1 GENERAL WARNINGS
1.1 Instruments measuring range
Verify that measuring range of the instrument is suitable with working conditions of the circuit/plant. Maximum process temperature must remain within the instrument measuring range, identified
on dial with two black triangles.
1.2 Admitted over temperatures
Accidental over temperatures are admitted for short periods, provided that they remain within the values indicated in the datasheet.
1.3 Ambient temperature
Instruments are designed to work with an ambient temperature
between -20°C/+60°C (-4°F/+140°F). Different ambient temperatures could give wrong indications or to damage the instrument. IT
is recommended to verify that the instrument installation position
is not subject to direct heat sources, both for convention or radiation: if this is not possible, please foresee a protective screen.
1.4 Mechanical vibrations
Verify that chosen position for the installation is not subjected to
continuous mechanical vibrations, because this will cause accuracy
errors and shorter working life of instrument. If this is not possible, it is recommended to use instruments filled with dampening
liquid.
2 INSTALLATION
2.1 Process installation
The thermometers must be installed in a thermowell that is suitable to the line or to the tank where the temperature is to be measured. It is also necessary to verify that the thermowell is suitable
for the installation of the thermometer. With reference to the figure N° 3 is necessary to verify that:
■ The minimum length “S” (see under table 3) guarantee that the
sensing element is totally immerged into dimension of thermowell and consequently into the piping, for best sensibility.
■ Bulb diameter d must be maximum 1mm less than the bore
diameter of thermowell.
■ Thermowell thread must be suitable with thermometer thread.
Then insert the instrument bulb till the bottom of thermowell,
screw the rotating connection of thermometer on thermowell
thread and orient the instrument in a position that permit an easy
reading. Then tighten with suitable key the rotating nut and verify
the solidity of the mounting.
NOTE: In the event that the tolerance between bulb diameter and
bore diameter of thermowell is more than 1mm fill the thermowell
bore with suitable transmission fluid such as silicone oil or graphite dust : this will improve the instrument performances.
3 Regulation of the instrument in
“every angle” version
Instrument case could be oriented before installation to the required inclination, up to a maximum of 90°. To do this it is necessary to
unscrew the two “A” screws, place the case in the required inclination and then screw again the two “A” screws. (see Fig. 2)
Fig. 2 – Regulation of case inclination
WARNING Absolutely do not use the instrument case to tighten it
on thermowell. This can cause damages to the instrument.