Tonghui TH2822, TH2822C, TH2822A Instruction Manual

Page 1
Model
Dual Handheld LCR Meter
TH2822/TH2822A/TH2822C
Instruction Manual
Page 2
1
Safety Summary
The following safety precautions are applicable to both operating and maintenance personnel and must be observed during all phases of operation, service, and repair of this instrument.
DO NOT OPERATE IN A FLAMMABLE OR
EXPLOSIVE ATMOSPHERE
Do not operate the instrument in the presence of much dust, direct sunlight, high humidity, strong electromagnetic radiation, etc.
NON-PROFESSIONALS SHUOLD NOT OPEN
THE REAR COVER
Maintaining, substituting parts or adjusting the instrument should be made by professional maintenance personnel. Please contact relevant distributor or Tonghui’s after-sale service department.
Page 3
2
DO NOT RE MOVE OR MODIFY THE
INSTRUMENT
Some replacements and unauthorized modifications might cause irreversible damage to the instrument.
SAFETY WARNING
Strictly follow the relevant safety statements in this manual involving safety, personnel injury, damage to the instrument, operation and environmental conditions causing poor test.
Page 4
3
Safety Guidelines
To ensure that you use this device safely, follow the safety guidelines listed below:
This meter is for indoor use, altitude up to
2,000 m. For short-time outdoor use, precautions should be taken to avoid direct sunlight, water and moisture, electromagnetic radiation, dust and explosion.
The warnings and safety precautions should
be read and well understood before the instrument is used.
Use the meter only as specified in this
manual.
Confirm that the circuits have been powered
off and all capacitors in the circuits been discharged before measuring in-circuit components.
Discharge all charging elements, such as
capacitors, before testing.
The power for the meter is supplied with a
single standard 9V battery. But also a line operation is possible using a power adapter
Page 5
4
of 12VDC/150mA. If a power adapter is selected, please be sure to meet the safety requirements of a relevant IEC standard.
The battery using in TH2822C is
rechargeable. Do not charge non­rechargeable batteries.
Safety Symbols
This symbol is a warning and indicates that the user should refer to the operating instructions located in the manual.
D C power
Indicates inside pin is positive (+),
outside is negative (-) .
Page 6
5
Environme ntal Con diti ons
Operating Enviro nment
0 °C to 40 °C
Storage Humidity
0 – 80% R.H.
Storage Enviro nment
-20 °C to +50 °C
Pollution degree
2
Page 7
6
Contents
Safety Sum ma ry ........................................................ 1
Safety Guideli n es ...................................................... 3
Introduction ............................................................... 9
Package Contents ................................................... 10
Front Panel Overvie w ............................................. 12
Front Panel Display Descriptions .................................. 13
Front Panel Buttons ...................................................... 15
Button Function Definition ............................................. 16
LCD Display Overview .................................................. 17
LCD Display Descriptions ............................................. 17
Special Display Indicators ............................................. 19
Test Port ....................................................................... 19
Powering Instrument .............................................. 21
Installing Battery ........................................................... 21
Connecting External Power Source .............................. 24
Low Battery Indication .................................................. 26
Backlight Display .......................................................... 27
Charge Display ............................................................. 28
Page 8
7
Operation Instruction.............................................. 30
Data hold mode (HOLD) ............................................... 30
Data Record Mode (REC) ............................................. 31
L/C/R/Z Select Mode .................................................... 33
D/Q/θ/ESR Select Mode ............................................... 34
Test Frequency (FREQ)................................................ 34
Tolerance Mode (TOL) ................................................. 35
Auto LCR Mode ............................................................ 37
Measurement Rate (RATE) .......................................... 39
Series/parallel Equivalent Mode ................................... 39
Utility Menu (UTIL) ........................................................ 41
Clear Functions (CLEAR) ............................................. 52
Remote Control (RMT) ................................................. 56
Fuse Detection .............................................................. 56
Quick Start Guide .................................................... 58
CAUTION...................................................................... 58
Inductance Measurement ............................................. 59
Capacitance Measurement ........................................... 62
Resistance Measurement ............................................. 64
Impedance Measurement ............................................. 66
Remote Communicati o n ......................................... 67
Page 9
8
Connecting Instrument to PC ........................................ 67
Virtual Serial Port Configuration .................................... 68
RMT Operation ............................................................. 69
Command Protocols ..................................................... 71
Command Reference .................................................... 77
Specifications .......................................................... 84
General Specifications .................................................. 84
Accuracy Specifications ................................................ 87
Maintenance ............................................................ 94
Service .......................................................................... 94
Cleaning........................................................................ 95
Limited Warranty ..................................................... 96
Page 10
9
Introduction
TH2822 series are des i gn e d f or m eas uring ind uct ance , capacitance and resistance components. The instrument can be powered by a 9V battery or external power adapter. The meter is not only applicable to the application occasion of bench meters but also conveniently used in the flow inspect and handheld measurement occasions.
TH2822 series provide the primary parameter of up to 40,000 readings, secondary parameter of 0.0001 reading resolution, the maximum measuring frequency of up to 100kHz, constant internal
resistance of 100Ω and testing level of 0.6Vrms. The auto range can rapidly display the measuring results
and automatically choose the desirable testing parameter s in ac c ordanc e with c omponents pr oper t ies . Its measuring accuracy is up to 0.25%.
Front panel push buttons maximize the convenience of function and feature selection such frequency, rate and L/C/R/Z selection. Tolerance mode can sort components, record mode aid to capture readings, convenient open/short clear function improve the measuring accuracy, utility menu help you easily take
Page 11
10
the selections of the key tone, auto power-off and storage.
All TH2822 series are equipped with the function of remote communication. The test data can be transferred to PC through a Mini USB connection, great for applications that require remote control and data acquisition.
Package Contents
TH2822 series are equipped with the following contents:
• TH2822/TH2822A/TH2822C handheld LCR meter
• Instruction manual
• Mini USB Interface Cable
• Red / Black Banana plugs—Crocodile clip
Test Leads
• Short-circuit plate
• *9V Alkaline battery or 8.4 rechargeable
battery
• *AC adapter
• *TH26027A Kelv in clip tes t lea ds
• * TH26009C SMD test tweezers
Page 12
11
• *TH26029C SMD test tweezers
* This can be purchased as optional accessories. All contents are subject to actual package list or box.
Please locate them from the original packaging to ensure nothing is missing. If in the case that an item is missing, please contact Tonghui or relevant distributor immediately.
Page 13
12
Front Panel Overview
Figure 1 - Front Panel Display (model TH2822C shown)
1 2 3
5
6
8 4 7
13
9
10
11
12
14
Page 14
13
Front Panel Display Descriptions
1. LCD display
2. USB communication / *Back light button
3. Power ON/OFF button
4. Frequency and record mode selection
5. Secondary display mode (D /Q//ESR, etc.)
6. Primary display mode (L/C/R/Z, etc.)/ auto LCR selection
7. Rate/equivalent mode selection
8. Hold mode/ utility menu
9. Tolerance mode/ utility arrow key
10. Open/short clear/ utility arrow key
11. 5-terminal test sockets (direct measurement on lead components or use of test fixture)
12. 3-terminal test jacks (for use of Banana plugs —Crocodile clip Test Leads)
13. Standard mini USB port (for remote control)
14. 12VDC external power input (use with an external power adapter)
*Backlight is not included in TH2822.
NOTE: Refer to the adapter
’
s label for input
parameters of it. Rated output parameters is 12VDC
,
150mA,4mm.
NOTE: Use with included power adapter only or
purchase the specified adapters from Tonghui. Use
Page 15
14
with improper power adapters may damage instrument.
NOTE: Internal battery supply will be automatically cut off since the normal supply of external power. If the battery is rechargeable (TH2822C), the external power will charge the battery simultaneously. TH2822C is installed an independent charging controller---charging control is still done even at the state of power-off.
WARNING: Before connecting an external power adapter, b e sure that the polarity ma tches the (+) and (-) labels as indicated inside the battery compartment. If it is not installed correctly and connected to an external power adapter, it might cause severe damage to the instrument. WARNING: If the battery is rechargeable (TH2822C), please be sure that the polarity matches the (+) and (-) labels as indicated inside the battery compartment ant the battery is rechargeable. DO NOT charge the non­rechargeable battery!
