Toshiba TIC-LF232A User Manual

Field Intelligent Device Series
Field Intelligent Device Series Electromagnetic Flowmeter Converter
Sanitary Electromagnetic Flowmeter
LF232
Introduction
The electromagnetic flowmeter uses Faraday’s Law of electromagnetic induction to measure the process flow. The device consists of two units: a detector, through
the fluid to be measured flows and in which
*1: HART protocol (Highway Addressable Remote
Transducer) is a communication protocol for industrial sensors recommended by the HCF Communication Foundation).
detector, and amplifies the signals from
LF232 can
1
(HART
Specifications
Overall Specifications
Measurement range in terms of flow velocity:
Conbination LF150, LF470
0 –1.0 ft/s to 0 – 32.8 ft/s (0 – 0.3 m/s to 0 –10 m/s)
Fluid conductivity: 5µS/cm or more
Ambient temperature: -4 to 140 °F (-20 to 60 °C)
(Storage temperature: -13 to 149 °F (-25 to 65 °C))
Structure: IP67 and NEMA 4X Watertight
Figure 1. LF232 Converter
Accuracy: the accuracy is dependent on the type of
detector combined with the LF232. See the following tables for accuracy when combined with specified detectors.
Detector combined
LF470 (Meter size 1/10", 1/6”, 1/4"
percent of range
050 ±0.8FS
50100 ±0.8FS ±0.8 of rate
Detector combined
LF150 (Meter size 28” to 120” (700 to 3000 mm))
0.3–1.0m/s 1.0–10m/s
(2.5 to 6 mm))
Accuracy Flow rate as a
±0.4FS
Accuracy: ±0.5 % of Rate*
* This pulse output error result is established under
standard operating conditions at Toshiba’s flow calibration facility, Fuchu Japan.
* Individual meter’s measurement error may vary up
to ± 0.8 % of Rate at 3.28 ft/s (1.0 m/s) or more and ± 0.4% of Rate ±0.157 inch/s (4mm/s) at 3.28 ft/s
(1.0 m/s) or less. * Current output: plus ± 8µA (0.05% of span). * Refer to individual calibration data for each
individual meter’s measurement error.
TIC-LF232A
TIC-LF232A
Power supply:
One of the following can be selected
100 to 240 Vac
(Allowable voltage 80 to 264 Vac, 50/60Hz)*
24 Vdc
(Allowable voltage 21 to 27 Vdc)*
*1 When the 7
th
digit of specification code is “B”,
2
1
power supply is 100 to 120Vac (Allowable voltage range is 80 to 132VAC, 50/60Hz) *2 When the 7
th
digit of specification code is “B”,
24Vdc cannot be selected.
Power consumption:
17W (27VA) or less
When 7th digit of specification code is “A”)
(
35W (120VA) or less
(When 7th digit of specification code is “B”)
Converter Specifications
Input signals
Analog signal — the voltage signal from detector,
proportional to process flow rate.
Digital input DI (opt.) (added when 9
th
digit of specification code is ”3”)
Voltage level: High level 20 to 30 V dc
Low level 2 V dc or less
Input resistance: 2.7 k
Number of inputs: 2 points (DI1 and DI2)
DI functions:
For digital input function, either of the following functions can be selected.
• Range switching One Digital Input is used: switching between large
and small ranges of forward/reverse, 2-range measurement.
Two Digital Inputs are used: switching between ranges of single direction, 4-range measurement.
• Totalizer control input
Totalizer Start/Stop control or Reset/Start
•Output hold input
Fixed with set value for current output and pulse output (loop check)
• Zero adjustment Starts zero adjustment (on-stream at zero flow rate)
when DI voltage level goes low after remaining high for 10 to 20 seconds.
Output signals
Current output:
4–20mAdc (load resistance 0 to 1K)
Digital output (DO) — One point (std.) and three
more point is optionally available as follows.
Digital output DO1 (std.):
Output type: Transistor open collector Number of outputs: One point Output capacity : 30Vdc, 200mA maximum
Digital output DO2DO4 (opt.) (Added when 9
th
digit of specification code is “3”)
Out put type: Solid-state relay output (non polarity) Number of outputs: 3 points Output capacity: 150Vdc, 150mA maximum or
150Vac (peak to peak), 100mA maximum
DO functions—Four digital output function, either of the following functions can be selected.
