trafag TMT501, TMT502 Instruction Manual

Trafag Italia Srl Via Cremona 1
20025 Legnano (Mi)
Tel. +39 0331 59 23 97 Fax +39 0331 59 98 15
www.trafagitalia.com info@trafagitalia.com
INSTRUCTIONS MANUAL
Instruction H21017a 05/2017
BIMETAL THERMOMETERS
TYPE: TMT501, TMT502
1 GENERAL WARNINGS
1.1 Instruments measuring range
Verify that measuring range of the instrument is suitable with wor­king conditions of the circuit/plant. Maximum process temperatu­re must remain within the instrument measuring range, identified on dial with two black triangles.
1.2 Admitted over temperatures
Accidental over temperatures are admitted for short periods, provi­ded that they remain within the values indicated in the datasheet.
1.3 Ambient temperature
Instruments are designed to work with an ambient temperature between -20°C/+60°C (-4°F/+140°F). Different ambient tempera­tures could give wrong indications or to damage the instrument. IT is recommended to verify that the instrument installation position is not subject to direct heat sources, both for convention or radia­tion: if this is not possible, please foresee a protective screen.
1.4 Mechanical vibrations
Verify that chosen position for the installation is not subjected to continuous mechanical vibrations, because this will cause accuracy errors and shorter working life of instrument. If this is not possi­ble, it is recommended to use instruments filled with dampening liquid.
2 INSTALLATION
2.1 Process installation
The thermometers must be installed in a thermowell that is sui­table to the line or to the tank where the temperature is to be me­asured. It is also necessary to verify that the thermowell is suitable for the installation of the thermometer. With reference to the figu­re N° 3 is necessary to verify that:
■ The minimum length “S” (see under table 3) guarantee that the
sensing element is totally immerged into dimension of ther­mowell and consequently into the piping, for best sensibility.
■ Bulb diameter d must be maximum 1mm less than the bore
diameter of thermowell.
■ Thermowell thread must be suitable with thermometer thread.
Then insert the instrument bulb till the bottom of thermowell, screw the rotating connection of thermometer on thermowell thread and orient the instrument in a position that permit an easy reading. Then tighten with suitable key the rotating nut and verify the solidity of the mounting.
NOTE: In the event that the tolerance between bulb diameter and bore diameter of thermowell is more than 1mm fill the thermowell bore with suitable transmission fluid such as silicone oil or graphi­te dust : this will improve the instrument performances.
3 Regulation of the instrument in “every angle” version
Instrument case could be oriented before installation to the requi­red inclination, up to a maximum of 90°. To do this it is necessary to unscrew the two “A” screws, place the case in the required inclina­tion and then screw again the two “A” screws. (see Fig. 2)
Fig. 2 – Regulation of case inclination
WARNING Absolutely do not use the instrument case to tighten it on thermowell. This can cause damages to the instrument.
A
2/2
H21017a
BIMETAL THERMOMETERS
TYPE: TMT501, TMT502
4 CALIBRATION AND CHECK
As good practice, is better to make a complete check of the instru­ments twice per year. The calibration check is normally made by using suitable thermo­static bath and test instruments. Check is carried out by checking the accuracy on the main divisions of the dial. Proceed as follow:
■ Insert both the thermometer bulb and the test instrument into
a stable and uniform thermostatic bath.
■ Verify that immersion length of both instrument into the ther-
mostatic bath are the same.
■ Wait for three minutes for a table indication, then verify that
thermometer shows the same temperature of test instrument, with foreseen tolerance.
■ Regulate, if necessary, the instrument indication using the ex-
ternal micrometric adjustment on the back side of the case or, in some cases, using the micrometric adjustment on the pointer.
■ Then remove the thermometer from thermostatic bath.
Figure 3 and table of minimum bulb diameters and length “S” are in mm
5 MAINTENANCE
The instrument doesn’t need any particular maintenance. Do not lubricate the moving parts because this could attract dust and form particles and impurities that could produce malfunctions or accuracy errors.
6 DISMOUNTING AND DISPOSAL
The instrument is essentially made in stainless steel. Therefore, after removing transparent, gasket, plugs and after removing all the process fluid traces from the wetted parts (especially if fluid is dangerous for persons or environment) the instrument could be recycled or scrape. Dispose of instrument components and packaging materials in an environmentally compatible way and in accordance with the coun­try-specific waste disposal regulations.
0... 60 150 135 130
0...80 135 110 105
0...100 115 110 105
0...120 110 100 95
0 ...160
100 90 90Up to
0...500
RANGE °C
Stem Ø
6 or 6.4
Stem Ø 8 or 10
Stem Ø
12
WARNING
In event of accidental breaking of instrument transparent, it is ne­cessary to replace it immediately, taking care of removing all the small pieces from the case.
Documents Instructions H21017 www.trafag.com/H21017
Data sheet H20017 H20018 www.trafag.com/H2001X