Titan Controls Atlas 5 Instruction Manual

Atlas 5TM
Handheld Indoor Air Quality Monitor
Instruction Manual
®
Measures CO2,
INTRODUCTION
Features:
Triple displays of CO level, temp.
2
and humidity
s
asy calibration humidity PC connect
Stable NDIR sensor for CO detection
2
Statistics of weighted average
TWA (8 hours weighted average)
STEL(15 minutes weighted average)
Backlight for working in dark areas Audible CO warning alarm
2
Battery and adaptor power supply E manual on CO and
2
via RS232 interface
1
Thank you for purchasing this portable
meter. This device measures CO
levels, air temperature and humidity. It is an ideal instrument for indoor air quality (IAQ) diagnostics.
ppm
The portable CO meter uses NDIR (non-
2
2
dispersive infrared) technology to ensure the reliability and long term stability.
MATERIAL SUPPLIED
Meter 4 AA batteries Operation manual Hard carrying case
This package contains:
POWER SUPPLY
The meter is powered by either 4 AA batteries or a DC adaptor(9V/1A output).
Install the batteries into the battery compartment on the rear and make sure they are in correct polarity and good contact.
nd “Lob” will appear on the LCD . And beeper sounds.
atter
When an adaptor is used, it will cut off the power supply from batteries. The adaptor can’t be used as a battery charger.
When battery voltage gets low, a
(Fig.1)
The CO sensor
2
can’t work under low voltage, so it beeps to indicate failed CO measurement
2
(press any key but to stop the beeps) and the readings won’t be displayed. Please replace with fresh b ies or connect with an adaptor.
IAQ55
%
O
CO
2
Fig.1
2
SET
3
LCD DISPLAY
TWA Time weighted average( STEL hort-term exposure limit (15 minutes weighted
MIN/MAX Minimum/Maximun readings Low battery indicator DP Dew point temperature AIR Air temperature WBT Wet bulb temperature % Unit of relative humidity
o
E (C/F) Celsius/Fahrenheit
AVG/ftm/m/s Vain icons in these models
8 hours)
S
average)
HOLD Readings are freezed unchanged
%
O
TWA
STEL
HOLD
ftm
m/s
CO
2
CO2 concentration
in ppm
Relative humidity in %
Air temp. Dew point Wet bulb temp. in
o
C or F
o
Symbols
Turns on and off the meter. Enters setup mode. Sets as non-sleep mode with .
SET
CAL Esc
DP/WBT
N X
M /AV
Exits setup page/mode. Enters CO calibration with .
2
Enters RH calibration with .
Freezes the current readings. Cancels data hold function.
Activates or cancels the backlight. Selects unit or increases value in setup.
Selects AIR, DP, WBT temps display. Selects unit or decreases value in setup.
Activates MIN,MAX,STEL,TWA function. Saves and finishes settings.
DP/WBT
KEYPAD
4
OPERATION
Press to turn the meter on and off. At power up, it emits a short beep and performs 30 seconds countdown(Fig.2) for meter warm up, then enters normal mode with current CO , temperatures,
2
and humidity readings displayed (Fig.3).
SET
%
O
CO
2
Fig. 2
Fig 3
Model IAQ55
DP/WBT
%
O
CO
2
%
O
CO
2
Fig.4
Fig.5
TAKING MEASUREMENT
Press to switch temperatures display. The lower left display will cycle from air temperature, dew point temp. (Fig.4), and wet bulb temp.(Fig.5).
The meter starts measurement when power on and update readings every second. In the condition of operating environment change (ex. from high to low temp.), it takes 30 sec to respond for CO sensor and 30 minutes for RH.
2
NOTE: Do not hold the meter close to face, breathing on meter affects CO
2 levels.
POWER ON/OFF
AIR,DP,WBT
DATA HOLD Press to freeze the readings, "HOLD” icon is displayed on the left top of LCD(Fig.6). All current readings are kept unchanged, except STEL and TWA. Press again to cancel the hold function.
