THORLABS MX Series, MBX Series, TLX Series, MX10A, MX40A User Manual

...
Remote Control User Guide
for the
MX / MBX / TLX Series
Remote Control User Guide for the MX / MBX / TLX Series
Table of Contents
Chapter 1
Chapter 2 Supported Commands ................................................................................................................... 4
Chapter 3 Remote Control Software ............................................................................................................ 14
Chapter 4 Thorlabs Worldwide Contacts ..................................................................................................... 20
Introduction ..................................................................................................................................... 1
1.1.Description ...................................................................................................................................... 1
1.2.Interfacing the Computer and the MX / MBX / TLX Instrument .................................................. 1
2.1.RF Amplifier Commands ................................................................................................................ 5
2.2.Laser Control Commands .............................................................................................................. 7
2.3.Mach-Zehnder EO Intensity Modulator Commands .................................................................... 9
2.4.System Commands ...................................................................................................................... 11
2.5.Variable Optical Attenuator (VOA) Commands ......................................................................... 12
3.1.Installing the Software and Opening a Connection with the Instrument ................................ 14
3.2.Using the Remote Control Application ...................................................................................... 15
3.3.Software Application Screenshots ............................................................................................. 16
3.4.Custom Application Development .............................................................................................. 19
Remote Control User Guide for the MX / MBX / TLX Series Chapter 1: Introduction
Chapter 1 Introduction
1.1. Description
The MX / MBX / TLX series of instruments may be remotely controlled via SCPI-type serial commands. This requires connecting a computer running the Microsoft instrument using a USB cable, or connecting a computer running any operating system to the RS-232 port on the instrument. Serial commands are sent to the USB or RS-232 ports located on the back panel of the instrument. The touchscreen interface remains active while the instrument is controlled remotely, and instrument functionality remains accessible through the touchscreen interface.
This document describes the process of interfacing the instrument with the controlling computer, defines the serial commands, and includes information about the remote control software tool that serves as an example and a tutorial for sending the serial commands. To download the remote control software tool and supporting files for your MX / MBX / TLX instrument, visit the following page and type in the instrument’s Item #:
https://www.thorlabs.com/manuals.cfm
®
Windows® 7 operating system, or later versions, to the
1.2. Interfacing the Computer and the MX / MBX / TLX Instrument
The following sections use the MX40G Electrical-to-Optical Converter as an example, but the procedure and guidelines apply to all of the MX / MBX / TLX series of instruments. These instruments may be connected with a computer via USB or RS-232 ports located on the back panel, as shown in Figure 1.
Please note that using both USB and RS-232 connections at the same time is not supported. If it is necessary to change from a USB to a RS-232 connection, or vice versa, first power off the instrument, change the connection, and then power on the instrument. This will ensure that the instrument recognizes the new connection.
B1
B5
B6
Figure 1 Back Panel of the MX40G
Label Description
B1
B2
B3
B4
B5
B6
B4
B3
B2
I/O Port (HDDB15 Connector)
Laser Interlock (2.5 mm Connector)
RS-232 Port (DB9 Connector)
USB Port (USB Type B Connector)
Power Connector
Power Switch
Supply On; Supply Off
Rev. D, September 29, 2017 Page 1
Remote Control User Guide for the MX / MBX / TLX Series Chapter 1: Introduction
Connecting to the USB Port on the Back Panel of the MX / MBX / TLX Instrument
A USB cable can be used to connect a computer running Windows 7, or later, with the port on the back panel of the instrument as shown in Figure 2. The USB port is type-B. A USB cable with a type-B connector on one end and a type-A connector on the other would allow connection with most computers.
USB Cable
RS-232
Port
USB Port
USB Port
Computer
Figure 2 Connecting the Computer and MX / MBX / TLX Instrument via the USB Ports
When discovered by the Windows PC, the MX / MBX / TLX instrument will appear as a human interface device (HID), rather than as a virtual serial port. The HID class includes the computer mouse and keyboard. Please note that PC terminal software, such as Tera Term, cannot be used to communicate with the MX / MBX / TLX instrument via USB cable. This is due to PC terminal software requiring the connected instrument to be recognized by the PC as a virtual serial port. The Thorlabs remote control application discussed in Chapter 3 can scan for, discover, and open a connection with the connected MX / MBX / TLX instrument.
Connecting to the RS-232 Port on the Back Panel of the MX / MBX / TLX Instrument
If the controlling computer has an RS-232 port, or if a USB to RS-232 adapter is connected to the computer’s USB port, an RS-232 cable can be used to make the physical connection between the computer and the RS-232 port on the back panel of the MX / MBX / TLX instrument. Any software capable of sending and receiving data using serial ports can then be used to control the MX / MBX / TLX instrument. Configure the serial port of the controlling computer for 115200 baud, 8 bits, no parity, 1 stop bit, and no flow control. The Thorlabs remote control application discussed in Chapter 3 can also control the MX / MBX / TLX instrument via an RS-232 connection. The pin assignments for the DB9 connector on the back panel of the MX / MBX / TLX instrument are described by Figure 3 and the table below.
