The Datex-Ohmeda Cardiocap/5 and accessories are indicated for indoor monitoring of hemodynamic (ECG,
impedance respiration, NIBP, temperature, SpO
rate, anesthetic agent, and agent identification), ventilatory (airway pressure, volume, and flow), and relaxation
status (NMT) of all hospital patients.
With the N-XOSAT option, monitoring of arterial oxygen saturation includes monitoring hospital patients during
conditions of clinical patient motion.
Cardiocap/5 is indicated for patients weighing 5 kg (11 lb.) or more.
Impedance respiration measurement is indicated for patients ages 3 years and older.
The monitor is indicated for use by qualified medical personnel only.
, and invasive pressure), respiratory (CO2, O2, N2O, respiration
2
CAUTION: US Federal law restricts this device to sale by or on the order of a licensed medical
practitioner. Outside the USA, check local laws for any restriction that may apply.
Classifications
IEC 60601-1:
• Type of protection against electric shock: Class I equipment.
• Degree of protection against electric shock (indicated by a symbol on the panel beside each connector):
Type BF applied part or Type CF applied part.
• The equipment is not suitable for use in the presence of a flammable anesthetic mixture with air or with
oxygen or nitrous oxide.
• Mode of operation: Continuous.
• CISPR 11: Group 1, c lass A
IEC 60529 (degree of protection against harmful ingress of water): IPX1
EU Medical Device Directive: IIb
Responsibility of the manufacturer
GE Healthcare Finland Oy (GE) is responsible for the safety, reliability and performance of the equipment only if:
• Assembly, operation, extensions, readjustments, modifications, service, and repairs are carried out by
personnel authorized by GE.
• Electrical installation complies with appropriate requirements.
• The equipment is used in accordance with the Cardiocap/5 User’s Guide and serviced and maintained in
accordance with the Cardiocap/5 Technical Reference Manual.
GE Healthcare assumes no responsibility for the use or reliability of its software on equipment that is not
furnished by GE Healthcare.
Trademarks
Datex®, Ohmeda®, and other trademarks (Cardiocap/5, AS/3, CS/3, S/5, S/5 Light, D-lite, Pedi-lite, D-fend,
D-fend+, MemCard, ComWheel, EarSat, FlexSat, OxyTip, Patient O
GE Healthcare Finland Oy.
Nellcor® is a registered trademark of Mallinckrodt Inc.
All other product and company names are the property of their respective owners.
, and Patient Spirometry) are trademarks of
2
Product availability
Some of the products mentioned in this manual may not be available in all countries. Please, consult your local
representative for the availability.
Cardiocap/5 Technical Reference Manual
Part I – General Service Guide
Overview
Monitor Structure
Safety Precautions
Product Specifications
Installation and Functional Check
Installation
Interfacing
Functional Check
Functional Check Form
Planned Maintenance
Planned Maintenance Instructions
Planned Maintenance Form
Troubleshooting
Messages
Troubleshooting Charts
Part II – Product Service Guide
1
2
3
4
Frames and Software
Hemodynamic Frame (F-MX)
Hemodynamic with Airway Gases Frame (F-MXG)
Anesthesia and Critical Care Software
Measurement Parameters
Parameter Unit (NESTPR)
Invasive Pressures and Second Temperature Option (N-XP)
The Technical Reference Manual is for use by service personnel who are qualified to perform service
and maintenance procedures on the Datex-Ohmeda Cardiocap/5. The information in this manual is
believed to be accurate and reliable, however, the manufacturer assumes no responsibility for its use.
The manual is organized as follows:
• Part I (chapter 1 – chapter 4) provides a general overview of the Cardiocap/5, including the
information needed to install, checkout, and maintain the monitor. Part I also includes
information for troubleshooting problems that may occur while using the monitor or that you may
encounter while perfoming procedures in this manual, such as the Functional Check, for example.
• Part II (chapter 5 – chapter 9) contains detailed functional descriptions of the Cardiocap/5
hardware and software, including measurement principles and components for each
measurement parameter. Procedures for replacing parts and making adjustments are also
included. Part II also contains illustrations and detailed descriptions of all service screens used
during checkout, maintenance, and other service-related activities. A list of parts with illustrations
is located at the end of Part II.
Overview
Read the entire manual and make sure you understand the procedures described before installing,
repairing, or adjusting the monitor. To avoid risks concerning safety or health, strictly observe all safety
precautions.
This manual relates to the monitor versions 6051-0000-164..00 and ..01.
