TraceableLIVE® Apps must be downloaded to a mobile
device.
ENABLE BLUETOOTH
1. Quick press twice START/STOP to enable Bluetooth, and
Bluetooth LCD symbol appears.
2. Device starts to advertise to be found, and to be
connected. Device Name shown on TraceableGO™ App
discovered list looks like CC653X-xxxx, where “CC653X”
indicates model number and “-xxxx” is last 4 digits of
device serial number.
3. If no connection has been made for ONE MINUTE,
Bluetooth will be disabled to save battery life, and
Bluetooth LCD symbol disappears, or quick press twice
again to disable Bluetooth.
TO VIEW PRE-CONFIGURED SETTINGS
1. Enable Bluetooth on device (see above).
2. Open TraceableGO™ or TraceableLIVE® App on any
bluetooth enabled mobile device.
Note: Mobile device must have Bluetooth enabled to
receive signal. To enable Bluetooth on a mobile device
see mobile device settings.
3. Use TraceableGO™ or TraceableLIVE® App to connect
to device. Once the app is open it will begin searching for
devices to connect to.
Scan for Nearby Devices
A list of available Dataloggers will appear.
Note: TraceableGO Bluetooth Dataloggers are serialized
and each logger name will match the sticker located on the
side of the unit.
4. Once device is selected select Congure.
5. The following parameters are congured through App:
START Mode, STOP Mode, ALARM Enable/Disable,
Celsius/Fahrenheit, Memory Mode, Data Logging Interval,
Alarm Setting.
6. TraceableGO™ comes pre-programmed. To view current
settings, tap on congure once device is selected in App.
7. Device comes pre-congured as shown below:
Device Name
Device Serial Number
Current Battery Status
Current Temperature and/or
Humidity Readings
Immediate Start: Once device is congured it will begin
Datalogging.
Push Button: Press START/STOP Button to begin
Datalogging.
Delayed: Select number of Hours, Minutes and Seconds
when device is to begin Datalogging.
2. STOP Mode
Never: Device will never stop Datalogging.
Push Button: Press START/STOP Button to Stop
Datalogging.
3. Memory Mode
Wrap When Full: Once memory is full, oldest data points
will be overwritten with new data points.
Stop When Full: Device will stop recording when Memory
is Full 64K (65,536) data points, 7.5 months on 5-minute
logging interval.
4. Unit Preferences
°F: Select Fahrenheit
°C: Select Celsius
5. Alarm Setting Alarm Low: TAP ON VALUE TO CHANGE
▪ Temperature: Set LOWEST Temperature before an
alarm is triggered.
▪ Humidity (6537 only): Set LOWEST Relative Humidity
before an alarm is triggered.
6. Alarm Setting High Alarm: TAP ON VALUE TO CHANGE
▪ Temperature: Set HIGHEST Temperature before an
alarm is triggered.
▪ Humidity (6537 only): Set HIGHEST Relative Humidity
value before an alarm is triggered.
7. Alarm Enable/Disable
Enabled: Alarm is on.
Disabled: Alarm is off.
8. Data Logging Interval
Slide to desired Logging Interval.
9. Save Congurations: Saves the current settings.
NOTE: Saving a Conguration will delete all data on
device.
Once device has been congured, device enters
STANDBY Mode.
Date/Time is also set to current to central time
(Automatically updates when connected to mobile
device).
NOTE: To download data either TraceableGO™ or TraceableLIVE
Apps must be downloaded to a mobile device.
ENABLE BLUETOOTH
1. Quick press twice START/STOP to enable Bluetooth, and
Bluetooth LCD symbol appears.
2. Device starts to advertise to be found, and to be
connected. Device Name shown on TraceableGO™ App
discovered list looks like CC653X-xxxx, where “CC653X”
indicates model number and “-xxxx” is last 4 digits of
device serial number.
3. If no connection has been made for ONE MINUTE,
Bluetooth will be disabled to save battery life, and
Bluetooth LCD symbol disappears, or quick press twice
again to disable Bluetooth.
TO DOWNLOAD RECORDED DATA
1. Device must be stopped. Enable Bluetooth on device (see
above).
Note: TraceableGO™ Bluetooth Dataloggers are serialized
and each logger name will match the sticker located on the
side of the unit.
4. Once device is selected select Download Data.
5. App will begin extracting the data from
TraceableGO™ Device.
2. Open TraceableGO™ or TraceableLIVE® App on any
bluetooth enabled mobile device. Note: Mobile device
must have Bluetooth enabled to receive signal. To enable
Bluetooth on a mobile device see mobile device settings.
3. Use TraceableGO™ or TraceableLIVE® App to connect
to device. Once the app is open it will begin searching for
devices to connect to.
Scan for Nearby Devices
A list of available Dataloggers will appear.
6. Once data download is complete, mobile device
will display options how to send PDF le, or CSV le
(TraceableLIVE only). Tap Done and download is complete.