Page 16
15
Front Panel Buttons
With the exception of the power button, all front panel buttons have specific colored labels on them. They are all marked in black, blue, or orange color. Each color has a specific representation, as described below:
Black—the primary function,meaning that
function will be set or configured upon pressing it.
Orange—the secondary function, it means that
the function will be set or configured if that button is pressed and hold down for 2 seconds.
Blue—the utility function, the function will be set
or configured if the UTIL button is pressed and hold down for a long time. See “Utility Menu” section for details.
NOTE: In the button operational instruction, we will use the button name to express the button operation without differentiating the type of button; Pay attention to the difference between “long press” and “press”.
Page 17
16
HOLD
FREQ
L/C/R/Z
/ESR
UTIL
CLEAR
RMT
P
S
TOL
RATE
AUTO
REC
D/Q/
θ
1 2 3 4
8
9 5
7 6
Figure 2 –Button Display (Model TH2822C shown)
Button Function Defi n ition
1. Power ON/OFF Button 2. Frequency/Record Mode Button 3. Remote Control/Bac klight Button 4. Readings Hold/Utility Menu Button 5. Tolerance mode/ Menu Selection Button 6. Clear/ Menu Selection Button 7. Rate/Equivalent Mode Button 8. LCR Primary Parameters/Auto LCR 9. Secondary Parameters Selection Button
Page 18
17
LCD Display Overview
LCR
ESR
MAXAVGMIN
θ
Q
AUTO
D
2105%
deg kHz PAL
SER
μ
H
p
F
μ
RMT
Mk
TOL
DH
n
n
@
OFF
Z
%
Ω
FASTSLOW
1
2 5 9 10
12
14
16
202127
28
29
8
3
23
24
25
4
15
11
13
6 7
22
17 18
19
Ω
26
Figure 3 - LCD Indicator Display
LCD Display Descriptions
1. LCRZ – Primary display indicator
2. MAX – Maximum reading indicator in the record mode
3. AVG – Average reading indicator in the record mode
4. MIN – Minimum reading indicator in the record mode
5. AUTO – Automatic LCR indicator
6. θ – Phase angle indicator for secondary display
Page 19
18
7. D – Dissipation indicator
8. Q – Quality factor indicator
9.
– Secondar y display
10.
– Beeper tone indicator for tolerance
mode
11. deg – Phase angle (θ) units indicator
12. Ω – ESR(ohm) units indicator
13. % - Percentage indicator (in tolerance mode)
14. kHz – Frequency units indicator
15. PAL – Parallel mode indicator
16. SER – Series mode indicator
17.
– Inductance units (L) indicator
18.
–Capacitance units (C) indicator
19. MkΩ – Resistance(R) /impedance units indicator
20. RMT – Remote mode indicator
21.
– Primary display
22. ESR – Series mode indicator for secondary parameters
23. DH – Data hold indicator
24. SLOW – measuring rate indicator
25. 2105% - Limits indicator in tolerance mode
26. FAST- Fast measuring rate indicator
27.
– Low battery/charging indicator
28. @OFF –Auto power-off indicator
Page 20
19
29. TOL –Tolerance mode indicator
Special Display Indicators
Indicates short clear if you press the
CLEAR button
Indicates open clear if you press the
CLEAR button
Error indication
Indicates correction (open/short clear)
mode
Indicates damaged or open fuse
AD converter error(UNK)
AD converter error(END)
Test Port
TH2822 series are creatively designed to combine 3­terminal port and 5-terminal port, which makes the
Page 21
20
convenient test and highly accurate test realized in the instrument.
+ -
!
H
CUR
L
CUR
H
POT
L
POT
1 2 3
Figure 4- test port
With the adoption of standard banana slots, the instrument can use inexpensive banana plug­crocodile clip as the test lead, which make the test quite convenient. However this configuration has low testing accuracy. For the improvement of accuracy when using external testing leads, TH2822 series are designed with 5­terminal testing slots and exclusive test fixture to ensure complete external 4-terminal test and measuring accuracy.
5-terminal
test slot
3-terminal
test port
1-Hight
Terminal 2-Low Terminal 3-Guard
Terminal
Page 22
21
NOTE: TH26027A, TH26009C, TH 260 29C 4-terminal test fixtures are optional for TH2822 series. Please refer to relevant instrument accessories.
Powering Instrument
There are two methods to supply the instrument: Battery and external power adapter. When the two power modes are available, the external DC adapter is prior to the battery. The two power modes can be automatical l y switche d with out inter ru pti on.
Installing Battery
TH2822 series can adopt battery for power supply so that you can take measurements whenever and wherever without many preparations.
9V alkaline batteries used in TH2822/TH2822A are non-rechargeable. Reference standard type is IEC6LR61.
Page 23
22
TH2822C applies 8.4V rechargeable battery with LH-200H7C as the reference type. Do not use non­rechargeab le batteries with the exception o f emergency cases. The reason is that the charging circuit will operate once the instrument is connected to external power.
To Install the Battery:
1. Open up the back-flip stand, and locate the
screw that tightens the battery compartment cover as indicated in Figure 5. Use a screwdriver to unscrew and remove the cover.
2. Insert proper battery into compartment. Note
the positive (+) and negative (-) terminals as indicated inside the battery compartment (See Figure 6). Be sure to insert the battery with matching polarity.
3. Place the battery compartment cover piece
by sliding it into the top slid first. Place screw at the bottom of the cover piece and tighten down with a screw driver.
4. Push and hold down the power ON/OFF
button for 2 seconds to turn on the instrument.
Page 24
23
Figure 5- Back Cover
Page 25
24
Connecting External Power Source
Some models of TH2822 series are equipped with standard external power adapters, which can use external source.
WARNING: Use the included or specified adapter only. Confirm p o wer p arameters be ones that adapters require before use.
Figure 6- Battery Compartment
Page 26
25
To connect the adapter, do the following:
1. If a battery is installed, please check the battery compartment again that the polarity of the battery matches the polarity as indicated by the labels inside the compartment.
WARNING: DO NOT, at any tim e, connect an external power adapter when a battery is installed incorrectly or a non-rechargeable battery is inserted in a rechargeable instrument. Doing so will damage the in strument and void its warranty.
2. Confirm that an appropriate power supply connects to the adapter.
3. Connect the AC adapter connector into the right side jack of 12VDC.
4. Connect the AC Adapter socket into an electrical outlet.
5. Press and hold down the power button for about 2s to turn on the meter.
Page 27
26
NOTE: The meter will automatically switch to consume power from the AC adapter instead of the battery when an AC adapter is plugged in and consume the power normally. In this event, if the battery insta lled c orrect ly is rec hargeab le (for inst ance, TH2822C), charging controller will be driven at the same time no matter the instrument is on or off.
Low Battery Ind ication
At the use of battery for power supply, if the display starts flashing the
indicator, it means that the battery voltage is below normal working voltage (below 6.8V). It is highly recommended that the battery be replaced as soon as possible before
12VDC Input
AC Adapter
Figure 7-Connecting AC Adapter to Meter
Page 28
27
continuing operation. See “Instal lin g Batt er y” for instructions.
If the battery is rechargeable, when the
indicator starts flashing, please charge the battery before continuing operation. When the external source is
plugged in, the flash of
indicator indicates the
charging state.
Backlight Display
Only TH2822A/C meters have Backlight display that can allow you to see the LCD display in dark conditions, TH2822 doesn’t support backlight.
To turn on the back light, you should press the button for a long time. To turn off the back light, you should press the button for a long time.
When Using Battery Power
When the meter is powered by using battery, the brightness of the back light will automatically decrease to conserve battery power. When the back light have lightened for about 15 seconds, the brightness will continuously decrease; and when the back light have
Page 29
28
lightened for approximate 30 seconds, the back light will automatic all y turn of f .
When Using Exter n al Power
When the meter is powered using an external AC adapter, once the back light is turned on, it will stay at its maximum brightness continuously and will not automatical l y turn off . Unplugg ing the ex ter na l po wer to use battery power, the back light will decrease its brightness and automatically turn off.
Charge Display
The power circuit of TH2822/A is non-rechargeable. When the external power adapter is plugged in, the power mode will automatically switched and the battery power circuit will be cut off. TH2822C is equipped with rec har gea ble circuit. When the external power adapter is connected, the built-in rechargeable battery will be charged automaticall y and the power mode will be switched to external power supply. Single charge circle is about 80 minutes and charge current is approximate 40mA. If a battery is full charged, the charge wi ll be automaticall y shut off ;
Page 30
29
instead, the battery will be charged again after a charge circle.