Related documentation
For information about using the monitor, refer to the following:
Cardiocap/5 for Anesthesia User’s Guide
Cardiocap/5 for Critical Care User’s Guide
Cardiocap/5 for Anesthesia User’s Reference Manual
Cardiocap/5 for Critical Care User’s Reference Manual
For PCA drawings, circuit diagrams, and component lists, order the PCA Drawings Service Kit. See the
Spare Parts chapter.
1-1
Cardiocap/5 Technical Reference Manual
1.2 Cardiocap/5 models and features
The Cardiocap™/5 is a configured monitor that is intended for indoor monitoring of the hemodynamic,
respiratory, relaxation, and ventilatory status of the patient. Two models of the monitor are available:
hemodynamic (F-MX) and hemodynamic with airway gas measurement (F-MXG). Both models can be
equipped with built-in options.
• All measurement parameter options (and the Recorder option, N-XREC) are factory-configured
and cannot be added after purchase.
• Data collection and management options (N-XNET and N-XDNET) can be added later, if the CPU
board supports Network functionality.
1.2.1 Options for hemodynamic model (F-MX)
The F-MX measures NIBP, ECG (3-lead and 5-lead), pulse oximetry (SpO2), temperature (T1), and
impedance respiration. The F-MX can be configured with the following built-in options:
N-XP Two invasive pressure channels and second temperature (T2)
N-XREC Recorder
The F-MX model can also be configured with one of the following built-in options:
N-XNSAT Nellcor® compatible pulse oximetry (SpO
N-XOSAT Datex-Ohmeda enhanced pulse oximetry (SpO
)
2
)
2
1.2.2 Options for hemodynamic model with airway gas measurement (F-MXG)
The F-MXG measures NIBP, ECG (3-lead and 5-lead), pulse oximetry (SpO2), temperature (T1),
impedance respiration, and airway gases. Gas measurement depends on which airway gas option is
installed (N-XC, N-XCO, or N-XCAiO):
N-XC Carbon Dioxide (CO
N-XCO CO
N-XCAiO CO
, N2O, and Patient Oxygen (O2)
2
, anesthetic agents, agent identification, N2O, and O2
2
The F-MXG can also be equipped with each of these built-in options:
N-XP Two invasive pressure channels and second temperature (T2)
N-XV Patient Spirometry (N-XCO or N-XCAiO option required)
N-XREC Recorder
The F-MXG can also be configured with one of the following options:
N-XNSAT Nellcor® compatible pulse oximetry (SpO
N-XOSAT Datex-Ohmeda enhanced pulse oximetry (SpO
N-XNMT NeuroMuscular Transmission (NMT) for relaxation measurement (N-XCAiO option required)
)
2
)
2
)
2
1-2
1.2.3 Data collection and management options (for F-MX and F-MXG)
For both models, these options can be factory-built or added later as upgrades, if the CPU board
supports Network functionality:
N-XNET Network
N-XDNET Data card and Network
1.3 Monitor structure
The Cardiocap/5 can be equipped with several factory-configured options. The block diagram and
descriptions that follow represent the maximum functionality of the monitor with all options installed.
Overview
Figure 1-1. Cardiocap/5 monitor structure
The main software and measurement technologies are based on AS/3 hardware and software. Some
parameter-measuring unit boards are interchangeable with AS/3 module boards, however, the units
cannot be replaced with the corresponding modules as the hardware of the assemblies is different.
1.3.1 Measurement parameter units
The maximum Cardiocap/5 parameter measurement configuration consists of the NESTPR, PVX, and
CAiO units plus ONE of the following units: NMT or NSAT or OSAT.
The NESTPR, CAiO, and NMT units are connected to the CPU through the Mother board and
communicate with the CPU over a standard AS/3 module bus. The NSAT or OSAT pulse oximetry unit is
also connected to the CPU through the Mother board and communicates with the CPU over a standard
AS/3 module bus, which is located on the SpO
Each parameter measurement board contains a CPU that processes measurement data for the
parameter(s) associated with that board before sending the data to the main CPU.
NESTPR unit
The NESTPR unit contains three boards for measuring hemodynamic parameters:
• The ECG board measures ECG (3-lead and 5-lead) and impedance respiration.
• The STP board measures oxygen saturation, temperature, and invasive blood pressure.
interface board.
2
• The NIBP board measures noninvasive blood pressure.
1-3
Cardiocap/5 Technical Reference Manual
PVX unit for Patient Spirometry (N-XV option)
The PVX unit connects to the CAiO unit and measures the patient’s airway flow and pressure (Patient
Spirometry).
CAiO unit (N-XC, N-XCO, and N-XCAiO options)
The CAiO unit measures airway gases. It is capable of measuring CO2, N2O, O2, anesthetic agents (AA)
and also identifying the present anesthetic agent.
