• Operation
Switch the unit on using the ON/OFF key. The display shows “CL F”. Select the test required using the MODE key: CLF (free Chlo­ rine) → CLt (total Chlorine) → Br → PH → CYS → tA → CLF ... (Scroll). The display shows the desired method.
Open the grey lid on the top of the sample chamber. Swirl the instrument several times under the water surface
allowing to fill the sample chamber completely. Take the instrument out of the water and close the grey lid. Hold the lid firmly down. Press the ZERO/TEST key. The method symbol flashes for approx. 8 seconds. The display shows “0.0.0”. Open the grey lid and add the approriate tablet: Method appropriate tablet Free chlorine DPD free Total chlorine DPD total pH value pH 6.5 - 8.4 Bromine DPD free Cyanuric acid Stabilizer Alkalinity Alkalinity { Close the lid and swirl the instrument gently for approx. 30 seconds (Alkalinity 1 min) holding the lid firmly down. | Afterwards keep the instrument steady and wait for 30 seconds when testing CLF/pH/CYS and 2 minutes when testing CLt/tA. }Press the ZERO/TEST key. The method symbol flashes for approx. 6 seconds. The result appears in the display. Rinse the sample chamber and lid area thoroughly after each test. It is not possible to repeat the reading because the measurement value is correctly displayed only during the waiting period stated above.
New zero calibration
Press the MODE key repeatedly until the desired method symbol appears in the display again.