1 2 3 4 5
6 7 8 9
Figure 3 RS-232 DB9 Connector on the
Back Panel of the MX / MBX / TLX
Instrument
Page 2 TTN116413-D04
Pin # Description
1
2
3
4
5
6
7
8
9
RS-232 Connector
Not Connected
RS-232 Input
RS-232 Output
Not Connected
Digital Ground
Not Connected
Not Connected
Not Connected
Not Connected
Remote Control User Guide for the MX / MBX / TLX Series Chapter 1: Introduction
Connecting to the MX /MBX / TLX Instrument Using the RS-232 Port on the Controlling Computer
Figure 4 illustrates the physical connection when the RS-232 ports on the MX / MBX / TLX instrument and the controlling computer are used. The details of the connection depend on the controlling computer’s RS-232 port. When it is a:
9-pin DB9 male connector, an option is to use a female-to-female 9-pin “null modem” cable. National
Instruments http://digital.ni.com/public.nsf/allkb/1EE0DD8AF67922FA86256F720071DECF
25-pin DB25 male connector, a null modem cable that converts between 25 and 9 pins can be used.
In either case, if a null modem cable is not available, a null modem adapter could be used with a standard cable.
®
explains the difference between a standard and a null modem cable here:
Null Modem Cable
RS-232
Port
USB Port
RS-232 Port
Computer
Figure 4 Connecting the Computer and MX / MBX / TLX Series Instrument via the RS-232 Ports
If the USB port on the controlling computer is used, a USB cable terminating in a USB to RS-232 adapter can be can used to allow connection to a null modem cable. When the connector on the adapter is a 9-pin DB9 male, and the null model cable has two 9-pin female DB9 connectors, it is possible to use the adapter and null modem cable to make a connection between the computer’s USB and the MX / MBX / TLX instrument’s RS-232 ports. This is illustrated in Figure 5.
Null Modem
USB Cable
Cable
USB to
RS-232
Port
USB Port
RS-232
Adapter
RS-232 Port
Computer
Figure 5 Connecting the USB Port on the Computer to the RS-232 Port on the Instrument
Please note that the USB port on the MX / MBX / TLX instrument should not be connected to the RS-232 port on the controlling computer; this configuration is not supported.
Rev. D, September 29, 2017 Page 3
Remote Control User Guide for the MX / MBX / TLX Series Chapter 2: Supported Commands
Chapter 2 Supported Commands
This section describes the commands supported by firmware version V1.6.1. The commands are modeled on the IEEE 488.2 standard commands for programmable instruments (SCPI) specification.
Following this convention, many commands in the “Syntax” columns of the following tables include both uppercase and lowercase letters. Uppercase letters identify the shortest acceptable form of the commands. A contiguous subset of the lowercase letters can be included in the command syntax as desired. (For example, VOA:SET?, VOA:SETP?, and VOA:SETPO? are all acceptable variations of the single VOA:SETpoint? command.)
The commands described below have been developed as a custom set, as no device in the SCPI standard is an exact analogue.
Please note that not every command described in the following sections applies to every MX / MBX / TLX Item #. In order for a command to apply to a specific instrument, the instrument must have the related built-in component. For example, the Laser Control Commands do not apply to instruments that do not include an internal tunable laser. The following table lists the commands supported by each instrument.
The command descriptions are grouped in the following sections:
Section 2.1: RF Amplifier Commands
Section 2.2: Laser Control Commands
Section 2.3: Mach-Zehnder EO Intensity Modulator Commands
Section 2.4: System Commands
Section 2.5: Variable Optical Attenuator (VOA) Commands
Item #
MBX
MX10A, MX40A
MX10B, MX40B
MX10C, MX40C
MX35E
MX40G
TLX1, TLX2
RF Amplifier
Commands
- -
-
-
Laser
Control
Commands
-
Mach-Zehnder Modular
Bias Control
Commands
System
Commands
-
-
Variable Optical
Attenuator
Commands
Figure 6 The Available Command Set for Each Instrument Corresponds to its Built-In Functionality
Page 4 TTN116413-D04
Remote Control User Guide for the MX / MBX / TLX Series Chapter 2: Supported Commands
2.1. RF Amplifier Commands
The MX series of instruments include several that have a built-in RF amplifier, and the following commands apply only to these instruments. The amplifier type is digital (limiting) in some of these products, such as the MX10A, MX40A, MX10B, MX40B, MX10C, and MX40C. In other products, such as the MX35E, the amplifier type is linear (analog). Only digital RF amplifiers have crossing point and swing settings, and only these RF amplifiers can operate in either analog or digital modes. In contrast, the gain may be set only for linear RF amplifiers. Unless otherwise noted, the following RF Amplifier commands apply to both amplifier types.
Command Syntax Description
Set Crossing Point (Analog Mode)
Get Crossing Point (Analog Mode)
Set Crossing Point (Digital Mode)
Get Crossing Point (Digital Mode)
Set Gain
Get Gain
AMP:CROSSing:ANAlog: N
AMP:CROSSing:ANAlog?
AMP:CROSSing:DIGital: N
AMP:CROSSing: DIGital?
AMP:GAIN: N
AMP:GAIN?
N is a floating point value between -1.0 and 1.0, inclusive. This parameter applies to the digital amplifier operating in analog mode and determines the location of the crossing point. N has no units; -1.0 corresponds to the maximum possible negative offset, and 1.0 corresponds to the maximum possible positive offset. The available range, which depends on the amplifier’s voltage limits and the current AMP:SWING setting, may be smaller than this full -1.0 to 1.0 range. When this is the case, N values within ±1.0 but exceeding the currently available range will set the crossing point to the closest permitted value. This command is not available for instruments with linear (analog) amplifier types.
