Thank you for purchasing TPI brand products. The TPI 1010a Indoor
Air Quality (IAQ) Meter is a state of the art, easy to use tester designed
to provide you with the necessary measurements to monitor and make
adjustments to air handling devices. The instrument is ruggedly constructed and comes with a 3 Year unit and 2 Year sensor Guarantee.
2. General Overview
The 1010a IAQ meter uses state of the art sensors to measure humidity, carbon monoxide (CO), carbon dioxide (CO2), and temperature. The
CO and CO2 sensors in your meter will need to be replaced periodically
and calibration is recommended once every year.
The CO sensor is electrochemical and this type of sensor is always
active once installed in the unit. Therefore the time the unit is off and
not being used must be taken into account when determining sensor
life. The sensors in your analyzer are warranted for two years. This
warranty does not cover sensors damaged through misuse of the
meter.
You should keep battery power applied to your sensors at all times.
The following guidelines will help prevent damage to your sensors:
Always store your unit in a place where the temperature does not get
down to or below freezing.
Always maintain battery power to the sensors. When the batteries get
low replace them as soon as possible.
Never allow foreign objects or material to enter the sensor holes, damage to the sensor may result.
Never over saturate your sensors by performing tests on equipment
with gas levels beyond the capability of you analyzer.
4.1 Turning On & Off / Performing a Test / Main Display
Always:
unit is in fresh air (no carbon monoxide present) prior to turning on. This will
allow the CO sensor to set to zero properly.
Press thekey and the 1010a will begin a 30 second countdown. During
this time the 1010a performs a self diagnostic and sets the CO sensor to zero.
Once the countdown ends the 1010a will display CO2, CO, Temperature, and
Humidity. NOTE: If the CO2 sensor seems to read out of specification perform
the Ambient CO2 Sensor Reset as described in section 4.9.
Enter the area to be tested, the measured values will be displayed.
Carbon dioxide (CO2) in parts per million (ppm) is displayed at the top, carbon monoxide (CO) in parts per million (ppm) is displayed in the middle,
temperature in °F or °C (selectable), and humidity in percent are displayed at
the bottom of the main display.
- Turn the 1010a on outside of the area to be tested. Make sure the
The battery indicator is located at the bottom left of the display. When the
batteries become low they must be replaced to maintain proper operation.
Carbon dioxide (CO2)
Carbon monoxide (CO)
Temperature
Battery indicator
To turn the 1010a off, press and hold the
off.
If 30ppm or more CO is measured the 1010a will beep and the display will
flash red as a warning. This alarm point is adjustable. Additional functions
can be activated during testing. Please see next sections.
Dew Point Temperature
Indicated by “DP” in the
display
Wet Bulb Temperature
Indicated by “WB” in
the display
Repeatedly pressing the
key from the main display cycles through dew
point temperature, wet bulb temperature, and ambient temperature displays.
Pressing thekey switches the temperature display between °F and °C.
Pressing thekey from the main display cycles between temperature
and clock display. To set the clock please refer to section 4.6.
Pressing the
key turns the display backlight on and off. Power
consumption is much higher when the backlight is activated. The backlight
will automatically turn off 30 seconds after it is activated. To conserve battery life the backlight should only be used when necessary.
Align the infrared window of the A740 printer with the infrared window of
the 1010a located on the upper right side of the meter. Keep the printer in
clear line of sight approximately 6 to 8 inches away from the 1010a. Press
and hold the
key for approximately 2 to 3 seconds and the 1010a will
send test data to the printer.
4.3 Ambient / Dew Point / Wet Bulb Temperature Display
Using the Arrow keys set the alarm point to the desired level. The
keys select the digit and the
keys increase or decrease the value.
Once the desired alarm point is set, press the
key to return to normal
operation.
1) With the 1010a turned on, press and hold down the
CO alarm level in ppm
4.8 Setting CO Alarm Level
and
keys
until the CO alarm screen is displayed.