The NSAT unit measures oxygen saturation and pulse rate using signal processing electronics and
software that are based on Nellcor stand-alone oximeters.
NeuroMuscular Transmission (N-XNMT option)
The NMT unit measures the relaxation status (TOF, DBS, and ST) of patients. When used with a regional
block cable, the unit acts as a nerve locator.
1.3.2 Communication
The CPU communicates with the hemodynamic parameters measuring unit (NESTPR) and airway gas
measuring unit (CAiO) over a standard AS/3 module bus. It is based on the widely used industry
standard RS485, which uses a differential serial method to transfer data and is quite robust.
RS485 serial communication supports multidrop or party line connections. This means all units
connected to the module bus use the same two physical wires for communication purposes. The
module bus uses a 500 kbps data transfer rate.
Communication with the I/O board and the DC/DC board takes place over an internal synchronous
serial bus. The same bus also controls display brightness and audio signals by means of D/A
converters located on the CPU board.
Communication with the Net takes place over a separate synchronous serial channel. Communication
with the Net is possible, if the CPU board is equipped with an Ethernet controller.
Communication with the Recorder takes place over an asynchronous serial channel.
1.3.3 CPU board
The control functions of the monitor are centralized on the CPU board. The CPU:
• Controls the power on/off sequencing.
• Controls the brightness of the LCD screen by means of the Backlight board.
• Controls the Inverter board that provides the high voltage for the display backlights.
1-4
• Reads input from the keyboard and ComWheel.
• Controls the serial channels and I/O functions of the monitor.
Two PCMCIA-compatible data card slots on the board are for loading software and transferring data.
Distributed processing
The parameter and airway measuring units contain their own microprocessor systems for performing
low level functions, such as waveform filtering and pneumatics control. At the same time, the main
CPU performs higher level tasks (trending and alarm control, for example).
1.3.4 Display
The main CPU directly controls the monitor display, a 10.4 inch color LCD. Supply voltages for the
display are connected via the CPU board and the Backlight board. The Inverter board provides high
voltage for the display backlights. The CPU controls the display brightness by adjusting the backlight
voltage.
1.3.5 I/O board
The I/O board contains D/A converters for analog outputs and the audio output amplifier. It also
contains connectors for the network identification plug, serial I/O, analog output, and external
keyboard.
Analog outputs are created by transferring digital data from the CPU to the D/A converters on the I/O
board through the internal synchronous serial bus. The network identification plug is connected to the
CPU over a separate synchronous serial channel.
Overview
1.3.6 DC/DC board
The DC/DC board converts 15 VDC coming from the AC/DC unit to different supply voltages for the
monitor. All outputs are short-circuit and over-voltage protected. The CPU controls the output voltages.
If the mains voltage drops, the 12 VDC back-up battery automatically supplies power to the monitor.
The battery will run the monitor for at least 15 minutes. The battery is always charged when mains
voltage is connected. The temperature sensor that measures the monitor’s internal temperature is
located on the board. The DC/DC board communicates with the CPU over the internal synchronous
serial bus.
1.3.7 AC/DC unit
The AC/DC unit converts the mains voltage to 15 VDC that is fed to the DC/DC board. The input
voltage range of the unit is 100 to 240 VAC. The DC/DC board shuts down the AC/DC unit when there
is over voltage detected on the 15 VDC output. The shutdown mode is reset by detaching the mains
power cord for 30 seconds.
1.3.8 Recorder (N-XREC option)
The Recorder prints trend data and record parameter waveforms. It connects to the CPU through an
asynchronous serial channel. Recorder supply voltages connect through a Recorder board that is
permanently attached to the recorder mounting box. The board contains a voltage filter and a delay
circuit for 12 V.
1-5
Cardiocap/5 Technical Reference Manual
1.4 Symbol definitions
Symbols on equipment
Attention! Read accompanying instructions, including all warnings and cautions,
before using this device.
This symbol has the following meanings when it appears on the screen:
• On the front panel indicates that protection against cardiac defibrillator discharge is
due in part to the accessories for pulse oximetry (SpO
invasive pressure (P) measurement.
), temperature (T) and
2
Pb
Pb
• When displayed beside the O
value, indicates that the FiO2 low-alarm limit is set
2
below 21%.
• When displayed next to the HR value, indicates that there is a risk that the monitor
counts pacemaker spikes (pacer is set ON R) or the monitor counts T-waves (a wide
QRS is selected).
Type BF applied part (IEC 60601-1). Defibrillator-proof protection against electric shock.
