Thunder Scientific 5A-1MP Owner's Operation And Maintenance Manual

THUNDER SCIENTIFIC CORPORATION
MODEL 5A-1MP
ELECTRONIC PSYCHROMETER
MEASUREMENT SYSTEM
OPERATION AND
MAINTENANCE MANUAL
THUNDER SCIENTIFIC CORPORATION
ALBUQUERQUE, NEW MEXICO 87123-3198
623 WYOMING BLVD. S.E.
ELECTRONIC PSYCHROMETER
MODEL 5A-1MP
RELATIVE HUMIDITY AND TEMPERATURE
MEASUREMENT SYSTEM
OPERATION AND
MAINTENANCE MANUAL
ALBUQUERQUE, NEW MEXICO 87123-3198
THUNDER SCIENTIFIC CORPORATION
623 WYOMING BLVD. SE
WWW.THUNDERSCIENTIFIC.COM
5A-1MP Laboratory Reference Psychrometer
TABLE OF CONTENTS
Section 1 - GENERAL INFORMATION
Warranty
1.1 GENERAL DESCRIPTION.........................................................................1-1
1.2 SPECIFICATIONS......................................................................................1-2
1.3 FEATURES.................................................................................................1-2
Section 2 - INSTALLATION
2.1 GENERAL..................................................................................................2-1
2.2 ELECTRICAL.............................................................................................2-1
2.3 COMPUTER/PRINTER INTERFACING......................................................2-1
2.4 IEEE 488 INTERFACING (OPTIONAL)....................................................2-3
Section 3 - OPERATION
3.1 GENERAL..................................................................................................3-1
3.2 STANDARD OPERATING PROCEDURES.................................................3-1
3.2.1 Power-up .............................................................................................3-1
3.2.2 Probe Preparation ................................................................................3-2
3.2.3 System Operation.................................................................................3-2
3.2.4 Probe # Selection.................................................................................3-3
3.2.5 Entering the Pressure ...........................................................................3-3
3.2.6 Entering the Flow Rate.........................................................................3-4
3.2.7 Enabling/Disabling the Blower .............................................................3-5
3.2.8 Enabling/Disabling the Printer..............................................................3-5
3.2.9 Displaying Desired Readings................................................................3-6
3.2.10 Changing the Time ..............................................................................3-7
3.2.11 Changing the Date ...............................................................................3-7
3.2.12 Changing the Time Interval..................................................................3-7
3.2.13 Changing the Amount of Averaging ....................................................3-8
3.2.14 Changing the Printer Options...............................................................3-8
3.2.15 Switching Between °C and °F................................................................3-9
3.2.16 Enabling the AUTO Mode.................................................................3-10
Section 4 - MAINTENANCE
4.1 PROBE WICK CHANGEOUT INSTRUCTIONS..........................................4-1
4.2 FACTORY CERTIFICATION AND RECERTIFICATION...........................4-2
4.3 NIST CERTIFICATION AND RECERTIFICATION ...................................4-3
Section 5 - RS-232 INTERFACE
5.1 GENERAL..................................................................................................5-1
5.2 RS-232 PROTOCOL ...................................................................................5-1
5.3 INITIALIZING THE HOST TERMINAL OR COMPUTER.........................5-1
5.4 CLEARING THE 5A-1MP I/O BUFFERS....................................................5-2
Section 6 - IEEE 488 INTERFACE
6.1 GENERAL..................................................................................................6-1
6.2 IEEE 488 ADDRESSING............................................................................6-1
6.3 CLEARING THE 5A-1MP BUFFERS .........................................................6-1
Section 7 - IEEE 488/RS-232 COMMAND REFERENCE
7.1 GENERAL..................................................................................................7-1
Section 8 - CABLING SPECIFICATIONS AND PINOUTS
8.1 DB25 PRINTER AND RS-232 CONNECTOR.............................................8-1
8.2 IEEE 488 CONNECTOR.............................................................................8-2
WARRANTY
Thunder Scientific Corporation warrants each instrument of its manufacture to be free from defect in materials and workmanship.
The limit of liability under this warranty is to repair or replace any part thereof which shall within one year after delivery to the original purchaser, and returned by the original purchaser, prove to be defective after examination by Thunder Scientific Corporation.
This warranty specifically excludes the probe sensors, which may become damaged or broken due to wick changes. Shipping damage is covered by insurance and will be honored if report is made within one week of receipt of the instrument. Warranty returns must be authorized by the factory.
Section 1
GENERAL INFORMATION
1.1 GENERAL DESCRIPTION
The Thunder Scientific Model 5A-1MP is a microprocessor based Laboratory Standard Psychrometer designed for precise measurement of air temperature and relative humidity.
