Temperature Controller Chapter 1: Warning Symbol Definitions
Symbol
Description
Direct Current
Alternating Current
Both Direct and Alternating Current
Earth Ground Terminal
Protective Conductor Terminal
Frame or chassis Terminal
Equipotentiality
On (Supply)
Off (Supply)
In Position of a Bi-Stable Push Control
Out Position of a Bi-Stable Push Control
Caution, Risk of Electric Shock
Caution, Hot Surface
Caution, Risk of Danger
Warning, Laser Radiation
Caution, Spinning Blades May Cause Harm
Chapter 1 Warning Symbol Definitions
Note: Throughout this manual, references to temperature are with respect to °C, even though the TC200 can be
set to display °C, °F, or K.
Below is a list of warning symbols you may encounter in this manual or on your device.
Rev N, July 31, 2017Page 1
Temperature Controller Chapter 2: Description
Chapter 2 Description
The TC200 Temperature Controller is a bench top controller intended for use with resistive heating elements rated
up to 18 Watts. This general purpose instrument can drive various types of heaters, including foil and resistive coil
types. It accepts feedback from either positive or negative temperature coefficient thermistors, has programmable
P, I, and D gains, and will display the temperature in °C, °F, or K. In addition, it can be programmed for up to five
sequential temperature settings along with associated ramp and hold times for each level. A user-programmable
maximum temperature limit provides protection to the device being heated, and a user-programmable power limit
protects the heating element from being over driven.
Capable of stand-alone operation from a simple keypad interface, the TC200 can be interfaced with a PC
using a standard USB Type B connector and our TC200 Application Program. Interfacing with a PC can
also be achieved by using LabView or LabWindows with a simple command-line interface from any
terminal window.
Contents of Package
Each TC200 is shipped with the following standard accessories:
TC200 User Guide
IEC 320-Compatible 120 VAC Line Cord for North American AC Wall Sockets (with the TC200) or
230 VAC Line Cord for Europe (with the TC200-EC)
CD-ROM Containing TC200 Application Software and USB Drivers
USB Standard B Cable
Page 2 TTN051989-D02
Temperature Controller Chapter 3: Safety
Chapter 3 Safety
SHOCK WARNING
HIGH VOLTAGE INSIDE
To avoid electrical shock, the power cord protective grounding conductor must be connected to
ground. Do not operate without cover installed. Refer servicing to qualified personnel.
Thorlabs provides the proper 120 VAC power input cable with each TC200 for use in the United States and the
proper 230 VAC power input cable with each TC200-EC for use in Europe. If using this unit anywhere else, the user
will need to supply a properly grounded power cable to power the unit.
ATTENTION
Do not obstruct the air ventilation slots in the housing.
All statements regarding safety of operation and technical data in this instruction manual will only apply when the
unit is operated correctly.
WARNING
Before applying power to your TC200 system, make sure that the protective conductor of the 3
conductor mains power cord is correctly connected to the protective earth contact of the socket
outlet! Improper grounding can cause electric shock resulting in severe injury or even death.
Make sure that the line voltage rating marked on the rear panel agrees with your local supply and that the
appropriate fuses are installed. Changing of the mains fuse can be done by the user (see page 28 for details).
ATTENTION
This unit must not be operated in explosive environments.
With the exception of the mains fuses, there are no user-serviceable parts in this product.
This device can only be returned when packed into the complete original packaging, including all foam packing
inserts. If necessary, ask for a replacement package.
Mobile telephones, cellular phones or other radio transmitters are not to be used within the range of three meters
of this unit since the electromagnetic field intensity may exceed the maximum allowed disturbance values according
to EN50082-1.
Rev N, July 31, 2017Page 3
Temperature Controller Chapter 4: Features
1
2 4 5 6 7
9
10
8
3
Chapter 4 Features
4.1. Front and Rear Panel
Figure 1 TC200 Front and Rear Panel
1. Power Switch – Turns the unit On and Off.
2. Power On Indicator – When this text is lit, power is applied to the unit.
3. LCD Display – Provides status and temperature information when in the main screen as shown above.
Various programmable mode screens are accessible via the Keypads (4). See Navigating the Front Panel on page 8 for more details.
4. Keypads – Used to adjust the temperature set-point (Temp Set), enable and disable the output, and
navigate through the various Mode screens.
5. Heater On Indicator – When this text is lit, the output relay is energized, and the heater controller is
enabled. Note: there are situations where the indicator (and relay) will NOT be on, yet the heater controller
will still be enabled – See Maximum Temperature Shutdown on page 6.
