TPP Pipettor Turbo-Fix - Operating instructions V13 Introduction
4 TPP Techno Plastic Products AG
1 Introduction
1.1 Intended use
TPP Pipettor Turbo-Fix is a pipette controller designed for aspirating and dispensing
aqueous solutions with plastic or glass pipettes of 1 to 100 ml volumes. It is intended for
measurement, control and laboratory use. Any use of this instrument in a medical or IVD
setting is under the sole responsibility of the user.
1.2 Safety notes
1) Do not use or charge TPP Pipettor Turbo-Fix in an atmosphere with danger of
explosion. Also, do not pipette highly flammable liquids such as acetone or ether.
2) When handling dangerous substances, comply with the material safety data sheet
(MSDS) and with all safety guidelines such as the use of protective clothing and safety
goggles. Never point a pipette in anyone’s direction.
3) Avoid pipetting of liquids whose vapours could attack the materials PA (polyamide),
POM (polyoxymethylene), FPM (fluor-rubber), NBR (nitrile-rubber), CR (chloroprene),
silicone. Corrosive vapours could also damage metallic parts inside the device.
4) Use an original TPP mains adapter only and protect it from moisture, otherwise
TPP Pipettor Turbo-Fix might be damaged.
5) Prolonged exposure of TPP Pipettor Turbo-Fix to UV-light can cause discolouration
and/or yellowing of the plastic housing. However, this will not affect the performance of
the device in any way.
6) Old Li-ion batteries may cause a safety risk. We recommend to replace the battery after
3 years of use. Also replace the battery if the charging intervals are unusually short or
if the charging takes much longer than usual (4 hours or more). These are indicators
that the battery has reached the end of its life-cycle.
7) Li-ion technology bears the risk of thermal runaway and cell rupture if the battery was
damaged. Do not expose the battery to heat (> 60°C) and avoid mechanical stress.
Batteries which were subject to deep discharges may develop internal short circuits,
leading to an increased self-discharge rate and heating during battery charging. This
may also result in thermal runaway and cell rupture.
8) To extend the battery life-cycle, it is recommended to charge the battery every 2
months if the pipette controller is not used regularly. If the pipette controller is not used
for more than 6 months, remove the battery from the instrument.
Regardless of the listed safety notes, additionally applicable regulations and guidelines of
trade associations, health authorities, trade supervisory offices, etc. must be observed.
Please visit our website www.tpp.ch
on a regular basis for up to date information regarding
REACH classified chemicals contained in our products.