%
O
HOLD
CO
2
Fig.6
BACKLIGHT Hold down for more than 1 second to activate and cancel backlight function.
MIN,MAX,STEL,TWA Under normal mode, press to see the minimum, maximum, and weighted average readings. Each press of , it displays MIN, MAX, STEL, TWA in sequence and returns to normal mode.
In MIN and MAX modes, it shows the minimum and maximum readings of CO on main display and of AIRor Dpor
2
WB temperatures and humidity on the lower displays. (Fig.7)
N X
M /AV
%
O
CO
2
Fig.7
In STEL and TWA modes, the main display shows the
(STEL) and 8 hours(TWA).
weighted average of
CO
readings for the past 15 minutes
2
The lower displays are the current AIR, DP/WB temperatures and humidity. (Fig.8)
5
N X
M /AV
%
O
STEL
CO
2
Fig.8
NOTE:
1.
While all readings are held unchanged,
STEL and TWA will keep updating every 5 minutes.
If the meter is turned on for shorter
than 15 minutes, the STEL value will be the weighted average of readings taken since power on. Same for TWA values that appear before 8 hours.
2.It takes at least 5 minutes to calculate STEL and TWA. The display shows “----” (Fig.9) during the first 5 minutes from power on.
3.
SET
%
O
STEL
CO
2
Fig. 9
The meter turns off automatically after 20 minutes of inactivity. To override the function, hold down and for 2 seconds to turn on the meter until “n” appears. NOTE: Auto sleep function will be disabled during calibration mode.
6
SET
AUTO POWER OFF
ALARM
The meter features audible alarm to give warnings when CO concentration
2
exceeds the limit. (See P1.0 in setup for setting alarm threshold). It emits beeps(Abt.80dB)when CO level goes
2
over the set value and stops when any key (but ) was pressed or readings fall below the set value. It beeps again when value goes over the limit. Restart the meter if beeper can't be stopped.
SETUP
Hold down un
setup mode.
To exit setup, press in P1.
0 or P3.0
and it returns to normal mode. Note: P2.
0 is not applicable in these
models.
P1.0 CO ALARM
2
When entering setup mode, P1.0 and “AL” (Fig.10) are displayed on the LCD. Press to go into P1.
1 for setting
CO
alarm threshold. The current set
2
value will be blinking on LCD
Press to increase the value or to decrease. Each press tunes
and the alarm range is from 100 to 9900ppm. When the preferred alarm value is set, press to save the setting or without saving and return to P1.
0.
P3.0 TEMPERATURE SCALE Press or in P1.
0 to access P3.0
for setting up temperature scale(Fig.12). Press and it goes into P3.
1 with
oo
blinking C or F current set(Fig.13) on the lower left display. To switch
press to save the setting or without saving and return to P3.
0.
der normal mode for
more than 1 sec to enter
(Fig11).
100 ppm
oo
C or F,
press and . Then
SET
CAL Esc
N X
M /AV
DP/WBT
N X
M /AV
CAL Esc
CO
2
CO
2
Fig. 10 Fig. 11
DP/WBT
N X
M /AV
DP/WBT
N X
M /AV
CAL Esc
Fig. 12
Fig. 13
7
O
8
CO CALIBRATION
2
CAUTION:
Do not calibrate the meter in the air with unknown CO concentration.
2
Otherwise, it will be calibrated as 400ppm by default and leads to inaccurate measurements.
CAL Esc
CO
2
Fig. 14
NOTE:
Ensure the batteries are with full voltage during the calibration to prevent from interruption or failed calibration.
The manual calibration is suggested to to be done in fresh outdoor air that is well ventilated.
Place the meter in the calibration site. Turn on the meter and hold down and simultaneously to enter CO
2
calibration mode (Fig.14). 400ppm and “CAL” are blinking on the LCD while performing calibration.
Wait about 5 minutes until the blinking stops and the calibration is completed automatically and back to normal mode.
To abort the calibration, turn off the meter at any time.
The meter is calibrated at standard 400ppm CO concentration in factory.
