Time Electronics
1061 Resistance Decade Box
Technical Manual
V1.1 12/08/11 Time Electronics Ltd
Botany Industrial Estate, Tonbridge, Kent, TN9 1RH
Tel: +44(0)1732 355993 Fax: +44(0)1732 770312
Email: mail@TimeElectronics.co.uk
Web Site: www.TimeElectronic s .com

2
Contents
1. S pecification ..................................................................................................... 2
2. Opera tion .......................................................................................................... 3
3. Guarantee & S ervicing ..................................................................................... 3
All Time Electronics' instruments are subject to continuous development and improvement and in
consequence may incorporate minor detail changes from the information contained herein.
1. Specification
Resistance Range:
Accuracy:
Number of Decades:
Increment:
Power Rating:
Voltage Rating:
Residual Resistance:
Connections:
1Ω to 1.2 MΩ
± 1%
6
1Ω steps
0.75 watt per resistor
Maximum 300V DC/AC RMS
Less than 150mΩ
4 mm terminals. A third terminal is to enable the
case to be earthed or connected to either output.
Dimensions:
Weight:
Optional Extras:
W355 x H63 x D82mm
0.75kg
Calibration certificates traceable to NPL and
UKAS (ISO 17025).
V1.1 Page | 2 1061 Technical Manual

3
2. Operation
The 1061 Decade Resistance Box is designed to meet the standard required in both
educational and industrial applications. Metal f ilm resistors, with the advantages of stability
and low temperature coefficient are used throughout. The case provides complete
electrostatic screening. Six large switches with a clear numerical readout make the
resistance box easy to use. The slim line design minimises bench space, while the 1%
accuracy and wide range makes it attractive for production or select-on-test uses.
Resistance is selected by dialing t he value required using the rotary switches. This enables
precise setting with a clear unambiguous indication.
For example - to select 3k83:
• set the 1kΩ bank to 3
• the 100Ω bank to 8
• and the 10Ω bank to 3
Connection to the box can be made via 4mm shrouded plugs, as
well as standard plugs, bare wires, and spade terminals
connections red and black connect to the resistance and the
grey terminal connects to the case for screening purposes.
. Two
Note: Remember that all resistance boxes have a residual resistance, i.e. when the dials
are set to zero a small resistance remains. If you are making precision measurements or
recalibrating the instrument, this residual value must be subtracted from all measurements.
V1.1 Page | 3 1061 Technical Manual

4
3. Guarantee & Servicing
Guarantee Period
This unit is guaranteed against defects in materials and workmanship for a period of one
year from its delivery to the customer.
We maintain comprehensive after sales facilities and the unit can, if necessary be returned
to us for servicing. During this period, Time Electronics Ltd will, at its discretion, repair or
replace the defective items. For servicing under guarantee, the instrument type and serial
number must always be quoted, together with details of any fault and the service required.
The purchaser of the instrument must prepay all shipping charges. Time Electronics Ltd will
pay return shipping charges.
This guarantee is void if servicing has been attempted by an unauthorised person or agent.
If, during the guarantee period, failure is due to misuse or abuse of the unit, the repair will
be put in hand without delay and charged unless other instructions are received.
Please note that if you require a new UKAS Certificate during the warranty period, this will
be charged at the current rate on our price list.
Service After Guarantee Period
Even after the guarantee period has expired, Time Electronics Ltd., can still service your
instrument. As the manufacturer, we have the specialised knowledge needed to keep your
instrument in peak condition and we also maintain a comprehensive spare parts service.
Please enclose details of the service required and your full company details including a
contact name when returning for servicing.
Returning Instruments
When returning instruments, please ensure that they have been adequately packed,
preferably in the original packing supplied. Time Electronics Ltd will not accept
responsibility for units returned damaged. Please ensure that all units have details of
the service required and all relevant paperwork.
Send the instrument, shipping charges paid to:-
Time Electronics Ltd
Botany Industrial Estate, Tonbridge, Kent, TN9 1RH
Tel: +44( 0)1732 355993 Fax: +44(0)1732 770312
Email:
Web Site:
mail@TimeElectronics.co.uk
www.TimeElectronics.com
V1.1 Page | 4 1061 Technical Manual
Disposal of your old equipment
1. When this crossed-out wheeled bin symbol is attached to a product it means the
product is covered by the European Directive 2002/96/EC.
2. All electrical and electronic products should be disposed of separately from the
municipal waste stream via designated collection facilities appointed by the
government or the local authorities.
3. The correct disposal of your old appliance will help prevent potential negative
consequences for the environment and human health.
4. For more detailed information about disposal of your old appliance, please contact
your city office, waste disposal service or return to Time Electronics.