The 5A-1MP uses a 2-bulb measurement system employing Wet and Dry calibrated Platinum Resistance Thermometers (PRT's) operating under a controlled air velocity. One PRT, referred to as the Dry Bulb, is used for sensing air temperature. The other PRT, the Wet Bulb, uses a specially woven wick (or sock) to facilitate an unimpeded capillary flow of distilled or high purity reagent water from a reservoir below. The Wet Bulb measures a depression temperature.
The signals from the PRT's are converted to digital format, read by the microprocessor, and mathematically corrected to temperature data. The data is then used to calculate Percent Relative Humidity and Dewpoint Temperature.
Front panel display of any of the above data is selected by user input at the keyboard. The upper display is the data readout device, and the lower display indicates which data or other function is currently enabled. Upon selection of data output to this real-time LED display, current readings are updated four to five times each second allowing for fast system response to any changes in the sampled environment.
Printer output, enabled from the keyboard, allows for hardcopy printouts of all current psychrometric data. Printer output is at user selected time intervals ranging from a one minute minimum to 24 hour maximum, or at any one minute increment in between.
An Auto Mode feature is also included which, when enabled, will put the probe in synchronous operation with the printer. In this mode, the probe aspirator will automatically run prior to any sample for hardcopy output. After printout, the aspirator will be disabled and will remain off until the next sample time is reached.
Bi-directional RS-232 and optional IEEE 488 communication capabilities allow for remote operation and data retrieval when connected to a computer or terminal. All functions available from the front panel keypad are also available over the RS-232 and IEEE 488 links through a series of more than 40 commands.
The Model 5A-1MP is a precision laboratory grade instrument, and when operated and handled accordingly, will provide years of excellent service.
1-1
1.2 SPECIFICATIONS
Relative Humidity (RH)
Measurement Range ....................................................................................1% to 100% RH
Resolution 0.01% RH
Guaranteed Accuracy ..............................................................................................±1% RH
Typical Accuracy ..................................................................................................±0.5% RH
Temperature
Measurement Range ...............................................................................0-100°C / 32-212°F
Resolution 0.01°C and °F
Guaranteed Accuracy .................................................................................. ±0.3°C / ±0.5°F
Typical Accuracy ......................................................................................... ±0.1°C / ±0.2°F
Data Interface
Printer Interface ........................................................................................Centronics Parallel
Computer Interface (standard)..............................................................Bi-directional RS-232
Computer Interface (optional) .................................................................................IEEE 488
Power Requirements ...............................................................................115 VAC, 50/60 Hz
(OPT) 220 VAC, 50/60 Hz
1.3 FEATURES
1) Z-80 Central Processing Unit, automatically computes percent relative humidity, temperature, and dewpoint in °C or °F.
2) Bi-directional RS-232 and IEEE 488 computer interfaces allow for remote operation/data collection.
3) Battery backed-up real time clock.
4) Centronics printer interface with user selectable input intervals from one minute to 24 hours in one minute increments.
5) AUTO mode capability enables probe aspirator for one minute prior to each printout.
6) Non susceptible to contamination by chemicals or long term exposure to high relative humidity and temperature.
7) Certification of Traceability to the National Institute of Standards and Technology.
8) Direct NIST certification available.
1-2
Section 2
INSTALLATION
2.1 GENERAL
The 5A-1MP is designed as a portable test instrument requiring no special physical installation. The instrument must, however, be connected to a source of AC power (Section
2.2) and may be connected to a printer and/or computer (Sections 2.2, 2.3, 2.4) for remote data collection and coefficient manipulation (also see Figures 2-1, 2-2, and 2-3).
2.2 ELECTRICAL
Using the 6' power cord provided, connect the 5A-1MP to a 115 VAC, 50/60 Hz, grounded outlet.
2.3 COMPUTER/PRINTER INTERFACING
The RS-232 computer interface shares a connector with the parallel printer interface (refer to Section 8.2 for the specific pinouts). Since the computer and printer interfaces are sharing the same connector, only the cables supplies by the factory should be used. When replacing or repairing factory supplies cables, the wiring connections described in Section 8.1 must be followed.
If a computer is to be connected via the RS-232 interface (but no printer is to be connected), use the small round interface cable with the hooded DB25 connectors at each end. Plug the cable end marked "5A-1MP" into the DB25 connector on the psychrometer. Plug the other end into an RS-232 port of a computer or terminal. Depending upon the connections at the computer terminal, a gender changer, null modem adapter, or other interface cables/connectors may be required.
Communicating with the 5A-1MP via the RS-232 interface is discussed in Section 5.