Returns a floating point value between -1.0 and 1.0, inclusive. This value has no units and corresponds to the current crossing point for the digital amplifier when it operates in analog mode. This command is not available for instruments with linear (analog) amplifier types.
N is a floating point value between -1.0 and 1.0, inclusive. This parameter applies to the digital amplifier operating in digital mode and determines the location of the crossing point. Please see AMP:CROSS:ANA for additional information. This command is not available for instruments with linear (analog) amplifier types.
Returns a floating point value between -1.0 and 1.0, inclusive. This value has no units and corresponds to the current crossing point for the digital amplifier when it operates in analog mode. This command is not available for instruments with linear (analog) amplifier types.
N is a floating point value between 10.0 and 23.0 in decibels. This values sets the gain of linear (analog) type amplifiers. This command is not available for instruments with digital (limiting) amplifier types.
Returns a floating point value between 10.0 and 23.0 in decibels, which corresponds to the current gain of the linear (analog) type amplifier. This command is not available for instruments with digital (limiting) amplifier types.
Rev. D, September 29, 2017 Page 5
Remote Control User Guide for the MX / MBX / TLX Series Chapter 2: Supported Commands
Command Syntax Description
Set Amplifier Mode to Analog
AMP:MODE: 1
Sets the operation of the digital amplifier to analog mode. This includes setting the crossing point to that currently specified for analog mode operation (AMP:CROSS:ANA) and applying the maximum possible swing. This command is not available for instruments with linear (analog) amplifier types.
Set Amplifier Mode to Digital
AMP:MODE: 0
Sets the operation of the digital amplifier to digital mode. This includes setting the crossing point to that currently specified for digital mode operation (AMP:CROSS:DIG) and applying the current value of swing (AMP:SWING). The default value of swing is the amplifier’s value for V
at 1 GHz, which is
determined at the factory and place in instrument memory. This command is not available for instruments with linear (analog) amplifier types.
Get Amplifier Mode
AMP:MODE?
Returns 0 if the amplifier mode is set to analog and 1 if the amplifier mode is set to digital. This command is not available for instruments with linear (analog) amplifier types.
Power Amplifier On
AMP:POWer: 1
Sets amplifier power to “on” and returns 1 on receipt of command.
Power Amplifier Off
AMP:POWer :0
Sets amplifier power to “off” and returns 1 on receipt of command.
Get Amplifier Power Status
AMP:POWer?
Returns 0 if the amplifier power is set to “off” and 1 if the amplifier power is set to “on.” This command returns the most recent requested power state, which may differ from the amplifier’s currently active power state. AMP:SET? can be used to verify the last requested power state is active.
Get Amplifier Status
AMP:SETpoint?
A 1 is returned and a steady green dot shows in the AMP field of the instrument’s touchscreen when the amplifier is fully ready and has reached the requested settings. Otherwise, a 0 is returned and the green dot blinks. Amplifier settings are updated quickly, typically within a second. However, it is possible that an immediate query after sending a command to adjust amplifier voltage will find the amplifier in a transient state and return a zero.
Set the Amplifier Swing (Digital Mode)
AMP:SWING: N
N is a floating point value with a standard range of 3.0 to
7.0 V
. This command sets the swing of the digital amplifier
pp
when it operates in digital mode, and this value can be set while operating in analog mode. Some factory-customized units have an extended swing range. Use the touchscreen GUI on the instrument to view the available range. The AMP:SWING:N command is not available for instruments with digital (limiting) amplifier types.
Get the Amplifier Swing (Digital Mode)
AMP:SWING?
Returns a floating point value with units of Vpp. This is the swing setting applied when the digital amplifier operates in digital mode. This value can be retrieved while operating in analog mode. This command is not available for instruments with linear (analog) amplifier types.
Set the Amplifier Swing to V
(Digital Mode)
AMP:SWING:VPI
Sets the swing to the amplifier’s V receipt of command. This V
instrument at the factory and saved to memory. It is the “optimal swing” setting applied when the V
at 1 GHz and returns 1 on
value is found for each
button on the
touchscreen interface is pressed. This command is not available for instruments with linear (analog) amplifier types.
Page 6 TTN116413-D04
Remote Control User Guide for the MX / MBX / TLX Series Chapter 2: Supported Commands
2.2. Laser Control Commands
These commands are available for the TLX series of instruments and those MX series products with a built-in tunable laser, which include the MX10B, MX40B, MX35E, and MX40G. In the following, “Laser” refers to the built­in tunable laser, unless otherwise specified. Please note that the ITU channels are defined using a 50 MHz grid.
Be advised that there will be a delay between when requested laser state is specified in a “set” command and when the requested state becomes the active laser state, as the instrument requires a finite period of time to execute commands. Typical execution times are given in the following table for each relevant command. After setting laser parameter(s), the LASER:SETpoint? command can be used to determine the laser’s active status.
Command Syntax Description
Set the ITU Channel Number
Get the ITU Channel Number
Turn Dither On
Turn Dither Off
Get Dither Status
Set Fine Tuning Frequency Offset
Get Fine Tuning Frequency Offset
Get Optical Laser Frequency
LASER:CHANnel: N
LASER:CHANnel?
LASER:Dither: 1 Sets the laser dither to “on” and returns 1 on receipt of command.
LASER:Dither: 0 Sets the laser dither to “off” and returns 1 on receipt of command.
LASER:Dither?
LASER:FINE: N
LASER:FINE?
LASER:FREQ?
