TPI 3433 User Manual

Temperature Tester 343
Instruction Manual
Test Products International, Inc.
9615 SW Allen Blvd., Ste. 104
Beaverton, OR USA 97005
503-520-9197 • Fax: 503-520-1225
info@testproductsintl.com
Test Products International, Ltd.
342 Bronte Road South, Unit 9
Milton, Ontario L9T5B7
905-693-8558 • Fax: 905-693-0888
Test Products International, Europe Ltd.
Longley House, International Drive, Southgate
Crawley, West Sussex RH10 6AQ
+44 (0) 1293 561212 • Fax: +44 (0) 1293 813465
L343 • 8/16/05 copyright © 2005 Test Products International, Inc.
Canada
England
EC DECLARATION OF CONFORMITY & GENERAL GUIDELINES 1
EC Declaration of Conformity
This is to certify that the 343 conforms to the protection requirements of the council directive 89/336/EEC, in the approxi­mation of laws of the member states relating to Electromagnetic compatibility by application of the following standards:
EN 50081-1 1992 Emissions Standard EN 50082-1 1992 Immunity Standard
To ensure conformity with these standards, this instrument must be operated in accordance with the instructions and speci­fications given in this manual.
General Guidelines ALWAYS
• Test the 343 before using it to make sure it is operating proper­ly.
• Inspect temperature probes before using to make sure there are no breaks or shorts.
• Double check all connections before testing.
NEVER
• Insert the probe into an electrically charged medium.
• Touch the probe tip and shaft after it has been inserted in hot material.
TECHNICAL DATA 2
Features and Benefits
Water Resistant Operates in any environment.
Auto Power Off Instrument automatically shuts off after 20 minutes of inactivity.
Field Calibration Perform ice bath calibration via the front keypad.
Specifications
Input Type : K-Type Thermocouple Probe Range Accuracy
General Specifications
Display LCD, 4 Digits, 9999 counts Update Rate 2.5 times per second Operating Temperature 32° to 122°F (0° to 50°C) Storage Temperature -13° to 158°F (-25° to 70°C) Power Supply 9 Volt Alkaline Battery Battery Life 200 hrs. Typical Size (HxLxW) 40mm x 190mm x 65mm (1.5” x 7.25” x 2.5”) Weight 315g (11.1 oz)
MEASUREMENT TECHNIQUES 3
WARNING: Please follow manufacturers test procedures whenever possible. Make sure probe tip and shaft are thoroughly cleaned before inserting it into any food product.
1. Controls and Functions:
Push Buttons
Turns the 343 on and off.
°C/°F Selects Centigrade of Fahrenheit
Min/Max Selects Min or Max reading measured
T1/T2/T Selects probe and differential mode
HOLD Freezes the reading on the LCD
2. Step by Step Procedures
a. Connect a temperature probe or probes to the 343.
b. Turn the 343 on.
c. Insert the tip of the probe into the substance being mea-
sured.
d. Read the temperature on the LCD, allowing approximately
30 seconds for the reading to stabilize. Use the push but­tons to switch between probes and access other funtions as described above.
FIELD CALIBRATION 4
The 343 has the ability to be field calibrated. Accuracy of +/-1°F within the 30°F to 120°F range can be achieved by performing an ice bath calibration.
The temperature of an ice bath is approximately 32°F(0°C). Fill a plastic or metal container with crushed ice and add clean water to a depth of at least 4 inches. Stir the ice and water for 2 to 3 minutes prior to performing calibration to ensure the water is completely chilled. Make certain there is plenty of ice in the mixture and always use clean water. Distilled water works well.
1. Step by Step Procedures
a. Connect both temperature probes to the 343.
b. Press and hold down the MIN/MAX and HOLD buttons and
turn the 343 on. CALF will be displayed and the 343 will enter calibration mode and begin cycling between T1 and T2.
c. Insert the probes into the ice bath and allow the readings to
stabilize.
d. Once the reading on both inputs has stabilized, press the
HOLD button. SAVE will be displayed and calibration is complete.
FOR FULL FACTORY CALIBRATION PLEASE RETURN YOUR THERMOMETER TO:
TPI / ATTN: SERVICE DEPT.
9615 SW ALLEN BLVD SUITE 104
BEAVERTON, OR 97005
Standard Accessories Part No.
9V Battery A009 Temperature Probe (2 each) GK13M Protective Boot A303 Soft Pouch A255 Instruction manual
Optional Accessories Part No.
Probes See Catalog
MAINTENANCE
1. Battery Replacement: The 343 will display battery symbol when the
internal 9 Volt battery needs replacement. Replace battery as follows:
a. Disconnect and remove all temperature probes from 343.
b. Loosen the 4 screws from the back of the 343.
c. Remove the back cover.
d. Remove old battery and replace it with new battery.
f. Reassemble instrument in reverse order from above.
2. Cleaning your 343
Use a mild detergent and slightly damp cloth to clean the surfaces of
343.
Loading...