RS-232 Terminal or Computer
Small, Round Serial Cable
Figure 2-1
Computer/Terminal Connection
2-1
If a printer is to be connected, use the flat ribbon type cable which contains a Centronics type 36 pin ribbon-D connector on one end, and two DB25 connectors on the other end. Plug the 36 pin Centronics connector into the mating connector of the printer. Plug one of the DB25 connectors into the mating connector of the psychrometer. Note that only one of the DB25 connectors is of the proper gender to mate with the psychrometer.
Unused Connector
Flat Ribbon Cable
Figure 2-2
Printer Connection
If both the printer and computer are to be connected to the psychrometer, connect the printer first by following the procedures previously outlined (see Figure 2-2). Then plug the RS-232 cable into the unused connector on the printer cable. This connection scheme allows both the printer and computer to communicate with the 5A-1MP simultaneously.
Small, Round Serial Cable
Flat Ribbon Cable
RS-232 Terminal or Computer
Figure 2-3
Computer/Terminal and Printer Connections
2-2
2.4 IEEE 488 INTERFACING (OPTIONAL)
The 5A-1MP may be connected to an IEEE 488 bus via its 24 pin D-type IEEE 488 connector located on the top panel above the computer/printer interface connector (see Figure 2-4). IEEE 488 communication with the 5A-1MP is discussed in Section 8. See Section 8.1 for wiring specifications.
IEEE 488 Cable
Flat Ribbon Cable
Figure 2-4
IEEE 488 Connection
IEEE 488 Computer
2-3
Section 3
OPERATION
3.1 GENERAL
The 5A-1MP is capable of operation in either manual or automatic mode, invoked either from the keypad, the serial interface, or the IEEE 488 interface. In general, operation consists of filling the probe reservoir with pure distilled water, turning the 5A-1MP ON, entering the current barometric pressure, and enabling the blower (or aspirator) motor. After taking data, the blower motor should then be turned off until another sample is required.
3.2 STANDARD OPERATING PROCEDURES
3.2.1 Power-up
Ensure that installation is performed in accordance with Section 2. Press the power switch ON. The LED Displays should indicate the time of day on the upper panel, and P--7 (meaning parameter 7, the time) on the lower panel. If for some reason the power-up mode appears incorrect, either cycle the power off and on again or execute a cold reset. A cold reset is executed by pressing the reset key while holding down the decimal point key. This procedure performs exactly the same function as cycling the power off and on. All values, functions, and modes of the system are restored to their power-up default status.
NOTE 1: Due to the use of CMOS circuitry within the 5A-1MP, peripherals such
as a printer may be supplying enough power to the 5A-1MP to prevent it from fully cycling down during a power-off sequence. This failure to fully cycle down may cause an invalid power-up sequence. To prevent this problem, one or both of the following should be performed when peripheral devices are connected to the 5A­1MP:
A) Power down peripherals when powering down the 5A-1MP. When powering up, power up the 5A-1MP first, then power up the peripherals.
B) Hold down the decimal point key on the keypad while either powering up or pressing RESET. This forces the 5A-1MP through the proper power up initialization sequence regardless of any previous failure to fully cycle down.
3-1
3.2.2 Probe Preparation
The black probe housing contains two Platinum Resistance Thermometers (PRT's) which are used for measuring the air temperature (called the Dry Bulb temperature) and the depression temperature (called the Wet Bulb temperature). The Wet Bulb sensor is the one which is covered by the woven cotton wick. During operation, the wick remains wet by drawing water (through capillary action) from the reservoir in the bottom of the probe housing. The reservoir must be filled prior to operation, and drained after operation allowing the wick to dry. (The probe should never be stored with a wet wick, as a stagnant wet wick provides an excellent media for the growth of micro-organisms (bacteria), thus altering the accuracy of the instrument).
Fill the reservoir with pure distilled water by removing the filler plug (the larger of the two screws) on the bottom plate of the probe housing. A dime usually works well for this purpose. Using the bottle and dropper provided, fill the reservoir to the bottom of the filling hole, while holding the probe so that the round white teflon shaft is vertical, and the motor end is downward. Replace the filler plug and tighten. Squirt several droppers full of water directly on the wick covering the Wet Bulb sensor. This flooding procedure is used to clean the wick and flush away possible contaminants. Dry the probe by gently shaking out excess water from within the teflon shaft portion of the probe. Avoid tapping or banging the probe. Cotton swab sticks (Q-tips) may be needed to thoroughly dry the inside of the white teflon probe shaft. Also dry the outside with a clean cloth while watching for leaks around the bottom of the probe at the edges of the reservoir cover. The reservoir cover is sealed by a rubber gasket and secured by the small screw in the center. Tighten this center screw as necessary.
Place the probe in the environment to be tested and connect the white power signal cable. If multiple probes were calibrated with this 5A-1MP, ensure that the probe number and cable number match, as they were calibrated from the factory as a set. Connect the other end of the cable to the mating connector on the 5A-1MP.