N is an integer from 1 and 96, inclusive, for the C-Band laser or from 1 and 93, inclusive, for the L-Band laser. This command sets the ITU channel of the laser and returns 1 on receipt of command. While tuning to the desired channel, the laser’s optical output may be temporarily reduced or turned off and then on.
Returns an integer value from 1 and 96, inclusive, for a C-Band laser or from 1 and 93, inclusive, for an L-Band laser. This command returns the most recent requested ITU channel, which may differ from the currently active ITU channel due to a typical channel set time duration of <10 s. The LASER:SET? command can be used to verify the last requested ITU channel is active.
Returns 0 if dither is set to “off” and 1 if dither is set to “on.” This command returns the most recent requested dither state, which may differ from the currently active dither state due to a typical dither set time duration of <10 s. The LASER:SET? command can be used to verify the last requested dither state is active.
N is an integer between -30,000 and 30,000, inclusive. The entered value is the frequency offset in MHz, and 1 is returned on receipt of command. Sending this command causes the laser frequency to be tuned to the sum of the current ITU channel frequency and this specified fine tuning frequency offset. The range of fine tuning frequency offsets spans the full frequency range between ITU channels.
Returns an integer between -30,000 and 30,000, inclusive, corresponding to the requested frequency offset in MHz. This value may differ from the current frequency offset due to a typical laser tuning time duration of <30 s. There is currently no accurate way to determine, via remote control, whether a fine-tuning operation has finished. However, other laser status information can be obtained using:
LASER:FREQ? LASER:SET? LASER:OOP?
Returns a floating point value with unit of GHz. the reported optical laser frequency, which is determined by the ITU channel (LASER:CHAN:N) and fine offset (LASER:FINE:N). As the resolution of the returned value is limited to 0.1 GHz (100 MHz), the values of the intermediate frequencies set using the fine­tuning feature are reported rounded to the nearest 0.1 GHz.
Rev. D, September 29, 2017 Page 7
Remote Control User Guide for the MX / MBX / TLX Series Chapter 2: Supported Commands
Command Syntax Description
Get Reported Optical Output Power (OOP)
LASER:OOP?
Returns a floating point value with units of dBm. This value corresponds to the optical output power from the laser module measured by an integrated photodiode. A typical operating value is 13.5 dBm. This is a different measurement than the one reported in response to the LASER:TAP:DBM? command, which measures the downstream optical power via optical tap. The two measurements may differ slightly. Optical output power will be unstable during laser warm-up and frequency tuning.
Power Laser On
LASER:POWer: 1
Sets the laser power to “on” and returns 1 on receipt of command. To ensure optimal stability and full optical output power, allow the laser to warm up for 15 minutes after powering on.
Power Laser Off
LASER:POWer: 0
Sets the laser power to “off” and returns 1 on receipt of command.
Get Laser Power Status
LASER:POWer?
Returns 0 if the laser power is set to “off” and 1 if the laser power is set to “on.” This command returns the most recent requested power state, which may differ from the laser’s currently active power state. The LASER:SET? command can be used to verify the last requested power state is active. Additional laser status and setpoint information can be obtained using:
LASER:SET? LASER:OOP? LASER:TAP:DBM?
LASER:TAP:MW?
Get Laser Status
LASER:SETpoint?
When the following conditions are all true, a 1 is returned and a steady green dot shows in the laser field of the instrument’s touchscreen. Otherwise, a 0 is returned.
The hardware key switch interlock on the front panel
is in the “on” position.
The instrument’s microcontroller, in response to
remote or touchscreen control, has commanded the laser to power on.
The laser’s currently active dither state (on or off)
matches the last requested dither state.
The laser’s currently active ITU channel matches the
last requested ITU channel.
The laser currently reports it is in the “ready” state.  The laser currently reports it is in the “enabled” state.
The laser currently reports its optical output power is
greater than 12.0 dBm.
Get Measured Optical Output Power in dBm
LASER:TAP:DBM?
Returns a floating point value with units of dBm. This is a measurement of the optical output power measured downstream from the laser module using an optical tap. The value returned by LASER:OOP is a measurement made by a photodiode integrated into the laser module. The two measurements may differ slightly. Optical output power will be unstable during laser warm-up and frequency tuning.
Get Measured Optical Output Power in mW
LASER:TAP:MW?
Returns a floating point value with units of mW, with the measurement details identical to those described for LASER:TAP:DBM?.
Page 8 TTN116413-D04
Remote Control User Guide for the MX / MBX / TLX Series Chapter 2: Supported Commands
2.3. Mach-Zehnder EO Intensity Modulator Commands
Several of the MX Series of instruments have built-in controllers for lithium-niobate-based, Mach-Zehnder EO intensity modulators (MZMs). The following commands apply only to these instruments, which include the MBX, MX10A, MX40A, MX10B, MX40B, MX35E, and MX40G.
Command Syntax Description
Get Calibration Status
Set Dither Amplitude
Get Dither Amplitude
Set Dither Frequency
Get Dither Frequency
Set Hold Ratio
Get Hold Ratio
Set Hold Voltage
Get Hold Voltage
MZM:CALibrating?
MZM:Dither:AMPLitude: N
MZM:Dither:AMPLitude?
MZM:Dither:FREQuency: N
MZM:Dither:FREQuency?
MZM:HOLD:Ratio: N
MZM:HOLD:Ratio?
MZM:HOLD:Voltage: N
MZM:HOLD:Voltage?