NOTE: A new charge circle will begin as soon as an external power is connected.
WARNING: If the instrument has rechargeable circuit, DO NOT connect to an external power when a non-rechargeable battery is installed. Doing so will cause the burst of the battery.
: It indicates low power of battery if not
connect to external power; if instrument is connected to external power, it means the charging state.
Page 31
30
Operation Instruction
Data hold mode (HOLD)
The data hold function allows the user to freeze the display data. The data displayed on LCD will not update upon the phase of test until data hold is turned off.
Turn On Data Hold
To use data hold, press the HOLD button. The “DH” indicator will display on the screen when data hold is active. At this moment, primary and secondary displayed on LCD are the testing results before the press of HOLD button.
Turn Off Data Hold
To disable the data hold, press the HOLD button again. The “DH” indicator wil l disappear on the scr een, and meter will remain in normal operation mode.
Page 32
31
Data Record Mode (REC)
If the data stability of tested components is poor and the data fluctuates in a range, data record mode can aid the reading of data. This mode is used for dynamically recording maximum, minimum, and average values in a range.
Enable Static Recording
Press and hold down the REC button for a long time to enter the data recording mode. The display should indicate “MAX AVG MIN” simultaneously, which indicates the meter is in static recording mode.
Using Static Recording
There are four different modes that can be selected in static recording. Per press of the REC button (in recording mode, FREQ will disable), the modes will change and repeat in the following order:
Recording Mode
Maximum Mode Minimum
Mode
Average Mode
Page 33
32
Recording Stat e
This is the default mode when enabling static recording. In this mode, LCD will display “MAX AVG MIN” indicator. In a relatively stable range of test data, a beep tone will sound once a recording has been stored.
NOTE: When the data fluctuation range is upwards of 1%, data record will dynamically be refreshed.
Maximum Display
Press REC button until the “MAX” indicator is shown on display. This indicates that the value in the primary display represents the recorded maximum value.
Minimum Display
Press REC button until the “MIN” indicator is shown on display. This indicates that the value in the primary display represents the recorded minimum value.
Page 34
33
Average Display
Press REC button until the “AVG” indicator is shown on display. This indicates that the value in the primary display represents the recorded average value.
Disable Static R eco rding
To exit this mode, press and hold the REC button for a long time. All the “MAX”, “MIN”, or “AVG” indicator will disappear on LCD.
NOTE: Changing the type of test parameters will automatically turn off static recording.
L/C/R/Z Select Mode
To select measurement mode, you should select primary parameter first. Each press of the L/C/R/Z button, the parameter will change and repeat as the following modes: L (inductance), C (capacitance), R (resistance), and Z (impedance).
NOTE: After changing primary mode, secondary display indicates the present frequency. If it is required to display corresponding secondary parameters, press the secondary button.
Page 35
34
D/Q/θ/ESR Select Mode
If necessary, press the D/Q/θ/ESR button to select secondary parameters. Each press of the D/Q/θ/ESR button, the following modes will be displayed on the screen: D (Dissipation factor), Q (Quality factor), θ (Phase angle), and ESR (Equivalent series resistance).
Test Frequency (FREQ)
TH2822 series handheld LCR meters apply AC signal to DUT for measurement. Frequency is among the main parameters of AC signal. By the presence of component’s non-ideality and distributed parameters, the effect of distributed parameters of test port and test lead, the test frequency used on the same component might ca use different tes t r es ult. Therefor e, a proper frequency should be selected before test.
Selecting Frequency
Change the test frequency, push the FREQ button. If the secondar y displa y does not indicate t he frequ ency, it will display the actual operating frequency when you press FREQ. If the secondary display indicates
frequency, at each press of the FREQ button, the
Page 36
35
meter will change among the following selectable frequencies:
TH2822: 100Hz/120Hz/1kHz TH2822A:100Hz/120Hz/1kHz/10kHz TH2822C:100Hz/120Hz/1kHz/10kHz/100kHz
Tolerance Mode (TOL)
The tolerance mode is specifically used for component sorting purposes. In tolerance mode, secondary display indicates the range of percentage. Tolerance mode, nominal value and sorting limit just come into play on prim ar y param eters . The selec table
ranges for sorting are as follows: 1%,5%,10%, 20% (20% is not for TH2822).
When entering tolerance mode, the data indicated in the primary display will be recorded as nominal value. Displayed value in percent age is:
100*(Mx-Nom)/Nom% Where, Mx is the test value displayed on the primary display; Nom is the recorded nominal value. The percentage value is used for sorting.
Page 37
36
Use Tolerance M o d e
To use the tolerance mode as the process shown below:
1. Select the desired primary measurement mode by pressing L/C/R/Z button.
2. Configure the proper test frequency and series/parallel equivalent mode.
3. Perform the operation of CLEAR appropriately if necessary.
4. Test standard implements or components with accurate and reliable measured value.
5. Once the desired measured reading is displayed, press the TOL button once to store the reading as the nom inal value. At this point,
the “TOL” will be displayed on the screen, indicating that th e toleranc e mode is ac tivated . A percentage m ode will be sho wn in the secondary mode to indicate the percentage deviation.
NOTE: Before the press of TOL button, the primary parameter indicated on LCD in any mode can be taken as the nominal value, including data ho ld, MAX, MIN , AVG data recording, etc.
6. If sorting is not necessary, you can skip this step. If it is necessary, by pressing the TOL button you can select the range of 1%, 5%,
Page 38
37
10% or 20%, which will be sho wn on LCD accordingly.
7. Changing test component, an audible tone will be heard. One single “beep” or tone means the component is within tolerance. Three “beeps” or tone means the component is out of tolerance.
WARNING: Be sure that the capacitor has been fully discharged before its test, or the instrument mig h t b e damaged.
Disable Tolerance Mode
Long press of TOL button will disable tolerance mode.
NOTE: Changing the test frequency, primary function, or secondary function will automatically disable tolerance mode.
Auto LCR Mode
Auto LCR function will automatically select the correspondi ng primary and secondary parameters and suitable series/parallel equivalent mode of L, C, R. The selection is done by judging the impedance
Page 39
38
property of component according to the test result. It is quite convenient for the measurements of mixed and unknown components.
Enable Auto LCR Mode
Long press of AUTO button will activate auto LCR mode. The “AUTO” indicator on LCD indicates that auto LCR mode is activated. In auto LCR mode, the match of secondar y param eter with primary parameter is shown as below:
Table1-Matching relations between primary and
secondary parameters in auto LCR mode
Primary Parameter
Secondary
Parameter
Capacitance (C)
Dissipation (D)
Inductance (L)
Quality Factor (Q)
Resistance (R)
Phase Angle (θ)
In auto LCR mode, series or parallel equivalent mode is selected in accordance with the magnitude of impedance. Parallel mode is selected at high impedance and series mode at low impedance.
Page 40
39
Disable Auto LCR Mode
Long press the AUTO button again will disable auto LCR mode. In addition, this mode will not continue through changing the primary and secondary modes, series/parallel equivalent mode and frequency mode. “AUTO” indicator on LCD will disapp ear when auto LCR mode is turned off.
Measurement Rate (RATE)
There are two selectable measurement rates in this instrument: fast and slow. The rate of fast measurement is about 4~5times/sec and slow measurement is approximate 1.5 times/sec. The stability of slow measurement is higher than fast measurement. The fast and slow rates can be directly switched by
pressing RATE button. “FAST” indicator will be displayed on LCD at fast rate and “SLOW” indicator at slow rate.
Series/parallel Equivalent Mode
For the presence of non-ideality and distributed parameters of components, actual components are usually equivalent to combined network of ideal components. In general, there are two simple
Page 41
40
equivalent models used in LCR meters, which are series model and parallel model.
Figure 8-Series and Parallel Equivalent Models of
Inductors and Capacitors
Appropriate equivalent modes could help to gain better measurement results. Generally, Series mode is better for components with low impedance (blow
100Ω),while parallel mode for components with high impedance (over 10kΩ). For the components with the
impedance between the two limits, equivalent mode has little effect on the testing result.
Selecting Measur em ent Mod e
Long press P<->S button, “PAL” on LCD ind icates parallel equivalent mode and “SER” means series equivalent mode.