3.2.3 System Operation
In order to measure relative humidity and other psychrometric data in the test environment, the user must
1) Prepare the probe. (Refer to Section 3.2.2).
2) Enter the probe number. (Refer to Section 3.2.4).
3) Enter the current pressure. (Refer to Section 3.2.5).
4) Aspirate the probe. (Refer to Section 3.2.7).
5) Wait approximately 1 minute for stabilization.
6) Display and/or print data. (Refer to Sections 3.2.8, ).
7) Stop probe aspiration. (Refer to Section 3.2.7).
If the psychrometer is being operated within a humidity generator (such as the Thunder Scientific Model 8500 Automated Two-Pressure Humidity Generator), fittings are available in order to aspirate the probe from the humidity generator rather than the blower motor. When using external sources such as this for aspiration, the flow rate of the source must be known and entered into the 5A-1MP. (Refer to Section 3.2.6, Entering the Flow Rate).
3-2
3.2.4 Probe # Selection
If the 5A-1MP has been supplied with more than one probe, the probe number which is currently connected to the system must be entered so that the appropriate calibration coefficients will be used in all mathematical calculations of Dry Bulb and Wet Bulb temperature. Failing to enter the probe number of the currently connected probe will result in erroneous data. Greatest accuracy will also be achieved when using a numbered probe cable which matches the number on the probe, since the probe and cable are factory calibrated as a set.
Enter the probe number by pressing
0
PROBE
SET
n
ENTER
where n is a number between 1 and 4 corresponding to the number on the connected cable and probe. The power-up default is always Probe 1. The probe number may also be changed from either the IEEE 488 or the RS-232 interface. (Refer to PROBe command, Section 7).
3.2.5 Entering the Pressure
For proper RH calculations, the absolute pressure of the test environment must be known and entered (in inches of mercury) into the 5A-1MP. The actual pressure may be measured using an absolute pressure transducer available from a number of different manufacturers. An alternate method is to obtain the approximate barometric pressure from the nearest airport or weather station. Be sure to obtain the actual pressure before any altitude adjustments have been applied. Since the pressure must be entered into the 5A-1MP in inches of mercury, some common conversions follow.
in Hg = PSIA * 2.03602
= Pa * 0.00014503774 = kPa * 0.14503774 = atm * 29.921 = mm Hg/25.4 = in H2O/13.6086
= mm H2O * 1.8665 = Torr/25.4
= Bar * 29.53 = mb/33.8638
3-3
Convert the pressure to inches of mercury if required and enter it in the following manner; however, use the appropriate pressure. For example, to enter a pressure of 30.12 inches of mercury, press
PRESS.
LED's display RH on top, P--1 on bottom.
SET
Enables SET mode. Displays Pressure setting (factory default 29.92) (Note Press. and Set are same color).
3
30.12 Displays 9.923
0
30.12. Displays 9230
.
30.12. Displays 9230.
1
30.12. Displays 230.1
2
30.12. Displays 30.12
ENTER
Ends SET mode. Display reverts to previous reading.
This new pressure value only remains valid until changed again from the keypad, serial interface or IEEE 488 interface, or until the unit is cold reset as in Section 3.2.1. Note that the decimal point is required but that it will automatically trail the second digit entered unless the decimal point key is pressed during the entry sequence. To change the pressure from either the RS-232 or IEEE 488 interface, refer to the PRESsure command, Section 7. To save the currently entered pressure as the power up default pressure, use the SAVE command, Section 7.
3.2.6 Entering the Flow Rate
When aspirating the probe from an external source such as an 8500 Automated Two­Pressure Relative Humidity Generator, the system mass flow rate may be entered into the psychrometer. As an example, a mass flow rate of 100 SLPM would be entered by pressing
3-4
FLOW
SET
1
0
0
.
0
ENTER
This number only remains valid until changed again from either the keypad, the IEEE 488, or the serial interface, or a cold reset is performed. (Refer to Section 3.2.1, Power-Up). Also, anytime the unit is aspirated by enabling the internal blower motor, the flow rate entered is ignored and all calculations are performed utilizing a preset flow rate determined during factory calibration. If probe aspiration is always accomplished with the internal blower, the flow rate need not be altered from its power-up default value. The flow rate may also be set from either the RS-232 or IEEE 488 interface. (Refer to FLOW command, Section 7).
3.2.7 Enabling/Disabling the Blower
BLWR
ON
Pressing the
OFF
key enables and disables the blower motor within the probe. When enabled, the leftmost dot of the lower LED display will illuminate. The blower should be enabled for at least 30 seconds to 1 minute prior to taking readings in order to allow the system to stabilize. After the desired readings have been taken, disable the blower by pressing the same key again. The blower may also be controlled from either the RS-232 or IEEE 488 interface. (Refer to BLOWer command, Section 7). Note that there is a normal 1 to 2 second delay associated with blower turnoff.
