Thank you for purchasing TPI brand products. The TPI 1008a Indoor
Air Quality (IAQ) Meter is a state of the art, easy to use tester designed
to provide you with the necessary measurements to monitor and make
adjustments to air handling devices. The instrument is ruggedly constructed and comes with a 3 Year unit and 2 Year sensor Guarantee.
2. General Overview
The 1008a IAQ meter uses state of the art sensors to measure carbon
dioxide (CO2), and temperature. The CO2 sensor in your meter will
need to be replaced periodically and calibration is recommended once
every year.
You should keep battery power applied to your sensors at all times.
The following guidelines will help prevent damage to your sensors:
Always store your unit in a place where the temperature does not get
down to or below freezing.
Always maintain battery power to the sensors. When the batteries get
low replace them as soon as possible.
Never allow foreign objects or material to enter the sensor holes, damage to the sensor may result.
On / Off key - Used to turn the IAQ meter on or off. Press to turn
on, press and hold to turn off.
Up Arrow key - Used to select and change parameters.
Down Arrow key - Used to select and change parameters.
Mode key - Used to switch between Celsius and Fahrenheit units
of display and to enter CO2 alarm level input mode. Press to
switch between Celsius and Fahrenheit units. Press and hold to
enter and exit CO2 alarm level mode.
Backlight key - Used to turn the display backlight on and off.
Min / Max key - Used to activate and cycle through minimum,
maximum, and average display modes. Press to activate Min/Max
mode, press repeatedly to cycle through min, max, and avg displays. Press and hold to deactivate Min/Max mode.
Log key - Used activate and deactivate data logging.
key and the 1008a will begin a 30 second countdown. During
this time the 1008a performs a self diagnostic. Once the countdown ends the
1008a will display CO2 and Temperature.
To turn the 1008a off, press and hold the
key down until the 1008a turns
off.
4.1 Turning On & Off / Performing a Test / Main Display
Enter the area to be tested, the measured values will be displayed.
Carbon dioxide (CO2) in parts per million (ppm) is displayed at the top and
Temperature in °F or °C (selectable)is displayed at the bottom of the main
display.
The battery indicator is located at the left center of the display. When the bat
teries become low they must be replaced to maintain proper operation.
-
The 1008a has a CO2 alarm. If the CO2 level measured is at or above the
set alarm level the 1008a will beep and the display will flash red as a
warning. This alarm point is adjustable. Additional functions can be activated during testing. Please see next sections.
key switches the temperature display between °F and °C.
To deactivate Min/Max mode and return to normal operation, Press and hold the
key until the 1008a beeps.
Pressing the
key turns the display backlight on and off. Power
consumption is much higher when the backlight is activated. The backlight
will automatically turn off 30 seconds after it is activated. To conserve battery life the backlight should only be used when necessary.
Pressing thekey will cause the 1008a to display “MAX”. This indicates the
Maximum CO2 and Temperature measured is displayed.
Press thekey again and the 1008a will display “MIN”. This indicates the
Minimum CO2 and Temperature measured is displayed.
Press thekey again and the 1008a will display “AVG”. This indicates the
Average CO2 and Temperature measured is displayed.
Pressing thekey will cause the 1008a to display “MAX MIN”. This indicates
the real time reading is displayed and the Min/Max mode is still active. The steps
above can be repeated to cycle through the Min, Max, and Avg displays.
Align the infrared window of the A740 printer with the infrared window of
the 1008a located on the upper right side of the meter. Keep the printer in
clear line of sight approximately 6 to 8 inches away from the 1008a. Press
and hold the
key for approximately 2 to 3 seconds and the 1008a will
send test data to the printer.
Pressing the
4.3 Min / Max Mode
Pressing the
In this mode the 1008a will keep track of the minimum, maximum, and average
CO2 and Temperature being measured. When “Min Max” is displayed the 1008a
will display the real time reading.
key activates Min/Max mode and “Max Min” will be displayed.
Using the Arrow keys set the alarm point to the desired level. The
keys adjust the alarm level from “Off” to 5000ppm in 100ppm increments.
Once the desired alarm point is set, press and hold the
key to return to
normal operation.
1) With the 1008a turned on, press and hold down the
CO2 alarm level in ppm
4.6 Setting CO2 Alarm Level
key until the CO2
alarm screen is displayed.