Returns a 0 if the MZM bias is not currently being calibrated and a 1 calibration is currently in progress. The MZM bias is calibrated automatically when the MZM Bias controller is turned on for the first time or when recalibration is triggered manually (MZM:RESET or using the RESET AUTO BIAS button on the touchscreen GUI). Calibration data are maintained when the bias controller is turned off and on without powering down the MX instrument, but MZM calibration data are not stored upon instrument shut-down.
N is a positive integer between 20 and 2,000, inclusive, with units of mV
. It sets the dither amplitude for those MZM bias modes
pp
that reference this value. Not all MZM bias modes use dither.
Returns the current dither amplitude setting in mV
. as a positive
pp
integer between 20 and 2,000. This will return the stored dither setting. It is not a measurement of the dither in the output from the MZM. Not all MZM bias modes use dither.
N is a positive integer between 1,000 and 10,000, inclusive, with units of Hz. Not all MZM bias modes use dither.
Returns the current dither amplitude setting in Hz as a positive integer between 1,000 and 10,000, inclusive. This will return the stored dither setting. It is not a measurement of the dither in the output from the MZM. Not all MZM bias modes use dither.
N is a positive integer between 100 and 10,000, inclusive. It has no units and represents the input power required to achieve the desired input to output power ratio. Determine N by first choosing the desired ratio of input to output powers, then multiply that ratio by 100. (e.g. If desired input and output power values are the same, their ratio is 1.0 and N = 100. If the input is 100.00 times greater than the output power, N = 10,000.) This setting is used only in Auto Power Ratio Positive and Auto Power Ratio Negative MZM bias modes.
Returns a positive integer between 100 and 10,000, inclusive. It has no units and represents the input power required to achieve the desired input to output power ratio, as described in the entry for MZM:HOLD:R:N. This will return the stored ratio setting. It is not a measurement of the MZM bias output. It is only used in Auto Power Ratio Positive and Auto Power Ratio Negative MZM bias modes.
N is an integer between -10,000 and 10,000, inclusive, and has units of mV. This setting is used only in Manual Voltage MZM bias mode.
Returns an integer between -10,000 and 10,000, inclusive, and has units of mV. This will return the stored ratio setting. It is not a measurement of the MZM bias output. It is only used in Manual
Voltage MZM bias mode.
Rev. D, September 29, 2017 Page 9
Remote Control User Guide for the MX / MBX / TLX Series Chapter 2: Supported Commands
Command Syntax Description
Set MZM Bias Mode
MZM:MODE: N
N is a positive integer between 0 and 9, inclusive. Set this value to select the MZM bias mode. The mode associated with each value is:
0: Bias control is off. 0 V is applied to the MZM bias.  1: Auto Peak bias mode, which uses dither to maintain
operation at the peak of the modulation function.
2: Auto Null bias mode, which uses dither to maintain
operation at the lowest point (null) of the modulation function.
3: Auto Quad Pos bias mode, which maintains operation at
the positive quadrature position on the modulation function. This is equivalent to using the touchscreen GUI to select Quad Mode and positive slope, with dither on.
4: Auto Quad Neg bias mode, which maintains operation at
the negative quadrature position on the modulation function. This is equivalent to using the touchscreen GUI to select Quad Mode and negative slope, with dither on.
5: Hold Quad Pos bias mode, which holds the bias at the last
voltage found in Quad mode. It is equivalent to using the touchscreen GUI to select Quad Mode and positive slope, with dither off.
6: Hold Quad Neg bias mode, which holds the bias at the last
voltage found in Quad mode. It is equivalent to using the touchscreen GUI to select Quad Mode and negative slope, with dither off.
7: Manual Voltage bias mode, which is equivalent to using
the touchscreen GUI to select Manual Mode while the “bias” setting is selected.
8: Auto Power Ratio Pos, which is equivalent to using the
touchscreen GUI to select Manual Mode while the “ratio” setting and positive slope are selected.
9: Auto Power Ratio Neg, which is equivalent to using the
touchscreen GUI to select Manual Mode while the “ratio” setting and negative slope are selected.
Trigger MZM Calibration
Get MZM Status
MZM:RESET
MZM:SETpoint?
Triggers a MZM bias calibration and returns a 1. While the calibration is running, the MXM:CAL? command returns a 1.
A 1 is returned and a steady green dot shows in the Bias field of the instrument’s touchscreen when the MZM bias is stable and at setpoint. A 0 is returned and the green dot in the Bias field blinks while the MZM is not at setpoint.
Get Post­MZM Power in dBm
Get Post­MZM Power in mW
MZM:TAP:DBM?
MZM:TAP:MW?
Returns the optical power, as a floating point value in dB, output by the MZM. This value is the scaled optical power measurement made at a tap located at the output of the MZM.
Returns the optical power, as a floating point value in mW, output by the MZM. This value is the scaled optical power measurement made at a tap located at the output of the MZM.
Page 10 TTN116413-D04
Remote Control User Guide for the MX / MBX / TLX Series Chapter 2: Supported Commands
2.4. System Commands
The system commands described in this section apply to all MX / MBX / TLX instrument Item #s. The LED brightness commands given below allow the user to individually set the brightness of the red, green, and blue LEDs that contribute to the under-chassis lighting integrated into the housings of the these instruments. The brightness can be set using a scale from 0.0 to 1.0, and the default combination is a value of 0.0 for the red LEDs, 0.0 for the green, and 0.75 for the blue. These LEDs can be used to provide a visual indicator of a change in instrument status. For example, the lighting could be changed to green to indicate the conclusion of a test driven by remote-control commands.