4.9 Ambient Air CO2 Reset
The CO2 sensor can be reset at ambient level as needed. This can be performed if
the CO2 reading seems to be too high or low or just to ensure your 1010A is
operating properly. Note: This does not take the place of a full factory calibration
which is recommended once per year.
1. Take the 1010A outside in ambient air. Allow the CO2 reading to stabilize. Make
sure you are not breathing into the sensor area.
2. Press and hold the key until the 1010A beeps and the display resets to
400ppm CO2 (+/- 75ppm). IMPORTANT: DO NOT PERFORM THIS PROCEDURE
INDOORS. IF THIS HAPPENS BY ACCIDENT, PERFORM THE PROCEDURE AS
INSTRUCTED ABOVE TO RESET THE CO2 READING.
The 1010a is equipped with an audible and visual alarm for carbon monoxide. The
alarm level is factory set at 30ppm. When CO above this level is measured, the
1010a will beep and the display will flash red.
The alarm level is adjustable and can be set from 10ppm to 500ppm.
key, "t =" will be displayed and the calculated percent outside
air will be displayed at the top of the display.
5) If you want to perform this measurement again, press thekey once and
return to step 1.
6) To exit and return to normal operation, press and hold the
key until “C-0”
is displayed. Press and hold thekey until a beep is heard and the 1010a
returns to normal operation.
HVAC systems use a combination of outside air, supply air, and return air to
maintain consistent and comfortable air quality. The 1010a can calculate outside
air either by measuring temperature or by measuring CO2.
The formula used to calculate % outside air is:
Percent outside air can be calculated using either temperature or carbon dioxide.
5.1 % Outside Air (Temperature)
To activate % outside air mode, press thekey about 2 seconds until “t-0” is
displayed in the middle display area.
1) Place the sensor of the 1010a perpendicular to the “Return” air flow. When the
reading stabilizes press thekey and capture the return air temperature,
"t-1" is displayed.
2) Place the sensor of the 1010a perpendicular to the “Supply” air flow. When the
reading stabilizes press the
"t-2" is displayed.
3) Place the sensor of the 1010a perpendicular to the “Outside” air flow. When
key, "C =" will be displayed and the calculated percent outside
air will be displayed at the top of the display.
5) If you want to perform this measurement again, press thekey once and
return to step 1.
6) To exit and return to normal operation, press and hold the
key until
a beep is heard and the 1010a returns to normal operation.
key until “C-0” is
displayed.
To activate % outside air mode, press the
displayed in the middle display area. Press and hold the
1) Place the sensor of the 1010a perpendicular to the “Return” air flow. When the
reading stabilizes press thekey and capture the return air temperature,
"C-1" is displayed.
2) Place the sensor of the 1010a perpendicular to the “Supply” air flow. When the
reading stabilizes press the
"C-2" is displayed.
3) Place the sensor of the 1010a perpendicular to the “Outside” air flow. When
the reading stabilizes press the
the Arrow keys set the interval to the desired level. The
3) Once the seconds have been entered, press the key to move to min / hr.
5) Once the logging interval is set press the
keys select the digit and the keys increase or decrease the value.
key and the 1010a will return
to the main display and “LOG” will flash indicating the logging function is active.
4) Using the Arrow keys set the interval to the desired level. The
6) To deactivate data logging press and hold down the
key until “LOG” is
no longer flashing.
For data retrieval please see the next section on how to retrieve logged data.
23 hr 59 min. Pressing the
keys switches between the two time
modes.
To activate data logging press and hold down the
screen is displayed. Once the logging function is activated all preciously logged
data is written over.
1) INT is displayed at the top to indicate the 1010a is waiting for the logging interval
to be set. The logging interval tells the 1010a how often to sample. For example,
a logging interval of 10 seconds (00:10) means the 1010a will log a reading every
10 seconds. The bottom indicates the current logging interval in minutes and sec
onds (MM:SS) or hours and minutes (HH:MM). Flashing “SEC” in the display
indicates the last two digits are seconds and a flashing “MIN” in the display indicates the last two digits are minutes. In seconds mode the time is adjustable from
1 sec to 59 min 59 sec. In the minutes mode the time is adjustable from 1 min to
It is recommended that the instrument be calibrated every 12 months.
