected by
setting A3 to “d” and A6 to “0”. The Demand mode allows control from a host device, usually
a PC. Figure B-1 shows a suggested cable diagram for interface to a PC. Figure B-2 shows
the serial data format for the Demand Mode.
INDICATORPC
TXD 3
S. GND 5
DSUB9DSUB9
FIGURE B-1. Cable Diagram for Indicator to IBM PC
2 RXDRXD 2
3 TXD
5 S. GND
4 DTR
6 DSR
8 CTS
<STX> <POL> xxxxx.xx<LB/KG><GR/NT> <CR> <LF>
Start
Transmission
Polarity:
<SP> = Positive
"–" = Negative
Weight Data
<SP><SP>
SpaceSpace
Units:
lb = pound
kg = kilogram
pc = pieces
pcs = pieces*
cu = cusotm unit
* Available on TI-500E and TI-500E-SS only
Gross/Net:
GR = Gross
NT = Net
Carriage
Return
Line
Feed
FIGURE B-2. Consolidated Controls Demand Mode
Page B-1
Page 29
B.1.1.1 RECOGNIZED HOST COMMANDS
“P” - This command is sent to the indicator to print the indicated display. The
indicator will not respond if the scale is in motion, positive overload or
negative overload.
“Z” - This command is sent to the indicator to zero the scale. The indicator will not
respond if the scale is in motion, positive overload or negative overload. The
indicator will also not respond if it is not in gross mode or within the zero
range specified in F4 of the Setup Menu.
“T” - This command is sent to the indicator to tare the scale. The indicator will not
respond if the scale is in motion, positive overload or negative overload. The
indicator will also not respond if it displaying a negative gross value.
“G” - This command is sent to the indicator to revert to gross mode. The indicator
will not respond if the scale is in motion, positive overload or negative
overload. The indicator will also not respond if it is not in net mode.
“N” - This command is sent to the indicator to revert to net. The indicator will not
respond if the scale is in motion, positive overload or negative overload. The
indicator will also not respond if it is not in gross mode or a tare has yet to be
established.
“C” - This command is sent to the indicator to toggle among the configured units.
B.1.2 PRINT TICKET MODE (NOT AVAILABLE)
The Print Ticket Mode is designed specifically for a serial printer and is selected by setting
A6 to “1”. Figure B-3 shows the fixed format of the print ticket.
For printers with limited buffers, this mode supports DTR pin handshaking. The DTR pin
from the serial printer is wired to the indicator’s RXD pin which then functions as a CTS pin.
Figure B-4 shows a suggested cable diagram for interfacing to a serial printer. Refer to the
printer’s user manual to confirm which pin is the DTR pin.
NOTES:
1. The TARE and NET fields are not printed unless a tare has been established in the
system.
2. The ID number field is not printed if it is disabled in A7 of the User Menu.
FIGURE B-3. Print Ticket FIGURE B-4. Cable Diagram for Indicator to Printer
Page B-2
INDICATORPRINTER
TXD 3
S. GND 5
DSUB9DSUB25
3 RXDCTS 2
20 DTR
7 S. GND
Page 30
B.1.3 SIMPLEX MODE
The Simplex Mode provides a continuous serial transmission mode and is selected by
setting A3 to “C” and A6 to “0”. The Continuous mode is used to interface to computers,
scoreboards, and other remote devices requiring constant data updating. The transmission
occurs at the end of each display update. Figure B-5 shows the serial data format for
Continuous Mode.
<STX> <POL> xxxxx.xx<L/K> <G/N> <STAT> <CR> <LF>
Start
Transmission
Polarity:
<SP> = Positive
"–" = Negative
Weight Data
Units:
L = pound
K = kilogram
P = pieces
PCS = pieces*
C = custom units
Gross/Net:
G = Gross
N = Net
Status:
<SP> = Valid
M = Motion
O = Over/under range
* Available on TI-500E and TI-500E-SS only
Carriage
Return
Line
Feed
FIGURE B-5. Consolidated Controls Continuous Mode
Page B-3
Page 31
APPENDIX C: DETERMINING PROPER SPAN GAIN (F2)
C.1 SPAN GAIN OVERVIEW
The Span Gain parameter found in F2 of the Setup Menu is directly related to the ADC (Analog to
Digital Converter) integration time. This means that the lower the setting, the higher the number of
measurements per second. A span gain setting of 25 produces about 25 to 30 measurements per
second, while a span gain of 200 produces only about 3 or 4 measurements per second.
There is really no wrong setting for span gain – except in two cases. Using a low setting for a high
resolution, low output system could yield instability. Using a high setting in a high output system
could yield non-linearity.
C.2 SETTING THE INITIAL VALUE FOR SPAN GAIN
1. Determine the number of desired external graduations and choose the corresponding value
listed in Table C-1 or Table C-1a under the number closest to your full-scale input range in
millivolts.
2. Enter the Setup Menu and save this number for the Span Gain parameter in F2.
3. Perform a system calibration. If the calibration proves unsuccessful, or you wish to view the
internal counts, proceed to the next set of instructions.