6535: Ambient Range: –20.0 to 70.0°C (–4.0 to 158.0°F)
6536/6538 Probe Range: –50.0 to 70.0°C (–58.0 to 158.0°F)
6539 Probe Range: –90.00 to 100.00°C (–130.00 to 212.00°F)
Resolution: 0.1°C
Accuracy:
6535: ±0.4°C between −10 and 70°C, otherwise ±0.5°C
6536/6538: ±0.3°C
6539: ±0.2°C
RELATIVE HUMIDITY & TEMPERATURE
Temperature—
Ambient Range: –20.0 to 70.0°C (–4.0 to 158.0°F)
Resolution: 0.1°C
Accuracy: ±0.4°C between –10 and 70°C, otherwise ±0.5°C
Relative Humidity—
Ambient Range: 0% to 95%RH, non-condensing
Resolution: 0.1% RH
Accuracy: ±3% RH between 5 to 75%, otherwise ±5% RH
EXTERNAL PROBE
6536 Bullet Probe: Standard plastic probe sensor with cable. Designed for use in air and
liquids, the sensor and cable may be completely immersed. Probe size: 3/16” diameter,
4/5” length; 10 foot cable
6538 Bottle Probe: Designed to simulate the temperature of stored liquids, use
in transportation coolers, refrigerators and freezers. Bottle probes are lled with a
patented nontoxic glycol solution that is GRAS (Generally Recognized As Safe) by the
FDA (Food and Drug Administration) eliminating concerns about incidental contact with
food or drinking water. The included micro-thin probe cable permits refrigerator/freezer
doors to close on it. (Do not immerse bottle probes in liquid). Probe size: 1 x 2-1/2
inches; 10 foot cable.
6539 Stainless-Steel/Platinum Probe: Detachable stainless steel 316 Probe with
Platinum Sensor and 9 feet of cable are supplied with the unit. Probe has a diameter of
1/8-inch, stem length of 6-1/4 inches, overall length of 9 inches.
Battery: 2 AAA Alkaline batteries (3.0V)
Dimension: L x H x D: 3.5 x 2 x 0.79” (89 x 51 x 20 mm)
Battery Level Indication:
Battery levelLCD symbol
≥ 80% (2.78V)
≥ 60% (2.56V), < 80%
Note: Battery level is updated every 5 minutes.
Note: Once battery level drops to below 10%, device may not work properly. Replace
batteries immediately.
Note: When replacing batteries, after removing the old batteries, wait 10 seconds before
inserting new batteries. Otherwise, Bluetooth may not work properly.
MEASUREMENT READING UPDATE FREQUENCY
Temperature and Humidity: 5 seconds;
Note: If a reading is out of operating range, corresponding location on LCD will display
‘--.-‘, and such out of range reading will not trigger alarm.
DATA LOGGING FREQUENCY:
5 min by default, user-adjustable between 1 minute and 12 hours with 1 minute step.
DATA STORAGE CAPACITY
Alarm: Most recent 90 Alarm events
Data: 64K (65536) data points, 7.5 months on 5-minute logging interval
DEVICE OPERATING MODES
• IDLE Mode: First time battery inserted, and device has not been congured
• STANDBY Mode: Device has been congured, but not started;
• RUN Mode: Device starts to measure and log data.
• STOP Mode: Device stops from RUN Mode. In STOP Mode, device does not update
measurements or log data, and the last measurements are displayed.
VIEW CURRENT READING
1. Temperature ONLY unit: Current reading, minimum/maximum reading, run time/alarm
time toggling, Memory mode are shown on LCD.
2. Humidity and Temperature unit: Current temperature/humidity reading toggling every 5
seconds, minimum/maximum reading since last clearing, run time/alarm time, Memory
mode are shown on LCD.
VIEW CURRENT MIN/MAX
Temperature ONLY unit: Current temperature Min/Max values are displayed on LCD.
Humidity and Temperature unit: Current temperature/humidity Min/Max values are toggled
display.
Note: Every time device is congured, or resumes to run from STOP Mode, Min/Max
values are reset.
VIEWING RUN TIME/ALARM TIME
Run Time/Alarm Time is toggled to display on LCD. If Run Time is displayed, LCD symbol
RUN TIME appears; if Alarm Time is displayed, LCD symbol ALARM TIME appears.
Note: Alarm Time is accumulated for both Low Alarm and Hi Alarm for each channel
If Memory Mode is set to WRAP WHEN FULL,
is set to STOP WHEN FULL, MEM appears on LCD.
ALARM
1. Once an alarm is triggered by any of temperature, humidity out of set alarm range, LCD
symbol LOW ALM and/or HI ALM starts to ash. An Alarm Event will be logged.
2. Pressing START/STOP once will clear alarm, LCD symbol stops ashing. An Alarm
Acknowledge Event will be logged.
3. If temperature, or humidity falls back to normal range, an Alarm Event will be logged. If
any of temperature, humidity measurement goes back to out of alarm range, alarm will
be triggered again.
