TIS CHECKBOX, TIS172ML, TIS173DB, TIS25 Instruction Manual

NOTE THE INSTRUMENT MUST ONLY BE USED BY SUITABLY
TRAINED AND COMPETENT PERSON(S).
TIS CHECKBOX
1
CONTENTS
Introduction & Safety
2-3
Checkbox Description & Maintenance
4-6
Checkbox Operation
7-9
Technical Specification
10
Recommended Verification Sequence Record Form
11
Warranty Information
12
Optional Accessories / Test Leads:
Part Number: TIS172ML Description: Universal three wire 4mm 13A Mains lead. Part Number: TIS173DB Description: Three wire distribution board lead set complete with crocodile
clips and probes. Part Number: TIS25 Description: Universal large padded carry case.
INTRODUCTION & SAFETY
2
INTRODUCTION & SAFETY
TIS Checkbox is a professional, multifunctional field checkbox, intended for use with all makes and models of installation test instruments.
In addition to its durable body and simple operation it boasts the following features:
 Insulation calibration with test voltages up to 1000 V,  LowΩ and continuity accuracy verification,  Continuity and insulation of measurement leads verification,  Hi Current and “No-trip loop” impedance measurement verification  Verification of RCD trip-out time measurement, including test current verification - 30 mA X1/X5.  Voltage and frequency calibration (A calibrated Multimeter to be used to verify mains voltage &
frequency). Recommended Multimeter(s): TIS201 TIS280
Please visit our web site: www.testinstrumentsolutions.co.uk to explore the full range of test instruments we provide.
The TIS Checkbox is designed for accuracy verification whilst on-site testing, this product does not replace the need to have the instrument calibrated periodically.
(Please refer to the manufacturers recommended calibration interval)
3
Safety Warnings and Precautions must be read and understood before the instrument is used. It is advised that the user is observed during use.
The instrument should not be used if any part of it is damaged. Replacement fuses must be of the correct type and rating, failure to fit the correctly rated fuse
may result in a safety hazard / damage to the TIS Checkbox.
The polarity of the mains input must be verified before powering the TIS Checkbox, failure to do
so may result in a safety hazard / damage to the TIS Checkbox.
Damaged leads / accessories must not be used in conjunction with the TIS Checkbox as this
may result in a safety hazard / damage to the TIS Checkbox.
Do not use the TIS Checkbox on supply systems with a voltage greater than 250V. Do not disassemble, risk of electric shock – service intervention or adjustment procedure is
allowed to be carried out only by a competent authorized person.
Caution: Risk of Electric Shock
Equipment protected throughout by Double Insulation (Class II)
Equipment complies with relevant EU Directives
Do not dispose in normal waste stream.
The TIS Checkbox is manufactured and tested to the following standards:
EN 61326 – Electromagnetic Compatibility Regulations. EN 61010 – Safety Regulations.