Tobii PCEye User’s Manual
59
Chapter 6 Appendices
Guidance and manufacturer’s declaration - electromagnetic immunity
Following is the description and manufacturer’s declaration of electromagnetic immunity for the Tobii PCEye.
Immunity test
Electromagnetic Immunity
- Operative and Compliance
Description
Electromagnetic
Environment - Guidance
Electrostatic Discharge
EN 61 000-4-2 (1995) + A1 + A2
Test level +/- 2 , +/- 4 and +/- 6kV contact
discharges and +/- 2, +/- 4 and +/- 8 kV
air discharges.
The EUT operated without any
degradation during the test.
The EUT complies with the performance
criteria.
Floors should be wood,
concrete or ceramic tile.
If oors are covered with
synthetic material, the relative
humidity should be at least 30%.
Mains power quality should be
that of a typical commercial or
hospital environment.
Radiated Electromagnetic Fields in
the Frequency Range 80 - 2500 MHz
EN 61 000-4-3 (2006)
Test level 3 V/m with 80% AM @ 1 kHz.
The EUT operated without any
degradation during the test.
The EUT complies with the performance
criteria.
Fast transient/burst
EN 61 000-4-4 (2004)
Test level: ±2 kV on AC power port.
The EUT operated without any
degradation during the test.
The EUT complies with the performance
criterion.
Surge
EN 61 000-4-5 (2006)
Test level: ±0,5 kV and ±1 kV line to line
on AC power port.
The EUT operated without any
degradation during the test.
The EUT complies with the performance
criterion.
Power Frequency Magnetic Fields
EN 61 000-4-8 (1993) + A1
The EUT has no parts sensible to a
magnetic eld. The product is deemed
to comply with the standard without
testing.