Type CF applied part (IEC 60601-1). Defibrillator-proof protection against electric shock.
Main Menu. Located beside the ComWheel to indicate you can open the Main Menu by
pressing the ComWheel when no other menu is displayed.
Power On/Standby.
This battery contains lead. Separate from other waste for disposal according to local
regulations.
This battery contains lead and can be recycled.
1-6
Dangerous voltage.
Gas outlet (in airway gas models only).
Ethernet connectors.
Equipotentiality. Monitor can be connected to potential equalization conductor.
Alternating current.
Fuse.
ESD warning symbol for electrostatic sensitive devices. Pins of connectors identified
with the ESD warning symbol should not be touched. Connections should not be
made to these connectors unless ESD precautionary procedures are used. See
"Safety precautions: ESD precautionary procedures" in the “User’s Reference
Manual” for details.
Symbol for non-ionizing electromagnetic radiation. Interference may occur in the
vicinity of equipment marked with the symbol.
This symbol indicates that the waste of electrical and electronic equipment must not
be disposed as unsorted municipal waste and must be collected separately. Please
contact an authorized representative of the manufacturer for information concerning
the decommissioning of your equipment.
Symbols on screens
When displayed on the upper left corner of the screen, indicates alarms are silenced.
When in the menu or digit fields, indicates that the alarm source has been turned off.
Sub menu. Selecting an alternative with this symbol in a menu opens a new menu.
Overview
The monitor is connected to the Monitor Network.
Data card (green) and/or Menu card (white) is inserted.
Indicates the beats detected.
Respiration rate is measured using impedance respiration measurement.
Back-up battery operation and remaining capacity.
Back-up battery charging.
Symbols on transport packaging
The contents of the package are fragile and have to be handled with care.
Indicates the correct upright position of the transport package.
The package must be kept in a dry environment.
The package should be kept within the indicated temperature limitations.
1-7
Cardiocap/5 Technical Reference Manual
1.5 Safety precautions
1.5.1 Warnings
Refer to the User’s Reference Manual for additional warnings to be observed while monitoring a patient.
A WARNING indicates a situation in which the user or the patient may be in danger of injury or
death.
Installation
The monitor or its components should not be used adjacent to or stacked with other equipment. If
adjacent or stacked use is necessary, the monitor and its components should be observed to verify
normal operation in the configuration in which it will be used.
Pins of connectors identified with the ESD warning symbol should not be touched. Connections should
not be made to these connectors unless ESD precautionary procedures are used. For details, see
section “1.5.3 ESD precautionary procedures”.
After transferring or reinstalling the monitor, always check that it is properly connected and all parts
are securely attached. Pay special attention to this in case of stacked mounting.
Do not use the monitor in high electromagnetic fields (for example, during MRI).
A printer or computer must be supplied from an additional transformer providing at least basic
isolation (isolating or separating transformer).
If you accidentally drop the monitor, have it checked by authorized service personnel prior to clinical
use.
To avoid explosion hazard, do not use the monitor in presence of flammable anesthetics. The monitor
measures only non-flammable anesthetics.
Do not touch the patient, table, instruments, modules or the monitor during defibrillation.
Power connection
Before connecting the power cord to the mains outlet, check that the local voltage and frequency
rating corresponds with the rating stated on the device plate on the rear panel of the monitor.
Use only hospital-grade grounded power outlets and power cord. Do not remove the grounding pin
from the power plug.
Use only an intact power cord. Replace the cord if it is cracked, frayed, broken, or otherwise damaged.
Do not apply tension to the power cord otherwise the cord may get damaged.
Do not use an additional multiple socket outlet, extension cord or adapters of any kind.
Before starting to use the system, ensure that the whole combination complies with the international
standard IEC 60601-1-1 and with the requirements of the local authorities. Do not connect any
external devices to the system other than those specified.
1-8
To avoid the risk of electric shock, this equipment must only be connected to a supply mains with
protective earth.
External connection
Do not connect any external devices to the monitor other than those specified.
Overview
Electrical shock hazard
When you connect equipment to the Cardiocap/5 input and output connectors, you are configuring a
medical system and are responsible for ensuring that the system complies with IEC/EN 60601-1-1
and with local requirements.
Do not touch any exposed wire or conductive surface while covers are off and the monitor is energized.
The voltages present can cause injury or death.
Always perform an electrical safety check and leakage current test of the monitor after service.
Fuse replacement
Replace a fuse only with one of the same type and rating.
Explosion hazard
To avoid explosion hazard do not use the monitor in the presence of flammable anesthetics.
Patient safety
Do not perform any testing or maintenance on the monitor while it is being used on a patient.
