TPI Scope Plus 460 User Manual

460 Handheld 20MHz
Digital Oscilloscope
User’s Guide
Test Products International, Inc.
9615 SW Allen Blvd.
Portland, OR 97005
Ph: 503-520-9197 ! Fax: 503-520-1225
E-mail: info@tpi-thevalueleader.com
www.tpi-thevalueleader.com
TM
Copyright © 2001
Test Products International, Inc.
All Rights Reserved
08/01
460 User’s Guide
ii
Table of Contents
C1 - Introduction........................................................................................C1-1
Features ..........................................................................................C1-2
Specifications ..................................................................................C1-3
Scope Modes ........................................................................C1-3
Multimeter Modes ................................................................C1-5
C2 - Getting Started ..................................................................................C2-1
Equipment Needed ..........................................................................C2-1
460 Setup ........................................................................................C2-1
Measuring Voltage ..........................................................................C2-3
Measuring Frequency ......................................................................C2-3
Measuring Resistance......................................................................C2-4
Testing Continuity ............................................................................C2-4
Diode Test ........................................................................................C2-4
C3 - Operation............................................................................................C3-1
Principal Connectors and Keys ........................................................C3-1
Key / Screen Menus ........................................................................C3-3
C4 - Advanced Functions ..........................................................................C4-1
Triggering ........................................................................................C4-1
Trend Mode......................................................................................C4-1
Glitch Capture ..................................................................................C4-2
Saving Screens ................................................................................C4-2
Using the 10:1 Probe ......................................................................C4-3
10:1 Probe Compensation Adjustment ............................................C4-3
C5 - Using Adapters ..................................................................................C5-1
Temperature Adapter - A301............................................................C5-1
High Current Adapter - A251 / A256 / A296 ....................................C5-1
Low Current Adapter - A254 ............................................................C5-2
460 User’s Guide
iii
Current Shunt Adapter - A130 ........................................................C5-2
Pressure Adapter - A620 ................................................................C5-2
PC Software & RS232 Cable - A404................................................C5-3
C6 - Safety Considerations ........................................................................C6-1
Instrument ......................................................................................C6-1
Environmental..................................................................................C6-2
Appendix A - Test and Calibration................................................................A-1
Appendix B - Maintenance Considerations ..................................................B-1
Appendix C - Glossary ................................................................................C-1
Appendix D - Optional RS-232C / PC Interface............................................D-1
Appendix E - Troubleshooting ......................................................................E-1
iv
460 User’s Guide
Chapter 1
Introduction
This manual describes the TPI Scope Plus 460 and its functions. Normal and advanced functions are covered in detail, as well as interactions with computers and other instrumentation. A full chapter is also devoted to safety considerations, both instrument based and interactively. The fol­lowing is a brief chapter and appendix summary:
Chapter 1 - Covers the introduction, features and Specifications.
Chapter 2 - Getting Started gets the new user up and running in the shortest time possible.
The basic functions are introduced in a hands on style designed to familiarize the user with the keypad and screen display.
Chapter 3 - Explains all the normal functions of the TPI Scope Plus 460, including buttons,
menus, screen displays, and a sequencing guide for both external key actions and their interaction with software driven menus.
Chapter 4 - Explains more advanced functions such as triggering, glitch capture, trend mode,
and saving screens.
Chapter 5 - Explains and illustrates uses of probes and adapters including pressure,
temperature, and current shunt adapters. This chapter also introduces general aspects of the optional optically isolated RS-232 interface.
Chapter 6 - This important chapter details operating methods and tips designed to preserve
both the life of your instrument, and safety factors to prevent human injury. The text is divided into Instrument and environmental safety subsections.
Appendix A - This appendix covers what to look for, and how to make simple tests to find out if
your TPI Scope Plus 460 requires replacement parts or calibration.
Appendix B - This Maintenance appendix gives tips on how to keep your TPI
Scope Plus 460 in top operating condition by assuring constant peak performance. The appendix also covers battery servicing and replacement.
