TIF 3340 User Manual

Page 1
Professional 4-channel
Thermocouple Thermometer
plus Infrared Thermometer
BUTTONS
= Power ON or OFF.
= Steps through maximum (MAX), minimum (MIN), and average (AVG) readings. To exit MAX / MIN / AVG mode, press this button for three (3) seconds.
= Switch between Celsius (°C), Fahrenheit (°F), and Kelvin (K) temperature units.
TIF3340
INFRARED
THERMOMETER
FEATURES
Large backlit display shows any combination of T1, T2, IR, T3, T4, T1-T2, T1-IR, T2-IR, IR-T3, and IR-T4,
plus MAX, MIN, and AVG.
Relative time clock on MAX, MIN, and AVG
provides a time reference for major events.
Electronic offset function allows compensation
for thermocouple errors.
Readout in °F, °C, or Kelvin (K).
Auto shutdown.
= Freeze and unfreeze display readings.
= Displays T1, T2, IR, T1-T2, T1-IR, and T2-IR (differential temperature measure­ ment) in the primary, secondary, and tertiary displays. Press button for two seconds to show T3, T4, IR, T3-T4, IR-T3, and IR-T4. Press button for two more seconds to exit.
= Press once to turn on backlight; press again to turn off backlight. Hold for three (3) seconds to access or exit Setup menu.
= Scrolls to the setup option you want to change; press again to increase the displayed setting.
= Press once to record; press again to stop
recording. In Setup menu, scroll to the setup option you want to change; press again to decrease the displayed setting.
= Press once to enter a setup option; press again to store the displayed setting in memory.
Sheet 1
Page 2
OPERATING INSTRUCTIONS
Displaying Temperatures
1. Plug a thermocouple(s) into the T1, T2, T3, or T4 input connector located at the top of the thermometer.
2. Turn the thermometer ON by pressing .
3. Press to select the preferred temperature scale (Celsius, Fahrenheit, or Kelvin).
4. Hold or attach the thermocouple(s) to the location requiring a temperature measurement. The temperature reading appears in the primary display.
5. Press to toggle the T1, T2, T3, T1-T2,T1-T3, and T2-T3 readings to show in the primary or secondary displays.
Notes:
“ - - - -” is displayed when a thermocouple is defective or not connected.
 OL(overload)isdisplayedwhenthetemperaturebeingmeasuredisoutsidethethermometer’srange.
Display Elements
1. Primary display: T1, T2, IR, T3, T4,
T1-T2, T1-IR, T2-IR, T3-T4, IR-T3, or IR-T4 reading.
2. Tertiary display: The elapsed time or IR
temperature reading.
3. MIN:SEC or HOUR:MIN display.
4. Fourth display: T1 or T3 reading.
5. Thermocouple type.
6. IR temperature.
7. Displays while Setup menu is activated.
8. Setup menu; offset of T1, T2, IR, T3, T4.
9. Auto power OFF.
10. Low power.
11. Fifth display: T2 or T4 reading.
12. MAX, MIN, AVG display.
13. Secondary display: T1, T2, T3, or T4
reading.
14 Unit of temperature.
15. Displays while recording data.
1516
T1-T2-IR-T3-T4 HOLD REC
1
T1-IR
IR-T4
-888.8
T1-T2-T3-T4 T1-T2-T3-T4
IR IR
2
K
F
C
14
13
-8888 -8888
3
4
5
hour : min : sec MAX MIN AVG
T1 T2
-8888 -8888
T3 T4
K IR TEMP SETUP OFFSET
12
11
16. Displays while freezing data.
Back, Sheet 1
6 7 8 9 10
Page 3
OPERATING INSTRUCTIONS contd.
Holding Displayed Readings
1. Press to freeze the displayed readings. The display shows HOLD.
2. Press to toggle the T1, T2, IR, T1-T2, T1-IR, and T2-IR readings in the primary, secondary, or tertiary display.
3. Press for two seconds to toggle the T3, T4, IR, T3-T4, IR-T3, and IR-T4 readings in the primary, secondary, or tertiary display.
4. Press for two seconds again to return to the original display.
5. Press HOLD to turn off the HOLD option.
Viewing MIN, MAX, and AVG Readings
Using the Infrared (IR) Thermometer
CAUTION: To prevent personal injury, do not stare into the laser beam. This is a Class 2 laser product.
1. Plug the output of the IR thermometer into the IR input on the side of the TIF3340. IR TEMP appears on the display.
2. Aim the IR light at the location to measure.
3. Press . The temperature appears in the tertiary display.
Shutting OFF the TIF3340
1. Press and hold the button while the secondary display counts down from three to zero. The thermometer shuts off after the display reaches zero.
1. Press to step through maximum (MAX), minimum (MIN), or average (AVG) readings. The display shows elapsed time since entering MAX / MIN / AVG mode, or the time at which the minimum or maximum occurred.
2. Press for three (3) seconds to exit.
Recording Data
1. When the meter is turned ON, the number of events that may be recorded is shown on the secondary display.
Note: If the maximum number of events have been recorded, the meter displays FULL. Refer to the Setup Menu: Erase the Memory procedure to clear the events.
1 0000
NOTE: If the sleep timer (SLP) is enabled, the thermometer automatically shuts down after twenty (20) minutes of inactivity.