Never install the monitor so that it is above the patient.
The monitor must not be used without manufacturer approved mounting attached.
Operation of the monitor outside the specified values may cause inaccurate results.
To prevent erroneous readings, do not use physically damaged sensors or sensor cables. Discard a
damaged sensor or sensor cable immediately. Never repair a damaged sensor or cable; never use a
sensor or cable repaired by others. A damaged sensor or a sensor soaked in liquid may cause burns
during electrosurgery.
PATIENTS WITH PACEMAKERS OR ARRHYTHMIAS: Monitor may count the pacemaker pulses as heart
beats during cardiac arrest, some arrhythmias, and with certain types of pacemakers particularly in
ON R mode. Do not rely entirely upon rate meter alarms. Keep patients with pacemakers and
arrhythmias under close surveillance.
PACEMAKER PATIENTS: The impedance respiration measurement may cause rate changes in Minute
Ventilation Rate Responsive Pacemakers. In this case, set the pacemaker rate responsive mode off or
turn the monitor impedance respiration measurement off.
Temperature probes
To prevent injury, use Datex-Ohmeda temperature probes only.
Cleaning and service
Only trained personnel with proper tools and test equipment should perform the tests and repairs
described in this manual. Unauthorized service may void the monitor warranty.
Always unplug the monitor before cleaning or service. After cleaning or service ensure that every part
of the monitor is dry before reconnecting it to the power supply.
Do not touch any exposed wire or conductive surface while any cover is removed and the monitor is
energized. The voltages present can cause injury or death.
Pins of connectors identified with the ESD warning symbol should not be touched. Connections should
not be made to these connectors unless ESD precautionary procedures are used. For details, see
section “1.5.3 ESD precautionary procedures”.
1-9
Cardiocap/5 Technical Reference Manual
Always perform an electrical safety check and a leakage current test on the monitor after service.
Handle the water trap and its contents as you would any body fluid. Infectous hazard may be present.
Accessories
Use only accessories, mounts and defibrillator-proof cables and invasive pressure transducers
approved by GE Healthcare. For a list of approved supplies and accessories, see the "Supplies and
Accessories" catalog. Other cables, transducers and accessories may cause a safety hazard, damage
the equipment or system, result in increased emissions or decreased immunity of the equipment or
system or interfere with the measurement. Protection against cardiac defibrillator discharge is due in
part to the accessories for pulse oximetry (SpO
measurement.
Single-use accessories are not designed to be reused. Reuse may cause a risk of contamination
and/or affect the measurement accuracy.
1.5.2 Cautions
Refer to the User’s Reference Manual for additional cautions to be observed while monitoring a patient.
A CAUTION indicates a condition that may lead to equipment damage or malfunction.
), temperature (T) and invasive pressure (P)
2
General
US Federal law restricts this device to sale by or on the order of a licensed medical practitioner.
Do not apply pressurized air to any outlet or tubing connected to the monitor. Pressure may destroy
sensitive elements.
Turn off the power before making any rear panel connections.
Use only cables and accessories approved by GE Healthcare. Other cables and accessories may
damage the system or interfere with measurement.
Vibrations during transport may disturb SpO
, ECG, impedance respiration, and NIBP measurements.
2
Installation
Leave space for air circulation to prevent the monitor from overheating.
Before use
Allow two minutes for warm-up and note any error messages or deviations from normal operation.
Airway gas measurement
Strong scavenging suction may change the operating pressure of the monitor and cause inaccurate
readings or internal damage.
Autoclaving and sterilizing
Do not steam autoclave or gas sterilize the monitor.
1-10
Cleaning and service
Do not use hypochlorite, ammonia-based, phenol-based, or acetone-based cleaners. These cleaners
may damage the surface of the monitor.
Do not immerse any part of the monitor in liquid or allow liquid to enter the interior.
Overview
Do not apply pressurized air to any outlet or tubing connected to the monitor.
Clean the fan dust filter on the rear panel once a month or whenever necessary.
Electrostatic discharge through the PC boards may damage the components. Before handling printed
circuit boards, wear a static control wrist strap. Handle all boards by their nonconductive edges and
use antistatic containers when transporting them.
Do not break or bypass the patient isolation barrier when testing PC boards.
If liquid has accidentally entered the system or its parts, disconnect the power cord from the power
supply and have the equipment serviced by authorized service personnel.
Batteries
• A lithium battery on the CPU Board. Dispose of the faulty IC containing the battery
according to local regulations.
The battery package of the power supply unit in this device contains lead, which is hazardous to the
environment. Dispose of the battery according to local regulations.