Appendix C - A glossary describing the meanings of terms associated with the
460 and its working environment.
Appendix D - This appendix introduces and explains the powerful RS-232C
Interface option. All aspects are covered from Installation to
460 User’s Guide Introduction
C1-1
Special Applications.
Appendix E - This Troubleshooting appendix covers problems that may arise
during installation or operation of your new TPI Scope Plus 460.
Features
The following features exist for the TPI Scope Plus 460:
"
20 MHz Bandwidth
Enough bandwidth to capture signals from AC/DC drive motors, sensors, actuators, line and control voltages, UPS and industrial machines.
"
Sample Rate
A real time sampling of 25 megasamples/sec to capture spikes and dropouts of industrial signals.
"
Dual Input
View two waveforms on the display for comparison and troubleshooting.
"
True RMS DMM
Includes a 4000 count True RMS DMM that measures AC/DC volts to 600V, frequency to 20MHz, resistance to 20M Ohm, diode test and continuity audible tone.
"
Autoranging
Autoranging capability across Voltage, Timebase and Frequency.
"
Bright LCD Backlight
Easy-to-see display in any light condition with user-adjustable backlight brightness levels
"
Optically Isolated RS232 Output
Transfer data safely without any direct connections to your computer’s systems security.
"
Optional Accessories
Temperature Adapter to measure temperature from -40 to 500
o
F (A301). Pressure Adapter to measure pressure from -30 to 500psi (A620). Protective Boot (A405). Soft Carrying Case with Shoulder Strap (A905). RS232 cable and software (A404). Current Adapters: o A254 - For measuring low current in milliamps o A256 - Up to 400 AC/DC amps
460 User’s Guide Introduction
C1-2
o A296 - Up to 1000 AC/DC amps
Specifications
Display:
Size 76mm X 76mm (3" X 3") Pixels 240 X 240 pixels LCD type STN Normally Gray or Black Back Light CCFL
Memory:
Waveform Screens (2) Setup User setup (1)
Power:
External Power Adapter (5.5V / 2A) Internal Battery Ni-MH 1.2V X 4 Battery Operating Time More than 2.5 hours with bright backlight Charging Time 10 hours
SCOPE MODES
Horizontal:
Samples per division 25 samples / div. Time/division Equivalent sampling 50ns to 500ns/div
Real time sampling 1us to 2s/div Modes Single, Normal, Auto, Glitch Accuracy Real time sampling +/- (0.1% + 0.04 time/div)
Equivalent sampling +/- (0.5% + 0.08 time/div)
Vertical:
Sampling Rate 25 Mega samples per second Bandwidth DC Coupled 1:1 shielded test leads ; DC to 10MHz
10:1 probe ; DC to 20MHz
AC Coupled 1:1 shielded test leads ; 10Hz to 10MHz
10:1 probe ; 10Hz to 20MHz Resolution 8 bits Channel 2 Coupling AC,DC,GND Input impedance Approx. 1M Ohm Sensitivity 50mV to 200V/div 1,2,5 sequence
460 User’s Guide Introduction
C1-3
Accuracy +/- (3% + 0.05 range/div) Display modes CH1, Dual Maximum Input Voltage 600V
Trigger:
Position Ver 1; Fixed Position / Ver 3; Adjustable Position Mode Free run, normal Source Internal (CH1/CH2) Slope Positive/Negative
