PGM100
Carbon Monoxide
Single Gas Monitor
The Value Leader
TM
www.tpi-thevalueleader.com
![](/html/3b/3b4e/3b4e8fc8f1eca1824abf7eaaf9a8349e54432275b0ec1b5494029946c42735fc/bg2.png)
Contents
Agency Listings................................................... 1
Safety Information................................................ 2
General Description & Physical Specifications.............. 3
Product Features..................................................4
Sensor Specifications, Alarm Settings, Bump Testing..... 5
Adjust Date / Time................................................6
Alarm Adjustment.................................................7
Calibration......................................................... 8
Operation and Use................................................9-10
Warranty and Repair Policy..................................... 11
![](/html/3b/3b4e/3b4e8fc8f1eca1824abf7eaaf9a8349e54432275b0ec1b5494029946c42735fc/bg3.png)
TPI PGM100
SINGLE GAS MONITOR
INSTRUCTION MANUAL
For use with Carbon Monoxide (CO)
Read and understand instructions before use.
Warning: To prevent ignition of flammable or combustible atmos-
pheres, disconnect power before servicing.
Warning: To reduce the risk of ignition of a flammable atmosphere, bat-
teries must only be changed in an area known to be nonflammable.
Avertissement: Pour reduire le risque d’fallumage d’fune atmosphere
inflammable, des batteries doivent seulement etre changees dans un
secteur connu pour etre inflammables.
IP65
EMC Directive (2004/108/EC)
EN 61000-4-2:1995, EN 61000-4-3:2002, EN 55011:2007
Class 1, Groups A, B, C & D
Exia, Temp code: T4
1
![](/html/3b/3b4e/3b4e8fc8f1eca1824abf7eaaf9a8349e54432275b0ec1b5494029946c42735fc/bg4.png)
Safety Information
1. IMPORTANT:
Read and understand this manual prior to use.
2. IMPORTANT:
Oxygen instruments will show high readings for up to
30 minutes if taken from warm air to cold temperatures and vice-versa. For
best results let them "soak" in the ambient conditions prior to use.
3. WARNING: Substitution of components may impair intrinsic Safety.
4. WARNING: This instrument contains a lithium battery which may
leak or explode if improperly used. DO NOT DISPOSE OF IN A FIRE.
5. WARNING: Only service in an area known to be free of combustible
gases.
6. WARNING: Instruments are not certified intrinsically safe in environ-
ments above 21% oxygen.
7. WARNING: To verify operation prior to each day's use a function test
(bump test) should be performed. If the instrument does not pass this
test, full calibration should be performed.
8. WARNING: Keep all openings free from dirt, debris and foreign
objects.
9. WARNING: Do not use a damaged or improperly operating instru-
ment. Contact a service representative immediately.
ONL
Y zero instrument in a gas free environment
SEULEMENT l'instrument zéro dans un gas libèrent l'environnenment.
WARNING: To maintain intrinsic safety, service must be performed by factory authorized technicians with approved replacement parts only.
AVERTISSEMENT:
Pour maintenir la sûreté intrinsèque, service doit être
exécuté par les techniciens autorisés par usine avec les pièces de rechange
approuvées seulement
2
![](/html/3b/3b4e/3b4e8fc8f1eca1824abf7eaaf9a8349e54432275b0ec1b5494029946c42735fc/bg5.png)
General Description
The TPI PGM100 is an exceptionally rugged personal gas monitor. The
PGM100 is available in single gas configurations only. Preset alarms provide visual, audible and vibration alerts for low and high alarm conditions.
Built-in "Time Weighted Average" (TWA) and "Short Term Exposure Limit"
(STEL) alarms provide even greater margin of safety for the user. A custom display provides the user with easy to view information including gas
concentration, time remaining in months to sensor replacement, battery
low alert and activation of calibration process. (See Figure 1)
FUNCTION INDICATORS
• Battery Low
• Operation Time Remaining (months)
• Calibration
• High/Low Alarm Set Points:
Factory preset to OSHA requirements
• TWA/STEL Alarms: Preset to OSHA limits
• Other alarm settings available
• Data logging: Up to 100 alarm events
• Continuous "ON" Option
PHYSICAL SPECIFICATIONS
Operational Temp: -4 to 122° F (-20 to 50° C)
Humidity: 15-90% RH (Non-condensing)
Duty Cycle: Continuous
Response Time: T90 < 30 seconds
Alarms: Sound: 90 Db @ 12" (30cm)
Sight: Display Alerts 270° viewable LEDs
Touch: Vibration
Power Source: 3.6V Lithium Battery
Size: 3.81" x 2.31" x 0.8" (9.67 x 5.87 x 2 cm)
Weight: 3.6 oz. (105 g)
Battery Life: 2+ Years
Sensor: Electrochemical
Construction: Environmental resistance IP65
Impact resistant rubberized housing
Durable belt clip
3
Figure 1: Custom Display