To replace the batteries safely, please refer to the instructions in this manual.
• Do not short-circuit the battery terminals. Short-circuiting the battery may produce a very high
current, which damages the battery and may cause injury to personnel.
• Do not dispose of the battery into open flame, nor put the battery near fire, as it may explode.
• Do not disassemble the battery. It contains electrolyte, which may damage clothing or cause
injury to skin or eyes. If exposed to electrolyte, wash the injured area with plenty of water and
contact a doctor.
See Symbols on equipment earlier in this chapter.
Special components and modifications
Special components used in this monitor are vital to assure reliability and safety. GE Healthcare
assumes no responsibility for damage if replacement components not approved by GE Healthcare are
used.
The manufacturer accepts no responsibility for modifications made to the monitor outside the factory.
Storage and transport
Do not store or transport the monitor outside the specified temperature, pressure and humidity
ranges:
Temperature -10 to +50 °C (14 to 122°F)
Ambient pressure 660 to 1060 hPa (500 to 800 mmHg)
660 to 1060 mbar
Relative humidity 0 to 85 % non-condensing
1.5.3 ESD precautionary procedures
• To avoid electrostatic charges building up, it is recommended to store, maintain and use the
equipment at a relative humidity of 30% or greater. Floors should be covered by ESD
dissipative carpets or similar. Non-synthetic clothing should be used when working with the
component.
• To prevent applying a possible electrostatic discharge to the ESD sensitive parts of the
equipment, one should touch the metallic frame of the component or a large metal object
1-11
Cardiocap/5 Technical Reference Manual
located close to the equipment. When working with the equipment and specifically when the
ESD sensitive parts of the equipment may be touched, a grounded wrist strap intended for
use with ESD sensitive equipment should be worn. Refer to the documentation provided with
the wrist straps for details of proper use.
ESD precautionary procedure training
• It is recommended that all potential users receive an explanation of the ESD warning symbol
and training in ESD precautionary procedures.
• The minimum contents of an ESD precautionary procedure training should include an
introduction to the physics of electrostatic charge, the voltage levels that can occur in normal
practice and the damage that can be done to electronic components if they are touched by
an operator who is electrostatically charged. Further, an explanation should be given of
methods to prevent build-up of electrostatic charge and how and why to discharge one’s
body to earth or to the frame of the equipment or bond oneself by means of a wrist strap to
the equipment or the earth prior to making a connection.
1.5.4 Disposal
Dispose of the whole device, parts of it, its packing material and this manual in accordance with local
environmental and waste disposal regulations.
1.5.5 Points to note
Medical electrical equipment needs special precautions regarding electromagnetic compatibility and
needs to be installed and put into service by qualified personnel according to the electromagnetic
compatibility information provided in Chapter 2.
Portable and mobile RF communications equipment can affect the medical electrical equipment.
Service and reparations are allowed for authorized service personnel only.
1.6 Specifications
All product specifications are subject to change without prior notice.
1.6.1 F-MX and F-MXG frames
Power supply
Rated voltages and frequencies: 100 to 240 VAC 60/50 Hz
Allowed voltage fluctuations: ± 10%
Maximum power consumption: 80 VA
Fuses (2): T2AH/250V
Environmental conditions
Operating temperature: +10 to +40 °C (50 to 104 °F)
Storage and transport temp: –10 to +50 °C (14 to 122 °F)
Relative humidity: 10 to 85 % noncondensing, in airway 0 to 100 % condensing
Atmospheric pressure: 660 to 1060 hPa (500 to 800 mmHg)
1-12
Mechanics
Dimension: 330 mm × 220 mm × 300 mm (width × depth × height)
Weight: <11.2 kg / <24.8 lbs (F-MXG with all options); <10.2 kg / <22.6 lbs (F-MX with all options)
LCD display
Display size: 10.4 inch
Display type: Active matrix color LCD display
Resolution: 640 × 480
Battery
Type: 12V 2.6AH, lead acid
Back-up battery time: at least 15 minutes when fully charged
Charging time: 5 hours (typical)
The green battery charge status LED is On when the battery is fully charged, on the holding voltage. The
LED flashes when the battery is being charged.
