Publish Date: 2000/03/17
.
.
BSI No.: SFB0003
Subject:
Changing the ROM
Model: TF610
Category Field Application Unit
Modification of Software (ROM) To be applied when necessary ROM
Factory Application: From the production of March, 2000
CONTENT
1. Reason for Modification
Change of the setting value of T1 timer
The default value of the T1 timer is 55 seconds. This applies to FCC (within 60
seconds).
However, it is possible to change this value in the option menu to over 60 seconds.
With this change, user will not be able to set it to over 60 seconds.
Parameter #15: T1 Timer
Current: 35 sec, 45 sec, 55sec (Default), 65 sec, 75 sec, 85 sec, 95 sec, 105 sec,
115 sec
New: 35 sec, 45 sec, 55 sec (Default)
.
.
.
.
2. Details of Modification
Location (IC No. / Unit name): ROM1 / MAIN PC board
Change of ROM version (Check Sum):
Written by N. Sasahara
Attached Files
NONE
ROM1 "1420K4TSUS09A41d"(D000) ---> "1421K4TSUS09A41e"(D100)