TPI 565C1 User Manual

The Value Leadertm
www.tpi-thevalueleader.com
TPI 565
Visit TPI, The Value Leadertm at:
www.tpi-thevalueleader.com
Test Products International, Inc. Headquarters: 9615 SW Allen Blvd. Beaverton, OR 97005 USA 503-520-9197 • Fax: 503-520-1225 e-mail: info@tpi-thevalueleader.com
Test Products International, Ltd. 342 Bronte St. South Unit #9 Milton, Ontario L9T 5B7 Canada 905-693-8558 • Fax: 905-693-0888 e-mail: info@tpicanada.com
Test Products International UK Ltd. Longley House, East Park Crawley, West Sussex RH10 6AP England Tel: +44 (0)1293 561212 Fax: +44 (0)1293 813465 e-mail: info@tpi-uk.com
Hotwire Air Velocity Meter
8
555M © 2004
1
Contents
3 Introduction
4 Instrument Overview
4 Operating Instructions
5 Volume Measurements CFM
5 Recording Data
6 RS232 Interface
6 Specifications
7 Calibration & Service
7 Guarantee
7 Troubleshooting Guide
CALIBRATION & SERVICE
It is recommended that the instrument be calibrated every 12 months. Please consult Test Products International for further details.
GUARANTEE
Your TPI 555 Vane Air Velocity Meter is guaranteed free from defects in materials and workmanship for 3 Years from the date of purchase.
Covered by TPI: - Repair parts and labour; or replacement of the product at the option of TPI. Normal transportation charges to the purchaser are also covered.
Not covered by TPI: - Damage to the product which are the result of abuse, im­proper use or maintenance are not covered. Any other expenses, consequential damages, incidental expenses including damages to property are not covered. Transportation expenses to the customer are not covered.
To obtain warranty performance: - Include with the product your name, address, phone number, written description of the problem and proof of purchase date. Carefully package and return to TPI.
This guarantee does not affect your statuary rights.
Trouble Shooting Guide
Problem Solution
Unit will not turn on Battery voltage is low, change batteries.
2
7
RS232 Output
Holding down "RCL" button will display "MAX, "MIN", "AVG". And then "RS-232" at the top right of the screen.
This function prints the reading currently being measured in ASCII code.
RS232 cable is needed in order to view the readings on your computer.
SPECIFICATIONS
Instrument General
Operating Temperature Range -10˚C to +50˚C
Operating Humidity Less than 95% non-condensing Battery 6 each 1.5Volty “AAA” Alkaline batteries Battery Life > 30 Hours Continuous Use Display Backlit Dual LCD with function annunciators Dimensions 74mm x 144mm x 29mm Weight 150g Casing Rubber Boot as Standard Switch Off Auto Power Off after 10 minutes
Sensors
Temperature Measurement -20˚C to +80˚C (-5˚F to 175˚F)
Temperature Accuracy ±1% of reading ±0.5°C ±1% of reading ±1.0°F
Velocity Measurement 0.2 to 20 m/s Velocity Accuracy ±5% of reading ±3 digits Units of Measure m/s, km/s, ft/min, knots, mile/h
1. Introduction
Thank you for purchasing TPI brand products. The TPI 565 Hot Wire Air Velocity Meter is a state of the art, easy to use instrument designed to provide temperature and air velocity readings. The instrument is ruggedly constructed and comes with a 3 Year Guarantee.
This manual will guide you through the functions of the TPI 565 which will give you many years of reliable service.
Your TPI 565 Hot Wire Air Velocity Meter comes complete with the following items as standard:
TPI 565 Instrument Rubber Boot Soft Carrying Case Batteries
Instruction Manual
Your TPI 565 Hot Wire Air Velocity Meter has the following options available:
Serial Computer Interface RS232
6
3
Instrument Overview
RS232 Output
Dual LCD
Unit Key
Hold Key
Power Key
Record Key
Recall Key
°C or °F key
Zero Key
Battery Compartment
Operating Instructions
1. Push ON button to turn meter on.
2. Pressing the 'C/'F button will toggle between 'C and 'F.
3. Push UNIT button to select the desired display units of the
air velocity. 5 units: - m/s, km/s, ft/min, knots, mile/h
4. Position the probe at desired location of measurement.
5. Read the temperature and velocity on the LCD.
6. It will take several minutes until the readings get stable after the probe is positioned.
7. Press "HOLD" button to freeze the display after taking a measure­ment is finished.
8. "DH" will be displayed on the LCD.
9. Press "HOLD" button again to return to normal operation.
Battery should be replaced when "LBT" is displayed at the top left of the screen.
Sensors
Remote Probe
Volume Measurements
To determine CFM (cubic feet per minute) in a duct, the area of the duct must first be measured (use the equations below). Then multiply an air velocity measurement by the area measurement to obtain CFM.
Area Equation for Square Ducts:
Height (H)
Area (A) = Height (H) X Width (W)
Width (W)
Area Equation for Round Ducts:
Radius (R)
Area (A) = π X R
Where R
π = 3.14
2
= (Radius x Radius)
2
CFM (ft3/min) = Air Velocity (ft/min) X Area (in square Inches)
144
Recording Data
If you want to record the reading changes, press "REC" button when the reading gets stable. Once activated the meter will begin recording.
-The REC annunciator will be displayed. The meter will record minimum, maximum and average temperatures.
-To recall the data after recording, press "RCL" button. The "MAX", "MIN", "AVG" values will be recalled sequentially.
If you press "RCL" button, both temperature and humidity will be dis­played at the same time on the LCD. The screen will display "MAX", "MIN", "AVG" sequentially.
Press "REC" button to return to normal operation. The meter has auto power off feature on normal operation.( 10 minutes) Auto power off is disabled on record mode.
4
5