Page 42
41
Default Equiva lent Mode
When select primary parameter, equivalent mode
varies with primary parameter:
For capacitors and resistors, default equivalent
mode is PAL; for inductors, is SER.
Utility Menu (UTIL)
The LCR meter has a built-in utility menu that allows you to configure some user preferences and settings. The buttons used to set and control the menu are
colored in blue. There are three buttons: UTIL, ,
. User can configure the beep tone, auto power-off timing, store/restore power-on state, view the battery voltage, etc.
Entering Utility Menu
Long press UTIL button wil l enter utility menu. Prim ar y display is menu option and secondary display is the current settings or parameters configured for the selected option. After the entrance into utility menu, the default menu option display will show “bEEP”.
Page 43
42
Configuration and Settings
The following contents are included in the utility menu:
Table 2-Utility Menu Options and Settings
Menu Options Settings/Parameters
bEEP
ON / OFF
AoFF
5 / 15 / 30 / 60 / OFF
PuP
PrE / Set
dEF yES / NO bAtt Battery Voltage
Uses of these menu options are as follows: Control beep sound( bEEP) Set auto power-off(AoFF) Set power-up state(PuP) Reset to default settings(dEF) Indicate battery voltage(bAtt)
By default, press the button UTIL to change or select a different menu option. To change the settings or parameters, use the and arrow keys. For each
UTIL button press, the meter will traverse through each menu options and will repeat itself in the following order:
bEEP AoFF PuP  dEF  bAtt
Page 44
43
NOTE: The change of settings has different application effects in accordance with different exiting mode. See “Exit Utility Menu” section (Saving and Exiting, Exiting without Saving) for details.
Beep Sound Setup (bEEP)
The “bEEP” menu option allows the user to enable or disable the beep sound for every key press.
NOTE: This option only disables the beep sound for each key press. It does not affect the beep sound for “Tolerance” mode and “Static Recording”, as well as the “auto power-off” warning.
Use the and arrow keys to choose ON or OFF. This setting will be immediately effective. But this state will not saved if choose “Exiting without Saving”; “Saving and Exiting” should be implemented if this setting needs to be effective after restarting.
Default setting: ON.
Auto power-off Setup(AoFF)
The “AoFF” menu option allows the user to select the auto power-off timer. The available timer settings are: 5min/15min/30min/60min/OFF. When the primary
Page 45
44
display shows “AoFF”, push theand  arrow keys to select the timer setting. The settings will be shown on the secondary display as Table 3. When AoFF is effective, this timer is always counting continuously; when the configured time is up, the meter will make an audible “beep” sound continuously to remind the user of prompt auto power-off. Before auto power-off, pressing any button will reset the timer count.
NOTE:
Auto power-off is effective only for battery
power.
NOTE: When auto power-off is efficient, the display of
“@OFF” indicates the operation of timer.
NOTE: Auto power-off will not work temporally in TOL
mode, REC mode and RMT mode. It will be activated after exiting of above modes.
Table 3- Auto Power-off Options
Secondary Display
Representation
5
5 minutes
15
15 minutes
30
30 minutes
60
60 minutes
OFF
Manual power off only
Page 46
45
The setting will be immediately effective. But this state will not saved if choose “Exiting without saving”; “Saving and exiting” should be implemented if this setting needs to be effective after restarting.
Default Setting: 5
Power-up State (PuP)
The “PuP” menu option allows user to configure the power-up state of the LCR meter, allowing user to
restore settings saved into internal EEPROM memory at power-up. The storable settings are as follows:
• Primary function mode (e.g. L/C/R)
• Secondary function mode (e.g. D/Q)
• State of auto LCR
• Series/para llel equ iv al ent mode
• Test frequency
• Tolerance mode
• Reference value for Tolerance mode
• Measurement rate
In the utility menu, Press the and arrow keys to select “PrE” or “SEt”. PrE means to preserve the previous setting while SEt means to save the current parameter, that is, cover the previous data.
Page 47
46
NOTE: Exiting mode decides whether to implement SEt. At “Exiting without saving”, SEt is ineffective; while at “Saving and Exiting”, SEt setting will be effective.
Default Setting: PrE
Configure and Sa ve Power-up State
Procedure to configure and save po wer -up state is as follows:
1. Before entering into the utility menu, configure all the measurement parameters firstly, such as frequency, primary and secondary parameters. If the meter is currently in the utility menu, exit first and enter into utility menu after measuring setup.
2. Press UTIL button for a long time to enter into utility menu.
3. Push the UTIL button to traverse through the utility options until you see “PuP” on the primary display.
4. In order to save the current meter settings for power-up state into internal memory, use
Page 48
47
either or button to change the settings so that the secondary display shows “Set”.
5. Press the UTIL button to check whether other desirable setups have already been set. With all settings done, exit the menu by long press of the UTIL button.
6. The meter has now saved all current settings into internal memory. At next power-up, the meter will turn on and recall the saved settings.
NOTE: The meter allows one set of settings to be stored into memory. Therefore, the same procedure is used to overwrite previously stored settings into memory.
Prevent Overwrite of Stored Settings
In the utility menu, the “PuP” option default setting is always “PrE”. If it is required to overwrite previously stored settings for power-up state, the option should change to “Set”. Therefore, when entering the utility menu, be sure not to change to “Set” to prevent overwriting any previously stored power-up settings.
Page 49
48
Reset Default Settings (dEF)
The “dEF” option is used to reset the current measuring setup and optional settings in utility menu to default settings. These default settings are as below:
Page 50
49
Table 4 - Instrument Default Settings
Settings Default Configuration Primary Function C (capacitance)
Secondary Function None (frequency) Auto LCR function Off Equivalent Met hod SER (series) Measurement Frequency 1kHz Measurement Speed Slow (SLOW) Tolerance Mode Off
Beep Sound On Auto Power-off Timing 5 minutes
Stored Measurement Setup
clear
Stored Utility Menu
Option
clear
Page 51
50
In “dEF” option, push either or button to choose “NO” or “yES”. “NO” means the instrument will not be set back to default. “yES” indicates to reset all settings to defau lt and to cl ear pre viousl y stor ed setup .
Default Setting: No
NOTE:
Exiting mode also dec id es whether or not to
perform “yES” function. At the use of “exiting without saving”, “yES” is ineffective; when “saving and exiting” is selected, restore operation will be efficient. NOTE: In the case where “PuP” is set to “Set” and “dEF” is set to “yES” simutaneously, the “PuP” setting has priority over the “dEF” setting. This means the instrument will not be set back to default upon saving and exiting the utility menu. Instead, the power-up settings will be stored and will be recalled upon the next power-up of the instrument.
Indicate battery voltage(bAtt)
When menu option changes to “bAtt”, the secondary display will indicat e battery voltage that is for reference instead of for operational function.
Page 52
51
Exit Utility Menu
There are two methods for exiting the utility menu: Saving and Exitin g, Exiting without Saving. The former saves all the changed settings before exiting, and the latter exits the menu without saving any changes.
Saving and Exiting
To save all utility menu option settings and to exit the menu, press and hold down the UTIL button for a long time. After this, the meter will exit the menu. Then
PuP and dEF will be performed and all settings will be saved. “Saving” means that corresponding settings will be saved in the built-in non-volatile memory. Therefore, data will not lose at the time of power-off and can be used at the time of next power-on.
Exiting Without Saving
If user decides to exit the utility menu without saving the optional settings, and gives up the operation of “PuP” and “dEF”, it can be done by simply pressing
Page 53
52
any front panel buttons except UTIL, , and
POWER. PuP and dEF operation will be ineffective. Settings, such as “bEEP” and “AoFF” will not be saved in the non-volatile memory but still be temporarily efficient before power-off.
Clear Functions (CLEAR)
There are two functions under CLEAR: Open Clear and Short Clear. Clear can decrease the distributed error caused by test leads, for instance, short clear can reduce the effects of contact resistors and test leads and open clear will minimize the influence of distributed capacitors and resistors on testing high impedance components.
Enter Clear Mode
For convenience, open clear and short clear are designed to share a button. By pressing the CLEAR button, the meter will automatically choose either
OPEN clear or SHORT clear.
Page 54
53
Open Clear
First select frequency to clear and keep test clip and test slot be open. Enter into clear by the press of
CLEAR, and a moment later the OPEN indicator will appear on secondary display after the automatic measurement judging. If user decides to perform open clear, another press of CLEAR should be done.