Sensitivity Real time sampling .... 2 divisions or more
Equivalent sampling .... 3 divisions or more
Trend Plot:
Plot Time 30sec/div to 1-hour/div Plot Data Type Max / Min Selectable
Temperature Measurement:
Optional Adapter A301 (k - Type thermocouple) Range -40 F to 500 F Accuracy +/- (1.5% + 5 digits)
Pressure / Vacuum Measurement:
Optional Adapter A620 Range Pressure 500 psi
Vacuum 30 inHg Accuracy Pressure +/- (5% + 5 digits)
Vacuum +/- (1.5% + 5 digits)
Current Measurement:
Optional Adapter 1 A251 (AC AMPS) Range 0 - 40A / 40 - 400A Accuracy ( 0 - 40A ) +/- (2.0% + 10 digits , 50 / 60 Hz)
(40 - 400A) +/- (2.0% + 10 digits , 50 / 60 Hz)
Optional Adapter 2 A256 (AC/DC AMPS) Range 0 - 40A / 40 - 400A Accuracy ( 0 - 40A ) +/- (2.0% + 10 digits , 50 / 60 Hz)
(40 - 400A) +/- (2.0% + 10 digits , 50 / 60 Hz)
Optional Adapter 3 A296 (AC/DC AMPS)
460 User’s Guide Introduction
C1-4
Range 0 - 400A / 400 - 1000A Accuracy ( 0 - 400A ) +/- (2.0% + 10 digits , 50 / 60 Hz)
(400 - 1000A) +/- (2.0% + 10 digits , 50 / 60 Hz)
MULTIMETER MODES
DC Volts (CH1/CH2)
Range Resolution Accuracy
400mV 0.1mV +/- (0.5% + 5 digits) 4V 0.001V +/- (0.5% + 5 digits) 40V 0.01V +/- (0.5% + 5 digits) 400V 0.1V +/- (0.5% + 5 digits) 600V 1V +/- (0.5% + 5 digits)
AC Volts (CH1/CH2)
Range Resolution Accuracy
400mV 0.1mV 20Hz to 50Hz; +/- (2% + 20 digits) 4V 0.001V 50Hz to 1kHz; +/- (1% + 10 digits) 40V 0.01V 1kHz to 20kHz; +/- ( 2% + 10 digits) 400V 0.1V 40Hz to 400Hz; +/- (1% + 10 digits) 600V 1V 40Hz to 400Hz; +/- (1% + 10 digits)
Frequency (CH1/CH2)
Range Resolution Accuracy
100Hz 0.01Hz +/- (1.0% + 10 digits) 1KHz 0.1Hz +/- (0.5% + 5 digits) 10KHz 1Hz +/- (0.5% + 5 digits) 100KHz 10Hz +/- (0.5% + 5 digits) 1MHz 100Hz +/- (0.5% + 5 digits) 10MHz 1kHz +/- (0.5% + 5 digits) 20MHz 10kHz +/- (2.5% + 5 digits)
OHM (CH1 ONLY)
Range Resolution Accuracy
400 Ohm 0.1 Ohm +/- (0.75% + 5 digits) 4k Ohm 1 Ohm +/- (0.5% + 5 digits) 40k Ohm 10 Ohm +/- (0.5% + 5 digits) 400k Ohm 100 Ohm +/- (0.5% + 5 digits) 4M Ohm 1k Ohm +/- (1.0% + 10 digits) 20M Ohm 10k Ohm +/- (3.0% + 20 digits)
460 User’s Guide Introduction
C1-5
Continuity (CH1 ONLY)
Range Resolution Accuracy
4 K ohm 1 ohm Beeps < 0.1 K ohm , +/- (2% + 5 digits)
Diode (CH1 ONLY)
Range Resolution Accuracy
4 V 1 mV Open circuit voltage < 5 V
Short circuit current < 5 mA Reading accuracy; +/- (2% + 5 digits)
460 User’s Guide Introduction
C1-6
Chapter 2
Getting Started
This chapter will get you started with the TPI Scope Plus 460. It will include initial power up, fol­lowed by a series of DMM and static tests. More detailed operations will be the subject of the next chapter. Refer also to the chapters and appendices covering specifications and maintenance if you run into problems outside the scope of this chapter. Also feel free to check with service support using the phone numbers given at the beginning of this manual.
