Time Electronics 1071 User Manual

Time Electronics
1071 Decade Capacitance Box
Technical Manual
V1.1 25/01/11 Time Electronics Ltd
Botany Industrial Estate, Tonbridge, Kent, TN9 1RH
Tel: +44(0)1732 355993 Fax: +44(0)1732 770312
Email: mail@TimeElectronics.co.uk
Web Site: www.TimeElectronic s .com
2

Contents

1. S pecification ..................................................................................................... 2
2. Opera tion .......................................................................................................... 3
3. Guarantee & S ervicing ..................................................................................... 4
All Time Electronics' instruments are subject to continuous development and improvement and in
consequence may incorporate minor detail changes from the information contained herein.

1. Specification

RANGE
9 x 10pF
9 x 100pF
9 x 1nF
9 x 10nF
9 x 100nF
9 x 1µF
9 x 10µF
TYPE
Silver mica 1% ± 5pF
Silver mica 1% ± 5pF
Silver mica 1% ± 5pF
Silver mica 1%
Poly-
carbonate
Poly-
carbonate
Polyester 5%
ACCURACY
@ 1kHz
1%
1%
MAX
VOLTAGE
300V dc 200V ac
300V dc 200V ac
300V dc 200V ac
100V dc
72V ac
100V dc
72V ac
100V dc 115V ac
63V dc 50V ac
TEMP
COEFF
ppm/degC
max
200 < 0.0025 > 50G ohms
200 < 0.0015 > 50G ohms
50 < 0.002 > 50G ohms
50 < 0.01 > 30G ohms
75 < 0.01 > 30G ohms
75 < 0.01 > 30G ohms
300 < 0.01 > 30G ohms
POWER
FACTOR
@ 1kHz
INSULATION TO CASE RESISTANCE
Zero Residual Capacitance:
55pF
Operating Temperature: 0 to 70°C Construction: The case is of robust metal construction finished in
two-tone blue and black. A removable protective boot houses the instrument
Dimensions: 215 x 100 x 120mm (incl. protective boot)
200 x 75 x 110mm (without)
Weight: 1kg (incl. protective boot), 0.6kg (without)
V1.1 Page | 2 1071 Technical Manual
3

2. Operation

The 1071 is a portable decade capacitance box, housed in a robust metal case and designed for industrial, laboratory and education requirements.
The unit incorporates a colour coding system which divides the setting into three groups -
µF = Red ηF = White ρF = Yellow
The high accuracy obtained in the 1071 is achieved by the use of accurate, stable capacitors. A thumbwheel switch provides precision selection of output capacitance.
A multiple gold contact switch arrangement ensures that should a contact fail, there is always a back-up contact available.
To select a value simply dial up with the digit switches. For example to set 2.746µF select:
2 on the 1µF bank
7 on the 100ηF bank
4 on the 10ηF bank
6 on the 1ηF bank
Connection to the box can be made via 4mm shrouded plugs, as well as standard plugs, bare wires, and spade terminals and the grey terminal connects to the case for screening purposes.
. Two connections red and black connect to the capacitance
Note: Remember that all capacitance boxes have a residual capacitance, i.e. even when the dials are set to zero there is still a capacitance value across the output. If you are making precision measurements or recalibrating the instrument this residual value must be subtracted from all measurements.
V1.1 Page | 3 1071 Technical Manual
4

3. Guarantee & Servicing

Guarantee Period This unit is guaranteed against defects in materials and workmanship for a period of one year from its delivery to the customer.
We maintain comprehensive after sales facilities and the unit can, if necessary be returned to us for servicing. During this period, Time Electronics Ltd will, at its discretion, repair or replace the defective items. For servicing under guarantee, the instrument type and serial number must always be quoted, together with details of any fault and the service required. The purchaser of the instrument must prepay all shipping charges. Time Electronics Ltd will pay return shipping charges.
This guarantee is void if servicing has been attempted by an unauthorised person or agent. If, during the guarantee period, failure is due to misuse or abuse of the unit, the repair will be put in hand without delay and charged unless other instructions are received.
Please note that if you require a new UKAS Certificate during the warranty period, this will be charged at the current rate on our price list.
Service After Guarantee Period
Even after the guarantee period has expired, Time Electronics Ltd., can still service your instrument. As the manufacturer, we have the specialised knowledge needed to keep your instrument in peak condition and we also maintain a comprehensive spare parts service.
Please enclose details of the service required and your full company details including a contact name when returning for servicing.
Returning Instruments
When returning instruments, please ensure that they have been adequately packed, preferably in the original packing supplied. Time Electronics Ltd will not accept responsibility for units returned damaged. Please ensure that all units have details of the service required and all relevant paperwork.
Send the instrument, shipping charges paid to:-
Time Electronics Ltd
Botany Industrial Estate, Tonbridge, Kent, TN9 1RH
Tel: +44( 0)1732 355993 Fax: +44(0)1732 770312
Email: mail@TimeElectronics.co.uk
Web Site: www.TimeElectronics.com
V1.1 Page | 4 1071 Technical Manual
Disposal of your old equipment
1. When this crossed-out wheeled bin symbol is attached to a product it means the product is covered by the European Directive 2002/96/EC.
2. All electrical and electronic product s s houl d be disposed of separately from the municipal waste stream via designated collection facilities appointed by the government or the local authorities.
3. The correct disposal of your old appliance will help prevent potent i al negative consequences for the environment and human health.
4. For more detailed information about dispos al of your old appliance, please cont act
your city office, waste disposal service or return to Time Electronics.
Loading...