3.2.8 Enabling/Disabling the Printer
If a printer is connected to the psychrometer. (Refer to Section 2.3, Computer/Printer Interfacing), current printouts of all relevant psychrometric data is available.
To enable the printer,
PRINT
ON
press the
OFF
key on the keypad. The second dot from the left on the lower display
will illuminate, and data will then be printed at the specified time intervals. (Refer to Section
3.2.12, Changing the Time Interval). Pressing the key again disables the printer.
PRINT NOW
The
key may be used at any time to immediately send one line of data to the
printer.
The start of each page is preceded with a header and the end of each page is appended with a formfeed (or top of form), provided that headers and formfeeds are enabled. (Refer to Section 3.2.13, Printer Options). The printer may also be controlled from either the RS-232 or IEEE 488 interface. (Refer to PRINter command, Section 7).
3-5
3.2.9 Displaying Desired Readings
With the 5A-1MP in the normal parameter display mode, indicated by a "P" on the leftmost digit of the lower LED display, select the desired display parameter by pressing any of the numeric keys. The keys and associated parameters are listed below:
Key Display Parameter
0
PROBE
Probe #
1
RH
RH
2
DP
Dewpoint
3
DRY TMP
Dry Bulb Temperature
4
WET TMP
5
DRY
6
WET
7
TIME
8
DATE
9
INTRVL
Wet Bulb Temperature
Dry Bulb Resistance
Wet Bulb Resistance
Time of Day
Date
Time Interval between printouts
Notice that the display parameters relate to the black text on the bottom of the keys.
3-6
SET
The blue text at the top of some of the keys indicates that when using the
key, it is
the parameter in blue which will be changed. (Refer to Sections 3.2.4-3.2.6, 3.2.10-3.2.14 for use of the SET key). Those keys without blue text have identical SET and DISPLAY parameters. (Exceptions are the Dry Ω and Wet Ω which have no corresponding SET parameters).
3.2.10 Changing the Time
7
TIME
h
key on the keypad. To change the time, press
m
m
ENTER
Display the time by pressing the
TIME
SET
h
where hh and mm are the hours and minutes desired. Time is entered in 24 hour military format. For example, the time 6:15 p.m. must be entered as 1815. The time may also be set from either the RS-232 or IEEE 488 interface. (Refer to TIME command, Section 7).
3.2.11 Changing the Date
8
Display the date by pressing the
DATE
key on the keypad.
Change the date by pressing
DATE
SET
m
m
d
d
ENTER
where mm and dd are the month and day respectively. For example, December 25 is entered as 1225. The date may also be set from either the RS-232 or IEEE 488 interface. (Refer to DATE command, Section 7).
3.2.12 Changing the Time Interval
9
Display the printout time interval by pressing the
INTRVL
key on the keypad. The
numbers shown on the upper display represent the hours and minutes between printed
3-7
readings whenever the printer is enabled (note that the printer is automatically enabled in AUTO mode).
To change the interval from its 1 minute default, press
INTRV L
where hh and mm represent the desired hours and minutes between printouts. Allowable intervals range from a minimum of 0001 (1 minute) to a maximum of 2359 (23 hours, 59 minutes). The interval may also be set from either the RS-232 or IEEE 488 interface. (Refer to INTErval command, Section 7).
3.2.13 Changing the Amount of Averaging
On power-up, the averaging amount is set to 0 and no internal averaging takes place. While averaging is not normally required for stable readings, its use may prove beneficial in unstable and/or electrically noisy environments. The 5A-1MP uses a weighted sum averaging technique where by the currently displayed value is multiplied by n, added to the new value, then the total is divided by n + 1. To change the averaging from its 0 default value, press
AVERAGE
where nnnn is a number between 0000 and 9999 inclusive, with reasonable values between 0 and 100.0 inclusive. Note that fractional numbers are also valid. The averaging amount remains valid until changed, or until a cold reset is performed. (Refer to Section 3.2.1, Power-Up). Averaging may also be changed from either the RS-232 or IEEE 488 interface. (Refer to AVERage command, Section 7).
SET
SET
h
n
h
n
m
n
m
n
ENTER
ENTER
3.2.14 Changing the Printer Options
Printer options available from the keypad include
A) Enabling/disabling formfeeds. B) Enabling/disabling the header. C) Setting the end of line terminator to
1) Carriage return only.
2) Line feed only.
3) Carriage return, line feed.
4) Line feed, carriage return.
D) Setting the number of lines per page.