Command Syntax Description
Set System Wavelength
Get System Wavelength
Set Red LED Brightness
Get Red LED Brightness
Set Green LED Brightness
Get Green LED Brightness
Set Blue LED Brightness
Get Blue LED Brightness
SYStem:WAVElength: N
SYStem:WAVElength?
RGB:RED: N
RGB:RED?
RGB:GREEN: N
RGB:GREEN?
RGB:BLUE: N
RGB:BLUE?
N is the positive integer value 1310, 1550, or 1590 with units of nm. This command sets the system wavelength, which is used to selected which calibration values to use when configuring the VOA and when evaluating the power measurements made at the taps. Setting the system wavelength optimizes the instrument for use at that wavelength. Setting this value is equivalent to using the touchscreen GUI to set the system wavelength.
Returns a positive integer value of 1310, 1550, or 1590 with units of nm that correspond to the current system wavelength setting.
N is an integer value between 0 and 100, which is used to set the brightness of the red LEDs that contribute to the under-chassis accent lighting.
Returns an integer value between 0 and 100, indicating the brightness of the red LEDs that contribute to the under-chassis accent lighting.
N is an integer value between 0 and 100, which is used to set the brightness of the green LEDs that contribute to the under-chassis accent lighting.
Returns an integer value between 0 and 100, indicating the brightness of the green LEDs that contribute to the under-chassis accent lighting.
N is an integer value between 0 and 100, which is used to set the brightness of the blue LEDs that contribute to the under-chassis accent lighting.
Returns an integer value between 0 and 100, indicating the brightness of the red LEDs that contribute to the under-chassis accent lighting.
Rev. D, September 29, 2017 Page 11
Remote Control User Guide for the MX / MBX / TLX Series Chapter 2: Supported Commands
2.5. Variable Optical Attenuator (VOA) Commands
The VOA commands described in this section apply to all MX / MBX / TLX instrument Item #s. These commands are provided for controlling the VOA and determining its status are described in the following tables.
Command Syntax Description
Set the Optical Attenuation Value
Get the Optical Attenuation Value
Get Difference between Actual and Setpoint Attenuation
Get Measured Attenuation Provided by VOA
Set VOA Mode to Constant Optical Output
Set VOA Mode to Constant Attenuation
Get VOA Mode
Set the Optical Output Power Value in dBm
Get the Optical Output Power Value in dBm
VOA:ATTen: N
VOA:ATTen?
VOA:ERRor?
VOA:MEASured?
VOA:MODE: 1
VOA:MODE: 0
VOA:MODE?
VOA:OUTput:DBM: N
VOA:OUTput:DBM?
N is a floating point value between 1.0 and 20.0, inclusive, with units of dB. This command sets optical attenuation of the VOA and returns 1 on receipt of command. If set while operating in constant power mode, the operation of the VOA will not be affected. Instead, the value will be cached and applied when constant attenuation mode is activated.
Returns the attenuation setting of the VOA as a floating point value between 1.0 and 20.0, inclusive, with units of dB.
Returns the difference between the attenuation provided by the VOA and the attenuation setpoint value as a floating point value with units of dB.
Returns the value of attenuation provided by the VOA as a floating point value with units of dB. The value of attenuation is the calculated ratio of the optical power measured before and after the VOA.
Sets the VOA mode to constant optical output power and returns 1 on receipt of command. While operating in this mode, the optical power measured after the VOA is monitored and the VOA bias voltage is adjusted to maintain the requested optical output power. Holding the optical output power constant requires adequate optical power input.
Sets the VOA mode to constant optical attenuation and returns 1 on receipt of command. While operating in this mode, optical power levels measured before and after the VOA are monitored. The VOA bias voltage is adjusted to maintain the specified ratio between the two.
Returns 0 if the VOA mode is set to constant optical attenuation and 1 if the VOA mode is set to constant optical output power.
N is interpreted as a floating point value between -20.0 and
20.0, inclusive, and the units are dBm. This command sets optical output power of the VOA and returns 1 on receipt of command. While this range exceeds the maximum power output of the built-in laser, the entire range is supported to accommodate the use of higher-power external laser sources. If this value is set while operating in constant attenuation mode, it will not affect the operation of the VOA. In this case, the value will be cached and applied when constant power mode is activated.
Returns the power setting of the VOA as floating point value between -20.0 and 20.0, inclusive, with units of dBm.
Page 12 TTN116413-D04
Remote Control User Guide for the MX / MBX / TLX Series Chapter 2: Supported Commands
Command Syntax Description
Set the Optical Output Power Value in mW
VOA:OUTput:MW: N
N is a floating point value between 0.01 and 100.0, inclusive, and the units are mW. This command sets optical output power of the VOA and returns 1 on receipt of command. While this range exceeds the maximum power output of the built-in laser, the entire range is supported to accommodate the use of higher-power external laser sources. If this value is set while operating in constant attenuation mode, it will not affect the operation of the VOA. In this case, the value will be cached and applied when constant power mode is activated.
Get the Optical Output Power Value
VOA:OUTput:MW?
Returns the power setting of the VOA as a floating point value between 0.01 and 100.0, inclusive, with units of mW.
in mW
Power VOA On
VOA:POWer: 1
Sets the VOA power to “on” and returns 1 on receipt of command. When the VOA is powered on, an active and automatic software control loop is engaged to vary the VOA bias voltage. This maintains the requested attenuation level, or optical output power, depending on the VOA operating mode.
