1.Congratulations: Thank you for purchasing TPI products. The SP341 is
easy to use and built to last. It is backed by a 3 year limited warranty
2.Product Description: The SP341is a K-Type thermocouple thermometer
“Smart Probe”. It connects to your mobile device and uses the TPI
Smart Probe App to display temperature readings. Data can be exported
to a CSV file as well.
B.SAFETY CONSIDERATIONS
1.Never attach the SP341 to a hot surface.
2.Always make sure the temperature probe used has the proper range for
the application.
3.Always use K-Type thermocouples only.
4.Do not expose the unit to damp environments for extended periods of
time.
K-Type probe connectors are polarized
and made so one prong is wider than the
other. This ensures connection with the
proper polarity is made. They are marked
+ and - just like the SP341 connector.
2
1.Input connection for K-Type probes. Connector is
recessed and polarized (marked + and -). Ensure the KType probe is connected properly (see below).
2.SP341 status LED indicator.
Orange and blinking = Bluetooth not connected
Green and blinking = Bluetooth connected
Blinking Red with either of the above = Low battery
3.Power Key. Press and hold to turn the instrument on.
Press and hold to turn the instrument off.
4.Built-in magnet. Attach to
metal surfaces for hands free
operation. Ensure surface is
not hot.
5.Serial number location. Last
three digits are the product
identification number and
marked in bold. Use this number to distinguish between
models when several SP341’s
are connected to a phone or
tablet.
6.Screw to open battery compartment. Turn the screw
counter clockwise until it is
just above the compartment
door. Lift on the screw to open
the compartment.
1. Download and install the TPI
Smart Probe App onto your
mobile device. Run the App. The
“Main-Real Time” screen will be
displayed.
The App will say “No Devices”
until a smart probe has been
connected.
2. Press and hold the SP341 On/Off
key down until it turns on. The
LED indicator will initially be
orange and blink.
The SP341 will auto connect to
your mobile device. The LED
indicator will turn Green and
blink to indicate a link has been
established. The App will display
the model and ID number of the
smart probe connected.
If the SP341 does not auto connect tap refresh on the App.
Connect a probe to the SP341
and the temperature reading will
be displayed.
If no probe is connected the display will read Open.
Refresh
This also clears any data
accumulated.
Displays the measured temperature if a probe is connected
Displays Open if no probe is
connected.
Note: Open will also be displayed if a probe is bad
(break in the probe or connector).
3. T
apping the list icon allows
optional functions to be
accessed. These include
Show/Hide, Unit, and Zeroing
Sensors.
Show/Hide : Allows a device
to be seen or hidden. Useful
when multiple devices are
connected.
Unit : Switch between °C
and °F.
Zeroing Sensors: Field calibration mode. See page 7
for more information.
1. Tap the Drop Down Menu icon to access additional features.
Main - Real Time: Tapping this returns you to the main screen.
Select Application: Tapping this pulls up sub applications for
variious smart probes. These include Temperature differential
(SP341), Air Volume Calulation (SP565), and Tightness Test
(SP620). Note: There are no sub applications for the SP1000.
Company Info: Allows you to enter your information to show up
on the CSV file report. You can enter your company name,
address, and email.
Customer Info: Allows you to enter the information of the customer the tests are being performed for. You can enter the customer name, address, and email. You can also enter the building
information in cases where the tests are being performed somewhere off-site from the customer.
Exit: Closes the Smart Probe App.
The Smart Probe App has a Drop Down menu allowing access to additional app features.
1. Tap the Drop Down Menu icon to access additional features.
2. Select “Company Info” to enter your information to show up on
the CSV file report. You can enter your company name, address,
and email.
3. Select “Customer Info” to enter the information of the customer
the tests are being performed for. You can enter the customer
name, address, and email. You can also enter the building information in cases where the tests are being performed somewhere off-site from the customer.
The Smart Probe App allows you to enter your company information and the customer
information who the test is being performed for. This information will be exported in the
CSV file (see page 4).
I. PERFORMING DIFFERENTIAL TEMPERATURE MEASUREMENTS
1. Tap the Drop
Down Menu icon
to access additional features.
The Smart Probe App allows you to perform differential temperature measurements with
one or two smart probes.
2. Tap “Select
Application”.
3. Select
“Temperature
(T1-T2).
Note: The other
applications shown
are for different
model smart probes
and do not work
with the SP341.
2. Tap the list icon to
bring up the options
menu.
3. Select “Zeroing
Sensors” to bring
up the field calibration screen.
6. Insert the temperature probe into
the ice water bath
and allow the temperature reading
to stabilize at or
near 32°F (0°C).
Note: The displayed temperature must be within +
5.4°F (+3°C)
of 32°F (0°C) for
field calibraton to
take effect.
7. Tap the Reset icon
to set the temperature to 32°F
(0°C). Field calibration is complete.
8. Tap Close to
return to normal
operation.
4. Tap the + icon to
select the SP341 to
be field calibrated.
5. Tap the SP341 to be
field calibrated.
Note: To return the SP3431 to factory default calibration tap the Default icon.
8
The Smart Probe App allows you to perform field calibration. This function enables a system accuracy (SP341 and Probe) of +/- 1°F (0.5°C) accuracy in the range of 30°F to
120°F (0°C to 48°C).
Preparation for Field Calibration:
Fill a plastic or metal container with crushed ice and add clean water to a depth of at least
4 inches. Stir the ice and water for 2 to 3 minutes prior to performing calibration to
ensure the water is completely chilled. Make certain there is plenty of ice in the mixture
and always use clean water. Distilled water works well. The temperature of an ice bath is
approximately 32°F (0°F).
Test Products International offers a variety of K-Type thermocouple probes for many
applications. Visit www.testproductsintl.com for additional information.
1. Turn the battery
compartment screw
counter clockwise
until it is just above
the compartment
door. Lift on the
screw to open the
compartment.
2. Replace the three
AAA batteries.
3. Re-install the battery cover and tighten the
screw by turning it clockwise.
N. WARRANTY
This product is warranted to the purchaser against defects in material and workmanship for three years from the
date of purchase.
Covered by Warranty: Repair parts and labor; or replacement of the product at company’s option. Normal transportation charges to the purchaser are also covered.
Not Covered by Warranty: Damages to the product which are the result of abuse, improper use or maintenance are
not covered. Any other expense, consequential damages, incidental damages, or incidental expenses including damages to property are not covered. Transportation expenses to the company are not covered.
Some states do not allow the exclusion or limitation of incidental or consequential damages, so the above limitations or exclusions may not apply to you.
To Obtain Warranty Performance: Include with the product: your name, address, phone number, written description of the problem and proof of purchase date. Carefully package and return to:
TPI, Inc.TPI CanadaTPI, Europe Ltd..
9615 SW Allen Blvd.342 Bronte Rd. S., Unit 6Longley House, International Drive
Beaverton, OR 97005Milton, Ontario L9T 5B7Southgate Crawley
USACanadaWest Sussex RH10 6AQ England
503-520-9197905-693-855844(0) 1293 530196
www.testproductsintl.comwww.tpicanada.comwww.tpieurope.com
Implied Warranties: Any implied warranties including implied warranties of merchantability and fitness for a par-
ticular purpose, are limited in duration to three years from date of purchase. To the extent any provision of this warranty is prohibited by federal or state law and cannot be preempted, it shall not be applicable. This warranty gives
you specific legal rights, and you may also have other rights which vary from state to state.