1. HOLIDAY DETECTOR FUNCTIONThe Tinker & Rasor Model AP/AC In-Plant Holiday Detector is an all-purpose electrical
inspection instrument which maintains a given inspection voltage in spite of the
electrical load on the circuit. It is recommended for use on any pipe diameter as well
as on flat surfaces when such surfaces are coated with a highly electrical resistance
material, and when the surface beneath the coating is electrically conductive. The
detector works equally well on damp or dry surfaces, and is especially desirable where
humid conditions prevail. The instrument can be delivered for operation on either 110v
or 220v AC power.
2. INSTRUCTIONS FOR UNPACKING & INSPECTION
A. Open the shipping carton and make note of the various accessories.
B. INSPECTION should be made upon receipt. If damage has occurred during
shipment, file a claim with the carrier immediately.
C. If it is necessary to contact your supplier or the manufacturer concerning
damaged or missing items, be sure to include the serial number, purchase order
number, and invoice number of the instrument in question.
3. CHECK-OUT INSTRUCTIONS
• Remove all accessories from the shipping carton.
• Make sure that the detector is turned OFF.
• Connect high voltage cable and ground cable to detector (insert and twist
clockwise). Never connect or disconnect the cables when the instrument is
turned on.
• Attach probe handle to voltage cable.
• Attach electrode to the voltage probe assembly and apply to the structure to be
inspected. The electrode should always make intimate contact with the surface
under inspection.
• Plug AC power cord into 110v power supply (220v on specially configured
instruments). See label on outside of case for type.
• The Model AP/AC Holiday Detector is now ready to operate.
• SELECT the HIGH or LOW position. Adjust the voltage setting, using the
digital display for reference.
NOTE: Digital display and actual output may differ as much as 5% due to manufacturer’s
specifications of display.
4. OPERATING METHODSA good ground return system for both the pipe and the detector will always provide the
best and most reliable inspection. The pipe to be inspected must be grounded from
the bare pipe to earth at some point along the pipe. If individual joints of the pipe are
to be inspected that are not electrically connected, each joint must be grounded. The
speed of the electrode's travel along the pipe should never be excessive, since faulty
inspection may result. Occasional checks of the detector operation should be made,
particularly if no holidays are being found. This can be accomplished by testing for the
spark and signal at the edge of the coating where bare pipe exists or by touching the
probe end to the bare pipe and noting the length of the spark and the signal
effectiveness. If the signal does not occur when the spark discharges from the
electrode into a known holiday, the ground return (i.e. the path between the metallic
pipe and earth and the earth to the ground trail of the detector) is of high resistance.
In this case, a better ground is required and a direct connection between the metal
pipe and the ground wire may be necessary. This type of grounding is extremely
uncommon unless the soil is very dry (if using a grounding rod) or the detectors output
is low.
Correct voltage output for a given thickness of coating has long been a matter of
controversy. However, recent formulas have been suggested which may be used as a
guide for correct peak voltages on various coating thickness. The calculation is as
follows:
V =K √ T
WHERE V =Peak voltage in volts.
T =Thickness of bonded coating in mils (not
including outer wrappers).
K =Constant of 1,250 for coatings over 30 mils
and 525 for fusion bonded epoxy coatings
under 30 mils.
These formulas, when applied to a coating of 3/32” thickness and with a constant of
1,250, would indicate an applied voltage of 12,500 volts peak or a coating of 16 mil
thickness with a constant of 525 would indicate an applied voltage of 2,100 volts. A
common practice used in setting inspection voltages in the field is to adjust the output
voltage by visual observation. It is the general consensus that spark discharge at
least twice the thickness of the coating will give adequate inspection voltage and
compensate for any irregularity in coating thickness and grounding conditions. If this
practice is desired for determining inspection voltage, it should be done while the
electrode is in the normal operating position and under actual grounding conditions.
The formulas and suggestions for setting voltage are supplied as a guide, and are not
necessarily recommended by the manufacturer of this detector.
Consult the coating manufacturer for recommended voltage applications.
5. INSTRUMENT SERVICING INSTRUCTIONS
A. Cleaning: Keep the instrument clean and dry. Clean the instrument case with a
soft cloth dampened with kerosene, then wipe dry. Do not use solvents such as
lacquer thinner, methyl ethyl keytone, etc.
KEEP ALL ELECTRICAL CONTACTS CLEAN
B. Voltage Output Checks
A. In case of LOW output voltage:
a. Check position of voltage selector switches.
b. Check for “parted” conductor in voltage probe and/or ground trail.
B. In case of NO output voltage:
a. Check power cord and plug.
b. Check for “parted” conductor in voltage probe and/or ground trail.
6. FACTORY REPAIRSHoliday Detectors returned to the factory for repairs should be sent
TRANSPORTATION PREPAID. In most cases the detector can be repaired and
returned the same day it is received at the factory.
WHEN ORDERING PARTS FOR YOUR DETECTOR OR REQUESTING FURTHER INFORMATION
ALWAYS GIVE THE DETECTOR’S SERIAL NUMBER.
Mailing AddressShipping Address
Tinker & Rasor Tinker & Rasor
P.O. Box 1667 791 S. Waterman Ave
San Bernardino, CA 92412 San Bernardino, CA 92408
This addendum to the Model AP/AC Operating instructions covers the customized features of the
dual Model AP Cabinet, designed to specific specifications from the customer.
This customized instrument cabinet contains two (2) Model AP High Voltage Holiday Detectors,
which have been designed to operate independently from one another. Each instrument has been
factory set to have a maximum high voltage output of 15kV (15,000 volts).
Although these instruments are meant to be operated independently of one another, there is only
one power source for the cabinet, a 110v AC power supply. When the cabinet has been connected
to an 110v AC power source, the green LED located in the center of the cabinet will light.
Each instrument, marked as Unit A and Unit B, has its own connections, located on the side of the
cabinet. These connections include:
(1) High Voltage cable connector
(1) Ground cable connector
(2) 110v Holiday Indication output connectors (external audio and visual indicators)
A ninth connection point is located near the bottom of the side panel, towards the front of the
instrument panel. This connection is for the 110v AC power supply.
Each instrument has its own controls, as described in the main body of these instructions. It is
important to point out the each digital display has its own DC battery power supply (9v alkaline)
which will need to be maintained for proper voltage indication.
Special Notes
The instrument has been supplied with extra long high voltage cables. (100’/each). It is important
that these cables are not coiled upon themselves at any time during operation of the instruments.
Also, if the cables are to rest upon a floor during operation, an underground or buried AC power
line beneath the instrument high voltage cables can cause coupling resulting in a high voltage
spike being sent back into the instrument and causing damage to the holiday detector. It is
recommended that the high voltage cables of each instrument be lifted off the floor to avoid this
possible issue.