BSINo. G00C977 Page 1
BSI No.: G00C977 Publish
2000/11/06
Date: Subject: Note for Controller Connection <Correction: Oct. 23, 2000>
Model: FC15, FC22
Category Field Application UNIT
FactoryApplication: -
CONTENT
When the controller (KR7037 or KR7038) is connected to FC-22 made in Dec. and Jan. at the factory, be sure to change the following ROM.
Note
If the controller is connected without ROM change, abnormal signals and image errors may occur.
z
The ROMs should be used in combination as in the following list. If they are used other than in the
z
following combination, errors (abnormal signal, etc.) may occur. However, when there is ROM improvement, it is recommended to use the new combination of the BSI information.
z
When the ROM is changed, be sure to perform
System ROM Main ROM IMC-ROM Scanner ROM AI-ROM
-----------------------------------------------------------------------------------------­Oct. 23, 2000 Change: Addition of model,
[Work after ROM change]
ROM
PMA-T310Sxx-ZZZ PRA-T310M-ZZ PRA-T310IMC-ZZ PRA-T310SCM-ZZ PRA-T511MAC-ZZ
Name
Addition of note
--- Refer to BSI #.G00B931,G00C964
[Work after ROM change]
Feb. at the
factory
150 160
12 13 12 13 11 <-­07 <--
Mar. at the
factory
.