C.3 VIEWING THE INTERNAL COUNTS
1. Enter the zero calibration menu (F16) and follow steps 1 to 3, but do not save the zero point.
2. After pressing ZERO to zero the offset, place the test weight(s) on the platform. The displayed
count is the internal count. If the count remains on zero, check your load cell connections.
3. At full scale, the displayed count should be a minimum of 2 times the desired external
graduations. However, for maximum stability, a ratio of 6:1 or higher is recommended.
4. If the displayed count is large enough, remove the test weight(s), re-zero the indicator if
necessary, and proceed with the calibration. If the displayed number is not large enough,
increase the Span Gain to the next highest choice in the Setup Menu and re-calibrate.
Table C-1: Minimum Recommended (6:1) Span Gain Table
Page C-2
Page 33
APPENDIX D: DISPLAYED ERROR CODES
CODE MODE MEANING / POSSIBLE SOLUTION
!!!!!!
Err 0
Err 1
Err 2
Err 3
Err 4
Err 5
Err 7
Normal Operating
Mode
Span Calibration
Mode (F17)
Span Calibration
Mode (F17)
Span Calibration
Mode (F17)
All Modes Non-volatile memory read error. One or more setup parameters have
All Modes Non-volatile memory write error. Indicator needs service.
Key-in Span
Calibration Mode
(F20)
Initialization No reading from the ADC. Make sure there is a load cell(s) connected to
Gross Overload. A weight greater than the rated capacity has been
applied to the scale. Remove the weight from the platter or try recalibrating the scale. Otherwise, check for a bad load cell connection or
possible load cell damage due to overloading.
Keyed-in weight value is larger than full-scale capacity. Use a smaller test
weight or check keyed-in value.
Keyed-in weight value is less than 1% of full-scale capacity. Use a larger
test weight or check keyed-in value.
There is not enough load cell signal to produce the internal counts
necessary to properly calibrate the scale. First check all load connections.
Use F16 mode to view internal counts. See Appendix C for more
information.
been lost.
You have attempted to enter a zero value for C1. Enter a known
calibration value greater than zero.
the indicator at start-up.
Err 9
Normal Operating
Mode
Span calibration value has been lost. Re-calibrate the scale.
Neptune 4500
Page D-1
Page 34
LIMITED WARRANTY
Seller warrants that the NEPTUNE Series Digital Indicator line will conform to written specifications, drawings, and other
descriptions made by the manufacturer, including any modifications thereof. The Seller warrants the goods against faulty
workmanship and defective materials. If any goods fail to conform to these warranties, Seller will, as its sole and exclusive
liability hereunder, repair or replace such goods if they are returned within the following warranty period:
Twelve (12) months from date of shipment from manufacturer.
These warranties are made upon the following TERMS and CONDITIONS:
This warranty is limited to the original equipment manufactured by TRANSCELL TECHNOLOGY, INC. Items not covered
under this warranty are batteries and normal wear items like connectors, shrouds, front panels and fuses.
For the first sixty (60) days from the date of installation, the warranty covers parts, on-site labor, and limited travel time
and mileage. (3 hrs/150 miles maximum per occurrence).
After sixty (60) days, the warranty covers the cost of replacement parts only.
However, at the discretion and prior approval of TRANSCELL, certain equipment may be returned , freight pre-paid, for
repair, free of any parts or labor charges.
TRANSCELL’s responsibility is confined to repair, replacement or credit of equipment of parts. The warranty does not
extend coverage to labor, material, freight or service charges involved in removal, shipping or reinstallation of
equipment or parts.
CONDITIONS WHICH VOID LIMITED WARRANTY:
This warranty shall not apply to equipment which:
A. Examination of such goods discloses that the nonconformity exists and was caused by accident, misuse, neglect,
alteration, improper installation improper or unauthorized repair, improper testing, or an act of GOD including lightning
and such goods have not been modified, altered, or changed by any person other than the Seller or its duly authorized
repair agents.
Transcell Technology, Inc. will have a reasonable time to repair or replace such goods.
THESE WARRANTIES EXCLUDE ALL OTHER WARRANTIES, EXPRESS OR IMPLIED, ORAL OR WRITTEN,
INCLUDING WITHOUT LIMITATION WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR
PURPOSE. SELLER WILL NOT IN ANY EVENT BE LIABLE FOR INCIDENTAL OR CONSEQUENTIAL DAMAGES.
IN ACCEPTING THIS WARRANTY, THE PURCHASER OR BUYER AGREES TO WAIVE ANY AND ALL OTHER
CLAIMS FOR RIGHT TO WARRANTY FROM TRANSCELL TECHNOLOGY, INC. SHOULD THE SELLER BE OTHER
THAN TRANSCELL TECHNOLOGY, INC., THE BUYER AGREES TO LOOK ONLY TO THE SELLER FOR WARRANTY
CLAIM OR CLAIMS.
No terms, conditions, understanding, or agreements purporting to modify the terms of this warranty shall have any legal
effect unless made in writing and signed by a corporate officer of the Seller.
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.