3-8
The factory default printer options are set at and broken down as follows:
0360
ABCC
A =0for a carriage return only at the end of each line (factory default).
1 for a line feed only. 2 for a carriage return then line feed. 3 for a line feed then carriage return.
B =0for no header and no formfeeds.
1 for header, but no formfeeds. 2 for formfeeds, but no header. 3 for header and formfeeds (factory default).
CC = 00-99 lines per page. (factory default=60).
To change the printer options, first determine the appropriate 4 digit number ABCC from the previous descriptions, then enter the number by pressing
RT O PT
If the printer was enabled, or if the
SET
A
B
PRINT NOW
C
C
ENTER
key had been used prior to setting the printer
options, the printer may need to be disabled then re-enabled to make the new printer options become valid (primarily when changing the lines per page). The options entered remain valid until changed again or a cold reset is performed. (Refer to Section 3.2.1, Power-Up). Printer options may also be set from either the RS-232 or IEEE 488 interface. (Refer to FORMfeed, HEADer, ENDLine, and LINEs commands, Section 7). Current printer options may also be stored as power-up defaults through the use of the SAVE command, Section 7.
3.2.15 Switching Between °C and °F
C / F
The 5A-1MP powers up in °F mode. To switch to °C, press the
key. The
rightmost dot on the lower LED display will illuminate. Return to °F with the same key. This key affects the display and printer temperature scales only, and does not affect the temperature scale of the serial and IEEE 488 interfaces which are always °C. The temperature scale is also selectable through either the RS-232 or IEEE 488 interface. (Refer to DEGRees command, Section 7).
3-9
3.2.16 Enabling the AUTO mode
AUTO
ON
Enable the AUTO mode by pressing the
OFF
key. The third dot from the left on the
lower display will illuminate indicating the AUTO mode is now enabled. The AUTO mode also enables the printer (illuminates the second dot) and controls the operation of the blower motor (the first dot). In AUTO mode operation, the 5A-1MP will automatically enable the blower exactly 60 seconds prior to each printout of data which occurs at the specified time interval. (Refer to Section 3.2.12, Changing the Time Interval). The blower will then be disabled immediately following each print, and will remain off until 60 seconds prior to the next print time. In this way, the blower remains on for only 1 minute during each sampling time interval, greatly extending the service of the blower motor which has a 300 hour nominal run life. Note that if the interval is set to 1 minute, the blower will remain on continuously.
NOTES:
1) While in AUTO mode, the printer remains enabled and cannot be disabled by the user.
PRINT
ON
The
OFF
key is therefore locked out and has no effect.
2) Once within the 60 seconds prior to each print time when the blower is enabled, it may not be disabled by the user.
BLWR
ON
The
OFF
key is therefore locked out and has no effect during this 60 second
time interval.
AUTO
ON
The AUTO mode is disabled by pressing the
OFF
key again. Disabling AUTO mode
also disables the printer and blower motor, restoring them back to user control from the keypad. The AUTO mode may also be controlled from either the RS-232 or IEEE 488 interface. (Refer to AUTO command, Section 7).
3-10
Section 4
MAINTENANCE
4.1 PROBE WICK CHANGEOUT INSTRUCTIONS
The CK-1 Wick Changeout Kit is provided to easily change wicks in the sensing probe. Wick changes should be performed after approximately 40 hours of operation, or more often, particularly if the instrument is in a dirty environment. A replacement wick supply should be kept on hand and can be ordered from the factory in advance of the need. Consult the factory for current price information in the event a replacement supply is needed.
To change the wick, perform the following:
1) Place probe upside down upon two blocks of wood or other material, one on each side of the probe handle.
2) Remove reservoir plug, cover screw, and rubber feet.
3) Gently pry reservoir cover with plastic tool and remove.
4) Carefully remove the four housing retainer screws located in the corners of the probe; they join the two halves of the probe.
5) Gently remove bottom half of probe housing, feeding the wick through the vertical teflon wick tube. Keep the large horizontal teflon air sleeve in place during removal.
Note - Blower motor is also loose once housing is removed.
6) Lift teflon air sleeve up and off the sensor mounts taking care to feed the wick through the air sleeve avoiding any lateral movement and undue strain upon the sensor.
7) The sensor and wick are now in the open where the wick may be removed. Using the dropper provided, apply several drops of ethyl alcohol to the portion of the wick covering the sensor. Allow to soak for approximately one minute, replenishing the alcohol as evaporation occurs.
8) Using one of the picks provided, gently probe the wick covering the sensor to loosen.
9) Using the curved tweezers provided, very gently push the wick up on itself or up on the sensor, assisting with the other hand from the reservoir side. (The wick construction is very similar to a "Chinese finger puzzle" in that pushing tends to release it where pulling only tightens the wick and compounds the problem of removal.) The wick should expand. When this occurs, use the tweezers to slide the wick from the sensor.
