3.1 Introduction of functions of each part
1) LCD: Display the test content and data
2) F key: Function key. Complete the combined function in conjunction with other
keys
3) ▲ key: UP key. Turn pages when viewing the test results
4) A key: Start the CLASS I test
5) POWER key: Turn on/off the machine
6) B key: Start the CLASS II test
7) ☀ Key: Turn the backlight on/off
8) ▼ key: DOWN key. Turn pages when viewing the test results
9) C key: Start the CORD test
10) Output socket: For the output of the test voltage during CLASS I, CLASS II and
CORD tests
11) Input socket: For the input of the test voltage during CORD and out-phase tests
12) Input Earth Bond Lead terminal (A conductor for connecting external earth points
on CLASS I appliances).
13) 12VDC power connector: A power connector when an external 12VDC power is
applied
I. Operation of instrument
4.1 Turn on / off
Press POWER to turn the machine on/off;
The instrument will be automatically shut down if there is no operation within 5 minutes;
Note: When the instrument displays , this indicates a lack of power and it is
necessary to replace the battery.
4.2 Calibration
During the replacement of the test leads, including the IEC lead (used when testing
extension cables / C Type functionality) or Earth Bond leads these functions require
recalibration of the TIS 600 before use. (This calibration will compensate for the lead
resistance during testing)
① Calibrate the CLASS I test grounding resistance.
Insert the Earth bond lead into the output socket (12) with the probe supplied, insert the
probe into the earth terminal and the input socket (10). Press and hold the F (2) and A (4)
function key together until the calibration interface is enabled.
Press the A (4) key, this will then perform the lead calibration/ adjustment, Press the F (2)
to exit the calibration process.
② Calibrate the CORD/IEC test ground resistance.
Insert the CORD/IEC mains test cord into the output socket (10) and the other end into
input socket (11). Press the POWER key to turn on the instrument, and press and hold the
combination key F + A until entering the calibration interface. Press the C key to calibrate
the CORD test grounding resistance; after the calibration is completed. The resistance of