Power VOA Off
VOA:POWer: 0
Sets the VOA power to “off” and returns 1 on receipt of command. As VOA is in-line with the optical path, there is an insertion loss associated with it. When the VOA is powered off, the VOA bias voltage is set to 0.0 V, which minimizes the optical attenuation through the VOA.
Get VOA Power Status
Get VOA Status
VOA:POWer?
VOA:SETpoint?
Returns 0 if the VOA power is set to “off” and 1 if the VOA power is set to “on.”
A 1 is returned, and a steady green dot shows in the VOA field of the instrument’s touchscreen, if the attenuation provided by the VOA is within 0.1 dB of the attenuation setpoint. The attenuation provided by the VOA is the calculated ratio of the optical power measured before and after the VOA. The attenuation setpoint references the last requested power setting if operating in constant power mode, and it references the last requested attenuation setting if operating in constant attenuation mode. If the difference between the attenuation provided by the VOA and the attenuation setpoint is >0.1 dB, a 0 is returned. This command returns a valid value whether or not the VOA is powered on.
Get Optical Power Output by the VOA in
VOA:TAP:DBM?
Returns the measured optical output of the VOA as a floating point value with units of dBm.
dBm
Get Optical Power Output by the VOA in
VOA:TAP:MW?
Returns the measured optical output of the VOA as a floating point value with units of mW.
mW
Rev. D, September 29, 2017 Page 13
Remote Control User Guide for the MX / MBX / TLX Series Chapter 3: Remote Control Software
Chapter 3 Remote Control Software
Thorlabs provides a software tool for computers running Windows operating systems (Windows 7 and later). The Remote Control application is primarily intended to be used as a reference example as well as a tool for exploring the behavior of and interactions between the various remote control commands. Use it to experiment with all of the currently supported laser and VOA commands, as well as to see examples of the outgoing commands and replies from the MX / MBX / TLX instrument. This tool sends serial commands to the instrument, but this application software is not intended for system integration, as it does not support running scripts or processes.
3.1. Installing the Software and Opening a Connection with the Instrument
Begin by downloading the application software by visiting the following link, typing in the Item # of the MX / MBX / TLX instrument, and clicking on the Software Download Icon shown in the list of files available for download.
https://www.thorlabs.com/manuals.cfm.
Unzip the file whose name begins with “RCUP.” Keep the “Remote Control Utility Program.exe” file in the same directory with the “platforms” folder and other supporting files, as the application will not run if these are separated.
Before running the application, connect the MX / MBX / TLX instrument and the controlling computer using the appropriate cabling as described in Section 1.2. Power on the instrument and ensure the power button on the front panel is glowing green and the touchscreen is active. Wait until the computer discovers the instrument, and then launch the application. The application window is shown in Figure 7. The actual application window does not include the red, green, blue, and violet outlines shown in Figure 7, these outlines were drawn on an image of the window for illustrative purposes.
Figure 7 Laser Tab of the Remote Control Tool’s Application Window
Page 14 TTN116413-D04
Remote Control User Guide for the MX / MBX / TLX Series Chapter 3: Remote Control Software
Select the “Connection Type” radio button corresponding to the utilized connector on the instrument. (The location of the radios buttons is indicated by the red outline in Figure 7.) Then click the “Connect” button to open a connection to the instrument.
After clicking the “Connect” button, the “Status” field, which is located at the bottom of the region enclosed by the green outline, should show a number of lines of text. If the application established a connection with the instrument, the text shown in the status field will conclude with “Device Opened Successfully.”
If a connection was not established between the computer and instrument, the text in the “Status” field will note the failure. A common reason for a failure to connect is that the instrument has not been recognized by the computer; the software cannot find the instrument unless it has been discovered by the computer. It may take up to 30 seconds for the computer to recognize the instrument. If the device fails to open successfully, we recommend closing the Remote Control application, waiting a few seconds, re- opening the application, and attempting again to connect to the instrument. Other reasons the application may fail to successfully open a connection to the instrument include the connection already being open or another copy of the application running. If none of these are the source of the problem, and the problem persists when a different USB cable is used, please contact Thorlabs’ Technical Support for assistance.
3.2. Using the Remote Control Application
When using the application, clicking the buttons located at the top of the application window sends serial commands to the MX / MBX / TLX instrument. The results of clicking the buttons are logged in the three rectangular text fields, which are bordered by the green outline in Figure 7. The commands sent by the computer are written to the PC Output field, the data sent back from the instrument in response to the commands are written to the Remote Instrument Output field, and the status of the software is written to the Status field.
The following examples illustrate the operation of the software and the interaction between the software and the touchscreen interface on the front panel of the MX / MBX / TLX series instrument that includes a built-in laser, such as the MX40G.
Example: Using the Application to Change the Laser’s ITU Channel
Adjust the “Channel” control (within the blue outline) to 5.
Click the “Set Channel” button (within the violet outline).
The text “LASER:CHAN 5”, which is the command sent to the instrument, is printed to the “PC Output”
field (within green outline).
After a short delay, the text “1” is printed to the “Remote Instrument Output” field (within green outline).
This indicates the command was received by the instrument.
The text printed to the “Status” field indicates the response was received.
Click the “Get Channel” button (within the violet outline).
The text “LASER:CHAN?” , which is the command sent to the instrument, is printed to the “PC Output”
field (within green outline).