4.7 Ambient Air CO2 Reset
The CO2 sensor can be reset at ambient level as needed. This can be performed if
the CO2 reading seems to be too high or low or just to ensure your 1008A is
operating properly. Note: This does not take the place of a full factory calibration
which is recommended once per year.
1. Take the 1008A outside in ambient air. Allow the CO2 reading to stabilize. Make
sure you are not breathing into the sensor area.
2. Press and hold thekey until the 1008A beeps and the display resets to
400ppm CO2 (+/- 75ppm). IMPORTANT: DO NOT PERFORM THIS PROCEDURE
INDOORS. IF THIS HAPPENS BY ACCIDENT, PERFORM THE PROCEDURE AS
INSTRUCTED ABOVE TO RESET THE CO2 READING.
The 1008a is equipped with an audible and visual alarm for carbon dioxide. The
alarm level is factory set to off. When CO2 above the alarm level is measured, the
1008a will beep and the display will flash red.
The alarm level is adjustable and can be set from 1100ppm to 5000ppm in
100ppm increments.
screen is displayed. Once the logging function is activated all previously logged
data is written over.
1) The log time interval is displayed in minutes at the bottom of the display.
The logging interval tells the 1008a how often to sample. For example, a logging
interval of 1 minute (01) means the 1008a will log a reading every minute. A logging interval of 10 minutes (10) means the 1008a will log a reading every 10 minutes. The log interval time is adjustable from 1 minute to 60 minutes in the following increments: 1, 2, 3, 5, 10, 20, 30, 40, 50, 60 minutes. A maximum of 50
readings can be logged.
2) Use the
3) Once the logging interval is set press the
key and the 1008a will return
to the main display and “LOG” will flash indicating the logging function is active.
The1008a will beep each time a reading has been logged.
4) To deactivate data logging press the
key and “LOG” will no longer flash.
For data retrieval please see the next section on how to retrieve logged data.
Log time interval in
minutes.
keys to increase or decrease the log interval time.
key, “LOG 01” and the logged CO2 and temperature reading
will alternately flash. Press the
to cycle through all of the logged
data.
2) To deactivate recall mode and return to normal operation, press the
key.
Logged data can be displayed on the 1008a display or sent to a PC by using the
USB cable and software included. The PC software also provides the capability of
real time monitoring from a PC.
Recalling and Displaying Logged Data
1) Press the
Retrieving Logged Data via PC
To retrieve logged data the USB cable and PC software (both supplied) must be
used.
Install the PC software on your PC and open it.
Connect the 1008a to the PC using the USB cable.
Run the software and select 1008a when prompted. Click on the connect to PC
button in the software window.
The software can be used to retrieve logged data and for real time monitoring.
It is recommended that the instrument be calibrated every 12 months.
Please consult
Test Products International for further details.
8.1 Battery Replacement
When the batteries become low they will require replacement.
1. Turn the meter over so the back is facing you.
2. Loosen the screw holding down the battery cover located under the tilt
stand.
3. Lift the tilt stand up and remove the battery cover.
4. Replace the batteries (3 x AA Alkaline) and install the cover and tighten the
screw.
8.2 Service
To obtain warranty and non-warranty performance or maintenance on your
analyzer: - Include with the product your name, address, phone number, written description of the problem and proof of purchase date. Carefully package
and return
American Society of Heating, Refrigerating and Air Conditioning Engineers
3
Occupational Safety and Health Administration Permissible Exposure Limit -- this level
is a time-weighted average and is an enforceable standard that must not be exceeded
during any eight-hour work shift of a 40-hour work week.
4
American Conference of Governmental Industrial Hygienist Threshold Limit Value -- this
level is a recommended time-weighted average upper limit exposure concentration for a
normal eight to 10-hour workday and a 40-hour work week.
If the CO2 sensor fails to initialize or is faulty the
1008a will display an error code. Try turning the
1008a off and on again. If it displays the error code
again the 1008a is in need of service. Please return
your instrument to TPI. See section 9.2 for instructions on returning your meter.
Temperature Sensor Error
If the temperature sensor fails to initialize or is faulty
the 1008a will display an error code. Try turning the
1008a off and on again. If it displays the error code
again the 1008a is in need of service. Please return
your instrument to TPI. See section 9.2 for instructions on returning your meter.