NOTE:“----” indicator on secondary display indicates that test terminal is out of open state and open clear cannot be performed.
Figure 9 - Open Clear
Page 55
54
Short Clear
First choose test frequency to clear and then insert a short plate to test slot. If SMD test tweezers or test clip is used, the short plate should short the test terminal. Enter into clear by the press of CLEAR, and a moment later the SHrT indicator will appear on secondary display after the automatic measurement judging. If user decides to perform short clear, another
press of CLEAR shou ld be done .
NOTE: “----” indicator on secondary display indicates
that test terminal is out of short state and short clear cannot be performed.
Figure10-Short Clear
Page 56
55
Quick Clear Procedure
Below is an example of steps to do open or short clear:
1. Select the primary and secondary function
mode for measurement; 2. Select test frequency; 3. Select equivalent mode; 4. Keep the test terminal open to perform open
clear; 5. Short the test terminal to perform short clear; 6. Connect DUT to start testing after clear.
NOTE:
1. Clear data is just temporarily stored in RAM, which means that the data will loss after power-off. Therefore, “clear”, prior to measurement, should be first done after power-on.
2. Clear data is stored under different frequency, thus they will be still valid at the change of test frequency (for instance, under the frequency of 1kHz, the meter has been cleared; and when the frequency is back to 1kHz, there is no need to clear again.)
3. Clear is not concerned with test parameters and series/parallel equivalent mode. In
Page 57
56
accordance with adv anc e d impedance network principle, the instrument performs clear operation. Though the complex impedance is cleared, the displayed parameter is just one of the elements of impedance, so it is cleare d too.
4. After a long time of continuous use, the meter will be affected by the temperature environment and test fixtures, test leads and contact resistance will change. It should be necessary to clear once more according to specific conditions so as to meet the requirement of accuracy.
Remote Control (RMT)
When the RMT button is used for remote communication, please see “REMOTE COMMUNICATION” section for details.
Fuse Detection
The meter has an internal fuse in the test signal terminal that protects the internal components from severely damaging the instrument. When the fuse is
Page 58
57
burned out, the “FUSE” indi c ator will app ear on the primary display and an internal “beep” will sound continuously. In this situation, none of the function buttons can be operated and all other meter functions will be disabled.
Figure 11-Fuse Display
In the event that the above screen is displayed, the instrument should be powered off. If this does not power off the meter, remove external AC adapter if that is used and/or remove the battery from the battery compartment. Please contact after-sale department of our company or appointed distributor for the change of fuse or maintenance.
Page 59
58
NOTE: Both fuse damage and none output of test signal caused by signal source fault will make “FUSE” alarm.
Quick Start Guide
CAUTION
• Do not measure a capacitor that is not fully discharged. Connecting a charged or partially charged capacitor to the input terminals will damage the instrument.
• When measuring an on-board component within a circuit, the circuit must be powered off before connecting the test leads.
• When used in a dusty environment, the instrument should be wiped and cleaned regularly. The electrical conductivity carried by the accumulated dust will eventua lly have an eff ect on th e use of the meter.
Page 60
59
• Do not leave the instrument exposed to explosive, direct sun-lighting and overheating environments.
• Before removing the cover, ensure that none DUT is connected to the meter and the instrument is disconnected from any circuit and is powered OFF.
NOTE: To achieve optimum precision, please refer to “Clear Function” for details bef ore tes ti ng.
Inductance Measurement
1. Press the POWER button for a long time to turn on the instrument.
2. Press the L/C/R/Z button until “L” is displayed on the screen to select inductance measurement
3. Insert an inductor into test slots or connect a tested inductor through a proper test accessory (e.g., banana plug-crocodile clip test leads, test fixture or S MD t es t t we e zer s ). See figures 12. And figure 13 shows the optional 4-terminal test accessories.
4. Press the FREQ button until the desired test frequency is displayed on screen.
Page 61
60
5. Press the D/Q/θ/ESR button to select the desired secondary parameter.
6. Read the readings on LCD for inductance measured values.
+ -
!
H
CUR
L
CUR
H
POTLPOT
Handheld LCR Meter
100Hz - 100kHz
L
Q
AUTO
SER
μ
H
@
OFF
HOLD
FREQ
L/C/R/Z
/ESR
UTIL
CLEAR
RMT
P
S
TOL
RATE
AUTO
REC
D/Q/
TH2822C
θ
Figure 12- Inductance Measurement
Below are TH26027A,TH26009C,TH26029C 4­terminal test accessories:
Page 62
61
TH26029C
SMD KELVIN TWEEZER
H
CURLCUR
GUARD
TH26009C
SMD KELVIN TWEEZER
H
CURLCUR
GUARD
H
CUR
L
CUR
GUARD
TH26027A
KELVIN CLIP
Figure 13- 4-termin al Test A ccess or ies
Page 63
62
Capacitance Measurement
WARNING: Before testing, ensure that the
tested capacitor has been fully discharged.
1. Press the POWER button for a long time to turn on the meter.
2. Press the L/C/R/Z button u ntil “C” is displayed on the screen to select capacitance measurement.
3. Insert a capacitor into test slots or connect a tested capacitor through a proper test accessory (e.g., banana plug-crocodile clip test leads, test fixture or SMD test tweeze rs). See figure 14. And figure 13 shows the optional 4-terminal test accessories.
4. CAUTION: Before inserting a capacitor or capacitive component into the input slots or terminals, be sure to fully discharge the component. Some larger capacitive components may take longer to discharge. In these cases, please allot enough time for a full discharge. If improper discharging of the component is done, it will damage the input terminals of the meter.
5. Press the FREQ button until the desired test frequency is displayed on screen.
Page 64
63
6. Press the D/Q/θ/ESR button to select the desired secondary parameter.
7. Read the readings on LCD for capacitance measured values.
+ -
!
H
CUR
L
CUR
H
POT
L
POT
Handheld LCR Meter
100Hz - 100kHz
C
AUTO
D
PAL
p
F
@
OFF
HOLD
FREQ
L/C/R/Z
/ESR
UTIL
CLEAR
RMT
P
S
TOL
RATE
AUTO
REC
D/Q/
TH2822C
θ
Figure 14- Capacitance Measurement
Page 65
64
Resistance Measurement
1. Press the POWER button for a long time to turn on the instrument.
2. Press the L/C/R/Z button u ntil “R” is displayed on the screen to select inductance measurement.
3. Insert a res istor into test sl ots or connect a tes ted resistor through a proper test accessory (e.g., banana plug-crocodile clip test leads, test fixture or SMD test tweezers). See figure 15. And figure 14 shows the opt ion al 4-terminal test accessories.
4. Press FREQ button until the desired test frequency is displayed on screen.
5. Press the D/Q/θ/ESR button to select the desired secondary parameter.
6. Read the readings on LCD for resistance measured values.
NOTE: The meter uses an AC signal for measurement of resistance, so what the test result shows is AC resistance property instead of DC resistance property.
Page 66
65
+ -
!
H
CUR
L
CUR
H
POT
L
POT
Handheld LCR Meter
100Hz - 100kHz
R
AUTO
D
kHz
PAL
HOLD
FREQ
L/C/R/Z
/ESR
UTIL
CLEAR
RMT
P
S
TOL
RATE
AUTO
REC
D/Q/
TH2822C
@
OFF
θ
Ω
Figure 15- Resistance Measurement
Page 67
66
Impedance Measurement
1. Press the POWER button for a long time to turn on the instrument.
2. Press the L/C/R/Z button until “Z” is displayed on the screen to select impedance measurement.
3. Insert impedance (resistor, capacitor or inductor) into test slots or connect tested impedance through a proper test accessory (i.e., banana plug-crocodile clip test leads, test fixture or SMD test tweezers). Figure 13 shows the optional 4­terminal test accessories.
4. Press the FREQ button until the desired test frequency is displayed on screen.
5. Press the D/Q/θ/ESR button to select the desired secondary parameter.
6. Read the readings on LCD for impedance measured values.
Page 68
67
Remote Communication
The meter has the capability to communicate with a PC over the mini USB interface. Upon installation of a USB driver, PC can control the instrument and collect test results over virtual serial interface.
Connecting Instrument to PC
Follow the below procedures for connection setup. 1. Install USB driver from CD or download the USB
driver from www.tonghui.com.cn.