2. Press to record an event. Press the button again to stop recording.
Sheet 2
Page 4
SETUP MENU
CHANGING SETUP OPTIONS
1. Press for three (3) seconds to enter Setup Menu. SETUP appears at the bottom of the display.
2. Press or to scroll to the desired setup option.
3. Press ENTER to change this setting.
4. Press or
5. Press ENTER to store new setting in memory.
6. Press SETUP for three (3) seconds to exit Setup Menu.
NOTE: Setup is disabled in MAX MIN AVG mode.
Option Menu Item Settings
Offset T1, T2, IR, T3, T4 T1, T2, IR, T3, or T4 offset
Sample Rate rAt Recording sample rate
Time Setting tiE Show / set the time
Clear CLR Clear memory of events
Sleep Mode SLP ON or OFF
until desired setting displays.
Set the Sample Rate of Temperature Data Recording
000 1
The default sample rate is one second, which is displayed on the secondary display as 0001 (minute–second).
1. While in Setup Menu, press to select which reading (minute—second) to adjust.
2. Press or
3. Press ENTER to store the new setting in memory.
until desired setting displays.
Set the Time
The secondary display shows year, month, and day. The tertiary display shows hour and minute. June 10, 2009, 1:31 PM is displayed in the graphic.
1. While in Setup Menu, press to select which reading to adjust: year—month—day— hour—minute.
K SETUP
2009 06 10 1331
K SETUP
Using Offset to Adjust for Probe Errors
The offset option can adjust thermometer readings to compensate for the errors of a
specic thermocouple and IR temperature. The
adjustment range allowed is: ± 9.0°F or ± 5.0°C.
1. While in Setup Menu, plug the thermocouple into the input connector.
2. Place the thermocouple in a known, stable temperature environment (such as an ice dry well calibrator). Allow the readings to stabilize.
4. Press or until the primary reading matches the calibration temperature.
3. Press ENTER to store new setting in memory.
Note: The primary display shows the temperature plus the offset, and the secondary display shows the offset. Individual offsets can be stored for T1, T2, IR, T3, or T4.
2. Press or
3. Press ENTER to store the new setting in
memory.
until desired setting displays.
Erase the Memory
The number of events stored is shown on the secondary display. This procedure outlines how to clear the memory.
1. While in Setup Menu, press or until CLR is displayed.
2. Press ENTER.
3. Press or until YES is displayed.
4. Press ENTER to clear the memory.
K SETUP
8888
Back, Sheet 2
Page 5
SETUP MENU contd.
MAINTENANCE
SLEEP MODE
When in sleep mode, the meter automatically shuts down after 20 minutes if no buttons have been pressed during that time.
is displayed if sleep mode is enabled.
1. While in Setup Menu, press or until SLP is displayed.
2. Press ENTER until ON or OFF is displayed.
3. Press or until desired setting is displayed.
4. Press ENTER to store the setting in memory.
Note: If sleep mode is not enabled, the meter does not automatically shut OFF, and will not show on the display.
K SETUP
REPLACING THE BATTERY
The thermometer must be OFF when changing the battery.
1. Press to power OFF the thermometer.
2. On the back side of the thermometer, locate the symbol.
3. While holding the thermometer, place your thumb on the . Using gentle pressure, push downward and pull backward; the battery cover will begin to open. Use your other hand to grasp the cover and pull downward. Lift the cover up and off.
4. Lift out the 9-volt battery and disconnect it from the harness.
5. Connect the new 9-volt battery to the harness, and place it into the battery cavity.
6. Install the battery cover by placing it over the battery and pushing upward until it snaps into place.
SPECIFICATIONS
General
Battery 9V
Operating Temperature 32°F to 122°F (0°C to 50°C) Noncondensing
Storage Temperature 14°F to 122°F
-10°C to 50°C
Temperature Range K-type Thermocouples -328°F to 2501°F
-200°C to 1372°C
IR Temperature -22°F to 1022°F
-30°C to 550°C
Display Resolution
0.1(°C•°F•K)<1000 1(°C•°F•K)>1000
7. Dispose of the old battery according to local, state, and federal regulations.
Temperature Accuracy
T1, T2, T3, T4
Above -148°F (-100°C) ± [0.15% rdg + 1.8°F (1°C)]
T1, T2, T3, T4 Below -148°F (-100°C) ± [0.5% rdg + 3.6°F (2°C)]
T1-T2, T3-T4 ± [0.5% rdg + 1.8°F (1°C)]
IR IR above 14°F (-10°C) ± [2.0% rdg + 3.6°F (2°C)] IR below 14°F (-10°C) ± 9°F (5°C)
T1-IR & T2-IR IR above 14°F (-10°C) ± [2.0% rdg + 5.4°F (3°C)] IR below 14°F (-10°C) ± 11°F (6°C)
T3-IR & T4-IR IR above 14°F (-10°C) ± [2.0% rdg + 5.4°F (3°C)] IR below 14°F (-10°C) ± 11°F (6°C)
Sheet 3
Page 6
WARRANTY
This product has been produced to provide unlimited service. Should it become inoperable after the user has performed the recommended maintenance, a no-charge repair or replacement will be made to the original purchaser. This applies to all repairable units that have not been damaged or tampered with. The claim must be made within One Year of the date of purchase.
655 Eisenhower Drive Owatonna, MN 55060 Toll Free: (800) 327-5060 Fax: (866) 287-7222
www.TIF.com
Back, Sheet 3 Form No. 554118 Rev. B September 13, 2013 © Bosch Automotive Service Solutions LLC
Loading...