Equipment Needed
Required:
"
Fully operational TPI Scope Plus 460 with batteries fully charged
"
Set of probes with alligator clip attachments available
"
Set of fixed or variable resistances.
"
Set of fixed or variable inductances.
"
Set of fixed or variable capacitances.
"
Static or variable DC voltage source.
Optional:
"
Signal generator source with varied outputs
"
Random glitch or voltage spike generator
460 Setup
1. Use the tilt stand if necessary to set up the instrument in a readable position.
2. Switch the unit on by pressing POWER ON button for 3 seconds. If the unit will power
on, attach the power adapter. Initial charge of the battery must be made with 460 turned on.
3. If the Power adapter is attached, check the battery level by pulling out the adapter and
checking if there is any change to the brightness of the display (For optimal performance, the battery should be fully charged even with the Power adapter attached).
4. Check the Contrast. Refer to Figure 2-1, and press the USER button. Note the Contrast
menu button is displayed on the screen above F2. - Press F2 and note the menu item on the screen is now darkened. Press the Up arrow to increase the contrast and the Down arrow to reduce the contrast.
5. Check the Brightness (-there are ten levels). Refer to Figure 2-1, and press the Brightness button (Sun Icon) to decrease the brightness. To increase the brightness, continue pressing the brightness button until you have reached the bottom level, at which
460 User’s Guide Getting Started
C2-1
3. Press F2 (below the Contrast Label).
4. Press the Up or Down direction arrows to change contrast.
2. Press the User Button
1. Press the POWER ON Button
Figure 2-1. Contrast Adjustment with Low Visibility
460 User’s Guide Getting Started
C2-2
time it will revert to the maximum brightness level.
NOTE: The Brightness control effects the backlighting whereas Contrast effects the intensity of the liquid crystal display (both for the tracing beams and the text display).
6. You have now completed the basic setup. Press Ch A (Channel A), to exit the user setup mode and ready the instrument for Multimeter operations.
Measuring Voltage
1. With the 460 set for Input A (Ch A button), use the down arrow to set the highlight on DCV. Press Enter.
2. Press the green V key, and connect the probes to a DC Voltage source. The voltage will appear in alphanumeric text at the top left of the screen. The trace line A should be steady, but if there are any AC components present on the line it will also show these (especially in the case of partially rectified AC).
NOTE: Pressing the Green key V selection first will ensure you are not being limited by the millivolt range. If the voltage reading is then read as a small value, use the mV side of the green rocker key to display the millivolt range on the screen.
3. Disconnect the DC Voltage source and press Ch A (Channel A blue button). Use the Up arrow to move the screen highlight to ACV. Press Enter.
4. Press the green V key, and connect the probes to the AC source. The voltage can be read at the top left of the screen expressed as an RMS value. The trace line A will show the shape of the AC voltage source.
Measuring Frequency
1. With the 460 set for Input A (Ch A button), use the down arrow to set the highlight on Freq. Press Enter.
2. Connect the probes to a frequency source. The frequency will be shown in alphanumeric text at the top left of the screen. The trace line A will show the nature of the frequency source.
Notes on Trace Stability:
a). Ensuring that the Channel B trace is off when making frequency measurements on the A channel, will reduce clutter on the screen, and make the A trace easier to read.
460 User’s Guide Getting Started
C2-3
b). Using the HOLD/RUN button will freeze the trace and allow waveform characteristics to be measured precisely against the background grid.
Measuring Resistance
1. With the 460 set for Input A (Ch A button), use the down arrow to set the highlight on OHM. Press Enter.
2. Connect the probes across the resistance source. The resistance value is shown in two forms. First as alphanumeric text at the top of the screen. Secondly, the value is shown visually against a calibrated bar. The value of the bar is shown at the base of the screen. For instance, if a resistance of 12.00 K Ohms is being read, a bar of value 40K Ohms will typically be displayed, and the actual value of the resistance will show as 12.00 against that bar
Testing Continuity
1. With the 460 set for Input A (Ch A button), use the down arrow to set the highlight on Beep. Press Enter.
2. Connect the probes across the source to be tested. If the line is closed (equiv. of short circuit) or not offering more than 100 Ohms of resistance, the full beep will be heard, and the bar will show the degree of resistance (if any).