6. Cooling Fan
Page 4 TTN051989-D02
Temperature Controller Chapter 4: Features
7. Main Output – Use this output to connect to the heating system. See page 25 for connector information.
8. Auxiliary Output – This output is intended for use only with Thorlabs' GCH Gas Cell Heaters. It provides
an unregulated output current proportional to the Main Output current. See page 29 for more information.
9. USB Connector – Use this connector for all serial interface connections.
10. AC Input Connector
4.2. Operation
1. Programmable Proportional (P), Integral (I), and Derivative (D) gain settings – The P, I, and D gains
can each be set to values from 0 to 250 (I and D) and 1 to 250 (P gain only). These values indicate
representative shares of each gain stage, not absolute gain values (which are fixed internally). Adjust each
of the gains accordingly to affect the best response from the particular heating system being controlled.
See page 9 for more details.
2. Automatic Offset Tune – Often times P, I, or D controllers will not perform as one would hope, leaving
either an offset value or constant oscillations about the set-point. The TUNE function allows the user to
tune the output of the TC200 exactly to the set-point value without having to adjust the P, I, or D controls.
Ideally the I gain and D gain settings should be set to zero and only the P gain should be adjusted to provide
a stable temperature. Once the temperature has reached its maximum value (which should be slightly less
than the set-point), enabling the TUNE function will digitally calculate the offset and adjust the output to
meet the set-point.
3. Selectable Resistive Sensors – The TC200 allows the use of three different types of resistive temperature
sensors: platinum PT100 and PT1000 resistors with Positive Temperature Coefficients (PTC), and a
standard Negative Temperature Coefficient (NTC) TH10K 10 kΩ resistor. When selecting TH10K as the
sensor, you can also enter the Beta value for the device to provide more accurate operation. The default
Beta value shipped with the unit is 3970. See page 26 for more information on the sensors.
4. Maximum Temperature Limit – A value for a maximum temperature limit (TMAX) can be programmed
into the TC200. TMAX can range from 20.0 to 205.0 C. The set-point value is limited to a maximum of
TMAX or 200.0 C, whichever is lower. If TMAX is set to a value that is below the current set-point
temperature, the set-point will automatically be lowered to the TMAX value, along with all stopping
temperatures programmed into the cycle parameters that are greater than TMAX. Note: The capability to
attain Maximum Temperature is load dependent.
5. Maximum Power Limit – A value for a maximum output power (PMAX) can be programmed into the
TC200. Based on a maximum output voltage of 24 VDC, the PMAX value will provide the proper current
limit to the output current source that is required to limit the maximum power to the heating device.
6. Selectable Temperature Display Units – All temperature values displayed on the front panel LCD display
can be presented in either C, F, or K. Internally, the unit operates strictly on C formats. Due to the
conversion factor used, values displayed in F will not increment at even 0.1 steps as they do with C and
K.
7. NORMAL vs. CYCLE Modes – Two operating modes are possible with the TC200. In NORMAL mode, the
unit operates as a single point controller, maintaining the set-point for as long as the user requires. In
CYCLE mode, the user can program one to five distinct temperature levels. A “ramp” time (in minutes) can
be programmed for each step to determine how long it will take to get from the starting temperature to the
stopping temperature. A “hold” time (also in minutes) can be programmed to determine how long the system
will stay at the stopping temperature before moving on to the next step. During the course of any step,
CYCLE Mode can be paused, stopping the ramp or hold timers at their present count. Removing the paused
function will resume the cycle at the same point at which it was paused.
Rev N, July 31, 2017Page 5
Temperature Controller Chapter 4: Features
8. Sensor Alarms – The TC200 can determine whether a temperature sensing element is short circuited or
not installed (open circuit). If a short or open circuit occurs, the TEMP ACTUAL field in the main LCD display
screen will read !SENSOR! If the sensor alarm occurs during normal operation, the output will immediately
be disabled. The unit cannot become enabled again until the error is resolved.
ATTENTION
The TC200 can only determine shorted or open sensor connections; it cannot determine if the
installed sensor does not match the programmed sensor selection. Care must be taken to ensure
that the settings match the actual sensor used, or damage will occur to the heating system.
