Times Technology T100 instruction manual

T100 Vector Impedance Analyzer
timestechnology.com.hk
User Manual Ver. 1.1
T100 is a state of the art portable Vector Impedance Analyzer. This powerful yet handy instrument is specifi cally designed for the advanced amateur radio experimenters working at the VHF/UHF band.
TABLE OF CONTENTS
General Description T100 Menu Structure Key Defi nitions Single Frequency Sweep Frequency Plot Calibration Settings PC Mode Firmware Update Specifi cations Accessories
GENERAL DESCRIPTION
The T100 Vector Impedance Analyzer is essentially a device to inject an RF signal into whatever is connected to its test port, and measures the amplitude and phase of the refl ected RF signal.
T100 Vector Impedance Analyzer can operate from 100MHz to 170MHz (VHF), and from 400MHz to 470MHz (UHF), with a 25KHz minimum step size.
T100 MENU STRUCTURE
MAIN MENU SINGLE FREQ
•SWR
•IMPEDANCE (SERIAL)
•IMPEDANCE (PARALLEL)
•S11 (RECTANGULAR & POLAR) SWEEP FREQ PLOT SWR IMPEDANCE (Z) RESISTANCE (R) REACTANCE (X) RETURN LOSS (S11) PHASE ANGLE CALIBRATE CONNECT OPEN CONNECT SHORT CONNECT LOAD SETTINGS BACKLIGHT AUTO ON OFF AUTO POWER OFF ENABLE DISABLE BATTERY ALKALINE NIMH CAL DATA PROTECT LOCK CAL DATA UNLOCK CAL DATA INFO PC MODE
KEY DEFINITIONS
POWER : Press and hold this key
for two seconds to turn on or off T100 Vector Impedance Analyzer.
CONFIRM : Use this key to confi rm the desired selections and fre­quencies entered.
MODE : In SINGLE FREQ, this key is used to switch between the following modes cyclically:
• SWR
• IMPEDANCE (SERIAL)
• IMPEDANCE (PARALLEL)
• S11 In SWEEP FREQ PLOTS, this key is used to switch between the following func­tions cyclically:
• FREQUENCY
• Y SCALE
• ALIGNMENT
• STEP SIZE
ARROWS : These keys are used to se­lect the desired menu items or op­tions. They are also used to in­crease/decrease the frequency. Hold to increase/decrease frequency auto­matically. In SINGLE FREQ, the step size is fi xed at 25KHz. In SWEEP FREQ PLOT, the current step size can be selected with the MODE key.
0
~
9
NUMERIC : Use these keys to enter the desired frequency directly. Frequencies within the valid oper­ating range will automatically be aligned to the 25KHz steps. Fre­quencies outside the valid operat­ing range will be ignored.
BACK : Cancel the current opera­tion and/or go back to the previ­ous menu.
SINGLE FREQ
In this mode, T100 can display the measured data in any of the following four represen­tations. You may switch cyclically through these representations with the MODE key.
•SWR
•IMPEDANCE (SERIAL)
•IMPEDANCE (PARALLEL)
•S11 (RECTANGULAR & POLAR)
SWR displays the Standing Wave Ratio in large font for easy reading. It also displays the impedance (Z) for reference.
IMPEDANCE (SERIAL) displays the imped­ance seen from the test port. The im­pedance is resolved into the real (re­sistive) part and imaginary (reactive)
SWR
FREQ:435025 Z=138.08
IMPEDANCE
FREQ:435025
101.80 - j112.58
3.25pF SERIAL
part connected in series. The corre­sponding inductance or capacitance is also displayed ac­cording to the sign of the reactance.
IMPEDANCE (PARALLEL) displays the impedance seen from the test port. The impedance is resolved into the real (resistive) part and imaginary (reac­tive) part connected in parallel. The cor­responding inductance or capacitance is also displayed according to the sign of the reac­tance.
IMPEDANCE
FREQ:435025
240.10 - j194.53
1.88pF PARALLEL
S11 displays the rectangular and polar representa­tions of the S11 measured.
S11
FREQ:435025 RECTANGULAR:
0.58 + j0.34 POLAR:
|0.67| <-30.10
SWEEP FREQ PLOT
Under this menu, T100 can plot any of the fol­lowing measured data over the desired frequen­cy range:
•SWR
•IMPEDANCE (Z)
•RESISTANCE (R)
•REACTANCE (X)
•RETURN LOSS (S11)
•PHASE ANGLE Use the MODE key to switch cyclically between
any of the following parameters to change their value.
•FREQUENCY - use the ARROW keys to increment or decrement the frequency by the value of STEP SIZE. The cursor on the plot will move accordingly. The frequency can also be en­tered directly using the numeric keys.
•Y SCALE - use the ARROW keys to select the maximum value on the Y axis. Possible values are 3, 10, 30, 100, 300, and 1000.
•ALIGNMENT - use the ARROW keys to select if the directly entered frequency should be aligned to the CURRENT position of the cursor, BEGIN (leftmost on the plot), CENTER (of the plot), or END (rightmost of the plot). Ranges outside the valid frequencies will be shaded.
•STEP SIZE - use the ARROW keys to select the desired frequency step size. Possible values are 25KHz, 100KHz, 500KHz,and 1MHz.