2. With a Mini-USB cable, connect one end of it to
the LCR meter and the other end to an available USB port on PC. Press the POWER button to turn on the meter.
3. Skip the next process when the driver has been
stalled.
4. When Windows recognize the USB connection,
PC will prompt user to install driver. The following process is to cancel installation guide and directly run the setup program in the setup file.
5. When completed, Windows will create a virtual
serial port and distri bute a serial number. See Windows device manager for details.
Page 69
68
6. Open the control software and use the assigned
serial number to communicate with the handheld LCR meter. You can Download the FastAccess PC communication software from our web site (
www.tonghui.com.cn) for a demo use.
+ -
!
H
CUR
L
CUR
H
POT
L
POT
Handheld LCR Meter
100Hz - 100kHz
LCR
ESR
MAXAVGMIN
θQ
AUTO
D
2105%
deg kHz
PAL SER
μ
H
p
F
μ
RMT
Mk
TOL
DH
n
n
@
OFF
Z
FASTSLOW
HOLD
FREQ
L/C/R/Z
/ESR
UTIL
CLEAR
RMT
P
S
TOL
RATE
AUTO
REC
D/Q/
TH2822C
θ
%
Ω
Ω
Figure 16- Connection to PC
Virtual Serial Port Configuration
Below is the serial communication configuration of TH2822 series:
Page 70
69
• Baudrate:9600
• Data bits: 8
• Parity: None
• Stop bits: 1
• Flow Control: None
Upon the installati on of USB driver, if the default assigned by serial port does not match to above configuration, please modify it as the following process:
Open Windows device manager→ Port→ Used Ser ial Port→ Right Click → Property→ Port Set up
RMT Operation
In the case of communication, the RMT button is used to change the running modes:
Change to Local Operation in Remote Control Mode; Change to Auto Fetch state in Local Operation mode; Disable the Auto Fetch function in Auto Fetch state.
NOTE: Auto Fetch state means to send testing result automatically without needing the initiative inquiry of PC.
Page 71
70
Remote Mode
Upon the acceptance of any commands from PC, the meter will be automatically set into remote mode. In this mode, the LCD display will show “RMT” indicator. When this is shown, all front panel button will be
locked and disabled, except for RMT and POWER buttons. If the meter has been in Auto Fetch mode before receiving remote control commands, Auto Fetch mode will be disabled. That is to say, among panel operation, Auto Fetch and Remote Control, remote control has the highest priority. To exit remote control, press the RMT button. When the “RMT” indicator disappears on the LCD display, the meter goes back to local operation.
NOTE: If the local operation is locked, RMT button will be ineffective as well. See *LLO common command in Command Reference section for details.
Press RMT button to change to Auto Fetch state, when it is not in the remote mode. See the description below.
Auto Fetch
When the meter is out of remote control, it can be changed to Auto Fetch mode. That is, the meter will
Page 72
71
automatically send data to the interface bus upon every measurement. Therefore, PC will obtain data by direct reading with no need to send any commands. It is quite useful in the recording data simply.
Enable/Disable Auto Fetch
To toggle between enabling and disabling auto fetching when the meter is out of remote control,
press the RMT button. In Auto Fetch, every flash of “RMT” means send a measurement result.
NOTE: Auto Fetch can be disabled by remote control. When a remote command is sent to the meter, Auto Fetch will be disabled. To re-enable Auto Fetch after
remote command, first press the RMT button to return panel operation and then enter into Auto Fetch mode by another press of RMT.
Command Protocols
Overview
TH2822 series support SCPI commands with ASCII character strings transmitting control commands and returning query information and data and with a
Page 73
72
specified terminator identifying the termination of a command line or query data line. The adoption of SCPI commands brings convenience for the interaction control of PC on the meter through programming. Command format is compliance with specifications and easy for understanding and use.
Common Command
The IEEE 488 standard defines the common commands for general use in all kinds of instrument. Common commands usually come with the asterisk “*” character, and may include parameters. Some examples of Common command like: *IDN?, *GTL, *LLO. TH2822 series support a few common commands. See the following command descriptions for more details.
Termination Character
A terminator is a character sent by a host, which identifies the end of a command line. Only when a termination character is received, the instrument will analyze and deal with command character string. Terminators can be any one of the following character strings:
Page 74
73
<CR> (Carriage Return, ASC(&H0D)) ; <LF> (Line Feed, ASC(&H0A));
<CR><LF>
Returned Result
After the meter executes a query command, the return of the result will be in the following format:
<Result> + <CR> <LF>
CR is Carriage Return and LF is Line Feed. For example, in inquiring measured results (FETCH?),
the format of the printed data will be shown as the following:
<Primary measured data, Secondary
measured data, Tolerance Result> <CR> <LF>
Data Types
Table 4 below explains the different data types transm itted on bus with ASCII char acter s :
Page 75
74
Table 4 - Data Type
Data Type
Explanation
Example
<NR1>
Integer
+800,-200,100,-50
<NR2>
Real
+1.56,-0.001,10.5
<NR3>
This representation
has an explicit radix point and an
exponent
+2.345678E+04
-1.345678E-01
<Boolean>
A parameter for Boolean setting. Always return “0” or “1” for Boolean
query command
ON or OFF
<Literal>
A string is used as command parameters with
short literal form
HOLD
Symbol Convention
Syntax Symbols in Commands
Below symbols is a part of commands:
Page 76
75
Syntax
Symbol
Explanation
:
Colon means to enter into next command level
;
Separate another command at the same command level
*
common command
,
multi-parameter delimiter
?
inquiry
Space, separate commands from parameters
“ ”
quoting part
Symbols in Command Explanations
Below symbols are used for the description of command format, which are not the constituting part of command.
Table 5-Command Specifier
Mark Symbol
Explanation
[ ]
Option; can be om itted
|
Exclusive OR
< >
Defined elem ent
( )
Comment
Page 77
76
Abbreviation, upper case and lower case commands
1. There are two command formats: complete
format and abbreviated format. In the following command description, capital s ect ion of the complete commands shows abbreviations format. It has the same effect of sending abbreviated and complete commands.
2. Abbreviation command is generally shown in
the form of 4-letter. Any abbreviation absent from command table is taken as incorrect command.
3. Either uppercase or lowercase letters, the
effects on the bus are the same.
Page 78
77
Command Reference
Common Command
*IDN?
Query instrument ID. Return: <instrument model>, <firmware version >, <serial number>
*LLO
Local Lockout. This means that all front panel buttons, including the RMT key is not avail able. ( POWER button is enabled.)
*GTL
Go to local and remove local lockout. If *LLO is sent, the only way to operate front panel is to go to *GTL.
*TRG
Trigger the instrument to take a measurement. Due to the automatically continuous test, *TRG command is of no use.
Page 79
78
SCPI Commands
FREQuency Subsystem
FREQuency <value>
Description: Set measurement frequency
Parameters are 100, 120, 1000, 10000,100000 or 100hz,120hz,1khz,10khz,100khz (model supported) Example: FREQuency 100hz
Set 100Hz frequency
FREQuency?
Description: Query the measurement frequency
Return: < 100hz,120hz, 1khz, 10khz,100khz>
FUNCtion subsystem
FUNCtion:impa < L | C | R | Z >
Description: Se lec t primary param eter
Example: FUNCtion:impa L Selects L as primary parameter
FUNCtion:impa?
Description: Qu ery primar y parameter Return: <L, C, R, Z, NULL >
FUNCtion:impb < D | Q | THETA | ES R >
Description: Select secondary parameter
Example: FUNCtion:impb D
Select D as secondary parameter
Page 80
79
FUNCtion:impb?
Description: Query secondary parameter Return: <D, Q, THETA, ESR, NULL>
FUNCtion:EQUivalent < SERies | parallel | PAL >
Description: Set equ i va lent mode
Parameters: SERies — series mode Parallel — parallel mode PAL — parallel mode
Example: FUNCtion:EQUivalent SERies Set series mode
FUNCtion:EQUivalent?
Description: Query the equivalent mode Return: <SER, PAL>
CALCulate subsystem
CALCulate:TOLerance:STATe < ON | OFF >
Description: Enable or disable tolerance mode Example: CALCulate: TOLerance:STATe ON
CALCulate:TOLerance:STATe?
Description: Query the tolerance mode Return: <ON, OFF >
CALCulate:TOLerance:NOMinal?
Description: Query the nominal value
Return: NR3 or -----(exceeding data range)
Page 81
80
CALCulate:TOLerance:VALUe?
