DIRECTIONS FOR USING THE FLOW CALIBRATOR 58-FMC
Connect the Quick Connect to the high pressure port of DP Transmitter under calibration.
• Leave the low pressure port of the transmitter open to atmosphere.
• Connect the output from the transmitter through required power supply to an accurate digital multimeter.
• Divide the range of the transmitter into five equal divisions, example 6000 mm WC =0, 1500 mmWc -25%, 3000 mmWC – 50%, 4500 mmWC – 75% and 6000 mmWC – 100%.
•  Check the Zero output of the transmitter without giving any pressure to the HP and LP side.
•  Give 6000 mmWC (full span 100%) to the HP side, check output correct span if required. Check zero for any shift and correct if necessary.
• Close the return valve of the squeeze bulb and pump pressure using the squeeze bulb to 25% (eg. 1500 mmWC) check the output.
• Go on in steps of 25% to check intermittent calibration.
•  Once again check zero and span for any shift or error.
•  Tabulate readings for future reference.
INSTRUCTION MANUAL
ISSUE 1217
www.tinsley.co.uk
3
KEY FUNCTIONS
ON/OFF : Using this key you can switch the gauge ‘ON’ and Turn ‘OFF’’.
MIN/MAX : It displays MIN/MAX pressure values.
CLEAR : This clears the previously stores min/max values.
ENGG UNITS : Using this key you can toggle through the different pressure Engineering units.
POWER & PRESSURE CONNECTION
POWER : 7.5 VDC internal alkaline battery AA size (5x1.5).
PRESSURE : ¼” BSP Male CONNECTION
PRESSURE RANGES AVAILABLE
: (-) 1000mmWC, (-) 1000mmWC
: 0 to 1000mmWC
: 0 to 5000mmWC
: 0 to 6000mmWC
: 0 to 7500mmWC
: 0 to 10000mmWC
: 0 to 30000mmWC
: 0 to 35000mmWC
Custom built ranges available upto 3500000mmWC
INSTRUCTION MANUAL
ISSUE 1217
www.tinsley.co.uk
4
PROCEDURE FOR CHANGING PASSWORD
Switch on the gauge & immediately Press the FnS key.
Password ‘00000’ (factory set password) is displayed.
Press ‘Up’ key when ‘C A L I’ is displayed.
‘PASC’ will be displayed.
Press enter when t h e above is displayed.
‘nEuPS’ is displayed as follows.
Press enter when th e above is displayed.
Enter the new password using shift & up keys.
Press enter.
Display will return to ‘PASC’ display.
Press Fns key to return the gauge to normal operation.
WARNING
Severe injuries can occur through improper use of pressure instruments. Do not exceed the recommended pressure limits of tubing and fittings. Never disconnect pressure instrument without relieving the system pressure first. Never insert any object into the pressure connection. The sensor diaphragm is very thin and will be damaged or destroyed by solid or sharp objects.
CONTACT US
ISSUE 1217
All information provided by Tinsley in this datasheet is believed to be accurate. Tinsley reserves the right to discontinue and change specifications and prices at any time without prior notice. For further details, please refer to our website www.tinsley.co.uk
1 Warner Drive Springwood Industrial Estate Braintree, Essex CM7 2YW
Tel: 01376 335271 E-mail: info@tinsley.co.uk
www.tinsley.co.uk
5
CONTACT US
Tinsley Instrumentation Ltd 1 Warner Drive Springwood Industrial Estate Braintree, Essex CM7 2YW
Tel: 01376 335271 E-mail: info@tinsley.co.uk
www.tinsley.co.uk
Loading...