1.6.2 NIBP
Measurement range:
Pulse rate range accepted: 30 to 250 bpm
Typical measuring time: adults 23 seconds, infants 20 seconds
Overview
adult 25 to 260 mmHg (3.3 to 34.7 kPa)
child 25 to 195 mmHg (3.3 to 26.0 kPa)
infant 15 to 145 mmHg (2.0 to 19.3 kPa)
1.6.3 Temperature
Measurement range: 10 to 45°C (50 to 113°F)
Measurement accuracy: 25 to 45.0 °C ± 0.1 °C (77 to 113 °F ± 0.2 °F)
10 to 24.9 °C ± 0.2 °C (50 to 76.8 °F ± 0.4 °F)
Probe type: Compatible with Datex-Ohmeda temperature probes only
1.6.4 ECG
Waveform display (with 50 Hz power supply frequency):
Monitoring filter: 0.5 to 30 Hz
ST filter: 0.05 to 30Hz
Diagnostic filter: 0.05 to 100 Hz
Waveform display (with 60 Hz power supply frequency):
Monitoring filter: 0.5 to 40 H
ST filter: 0.05 to 40 Hz
Diagnostic filter: 0.05 to 100 Hz
Heart rate
Measurement range: 30 to 250 bpm
Measurement accuracy: ± 5% or ± 5
Pacemaker pulse detection level: 2 to 500 mV
Pacemaker pulse duration: 0.5 to 2 ms
1-13
Cardiocap/5 Technical Reference Manual
1.6.5 Impedance respiration
Respiration range: 4 to 120 respirations/minute
Accuracy: ± 5% or ± 5 bpm
Calibration range: 50 to 100%
Calibrated against functional saturation
Measurement range: 40 to 100%
Measurement accuracy (% SpO
80 to 100% ± 2 digits;
50 to 80% ± 3 digits;
Below 50% unspecified
NOTE: SpO
measurement accuracy is based on deep hypoxia studies using Datex-Ohmeda FingerSat
2
sensors on volunteered subjects. Arterial blood samples were analyzed by a Radiometer OSM
CO-oximeter. Refer to the sensor instructions for specific SpO
±1 SD):
2
accuracy data.
2
Pulse rate
Measurement range: 30 to 250 bpm
Measurement accuracy: ± 5% or ± 5 bpm
Default alarm limits
: high Off, low 90%
SpO
2
Pulse rate: high 160, low 40
NOTE: Limits are adjustable.
Sensor emitter wavelength ranges
Red LED: 660 nm
Infrared LED: 900 nm
1.7 Specifications for options
1.7.1 Classifications
According to IEC 60601-1
• CLASS I EQUIPMENT and INTERNALLY POWERED EQUIPMENT according to the type of
protection against electrical shock.
• TYPE BF or CF equipment according to the degree of protection against electric shock is
indicated by a symbol beside each patient connector.
1-14
• Degree of protection against harmful ingress of water as detailed in the IEC 60529:
Monitor: IPX1, vertically falling water drops shall have no harmful effects (applicable when the
monitor is in upright position, or tilted backwards). In the protective case IPX4, splash proof,
only when the case is closed properly, the monitor is intact and operates on battery power.
Power adapter: IPX0, ordinary equipment.
Power adapter for Transport Vehicles: IPX1.
• EQUIPMENT not suitable for use in the presence of a FLAMMABLE ANAESTHETIC MIXTURE with
air or with oxygen or nitrous oxide.
• CONTINUOUS OPERATION according to the mode of operation.
• CISPR 11: Group 1, Class A.
Group 1 contains all ISM (Industrial, scientific and medical) equipment in which there is
intentionally generated and/or used conductively coupled radio-frequency energy which is
necessary for the internal functioning of the equipment itself.
Class A equipment is suitable for use in all establishments other than domestic and those
directly connected to the public low-voltage power supply network that supplies buildings used
for domestic purposes.
Classification according to EU Medical Device Directive
± (0.15 vol% + 5 % of reading) ± (0.15 vol% + 5 % of
0 to 6% 0 to 8% 0 to 20%
< 600 ms typically with
a 3 m sampling line
< 650 ms typically with
a 6 m sampling line
reading)
< 0.15 vol% N2O < 0.15 vol% N2O < 0.15 vol% N2O
< 600 ms typically with
a 3 m sampling line
< 650 ms typically with
a 6 m sampling line
± (0.15 vol% + 5 % of
reading)
Agent identification
Identification threshold: 0.15 vol% typically
Identification time: < 20 seconds (for pure agents)
Mixture identification threshold for 2nd agent: 0.2 vol% +10% of total conc.
1-17
Cardiocap/5 Technical Reference Manual
MAC
Range: 0 to 9.9 MAC
Equation:
MAC( AA) =
%(ETAA)
x(AA)
%
+
2
ETN O
100
where x(AA): Hal = 0.75%, Enf = 1.7%, Iso = 1.15%, Sev = 2.05%, Des = 6.0%.