Description: Qu ery the percentage value of
tolerance
Return: NR3 or ----- (exceeding data range)
CALCulate:TOLerance:RANGe < 1 | 5 | 10 | 20 >
Description: Set tolerance range as 1%,5%,10%
or 20% (20% not available for some models ) Example: CALCulate:TOLerance:RANGe 1
Set 1% tolerance range
CALCulate:TOLerance:RANGe?
Description: Query the tolerance range
Return: <BIN1, BIN2, BIN3, BIN4 or ---- > “----” means unsetting range
CALCulate:RECording:STATe < ON | OFF >
Description: Enable or disable recording function Example: CALCulate:RECording:STATe ON
CALCulate:RECording:STATe?
Description: Query the recording state Return: <ON or OFF>
CALCulate:RECording:MAXimum?
Description: Query the maximum value of
recording function
Return: <NR3, NR3> ( primary and secondary
parameters, when data exceeds limits or there is no data, what returns is“----”.)
Page 82
81
CALCulate:RECording:MINimum?
Description: Query the minimum value of
recording function
Return: <NR3, NR3> (primary and secondary
parameters, when data exceeds limits or
there is no data, what returns is “----“)
CALCulate:RECording:AVERage?
Description: Query the average value of
recording function
Return: <NR3, NR3> (primary and secondary
parameters, when data exceeds limits or
there is no data, what returns is “----“)
CALCulate:RECording:PRESent?
Description: Query the present value of
recording function
Return: <NR3, NR3> (primary and secondary
parameters, when data exceeds limits or
there is no data, what returns is “----“)
FETCh Subsyste m
FETCh?
Description: query the measurement result.
Return: <NR3, NR3, N R1 > that is primary parameter, secondary parameter and compared result.
Example: FETCh?
Page 83
82
Summary of Supported SCPI Commands
Table 2 - Summary of SCPI Commands
Command
Parameter
Explanation
FREQuency
<Value>
Set Test Frequency
FREQuency?
Query Test Frequency
FUNCtion
:impa
<Literal>
Select primary display parameter
:impa?
Query primary display parameter
:impb
<Literal>
Select secondary display parameter
:impb?
Query secondary display parameter
:EQUivalent
<Literal>
Set equivalent mode
:EQUivalent? Query equivalent mode
CALCulate
:TOLerance
:STATe <Boolean> Enable/disable tolerance
mode
:STATe? Query tolerance mode
:NOMinal?
Query nominal value
:VALUe?
Query percent of tolerance
:RANG <Value> Set tolerance range
:RANGe?
Query tolerance range
:RECording
:STATe
<Boolean>
Enable/disable recordi ng function
:STATe? Query recording state
:MAXimum?
Query max. value of recording
Page 84
83
:MINimum?
Query min. value of recording
:AVERage?
Query average value of
recording
:PRESent?
Query present value of
recording
FETCh?
Query measurement result
Error Codes
If the received commands or parameters from bus contain errors, the meter will terminate the analysis and execution commands. At the same time, error code will be d ispla yed on L CD and bee p will be sou nd. Below defines the error description based on the error code.
E10: Unknown command E11: Parameter Error E12: Syntax Error
Page 85
84
Specifications
Below is the general specification and the accuracy specification of TH2822 series and applicable to TH2822, TH2822A, TH2822C. Declaration: The specifications are subject to change without notice.
*NOTE: 120Hz is the rated frequency. The actual frequency is 120.048Hz.
General Specifications
Function
Measurement Parameters
Primary: L/C/R/Z Secondary: D/Q/θ/ESR
Equivalent Mode Series, Parallel Auto LCR Function Manual, Auto Ranging Mode Auto Test Terminals 3-terminal,5-terminal Measurement Speed 4 meas/sec,1.5meas/sec Correction Short, Open Tolerance Mode TH2822 1%,5% , 10%
Page 86
85
TH2822A/C 1%,5% , 10% , 20% Input Protection Fuse 0.1A / 63V Interface Mini-USB (virtual serial port)
Test Signal
Signal Frequency
TH2822 100Hz,*120Hz,1kHz
TH2822A 100Hz,*120Hz,1kHz,10kHz
TH2822C
100Hz,*120Hz,1kHz,10kHz, 100kHz
Test Signal Level 0.6Vrms Output Impedance 100Ω
Display
Display LCD primary-secondary dual display
Backlight (not available for TH2822)
Battery supply: when backlight is on, luminance is reduced by half 15s later and automatically turned off 30s later. Powered by adapter: backlight off until
manually turned off
Readings
Max. Counts of Primary Parameter:
40,000; D / Q / θ Min. resolution of
secondary parameter: 0.0001.
Basic Accuracy
0.25%
(see accuracy specifications for
details)
Primary Range and Resolution
See Accuracy Specifications
Secondary Range for Displ ay Resolution
Page 87
86
Parameter
ESR 0.0000W -- 999.9W 0.0001W D 0.0000 -- 9.999 0.0001 Q 0.0000 -- 9999 0.0001 θ -179.9° -- 179.9° 0.01°
Power Supply
Battery Model
TH2822/A IEC 6LR61,9V Alkaline battery TH2822C
LH-200H7C,8.4V Ni-MH 200mAH rechargeable battery
AC Adapter
Input:220V(1±10% ), 50Hz(1 ±5%) Output:12V -15V DC
Operating Current (with backlight off)
Max.:28mA Typical:25mA (@1kHz,100Ωload)
Standby(Power Off)Current
Max. :2μA(Non-rechargeable)
Battery Life
TH2822/A
16 Hours (typical) based on
backlight off and new alkaline
TH2822C
6 Hours (typical) based on backlight off and new fully charged Ni-MH battery
Charge Time and
Current(TH2822C)
Max.: 80min Max. Current: 42mA
Auto Power Off (valid for battery powered)
5min/15min/30min/60min/OFF available ;factory defa ult :5mi n
Low Voltage Indicator
When battery voltage drops below 6.8V, Low Voltage Indicator turns on
Page 88
87
General
Operation Condition
Temperature 0℃ -- 40℃
Relative Humidity ≤90% R.H. Weight 350g
Dimensions (H ´ W ´ D)
190mm ´ 90mm ´ 41mm
Safety and EMC compliance
IEC 61010-1:2001 IEC 61326-2-1:2005
Accuracy Specifications
Notices: 1. Environment temperature:23℃±5℃;
Humidity: ≤75% R.H.
2. Valid after 10 minutes of warm up time. 3. Test in measuring slots on front panel. 4. Measurements performed after correct open
and short correction. 5. Test in the recommended equivalent mode. 6. Percentage accuracy representation:
±(%reading+ least significant digits)
7. Actual measurement and display range may
exceed the ranges in below table, but we do
not specify its accuracy again. 8. Subscript Explanation:
Page 89
88
s—series equivalent; p—parallel equ ival ent;
e—accuracy 9. Some parameters cannot be expressed as
the way of data sheet; therefore they can just
be converted by formulas according to
corresponding results.
Inductance (L) and Quality Factor (Q)
Range Display Range
Accuracy
Equi-
valent
Mode
Le De *
100Hz/120Hz
1000H 400.0H∼1000.0H 1.00%+3 0.0100 Parallel
400H 40.00H∼399.99H 0.35%+2 0.0035 Parallel
40H 4.000H∼39.999H
0.25%+2 0.0025 Parallel
4H 400.0mH∼3.9999H 0.25%+2 0.0025 ----
400mH 40.00mH∼399.99mH 0.25%+2 0.0025 Series
40mH 4.000mH∼39.999mH 0.45%+2 0.0045 Series
4mH 0uH∼3.999mH 1.40%+5 ------ Series
1kHz
100H 40.00H∼100.00H 1.00%+3 0.0100 Parallel
40H 4.000H∼39.999H 0.35%+2 0.0035 Parallel
4H 400.0mH∼3.9999H 0.25%+2 0.0025 Parallel
400mH 40.00mH∼399.99mH 0.25%+2 0.0025 ----
40mH 4.000mH∼39.999mH 0.25%+2 0.0025 Series
4mH 400.0uH∼3.9999mH 0.45%+2 0.0045 Series
400μH 0.0uH∼399.9μH 1.40%+5 ------ Series
Page 90
89
10kHz
1000mH 400.0mH∼999.99mH 0.80%+3 0.0080 Parallel
400mH 40.00mH∼399.99mH 0.35%+2 0.0035 Parallel
40mH 4.000mH∼39.999mH 0.25%+2 0.0025 ----
4mH 400.0uH∼3.9999mH 0.30%+2 0.0030 Series
400μH 40.00uH∼399.99μH 0.45%+2 0.0045 Series
40μH 0.00uH∼39.99μH 1.40%+5 ------ Series
100kHz
100mH 40.00mH∼399.99mH 1.20%+5 0.0120 Parallel
40mH 4.000mH∼39.999mH 0.80%+2 0.0080 Parallel
4mH 400.0uH∼3.9999mH
0.50%+2 0.0050 ----
400μH 40.00uH∼399.99μH 0.50%+2 0.0050 Series
40μH 4.000uH∼39.999μH 0.80%+5 0.0080 Series
4μH 0.000uH∼3.999μH 2.50%+10 ------ Series
*Note: Accuracy of De is assessed when De <0.5 Quality factor Q and Accuracy Qe is calculated by the following formula:
when
ex
DQ ×
≤ 1,
ex
ex
e
DQ
DQ
Q
×
×
±=
1
2
Where,
x
Q
is the measurement value.