After a short delay, the text “5”, which is the most recent ITU channel requested, is printed to the “Remote
Instrument Output” field (within green outline).
Rev. D, September 29, 2017 Page 15
Remote Control User Guide for the MX / MBX / TLX Series Chapter 3: Remote Control Software
Example: Using the Software Application with the MX40G’s Front Panel Touchscreen
Navigate to the Laser Settings Page, which is
shown in Figure 8, from the Home Screen
Use the Laser Settings Page to change the ITU
Chanel to 6.
Click the “Get Channel” button (within the violet
outline) in the Remote Control Application.
The text “LASER:CHAN?” , which is the
command sent to the MX40G, is printed to the “PC Output” field (within green outline).
After a short delay, the text “6”, which is the ITU
channel entered using the touchscreen, is printed to the “Remote Instrument Output” field
Figure 8 Laser Settings Page of the MX40G
(within green outline).
3.3. Software Application Screenshots
The functions in the Laser tab (Figure 9) correspond to the commands described in Section 2.2: Laser Control Commands
Page 16 TTN116413-D04
Figure 9 Laser Tab
Remote Control User Guide for the MX / MBX / TLX Series Chapter 3: Remote Control Software
The functions in the VOA tab (Figure 9) correspond to the commands described in Section 2.5: Variable Optical Attenuator (VOA) Commands.
Figure 10 VOA Tab
The functions in the Amp tab (Figure 11) correspond to the commands described in Section 2.1: RF Amplifier Commands.
Rev. D, September 29, 2017 Page 17
Figure 11 Amp Tab
Remote Control User Guide for the MX / MBX / TLX Series Chapter 3: Remote Control Software
The functions in the MZM Bias tab (Figure 12) correspond to the commands described in Section 2.3: Mach­Zehnder EO Intensity Modulator Commands.
Figure 12 MZM Bias Tab
The functions in the System tab (Figure 13) correspond to the commands described in Section 11: System Commands.
Page 18 TTN116413-D04
Figure 13 System Tab
Remote Control User Guide for the MX / MBX / TLX Series Chapter 3: Remote Control Software
3.4. Custom Application Development
One option for users interested in developing custom software is to use our Remote Control application as a reference and starting point. Please contact us to request the source code. Using this source code and a development platform such as free version of the Qt expanded.
When the computer is connected to the RS-232 port on the MX / MBX / TLX series instrument, commands are sent directly to the instrument’s universal asynchronous receiver/transmitter (UART). This is not the case for applications communicating with the instrument over USB. In this case, commands sent by the computer address a Silicon Labs
®
USB to UART bridge chip built into the instrument. The UART is interfaced to the bridge chip. A
dynamic-link library (DLL) available from Silicon Labs’ website can be used to communicate with the bridge chip.
If you have questions or would like guidance as your develop a custom test configuration, please contact us. We are happy to assist you.
®
software, our example application can be modified and
Rev. D, September 29, 2017 Page 19
5HPRWH&RQWURO8VHU*XLGHIRUWKH0;0%;7/;6HULHV&KDSWHU7KRUODEV:RUOGZLGH&RQWDFWV
Chapter Thorlabs Worldwide Contacts
USA, Canada, and South America
Thorlabs, Inc. 56 Sparta Avenue Newton, NJ 07860 USA Tel: 973-300-3000 Fax: 973-300-3600 www.thorlabs.com www.thorlabs.us (West Coast) Email: sales@thorlabs.com Support: techsupport@thorlabs.com
Europe
Thorlabs GmbH Hans-Böckler-Str. 6 85221 Dachau / Munich Germany Tel: +49-(0) 8131-5956-0 Fax: +49-(0) 8131-5956-99 www.thorlabs.de Email: europe@thorlabs.com
France
Thorlabs SAS 109, rue des Côtes 78600 Maisons-Laffitte France Tel: +33 (0) 970 444 844 Fax: +33 (0) 825 744 800 www.thorlabs.com Email: sales.fr@thorlabs.com
Japan
Thorlabs Japan, Inc. 3-6-3 Kitamachi, Nerima-ku, Tokyo 179-0081 Japan Tel: +81-3-6915-7701 Fax: +81-3-6915-7716 www.thorlabs.co.jp Email: sales@thorlabs.jp
UK and Ireland
Thorlabs Ltd. 1 Saint Thomas Place Ely CB7 4EX Great Britain Tel: +44 (0) 1353-654440 Fax: +44 (0) 1353-654444 www.thorlabs.com Email: sales.uk@thorlabs.com Support: techsupport.uk@thorlabs.com
Scandinavia
Thorlabs Sweden AB Bergfotsgatan 7 431 35 Mölndal Sweden Tel: +46-31-733-30-00 Fax: +46-31-703-40-45 www.thorlabs.com Email: scandinavia@thorlabs.com
Brazil Thorlabs Vendas de Fotônicos Ltda. Rua Riachuelo, 171 São Carlos, SP 13560-110 Brazil Tel: +55-16-3413 7062 Fax: +55-16-3413 7064 www.thorlabs.com Email: brasil@thorlabs.com
China Thorlabs China Room A101, No. 100, Lane 2891, South Qilianshan Road Putuo District Shanghai 200331 China Tel: +86 (0) 21-60561122 Fax: +86 (0) 21-32513480 www.thorlabschina.cn Email: chinasales@thorlabs.com
Page 0 771'
www.thorlabs.com
Loading...