2
It’s suggested to do manual calibration regularly to maintain good accuracy.
9
33% calibration
Plug the sensor probe into 33% salt bottle. Hold down and under normal mode to enter 33% calibration (Fig.15). “CAL” and calibrating value (32.7% if at LCD with current temperature at the left.
Meter is now calibrating, and will finish in about 60 minutes when “CAL” and humidity stop blinking. (Fig.16)
o
25 C,77oF) are blinking on the
The meter defaults to have the humidity calibrated with 33% and 75% salt solution. The ambient condition is recommended
o
to be at 25 C (77oF) and stable humidity (better to be close to the calibrating value). To abort calibration, just turn off the meter.
RH CALIBRATION
CAL Esc
DP/WBT
%
O
Fig. 15
%
O
Fig. 16
%
O
Fig.17
75% calibration
After 33% calibration, plug the sensor probe into 75% salt bottle, then press to enter 75% calibration (Fig.17).
N X
M /AV
CAUTION:
Do not calibrate the humidity without the default calibration salt. Otherwise, it will cause permanent damage.
10
NOTE:
Users can also calibrate either point. To calibrate 33% only, press and exit when 33% calibration is completed. To calibrate 75% only, press or within the 5 minutes while initializing 33% calibration.
CAL Esc
DP/WBT
“CAL” and calibrating value (75.2% if
o
at 25 C, 77oF) are blinking on the LCD with current temperature at the left. Meter is now calibrating. Wait about 60 minutes until blinking stops, then calibration is
completed and it returns to normal mode.
? Error messages
E01: CO sensor damaged.
2
E02: The value is under range. E03: The value is over range. E04: The original data error results in this error (DP, WB) E07: Too low voltage to measure CO .
2
Replace batteries or use an adaptor. E11: Retry humidity calibration. E17: Retry CO calibration.
2
E31: Temperature sensor damaged. E34: Humidity sensor damaged.
? Can’t power on
Press for more than 0.3 seconds and try again. Check if batteries are in good contact and correct polarity or the adaptor is plugged in and getting 120 volt power.
TROUBLESHOOTING
SET
? Fixed readings
Check whether data hold function was activated. (HOLD icon at the left top)
? Slow response
Check whether the air flow channels on the rear are blocked.
11
PC CONNECTION
The meter can do PC link for
data analysis
on-line logging and via RS232 interface and software. The protocol is as follows.
Model IAQ55
Cxxxxppm:Txxx.xC(F):Hxx.x%: dxxx.xC(F):wxxx.xC(F) LRC CRLF
Description: $CO :Air:RH:DP:WBT LRC
2
CRLF
A.9600 bps, 8 data bits, no parity.
SPECIFICATION
SPECIFICATION
IAQ55
CO
2
Temperature
Humidity
Range 14°F-140°F (-10.0-60.0°C) Resolution 0.1°F / 0.1°C Accuracy +0.9°F / +0.6°C
Range 0.0-99.9% Resolution 0.1% Accuracy +3% (10-90%) +5% (others)
Warm Up 30 seconds Operating 32°F-106°F (0-50°C), 0-95%RH
(avoid condensation) Storage 68°F-140°F (-20-60°C), 0-99%RH (avoid condensation)
Power 4 AA batteries, DC adaptor Battery Life
10-24 hours (depending on battery type)
Range
Resolution 1 ppm Pressure +1.6% reading per kPa deviation Dependence from normal pressure, 100kPa
0-2000ppm, Accuracy +75ppm +5% Reading 2001-9999 Accuracy (not specified)
Notes:
13
WARRANTY
The meter is warranted to be free from defects in material and workmanship for a period of one year from the date of purchase. This warranty covers normal operation and does not cover misuse, abuse, alteration, neglect, improper maintenance, or damage resulting from leaking batteries. Proof of purchase is required for warranty. Warranty is void if the meter
has been opened.
VANCOUVER, WASHINGTON U.S.A.
www.titancontrols.net
DISTRIBUTED BY
Loading...