Page 91
90
Capacitance(C) and Dissipation (D)
Range
Display Range
Accuracy
Equi-
valent
Mode
Ce De*
100Hz/120Hz
20mF 4.000mF∼20.000mF 5.00%+5 ±0.0500 Series
4mF 400.0μF∼3.9999mF 1.00%+3 ±0.0100 Series
400μF 40.00μF∼399.99μF 0.35%+2 ±0.0035 Series
40μF 4.000μF∼39.999μF 0.25%+2 ±0.0025 Series
4μF 400.0nF∼3.9999μF 0.25%+2 ±0.0025 ----
400nF 40.00nF∼399.99nF 0.25%+2 ±0.0025 Parallel
40nF 4.000nF∼39.999nF 0.35%+3 ±0.0035 Parallel
4nF 0pF∼3.999nF 1.25%+5 ------ Parallel
1kHz
1000μF 400.0μF∼999.99μF 2.00%+5 ±0.0200 Series
400μF 40.00μF∼399.99μF 1.00%+3 ±0.0100 Series
40μF 4.000μF∼39.999μF 0.35%+2 ±0.0035 Series
4μF 400.0nF∼3.9999μF 0.25%+2 ±0.0025 Series
400nF 40.00nF∼399.99nF 0.25%+2 ±0.0025 ----
40nF 4.000nF∼39.999nF 0.25%+2 ±0.0025 Parallel
4nF 400.0pF∼3.9999nF 0.35%+3 ±0.0035 Parallel
400pF 0.0pF∼39.99nF 1.25%+5 ------ Parallel
Page 92
91
10kHz
100μF 40.00μF∼100.00μF 3.00%+5 ±0.0300 Series
40μF 4.000μF∼39.999μF 1.50%+3 ±0.0150 Series
4μF 400.0nF∼3.9999μF 0.35%+2 ±0.0035 Series
400nF 40.00nF∼399.99nF 0.25%+2 ±0.0025 Series
40nF 4.000nF∼39.999nF 0.25%+2 ±0.0025 ----
4nF 400.0pF∼3.9999nF 0.25%+2 ±0.0025 Parallel
400pF 40.00pF∼399.99pF 0.35%+3 ±0.0035 Parallel
40pF 0.00pF∼39.99pF 1.25%+5 ------ Parallel
100kHz
10μF 4.000μF∼10.000μF 6.00%+20 ±0.0600 Series
4μF 400.0nF∼3.9999μF 2.50%+10 ±0.0250 Series
400nF 40.00nF∼399.99nF 0.80%+5 ±0.0080 Series
40nF 4.000nF∼39.999nF 0.50%+2 ±0.0050 Series
4nF 400.0pF∼3.9999nF 0.50%+2 ±0.0050 ----
400pF 40.00pF∼399.99pF 0.80%+2 ±0.0080 Parallel
40pF 4.000pF∼39.999pF 1.20%+5 ±0.0120 Parallel
4pF 0.000pF∼4.999pF 3.00%+10 ------ Parallel
Page 93
92
Impedance(Z) and Phase Angle
(θ)
Range
Display Range
Accuracy
Equi-
valent
Mode
Ze
θe
100Hz -- 10kHz
10MΩ 4.000MΩ∼10.000MΩ 3.00%+5 ±1.75° Parallel
4MΩ 400.0kΩ∼3.9999MΩ 1.25%+3 ±0.75° Parallel
400kΩ 40.00kΩ∼399.99kΩ 0.35%+2 ±0.25° Parallel
40kΩ 4.000kΩ∼39.999kΩ 0.25%+2 ±0.15° Parallel
4kΩ 400.0Ω∼3.9999kΩ 0.25%+2 ±0.15° ----
400Ω 40.00Ω∼399.99Ω 0.25%+2 ±0.15° Series
40Ω 4.000Ω∼39.999Ω 0.35%+2 ±0.25° Series
4Ω 0.4000Ω∼3.9999Ω 1.00%+3 ±0.60° Series
0.4Ω 0.0000Ω∼0.3999Ω 3.00%+5 ----- Series
100kHz
10MΩ 4.000MΩ∼10.000MΩ 8.00%+20 ±4.60° Parallel
4MΩ 400.0kΩ∼3.9999MΩ 3.00%+10 ±1.75° Parallel
400kΩ 40.00kΩ∼399.99kΩ 1.20%+5 ±0.69° Parallel
40kΩ 4.000kΩ∼39.999kΩ 0.80%+2 ±0.46° Parallel
4kΩ 400.0Ω∼3.9999kΩ 0.50%+2 ±0.30° ----
400Ω 40.00Ω∼399.99Ω 0.50%+2 ±0.30° Series
40Ω 4.000Ω∼39.999Ω 0.80%+5 ±0.46° Series
4Ω 0.4000Ω∼3.9999Ω 2.50%+10 ±1.43° Series
0.4Ω 0.0000Ω∼0.3999Ω 6.00%+20 ----- Series
*Note: Accuracy of De is assessed when De <0.5
Page 94
93
Equivalent Series Resistance
Accuracy of equivalent series resistance is calculated by below formula:
Rse =
ex
X
φ
×±
Where,
x
X
is measured reactance,
xx
fLX
π
2=
or
x
x
fC
X
π
2
1
=
e
φ
is the phase angle accuracy,
180
π
θφ
×=
ee
Notice:the accuracies of ESR and Rs are same.
Equivalent Para ll el Resis ta nc e
Accuracy of equivalent series resistance is calculated according to the below formula:
Rpe =
ex
epx
D
R
φ
φ
×
±
R
px
is the measured value of Rp, Dx is the measured
value of dissipation.
Page 95
94
Maintenance
W
ARNING: Do not perform any service by
yourself. Service should only be done by quali fied personnel and trained technicians. WARNING: Beware of liquid and residues, especially conductive matter.
Se
rvice
I
f the instrument fails to power on, first check battery, external source and power sockets and the validity of buttons. If the measurement result is abnormal, first inspect the condition of test accessories and damage of reed in test slots. See operation manual for information of correct operation. Do not replace components. Contact the after-sale department of our company or relevant distributors for uncertain maintenance. The meter must be turned off or removed from external power before replacing battery. See Installing Battery section for details.
Page 96
95
Cleaning
B
efore cleaning this meter, make sure the power is OFF and remove external AC adapter if one is used. To avoid electrical shock or damaging the meter, prevent water from getting inside the case. In the case that water gets inside, remove the battery immediately and do not operate the instrument immediately. To clean the meter, wipe the dirty parts with a soft cloth soaked with diluted neutral detergent. Avoid having the instrument too wet to prevent the detergent from penetrating into the inside components of the meter, causing damages. After cleaning, make sure the instrument is completely dried before operating it again.
Page 97
96
Limited Warranty
The instrument, component parts and accessories will be free from defects in workmanship and materials for a period of two years from date of purchase. Tonghui will, without charge, repair or replace, at its option, defective product or component parts. Returned product must be accompanied by proof of the purchase date in the form of a sales receipt and the included accessories. This warrant y does not appl y in the event of mis us e or abuse of the product or as a result of unauthorized alterations or repairs. The warranty is void if the serial number is altered, defaced or removed.
Loading...