Note: The continuity test function can also be used to test the resistance in a path up to a value of 4K Ohms. The actual value will be shown alpha­numerically in the top left of the screen, and also shown graphically against a bar having a span of 0 to 4K Ohms
Diode Test
1. With the 460 set for Input A (Ch A button), use the down arrow to set the highlight on DIODE. Press Enter.
2. Connect the probes across the diode in the +/+ direction. A total resistance should be noted (The full bar shows no reverse current flow).
3 Connect the probes to the diode in the normal +/- direction. The bar display should show
an open or almost open condition. - Forward bias is shown against a scale of 2.0 volts and is also displayed alpha-numerically at top left of the screen.
460 User’s Guide Getting Started
C2-4
Chapter 3
Operation
This chapter provides a general overview of the tpi Scope Plus 460 from an operations point of view. The keys and adapter connectors will first be identified followed by a pictorial description of the key and screen menu functions.
Principal Connectors and keys
Figure 3-1 shows the general layout of the front panel and side connectors. Many of the func­tions are straight forward and identified in the illustration. The numbered items in the diagram need additional explanation and are detailed below:
1. Ch A
This key controls the display of oscilloscope Channel A. The first screen requires the user to first identify the nature of the display, e.g., AC Volts, DC Volts, Frequency, etc. Then, after pressing ENTER, the Channel A trace is displayed as well as the correct alphanumerics.
2. Ch B
This key controls the display of oscilloscope Channel B. The first screen requires the user to first identify the nature of the display, This list is different from
Channel A and includes specialized measuring scales such as
o
C, oF and Psi, etc.
After selecting the scale, press ENTER , and the Channel B trace is displayed.
3. S TIME nS
This rocker switch controls the oscilloscope timebase. Pressing the left side decreases the sweep rate in 1, 2, 5, increments down to 2-seconds per division. Pressing the right side speeds the sweep rate in increments up to 50-nano seconds per division. Sweep Rate is displayed above the TRIGGER label at the base of the screen
4. mV / V
This vertically orientated rocker switch controls the vertical sensitivity for the A Channel oscilloscope. The scale is in 1, 2, 5, increments from 50mV to 200V per vertical division. The readout is at the extreme left of the display just above the A MOVE label.
5 mV / V
This rocker switch is identical to Callout #4 except applicable to Channel B. The readout is at the extreme right of the display just above the B MOVE label.
6. <Sun Icon>
This is the brightness button. It controls the screen background intensity of the
460 User’s Guide Operation
C3-1
backlit LCD screen. The button has ten cycles of brightness, after which it resets to the first position.
Figure 3-1. 460 Front Panel layout
Ch A COM Ch B
F1 F2
F3
F4
Ch A
Ch B
S TIME nS
ENTER
mV
V
mV
V
AUTO
HOLD
RUN
SCOPE
USER SAVE
POWER ON
Channel B Connector
Channel A Connector
Common (Ground)
RS-232C
Connector
Power
Plug
Alphanumeric
Display Area
Dual Trace
Display Area
Menu Area
Menu Function
Keys (F1-F4)
Highlight Direction
Arrows
460 Power
Switch
Function
Acknowledge
Key
1
3
4
2
5
6
7
8
9
11
10
460 User’s Guide Operation
C3-2
7. AUTO
This is the Auto Ranging and optimal setting button. In the AUTO mode, the 460 will sense the parameter being measured and automatically adjust the oscilloscope timebase to suit. The button also toggles between AUTO and MANUAL modes. Note that the previous manual settings are lost once auto mode has activated.