• Totalizer pulse output
DO1 or DO2 can be selected (Forward direction pulse and reverse direction pulse can be assigned independently)
In the case of DO1
Pulse rate: 3.6 to 3,600,000 pulses/h
Pulse width:Settable within the range of 0.3 to
500ms
In the case of DO2
Pulse rate: 3.6 to 360,000 pulses/h
Pulse width: Settable within the range of 4 to
500ms
Rage switching output One Digital Output (DO) is used
Single direction, 2-range switching signal Forward/reverse direction switching signal
Two Digital Outputs (DO) are used
Single direction, 4-range switching signals Forward/reverse 2-range switching signals
High/low limit alarm output
An alarm is output when flow rate goes above or
below the set-point value
Setting range: -10 to 110% of the settable
maximum range
High limit 2 points, low limit 2 points can be set At the time of alarm output, Normally Open or Normally Closed contact can be selected
• Preset point output
Contact ON when totalizer count exceeds the set value
Setting range: 1 to 99999999 count
• Converter error alarm
An alarm is output when an operation error is detected by self-diagnosis. At the time of alarm output, Normally Open or Normally Closed contact can be selected
2
Communications signal:
A digital communications (HART protocol) signal is superimposed on 4 to 20 mA dc analog output signal.
Load resistance: 240 to 1 k Load capacitance: 0.25 µF or less
Load inductance: 4mH or less (For maximum cable length, about 2km is a guideline length when CVV-S 1.25mm
2
is used
under standard installation condition.)
Note: HART (Highway Addressable Remote Trans-ducer) is a communications protocol for industrial sensors recommended by the HCF (HART Commu-nication Foundation).
Output display:
16-character×2-line dot-matrix LCD (With back light).
2 units can be selected from the flowing units: flow velocity, instantaneous flow rate, total flow (forward/reverse/difference flow), total count, %, custom unit.
Parameter settings—Parameters can be set asfollows
IR Switches: Various parameters can be set without
opening the converter housing using 4 infrared switches (password can be set)
•Zero adjustment: Zero point adjustment can be started by pressing the switch in the converter.
Digital communication: The AF 900 hand-held
terminal is needed to set parameters.
Damping: 0.5 to 60 s (selectable in 1s increments)
“Field re-verification”
Mag-Prover – Toshiba’s
Zero span calibration tool allows unit to be re-calibrated and verified using internal software program. (For more information contact Toshiba International Corp.)
Conditions when power fails:
Parameter setting values are stored in non-volatile memory and the values will be restored when the power returns to normal condition. The output and display will stay as follows when power fails.
Current output: 0 mA Digital output: OFF (contact open)
LCD display : No display
TIC-LF232A
Surge protection: Surge protectors are installed in the
power supply, excitation circuit, current signal output and digital I/O circuit.
Terminal block structure: 21-pole, screw connection
type (M4 screw)
Housing: Aluminum alloy Coating: Acrylic resin-baked coating, pearl-gray
colored
Cable connection ports :
A cable gland is provided for each port.
OD of cable φ11 to 13 mm Material Nylon 66 G (PF) 1/2 male screws.
Vibration resistance:
No resonance to the following levels of vibration:
• 10 to 55Hz with amplitude of 0.07mm
• Vibration of 30Hz with 29.4 m/s
2
in 4h in each
direction will not cause any defect to unit.
Note: If the flowmeter is intended to be used in a
location where vibration is applied constantly contact Toshiba.
Dimension and Weights:
See Figure 2
3
TIC-LF232A
r
Installation
Dimensions
9.33 (237)
8.74 (222)
2.91 (74)
4-φ12 hole
6.10 (155)
Installation method
据付方法
Installed by panel or pipe
据付(面)
Appointed bout and nut fo
パイ取付け
installation is attached.
指定された取付用ナッ 属し
In the case of 2B pipe installation
2Bイプ取付けの場合
M10 U ボル
M10 U bolts
Signal cable ground
Grounding terminal
Excitation cable ground
2.99 (76)
2.99 (76)
2.99 (76)
Output cable ground
1.57
(40)
2.17(55)
I/O cable ground
Weight : Approximately18 lb (8 kg)
15.75 (400)
Power supply cable ground
(including a mounting bracket)
14.57(370)
2B pipe
2B(50mm)
In the case of panel installation
ネル取付穴加工寸
11 to 12 φhole
2.91 (74)
74
1112φ (4ヵ)
4 places
370
14.57 (370)
400
15.75 (400)
Figure 2. Separate type converter LF232
4
(
r
r
External Connections
Terminal box
wire
5.5mm
2
or more
Grounded with 100Ω o less ground resistance
Thick walled steel conduit
Signal cable (2-wire shielded hard-rubber sheathed cable)
(3-wire hard-rubber sheathed cable)
Detector
Excitation cable
Output cable (CVV-S)
Digital I/O cable
(CVV-S)
Figure 3. External Wiring Diagram
TIC-LF232A
Power supply
CVV)
Power supply
Current output (4 to 20mA)
Digital input 2 (opt.)