4. If device’s START Mode is congured as PUSH BUTTON TO START (default), device
LCD displays ‘push to start’. Press and hold START/STOP until LCD symbol appears.
Device enters RUN Mode. If device’s START Mode is congured as IMMEDIATE
START, device enters RUN Mode immediately. If device’s START Mode is congured
as DELAYED TIME START, delayed start time set by user is counting down on device
LCD. Once the counting time reaches 0, device enters RUN Mode.
5. If in RUN Mode, LCD symbol RUN appears, and logs data at user-dened interval.
If STOP Mode is congured as PUSH BUTTON TO STOP, press and hold START/
STOP until LCD symbol STOP appears. Device enters STOP Mode. If STOP Mode
is congured as NEVER STOP, device will ignore button press, and will stop when
memory is full if Memory Mode is set to STOP WHEN FULL, or will stop when
TraceableGO™ App is connected to device and downloads the data.
6. If in STOP Mode, press and hold START/STOP until LCD symbol RUN appears. Device
enters RUN Mode and resumes to log data at current setting. Anytime device resumes
to log data from STOP Mode, Min/Max values are reset.
Note: If device in STOP Mode, START Mode will be set to PUSH BUTTON TO START
regardless of previous Start Mode Setting. If Delayed Start Mode is needed while device in
STOP Mode, device must be re-congured.
Note: While device is connected to TraceableGO™ App, once Device has received
command from App to upload data to App, device will stop logging data and enter STOP
Mode if it is still in RUN mode.
VIEW MEMORY USAGE, CURRENT DATE/TIME, DEVICE’S SERIAL NUMBER
1. Press and release START/STOP button
2. Memory usage in percentage is shown on LCD. The percentage indicates how much
internal data memory storage has been used;
3. Number of days, hours, minutes left till memory is full are also displayed on the second
line;
appears on LCD; If Memory Mode
Note: If Memory Mode is set to WRAP WHEN FULL: LCD symbol
on the display when the memory is full. Once memory is full, the oldest data points will be
overwritten with new data points.
If Memory Mode is set to STOP WHEN FULL: LCD symbol MEM starts to ash on the
display when the memory is 95% full. Once memory is full, the device will stop logging new
data points.
CLEAR DATA MEMORY STORAGE
1. Data points stored in device internal storage can only be cleared through App or battery
removal.
2. Each time device is congured, all stored data points will be cleared.
3. Run Time/Alarm Time are also reset.
REGULATORY INFORMATION
This equipment has been tested and found to comply with the limits for a Class B digital
device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide
reasonable protection against harmful interference in a residential installation. This
equipment generates, uses and can radiate radio frequency energy and, if not installed
and used in accordance with the instructions, may cause harmful interference to radio
communications.
However, there is no guarantee that interference will not occur in a particular installation. If
this equipment does cause harmful interference to radio or television reception, which can
be determined by turning the equipment off and on, the user is encouraged to try to correct
the interference by one or more of the following measures:
• Reorient or relocate the receiving antenna.
• Increase the separation between the equipment and receiver.
• Connect the equipment into an outlet on a circuit different from that to which the
receiver is connected.
• Consult the dealer or an experienced radio/TV technician for help.
This device complies with Part 15 of the FCC Rules. Operation is subject to the following
two conditions: (1) this device may not cause harmful interference, and (2) this device
must accept any interference received, including interference that may cause undesired
operation.
Hereby, Traceable Products, declares that this digital thermometer is in compliance with
the essential requirements and other relevant provisions of Directive 1999/5/EC.
This device complies with Industry Canada licence-exempt RSS standard(s). Operation is
subject to the following two conditions: (1) this device may not cause interference, and (2)
this device must accept any interference, including interference that may cause undesired
operation of the device.
Le présent appareil est conforme aux CNR d’Industrie Canada applicables aux appareils
radio exempts de licence. L’exploitation est autorisée aux deux conditions suivantes
: (1) l’appareil ne doit pas produire de brouillage, et (2) l’utilisateur de l’appareil doit
accepter tout brouillage radioélectrique subi, même si le brouillage est susceptible d’en
compromettre le fonctionnement.
starts to ash
NOTE:
THE GRANTEE IS NOT RESPONSIBLE FOR ANY CHANGES OR MODIFICA-
4. Within 10 seconds, press START/STOP again, current date/time is displayed on device
LCD. The following gure shows 9/14/2017, 17:30
TIONS NOT EXPRESSLY APPROVED BY THE PARTY RESPONSIBLE FOR COMPLIANCE. SUCH MODIFICATIONS COULD VOID THE USER’S AUTHORITY TO OPERATE
THE EQUIPMENT.
W
5. Within 10 seconds, press START/STOP again, device S/N will be displayed on LCD.
6. To return to normal working status, press START/STOP again, or wait for 10 seconds
and device will automatically go back to normal working status.