4-1
CAUT ION!
IF WICK IS FROZEN TO THE SENSOR WITH DEPOSITS DUE TO CONTAMINATED WATER, DO NOT
FORCE. THIS WILL BREAK THE SENSOR LEADS, CAUSING EXPENSIVE REPAIR COSTS. IF ALCOHOL DOES NOT LOOSEN THE DEPOSIT, RETURN TO THUNDER SCIENTIFIC FOR REMOVAL. THUNDER SCIENTIFIC IS NOT RESPONSIBLE FOR LEAD BREAKAGE AND/OR SENSOR REPLACEMENT IF SUCH OCCURS.
If a wick changeout schedule is maintained and the water is drained from the probe after each use this problem should not occur.
10) Apply provided finger cots to the thumb and index finger of each hand. Soak a clean cloth with alcohol and remove all traces of talcum from the outer surfaces. If this is not done, the wick becomes contaminated.
11) The replacement wicks are supplied in a sealed vial of alcohol. Remove wick from the vial.
12) Using the taper tool provided, push the wick onto the tool approximately one inch or more, causing expansion by pushing on the wick.
13) Remove from tool and clip frayed end square.
14) Slide wick gently onto sensor and over the teflon to the shoulder.
15) Using the dropper, apply several drops of water on the wick at the sensor. Note wick expands slightly, indicating water retention. Gently form the wick around end of sensor and flatten between the sensor and wick tube to offer minimum wind resistance.
16) Feed wick through teflon air sleeve, slide sleeve over teflon sensor mounts, and seat it on the probe half.
17) Feed wick through wick tube in bottom probe half and mate the probe halves. Coil wick inside reservoir and replace cover.
18) Replace four corner screws, rubber feet, and reservoir plug.
4.2 FACTORY CERTIFICATION AND RECERTIFICATION
Factory certification is initially provided at no charge with each new instrument. Since certification is typically good for one year, the instrument should be recertified by the factory on an annual basis. Generally, three temperature calibration points and three humidity check points are sufficient. If the instrument requires more than general cleaning, a reconditioning charge will also be applied. Consult the factory for current pricing information.
4-2
4.3NIST CERTIFICATION AND RECERTIFICATION
If it is desired that the instrument be certified by the National Institute of Standards and Technology (NIST), all NIST charges are passed net plus 20% to the customer, plus F.O.B. charges to and from, plus insurance coverage.
In the case of NIST certification, it is recommended (although not required) that the instrument be sent to Thunder Scientific, so that NIST certification can be performed in conjunction with recommended annual factory calibration service.
4-3
Section 5
RS-232 INTERFACE
5.1 GENERAL
Virtually all functions available through the keypad are also available through the RS-232 interface. Each command available over the RS-232 interface is included in Section 7 and is provided with a description, normal syntax and usage, and both computer and terminal communication examples. There are basically two types of commands available; one for input, one for output. Input commands consist of the basic command word, followed by a question mark (?), and are used to input information from the 5A-1MP to the computer or terminal.
Output commands consist of the basic command word, followed by an equals sign (=), and the numeric or string data, and are used to output information from the terminal or computer to the 5A-1MP. Spaces are completely ignored by the 5A-1MP and are neither acknowledged nor echoed. Errors in command syntax do not generate error messages. The command in error is simply ignored.
For wiring/cabling instructions, refer to Section 2.3.
5.2 RS-232 PROTOCOL
All commands and data sent to and from the 5A-1MP are done so at the following factory preset parameters:
Baud Rate: 1200 (Factory selectable from 300-9600) Word Size: 8 Bits Stop Bits: 1 Parity: None
The 5A-1MP only requires one stop bit when receiving, but will use two stop bits while sending. The computer terminal may be set to either one or two stop bits, depending on user preference.
5.3 INITIALIZING THE HOST TERMINAL OR COMPUTER
Ensure that a computer/terminal is connected in accordance with Section 2.3. In order to communicate over the RS-232 interface, the terminal or computer must be configured to operate with the same protocol as that of the 5A-1MP. For the proper protocol parameters, refer to Section 5.2, RS-232 Protocol.
The following example illustrates the initialization for IBM BASICA when the 5A-1MP interface is connected to com1 of the computer.
