TPI SP341 Quick Start Manual

Page 1
SP341 K-Type Temperature Smart Probe
Test Products International, Inc.
Use standard K-type thermocouple probes with sub-mini input plug. GK41M air probe included. Other probes sold separately.
Visit www.tpieurope.com
Uses any K-Type probe with sub-mini connector
Connects to smart phone and uses the TPI Smart Probe App or the TPI View app.
Perform differential temperature measurements with one instrument
Up to 6 SP341 smart probes can connect to a single mobile device (most devices)
Auto field calibrate to +/-1°C system accuracy in the range of -1°C to 49°C using field calibrate mode
3 Year Limited Warranty
Page 2
TABLEOFCONTENTS
A. Introduction............................................................... 1
B. Safety Considerations.............................................. 1
C. Specifications........................................................... 2
D. Instrument Overview................................................ 3
E. Maintenance............................................................. 4
F. Trouble Shooting...................................................... 4
G. Accessories.............................................................. 4
H. Warranty................................................................... 5
I. Replacing Batteries.................................................. 5
Page 3
A. INTRODUCTION
1. Congratulations: Thank you for purchasing TPI products. The SP341 is easy to use and built to last. It is backed by a 3 year limited warranty.
2. Product Description: The SP341 is a K-Type thermocouple thermometer “Smart Probe”. It connects to your mobile device and uses the TPI Smart Probe App or TPI View App to display temperature readings.
Data can be displayed, saved, logged, allocated to a specific
job, exported & e-mailed in either pdf or csv depending on which app the SP341 is connected to.
3. Download either the SmartProbe or the View app from either the Apple App Store or the Google Play Store
B. SAFETY CONSIDERATIONS
1. Never attach the SP341 to a hot surface.
2. Always make sure the temperature probe used has the proper range for the application.
3. Always use K-Type thermocouples only.
4. Do not expose the unit to damp environments for extended periods of time.
1
SmartProbes App View App
Page 4
C. SPECIFICATIONS
2
Page 5
D. INSTRUMENTOVERVIEW
1. Input connection for K-Type probes. Connector is recessed and polarized (marked + and -). Ensure the K-Type probe is connected properly (see below).
2. SP341 status LED indicator. Orange and blinking = Bluetooth not connected Green and blinking = Bluetooth connected Blinking Red with either of the above = Low battery
3. Power Key. Press and hold to turn the instrument on. Press and hold to turn the instrument off.
4. Built-in magnet. Attach to metal surfaces for hands free operation. Ensure surface is not hot.
5. Serial number location. Last three digits are the product identification number and marked in bold. Use this
number to distinguish
between models when several SP341’s are connected to a phone or tablet.
6. Screw to open battery com­partment. Turn the screw counter clockwise until it is just above the compartment door. Lift on the screw to open the compartment.
K-Type probe connectors are polarized and made so one prong is wider than the other. This ensures connection with the proper polarity is made. They are marked + and - just like the SP341 connector.
3
Page 6
4
E. MAINTENANCE
1. Clean the surface of the instrument with a damp cloth.
2. Replace the three AAA batteries when the low battery indicator appears.
F. TROUBLE SHOOTING
Symptom Probable Cause
Temperature seems inaccurate. • Check thermocouple probe for
incorrect wiring or open connection.
• Low battery.
• Non K-Type probe being used.
• Ensure connection with phone is working.
SP341 does not turn on. • Dead or low batteries.
• Not holding down power switch until unit turns on.
• Defective POWER ON switch.
SP341 displays “OPEN”. • No probe connected.
• Temperature probe is bad.
SP341 measures temperature • Thermocouple probe wired backwards. backwards.
• Probe connected with reverse polarity.
G. ACCESSORIES
Included Accessories
Part Number Description
A341SP Soft pouch
GK41M General purpose
temperature probe
Optional Accessories
Part Number Description
A925 Carrying case for 4 smart
probes
A927 Carrying case for SP341 & temperature probes
Test Products International offers a variety of K-Type thermocouple probes for many applications. Visit www.tpieurope.com for additional information.
Page 7
This product is warranted to the purchaser against defects in material and workmanship for three years from the date of purchase.
Covered by Warranty: Repair parts and labor; or replacement of the product at company’s option. Normal transportation charges to the purchaser are also covered.
Not Covered by Warranty: Damages to the product which are the result of abuse, improper use or main­tenance are not covered. Any other expense, consequential damages, incidental damages, or incidental expenses including damages to property are not covered. Transportation expenses to the company are not covered.
To Obtain Warranty Performance: Include with the product: your name, address, phone number, written description of the problem and proof of purchase date. Carefully package and return to:
TPI, Inc. TPI Canada TPI, Europe Ltd..
9615 SW Allen Blvd. 342 Bronte Rd. S., Unit 6 Longley House, International Drive Beaverton, OR 97005 Milton, Ontario L9T 5B7 Southgate Crawley USA Canada West Sussex RH10 6AQ England 503-520-9197 905-693-8558 44(0) 1293 530196 www.testproductsintl.com www.tpicanada.com www.tpieurope.com
Implied Warranties: Any implied warranties including implied warranties of merchantability and fitness
for a particular purpose, are limited in duration to three years from date of purchase. To the extent any provision of this warranty is prohibited by federal or state law and cannot be preempted, it shall not be applicable. This warranty gives you specific legal rights, and you may also have other rights which vary from country to country.
5
I. REPLACING BATTERIES
1. Turn the battery compartment screw counter clockwise until it is just above the compartment door. Lift on the screw to open the compartment.
2. Replace the three AAA batteries.
3. Re-install the battery cover and tighten the screw by turning it clockwise.
H. WARRANTY
Page 8
Test Products International, Inc.
9615 SW Allen Blvd., Ste. 104
Beaverton, OR 97005
Tel: 503-520-9197
www.testproductsintl.com
Test Products International, Ltd.
342 Bronte Road South, Unit #6
Milton Ontario Canada L9T 5B7
Tel: 905-693-8558
www.tpicanada.com
Test Products International Europe Ltd.
Longley House, International Drive
Southgate, Crawley, West Sussex RH10 6AQ
Tel:: +44 (0) 1293 530196
www.tpieurope.com
Loading...