CALIBRATION
T100 Vector Impedance Analyzer employs indus­trial standard Open/Short/Load (OSL) cali­bration method to eliminate various system errors. To perform accurate measurements, T100 needs to be calibrated against the OPEN, SHORT, and LOAD standards. Three high qual­ity OPEN, SHORT, and LOAD SMA terminators are included for this purpose. Resulting calibra­tion data to be used in computations will be stored in non-volatile memory in T100.
Calibration data is locked upon power up of T100 to avoid ac­cidental erasure. It has to be unlocked in the SETTINGS menu,
CALIBRATE
DATA LOCKED GOTO SETTINGS MENU TO UNLOCK DATA FIRST
before CALIBRATION can be performed.
Connect securely the OPEN standard, and then press the OK key to start calibration. Likewise for the SHORT and LOAD standards in sequence to complete the calibration.
Calibration data will be locked again auto­matically after cali­bration. The instru-
CALIBRATE
CONNECT OPEN
CALIBRATE
CALIBRATION DONE
ment is now ready to use. If desired, T100 can be re-calibrated at any time.
SETTINGS
SETTING MENU
A number of general options and informa­tion are available in this menu.
BACKLIGHT AU TO - Turn off
backlight automat­ically if no key is pressed for 30 seconds.
ON - Leave backlight
always on
BACKLIGHT
AUTO POWER OFF BATTERY CAL DATA PROTECT INFO
BACKLIGHT MENU AUTO
ON OFF
OF F - Leave back-
light always off
AUTO POWER OFF EN ABLE - Automatic
power off if no key is pressed for 5 minutes.
DI SABLE - No auto-
matic power off.
AUTO POWER OFF MENU ENABLE
DISABLE
BATTERY Choose between ALKALINE and NIMH to refl ect the type of batteries being used. It only affects the threshold voltage dictating when a low battery sign on the upper right hand corner of the display will appear.
CAL DATA PROTECT Select UNLOCK CAL DATA before calibration. Calibration data will
BATTERY MENU ALKALINE
DISABLE
PROTECT MENU LOCK CAL DATA
UNLOCK CAL DATA
be locked automatical­ly after calibration. Alternatively, choose LOCK CAL DATA to lock calibration data manu­ally if necessary.
INFO Display general infor­mation of the device, as well as the bat­tery voltage. ( 5 blocks on the battery bar as battery full and 1 block as battery empty.)
SYSTEM INFORMATION
MODEL : T100 H/W VER: 1.00 F/W/VER: 1.00 BATTERY:
PC MODE
T100 is capable of communicating with a PC via an emulated serial link over the USB connection. Virtual serial port driver for Windows XP is included in the T100 CD-ROM and should be installed be­fore connecting your T100 to a PC. A sample host program for Windows XP is also included to demonstrate controlling T100 from a PC, and displaying T100 mea­sured data.
The (virtual) serial communication be­tween T100 and the PC should be confi gured as follows:
Baud Rate = 38400 baud Parity Bit = None Data Bits = 8 Stop Bit = 1
On entering the PC MODE, T100 will dis­play the “WAITING FOR HOST CMD” message and listen to the serial port.
PC MODE
WAITING FOR HOST CMD
To command T100 to tune to a specifi c fre- quency, the host program shall transmit a six-digit frequency in ASCII characters via the serial port.
To command T100 to take a measurement, send an ASCII character ‘S’. Then re­ceive from the serial port a null termi­nated ASCII string. This is the measured magnitude and argument in degrees, sepa­rated by a comma. The same result will be displayed on T100 at the same time.
Send another ‘S’ to repeat measurement at the same frequency, or issue another six digit frequency to tune to a new frequen­cy.
Sending an ASCII character ‘D’ will bring T100 back to the initial state, display­ing the “WAITING FOR HOST CMD” message and listen to the serial port.
PC MODE
FREQ : 144000 |S11| <S11
0.9577 -14.72
FIRMWARE UPDATE
T100 fi rmware can be updated as new releases are available. Updated fi rmware fi le can be downloaded from the Internet to a PC and programmed to the T100 via the USB interface.
Caution
1. Please use fresh batteries when updating fi rmware.
2. Do not turn power off or re­move batteries during fi rmware updating.
Hold down the “Mode”
Firmware Update
key while switching on T100 to enter the “Firmware Update” Mode. T100 will show the screen on the left.
On a PC running Windows XP operating system, open a DOS console in a directory with the two les update.exe and fi rmware.ENC.
Plug in the USB cable to connect the PC and T100. Find out which USB emulated COM port T100 is connected to from the Control Panel/ System/Hardware/Device Manager on the PC.
With T100 at “Firmware Update” mode, issue the following command in the DOS console to update the fi rmware with the fi le
update
rmware
.ENC -COMx -38400
rmware
.ENC :
T100 SPECIFICATIONS
Frequency Range: 100MHz ~ 170MHz (VHF) 400MHz ~ 470MHz (UHF) PLL Synthesized
Frequency Resolution : 25KHz Test Port Connector : SMA Output Power : > 0dBm (typical) Harmonics : < -30dB (minimum) Display :
128 x 64 dots graphical LCD with backlight Physical Dimensions :
140mm(length) x 68mm(width) x 25mm(thick) Net Weight : 130 gram (without batteries) Power : AA Size Battery x 2
ACCESSORIES
SMA connector for OPEN Calibration: P/N 202112 SMA connector for SHORT Calibration:P/N 132331 SMA connector for LOAD Calibration: P/N 132360
SMA male to BNC female adaptor SMA male to SO-239 adaptor USB A to mini-B cable CD-ROM
Loading...