5-1
10 'INIT ROUTINE FOR 5A-1MP TO COM1 CONNECTION. 20 'SET COM1 TO #1. 30 OPEN "COM1:1200,n,8,2,cs,ds,cd,rs" as #1 40 ' 50 ' 60 'NOW CLEAR 5A-1MP DATA BUFFERS 70 ' 80 PRINT #1, CHR$ (3); 90 ' 100 ' 110 'DISABLE 5A-1MP ECHOING. 120 PRINT #1, "ECHO=OFF" 130 ' 140 ' 150 'WAIT A BIT. 160 FOR X=1 TO 50:NEXT X 170 ' 180 'CHECK AND CLEAR COM1 INPUT BUFFER. 190 IF EOF(1) THEN 210 200 A$=INPUT$ (LOC(1), #1) 210 ' 220 ' 230 ' 240 'INITIALIZATION COMPLETE. START COMMAND PROGRAMMING. . . .
5.4 CLEARING THE 5A-1MP INPUT BUFFERS
After computer or terminal initialization and prior to actual communications between the terminal or computer and the 5A-1MP, the data I/O buffers of the 5A-1MP should be cleared. To clear the I/O buffers from a terminal, send a Control C or ASCII Code 3 (ETX).
To clear the I/O buffers from a computer, send a CHR$ (3). For an example of this, see lines 60-80 of the initialization program of Section 5.3.
5-2
Section 6
IEEE 488 INTERFACE
6.1 GENERAL
Virtually all functions available through the keypad are also available through the optional IEEE 488 interface. Each command used over this interface is included in Section 7 and is provided with a description, normal syntax and usage, and an HP BASIC example. There are basically two types of commands available; one for input, one for output. Input commands consist of the basic command word, followed by a question mark (?), and are used to input information from the 5A-1MP to the computer.
Output commands consist of the basic command word, followed by an equals sign (=), and the numeric or string data, and are used to output information from the computer to the 5A-1MP. Spaces are completely ignored by the 5A-1MP. Errors in command syntax do not generate error messages. The command in error is simply ignored.
6.2 IEEE 488 ADDRESSING
All 5A-1MP'S are initially set to an IEEE 488 address of 5 when originally shipped and during factory calibration/recalibration. The address is not hardware selectable, but is software selectable, and may be changed to any value between 0 and 9. Once changed, the new address becomes permanent (until changed again). Since this address is stored in non-volatile zero-power RAM, there is no limit to the number of times which the address may be altered. To change the address, refer to the ADDR command in Section 7. Note that the SAVE command does not affect the stored address.
6.3 CLEARING THE 5A-1MP I/O BUFFERS
After power-up, it is good practice to clear the 5A-1MP I/O buffers prior to beginning IEEE 488 communications.
Executing the HP BASIC command CLEAR 7 clears all devices on the bus, while CLEAR 705 clears the 5A-1MP only. Either is sufficient to clear the 5A-1MP I/O buffers.
The HP BASIC CLEAR command may also be used at any time to clear pending 5A- 1MP outputs and garbage in the 5A-1MP input buffer due to aborted OUTPUT 705 commands.
Refer to the following example:
10 ASSIGN @ PSY TO 705 !Set up communication path 20 CLEAR @ PSY !Clear 5A-1MP I/O buffers . . .
6-1
Section 7
COMMAND REFERENCE
7.1 GENERAL
The following is a comprehensive list of commands available over the RS-232 and IEEE 488 interfaces. Each of the commands is organized in alphabetical order on the remaining pages. Each command contains examples of its usage and proper syntax when connected to a computer on the IEEE 488 bus, or to a computer/terminal on the RS-232 interface. Note that only the first four characters of each command are actually required when communicating with the 5A-1MP, and additional characters are simply ignored.
Note also that each command ends with either a questions mark (?), (when requesting data from the 5A-1MP), or an equal sign (=) followed by data (when sending data to the 5A­1MP).
In general, commands which alter the psychrometer's configuration or calibration only remain in effect until power is removed or a cold reset is performed. However, all calibration coefficients, the printer configuration, the default pressure, and the IEEE 488 device address may be stored in non-volatile memory as the new power-up default values. Refer to the ADDRess, SAVE and RESTore commands for further details.
COMMANDS GROUPED BY CATEGORY
Temp, RH, Dew Point DRYTemp
WETTemp RH DP
Saturation Vapor Pressures DRYSvp
WETSvp DPSVp
SET Parameters AVERage
FLOW PRESsure PROBe
Real Time Clock TIME
DATE
Printer Control PRINt
FORMfeed HEADer LINEs ENDLine INTErval
7-1
Misc Control AUTO
BLOWer DEGRees DISPlay RESEt RESTore SAVE
Calibration Parameters DRYOhms
WETOhms DZERo DSPAn DLINearity WZERo WSPAn WLINearity DOZEro DOSPan DOLInearity WOZEro WOSPan WOLInearity
RS-232 Parameters ECHO
IEEE 488 Parameters ADDRess
7-2
Loading...
+ 67 hidden pages