Normal conditions
After 30-minute warm-up period:
Ambient temperature: 18 to 28 °C, within ± 5 °C of calibration
Ambient pressure: 500 to 800 mmHg, ± 50 mmHg of calibration
Ambient humidity: 20 to 80% RH, ± 20% RH of calibration
Non-disturbing gases:
Ethanol C
Acetone (< 0.3%)
Methane CH
Nitrogen N
Carbon monoxide CO
Nitric oxide NO (< 200 ppm)
Water vapor
Maximum effect on readings:
< 0.2 vol%
CO
2
O2, N2O < 2 vol%
Anesthetic agents < 0.15 vol%
Effect of helium: decreases CO
decreases O
Effect of Xenon: decreases CO
Effect of anesthetic agents to monitors without anesthetic measurement:
decreases CO
OH (< 0.3%)
2H5
(< 0. 3%)
4
2
readings < 0.6 vol% typically
2
readings < 3 vol% typically
2
readings < 0.4 vol% typically
2
readings < 0.5 vol% typically
2
1-18
Conditions exceeding normal
Accuracy specifications under conditions n o p q:
n Ambient temperature: 10 to 40 °C, within ± 5 °C of calibration
Ambient pressure: 500 to 800 mmHg, ± 50 mmHg of calibration
Ambient humidity: 10 to 98% RH, ± 20% RH of calibration (non-condensing)
Respiration rate: 35 to 60 breaths per minute
o During warm-up, 2 to 10 minutes (anesthetic agents 5-10 minutes) under normal conditions
p During warm-up, 10 to 30 minutes under normal conditions
O > 85%
q N
2
Overview
Parameter
CO2
O2
N2O
Agents (Des, Enf, Hal, Iso, Sev)
Parameter
CO2
O2
N2O
Agents (Des, Enf, Hal, Iso, Sev)
Parameter
CO2
O2
Accuracy under Condition
n
± (0.3 vol% + 4% of reading); at 5 vol% error ± 0.5 vol%
± (2 vol% + 2% of reading)
± (3 vol% + 3% of reading
± (0.2 vol% + 10% of reading)
Accuracy under Condition
o
± (0.4 vol% + 7% of reading); at 5 vol% error ± 0.75 vol%
± (3 vol% + 3% of reading)
± (3 vol% + 5% of reading)
± (0.3 vol% + 10% of reading)
Accuracy under Condition
p
± (0.3 vol% + 4% of reading); at 5 vol% error ± 0.5 vol%
± (2 vol% + 2% of reading)
N2O
Agents (Des, Enf, Hal, Iso, Sev)
Parameter
N2O
1.7.6 Patient Spirometry (N-XV)
Accuracy specifications apply in normal conditions.
After 10-minute warm-up period
Ambient temperature: 10 to 40 °C
Ambient pressure: 500 to 800 mmHg
Ambient humidity: 10 to 98% RH
Airway humidity: 10 to 100% RH
Respiration rate: 4 to 35 breaths/minute (adult); 4 to 50 breaths/minute (pediatric)
I:E ratio: 1:4.5 to 2:1
Intubation tube: 5.5 to 10 mm (adult); 3 to 6 mm (pediatric)
± (3 vol% + 3% of reading
± (0.2 vol% + 10% of reading)
Accuracy under Condition
q
± (2 vol% + 8% of reading)
1-19
Cardiocap/5 Technical Reference Manual
Detection through D-lite™ or Pedi-lite™ flow sensor and gas sampler:
Stimulus pulse: Square wave, constant current
Pulse width: 100, 200 or 300 µs
Stimulus current range (supramax and manual): 10 to 70 mA with 5 mA steps
Stimulus current accuracy: 10% or ±3 mA (whichever is greater)
Maximum load: 3 kΩ
Maximum voltage: 300 V
Regional block mode
Stimulation modes: Single twitch
Intervals: 1 second, 2 seconds, 3 seconds
Stimulus pulse: Square wave, constant current
Pulse width: 40 µs
Stimulus current range: 0 to 5.0 mA with 0.1 mA steps
Stimulus current accuracy: 20% or 0.3 mA (whichever is greater)
1.7.8 Recorder (N-XREC)
Principle: thermal array
Print resolution
Vertical: 8 dots/mm (200 dots/inch)
Horizontal: 32 dots/mm (800 dots/inch) at speed of 25 mm/second and slower
Paper width: 50 mm, printing width 48 mm
Traces: selectable; 1, 2, or 3 traces
Print speed: 1, 6.25, 12.5, 25 mm/second
Overview
1-21
Cardiocap/5 Technical Reference Manual
1-22
Loading...
+ 208 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.