Digital input 1 (opt.) (20 to 30 Vdc)
Digital output 4 (opt.)
Digital output 3 (opt.)
Digital output 2 (opt.)
Digital output
Grounded with 100Ω o less ground resistance
5
TIC-LF232A
Wiring Precautions
(1)Be sure to use thick walled steel conduit (22mm)
for signal and excitation cable wiring between the detector and converter. Use flexible conduit at the cable outlets of the detector and converter.
(2) Connect the grounding wire (IV wire 5.5mm² or
more) to a good earth ground (100 or less ground resistance). Make the wire as short as possible. Do not use a common ground shared with other equipment where earth current may flow. An independent earth ground is recommended.
(3) The allowable cable lengths between the detector
and converter for the separate type flowmeter depend on the electrical conductivity of the object fluid. See Figure 4 below.
(4) Digital output 1, digital output 2 to 4 , digital output
1 to 2 are insulated mutually.
300 200
100
ケーブル許容
50
30 20
10
〔m
5
Allowable cable length (m)
3
LF130形、LF131形、LF150形検出器と接続の場合
(1)When combined with LF150.
LF470形検出器と接続の場合
(2)When combined with LF470.
Figure 4. Electrical Conductivity and Cable length
53
Conductivity (µS/cm)
電率〔μS/cm
20 30 50 100 200
10
About establishment environment
Do not store or install the flowmeter:
• Where there is direct sunlight.
• Where excessive vibration or mechanical shock occurs.
• Where high temperature or high humidity conditions exist.
• Places where corrosive atmospheres exist.
• Places submerged under water.
• Place where there is sloped floor. To put the flowmeter temporarily on the floor, place it carefully with something, such as stopper, to support it so that the flowmeter will not topple over.
In areas like the following, there may be the case that infrared switches do not function correctly. (If these are unavoidable, use an appropriate cover.)
(1) Where unit (operation panel) is exposed to direct
sunlight, reflection of light onto window pane and
diffused light reflection. (2) Where smoke and steam may occur. (3) Where exposed to direct snow, ice or mud.
Ordering Information
1. When ordering the LF232 converter, refer to Tables 1 (Type Specification Code). An entry must be made for each of the columns in each of these tables.
2. Measuring range
3. I/O function specifications.
4. Ordering scope:
Actual flow calibration data: (required or not)
5. Specification of combined detector
6. Other scope: Specification other than standard items.
Consult a Toshiba representative before ordering.
6
Table 1. Specification Code for converters (LF232)
Model Specification Code
1 2 3 4 5 6 7 8 9 10 11 12 13 14
L F 2 3 2
1
A
A
A
B
C E
1
3
1 2
Separate type converter Purpose
Standard
Compatible detectors (Note1,2,3)
Medium to small-size detectors: Meter size 1/10” to 18” (2.5 to 450 mm) Large-size detectors: Meter size 28” to 120” (700 to 3000 mm)
Mounting nuts and bolts
Panel, wall mounting (BNP material : SUS304) Pipe mounting (BNP material : SUS304)
Digital input/output
Current output
+ Digital output points (1 point) Current output + Digital output points (4 points) + Digital input points (2 points)
Communication function HART communication Power supply
100Vac-240Vac, 50/60Hz (Note 2) 24Vdc (Note 3)
Standard
Contents
Note 1: For applicable detector code, select one from the following table.
To combine with an existing detector, select one depending on the type of replaceable converters. For combination with detectors not listed below, contact Toshiba.
Applicable
detector code
Applicable
detector
Applicable detector
(discontinued model)
334,
A
LF470 type
335 type (Converters of types shown right) 337,
B
LF150 type
335 type (Converters of types shown right)
Replaceable converters
372, LF200, LF220, LF420, LF230* A type
373, 378, LF230* B type
TIC-LF232A
Note 2: When the 7th digit is “B” and power supply specification code is “1”, power supply specification is 100 to
120VAC.
Note 3: 24VDC can be specified only when the 7th digit is “A”.
7
TIC-LF232
/
ISO9001 and ISO14001 are certified.
Misuse of this product can result in damages to property or human injury. Read related manuals carefully before using this product.
TIC-LF232A
Specifications are subject to change without notice.
Printed in Japan 2008-5 (TDOC)
© TOSHIBA Corporation 2008
All Rights Reserved.