Please see section 9.2 below..
Problem: CO2 reading out of specification
Possible Causes: Sensor needs calibration, T/C key held down indoors
Corrective Action: Perform ambient air reset outlined in section 4.9.
Replace batteries
Send unit to TPI for factory calibration and maintenance.
Problem: CO reading out of specification
Possible Causes: Sensor needs calibration or replacement
Corrective Action: Replace batteries
Send unit to TPI for factory calibration and maintenance.
Problem: Temperature or Humidity reading out of specification
Possible Causes: Sensor needs calibration or replacement
Corrective Action: Replace batteries
Send unit to TPI for factory calibration and maintenance.
9.1 Battery Replacement
When the batteries become low they will require replacement.
1.
Turn the meter over so the back is facing you.
2. Loosen the screw holding down the battery cover located under the tilt
stand.
Lift the tilt stand up and remove the battery cover.
3.
4.
Replace the batteries (3 x a Alkaline) and install the cover and tighten the
screw.
9.2 Service
To obtain warranty and non-warranty per
analyzer: - Include with the product your name, address, phone number, written description of the problem and proof of purchase date. Carefully package
and return to:
formance or maintenance on your
TPI / Attn. Repair
9615 SW Allen Blvd. Suite 104
Beaverton, OR 97005
% Outside Air - HVAC systems use a combination of outside air, supply air, and
return air to maintain consistent and comfortable air quality. The 1010a can calcu
late outside air either by measuring temperature or by measuring CO2. The formula for % Outside Air is:
If too little outdoor air enters a home, pollutants can accumulate to levels that can
pose health and comfort concerns. Unless they are built with special mechanical
means of ventilation, homes that are designed and constructed to minimize the
amount of outdoor air that can 'leak' into and out of the home may have higher pollutant levels than other homes. However, because some weather conditions can
drastically reduce the amount of outdoor air that enters a home pollutants can build
-
up even in homes that are normally considered 'leaky'.
Dew Point Temperature - This is the temperature at which condensation begins
Wet Bulb Temperature - This is the lowest temperature evaporating water can
reach
Recommended Levels by Agency
Parameter
Humidity20% - 60%30% - 60%N/AN/A
Temperature68 - 75 (winter)
73 - 79 (summer)
Carbon Dioxide1,000ppm
(<800ppm preferred)
Carbon Monoxide 9ppm9ppm50ppm25ppm
N/A-Not Applicable or Not Established
1
Illinois Department of Public Health (2009)
2
American Society of Heating, Refrigerating and Air Conditioning Engineers
3
Occupational Safety and Health Administration Permissible Exposure Limit -- this level
is a time-weighted average and is an enforceable standard that must not be exceeded
during any eight-hour work shift of a 40-hour work week.
4
American Conference of Governmental Industrial Hygienist Threshold Limit Value -- this
level is a recommended time-weighted average upper limit exposure concentration for a
normal eight to 10-hour workday and a 40-hour work week.
If the CO2 sensor fails to initialize or is faulty the
1010a will display an error code. Try turning the
1010a off and on again. If it displays the error code
again the 1010a is in need of service. Please return
your instrument to TPI. See section 9.2 for instructions on returning your meter.
CO Sensor Error
If the CO sensor fails to initialize or is faulty the
1010a will display an error code. Try turning the
1010a off and on again. If it displays the error code
again the 1010a is in need of service. Please return
your instrument to TPI. See section 9.2 for instructions on returning your meter.
CO Sensor Caibration Due
When the last calibration date of the CO sensor is
one year or older the 1010a will display a reminder
that calibration is due before returning to normal
operation. Please return your instrument to TPI for
calibration. See section 9.2 for instructions on
returning your meter.
Temperature and/or Humidity Sensor Error
If the temperature and/or humidity sensor fails to initialize or is faulty the 1010a will display an error
code. Try turning the 1010a off and on again. If it displays the error code again the 1010a is in need of
service. Please return your instrument to TPI. See
section 9.2 for instructions on returning your meter.