Toxalert Tox-ETO-ANA User Manual

TOXALERT
ETHYLENE OXIDE (ETO) SENSOR MODEL TOX-ETO/ANA
FEATURES:
•••• Linear 4-20 mA output signal •••• Interfaces directly to DDC Systems.
••• Automatic low sensitivity check •••• Output locked during warm up and
••• Low voltage circuits calibration.
••• Sensor failure indication •••• Explosion proof housing.
GENERAL DESCRIPTION:
The Model TOX-ETO/ANA, self-contained Ethylene Oxide sensor module, utilizes electrochemical technology to provide accurate monitoring of Ethylene Oxide (C
It is designed for fixed installations and use in ambient monitoring applications. The sensor module is a 4 to 20 mA transmitter designed for use in conjunction with Toxalert’s Models TOX-4ANA and ToxControl Controllers or any industry standard 4-20 mA loop controller.
The sensor module requires a three wire connection to a controller that provides a nominal 28 vdc power and the module sources current in proportion to the gas concentration.
GUIDE SPECIFICATION:
Provide an Ethylene Oxide (ETO) sensor(s) with a range of 0 to 20 ppm (parts per million) and a linear output of 4 to 20 mA over its range. The sensor shall be housed in an explosion proof housing and have explosion proof conduit connection to the housing to maintain system integrity. The sensing element shall be electrochemical and have a life expectancy of two years. The sensor response time shall reach 90% of level being sensed in 90 seconds. The sensor shall be powered by low voltage and have a self check capability and an LED to indicate sensor “Low Sensitivity.” If the sensor fails, its output shall drop to zero to signal the controller of failed condition. The sensor shall be as manufactured by Toxalert, Inc.
O) in the range of 0-20 ppm.
2H4
1 of 2
TOXALERT INTERNATIONAL, Inc., PO Box 159, Mound, MN. 55364 PHONE: (952) 472-4541 • FAX: (952) 472-4960
www.toxalert.com
MODEL TOX-ETO/ANA
SPECIFICATIONS:
•••• Type: Electrochemical cell-diffusion barrier type •••• Operating Temperature : 4° to 122°F
•••• Range: 0 to 20 ppm ethylene oxide in air -20° to 50°C
•••• Response Time: 90 seconds to 90% indication •••• Relative Humidity: 15 to 90% RH non-
•••• Typical Life: Two years in normal service condensing
•••• Electrical Data:
Input Voltage: 14 – 30 vdc Output Signal: Linear 4 to 20 mA Input Power: 1.4 Watt Signal Output-Trouble: 0 mA Input Current: 40 mA @ 28 vdc Wiring: 3 wire non-isolated Loop Resistance: 800 Ohms
•••• Unit Construction:
Mounting: Conduit ¾” NPT or flexible cable Physical: H: 7.6”, D: 4.6”, W: 4.2” Housing: Explosion proof (Nema 7) Weight: 2.7 pounds
(Div 1, Class 1, Groups C, D)
Gas Sensor Modules available from Toxalert include:
Ammonia (NH3)
Combustible Gas Carbon Monoxide (CO) Carbon Dioxide (CO2)
Chlorine (Cl2)
Oxygen (O2)
Hydrogen (H2)
Hydrogen Chloride (HCl) Hydrogen Cyanide (HCN) Hydrogen Sulfide (H2S)
Nitric Oxide (NO) Nitrogen Dioxide (NO2)
Ethylene Oxide (ETO) Refrigeration And more….
Tox-ETO 2 of 2 11/97
TOXALERT INTERNATIONAL, Inc., PO Box 159, Mound, MN. 55364 PHONE: (952) 472-4541 • FAX: (952) 472-4960