8. SCOPE
This is a dual function oscilloscope switch. Initial activation (Scope /S1) sets the oscilloscope program to display the SCOPE INPUTS interactive menu. Activating the SCOPE switch a second time, sets the program to display the Channel A waveform, the Channel B waveform (if it is switched ON), and the A / B MOVE, TRIGGER / TREND selection menu. Pressing the switch a third time will toggle the program back to SCOPE INPUTS.
9. HOLD / RUN
This is a toggle switch that allows the user to freeze the currently dis played waveforms. Once frozen, the user can choose to record the waveform using the SAVE key (see callout #11), or to press the HOLD / RUN key a second time to unfreeze the display.
10. USER
This is a toggle switch that allows a user to interrupt most ongoing programs to revert to a setup mode to adjust Contrast, RS-232C output, or Power Down, and then revert to the original program by pressing the button a second time. Note that the POWER DOWN menu selection allows an automatic timed power down for special functions.
11. SAVE
This is a toggle switch that allows a screen save (traces only) for a maximum of two saves and/or recalls. You can delete an existing capture, or automatically overwrite to the selected memory. Pressing the button a second time reverts to the original display.
Key / Screen Menus
The following tables provide a graphical explanation of the key and menu interactions and menu item sequencing. Most sequences begin with a key and are at the left of the diagrams. The col­ors of the keys are below the appropriate boxes.
Note that the two Scope entries (Scope / S1 in Table 3-1, and Scope / S2 in Table 3-2) represent the two sequential functions of the same switch.
460 User’s Guide Operation
C3-3
DIODECHA
ACV
DCV
FREQ OHM BEEP
CHB
OFF ON
Blue
Blue
ACV DCV FREQ
1mV/A 10mV/A
100mV
/A
1mV/A 10mV/A
100mV
/A
1mV
/
o
C
1mV
/
o
F
1mV
/Psi
1mV /inHg
ACA
DCA
SCOPE
/ S1
F1
F2
F4
A
ACV
DCV GND
GLITCH
MODE
AUTO NORM SINGLE
Brown
OFFB ACV DCV GND
PROBES SELECT
SEL B 1:1 10:1
SEL A 1:1 10:1
Probe Adjust
Enter
NOSAVE
SAVE
RECALL DELETE
SCREEN
1
SCREEN
2
Brown
Table 3-1. 460 User Function / Menu Interactions – Part 1
DIODECHA
ACV
DCV
FREQ OHM BEEP
CHB
OFF ON
Blue
Blue
ACV DCV FREQ
1mV/A 10mV/A
100mV
/A
1mV/A 10mV/A
100mV
/A
1mV
/
o
C
1mV
/
o
F
1mV
/Psi
1mV /inHg
ACA
DCA
SCOPE
/ S1
F1
F2
F4
A
ACV
DCV GND
GLITCH
MODE
AUTO NORM SINGLE
Brown
OFFB ACV DCV GND
PROBES SELECT
SEL B 1:1 10:1
SEL A 1:1 10:1
Probe Adjust
Enter
NOSAVE
SAVE
RECALL DELETE
SCREEN
1
SCREEN
2
Brown
Table 3-1. 460 User Function / Menu Interactions – Part 1
460 User’s Guide Operation
C3-4
USER
F1
F2 F3 F4
EXIT
Brown
SCOPE
/ S2
F1
A
MOVE
F2
TRIGG-
ER
F1 EXIT
Brown
F2
SOURCE
A, B
F4
TRIG. SLOPE
F3
TRIG. LEVEL
Table 3-2. 460 User Function / Menu Interactions – Part 2
CON-
TRAST
RS-
232C
POWER
DOWN
F3
TREND
F1 EXIT
F2
MAX
F4
RE-
START
F3
MIN
F4
B
MOVE
AFTER 10 MIN
AFTER 20 MIN
AFTER 30 MIN
DIS-
ABLE
460 User’s Guide Operation
C3-5
Loading...
+ 44 hidden pages