9. Maximum Temperature Shutdown – Based on the setting of the TMAX parameter discussed above, if
the TC200 reads back a temperature value that equals or exceeds TMAX, the output relay will be opened
and !TMAX! will be displayed in the MODE field of the main LCD display screen. The controller, however,
will remain enabled. Once the system temperature lowers down to the set-point value, the output relay will
be closed, connecting the controller to the heater again. If the temperature increases past TMAX again, the
relay will open. If this cycle occurs three times, on the third time, the relay will open and the controller will
automatically disable. The TMAX text will continue to be displayed until it is cleared manually by the user
by either pressing the SELECT keypad while in the main screen, or by pressing the ENABLE keypad.
Page 6 TTN051989-D02
Temperature Controller Chapter 5: Stand-Alone Operation
Chapter 5 Stand-Alone Operation
5.1. Initial Setup
1. Place the unit on a dry, level working surface.
2. Make sure the POWER switch on the front of the unit is in the OFF position (0).
3. Plug the female end of the provided AC line cord into the AC Input Receptacle on the rear of the unit. Plug
the male end into a properly grounded AC socket.
4. Connect the heater to be controlled to the MAIN OUTPUT connector on the rear panel of the unit. Refer to
page 25 for information on proper connections to the output connectors.
5.2. Turning the Controller On
1. Press the POWER switch to the ON position (|). If the AC line voltage is sufficient and the internal power
supply in the unit is fully functional, the PWR ON indicator will light up and the LCD screen will display a
brief message stating the model number and the firmware revision. After approximately 2 seconds, this
message will scroll off of the screen and the main LCD screen will appear.
2. The default power-up settings for the main screen are always NORMAL Mode, heater disabled, the last
active sensor type used, and the last temperature set-point that was active on the unit. To change these
settings, see Navigating the Front Panel on page 8.
3. All parameters from the last working session with the controller will be loaded during unit initialization. These
include: PID gain settings, “Tuning” offset, sensor selection (including the TH10K Beta value if applicable),
Maximum Temperature TMAX setting, Maximum Power PMAX setting, the last display units used (°C, °F,
or K), and all cycle mode parameters for the five programmable cycle steps.
4. If the heater has not been operated yet, the TEMP ACTUAL field on the LCD display should show a value
appropriate for a “cold” system (somewhere around room ambient temperature). If it is noticeably different,
you should double-check that the selected sensor (PTC100, PTC1000, TH10K) matches the actual sensor
used in your heating system.
5.3. Enabling the Heater
1. Press and release the ENABLE keypad on the front panel to turn on the heater. At this time, the Heater ON
indicator will immediately light up. If not, seepage 27 for troubleshooting suggestions.
2. If successfully enabled, the STATUS field in the LCD screen will read “ENABLED” and the MODE field will read either “NORMAL” or “CYCLE” depending on the particular mode chosen.
5.4. Adjusting the Temperature in NORMAL Mode
The set-point temperature can be changed at any time while in NORMAL Mode, regardless of whether the controller
is enabled, simply by pressing the UP or DOWN arrow keypads. Pressing the UP arrow will increment the
temperature by 0.1°, and pressing the DOWN arrow will decrement the temperature by 0.1°. As with all of the
numeric adjustments that can be done on the TC200, holding the arrow keypads down will accelerate the rate of
increase or decrease the longer the arrow keypad is held down.
Rev N, July 31, 2017Page 7
Temperature Controller Chapter 5: Stand-Alone Operation
5.5. Turning the Heater Off
When necessary, the heater output can be turned off by either pressing and releasing the ENABLE keypad
or pressing the POWER switch to the OFF ( 0 ) position, which will turn off the entire unit.
The heater will also turn off immediately if a shorted or open temperature sensor is detected. In that case,
!SENSOR! will be displayed on the LCD screen.
Whenever the unit is ON and the heater is not enabled, the status will be indicated as “DISABLED” on the
LCD and the HEATER ON indicator will be turned off.
5.6. Navigating the Front Panel
5.6.1. Overview
The front panel contains six keypads that can be used for various functions depending on what menu screen
is actively displayed.
The Main Display screen displays the selected temperature sensor, the basic operating mode (NORMAL
or CYCLE), the current temperature set-point, the current actual temperature, and various status messages
such as ENABLED, DISABLED, !TMAX!, and !SENSOR! alarms.
There are five MODE screens: PID parameters, Sensor Select, TMAX/PMAX values, Units Display
(°C, °F, or K), and NORMAL vs. CYCLE Mode.
There is a sixth sub-menu for programming the CYCLE Mode parameters: Stop Temp, Ramp Time, and
Hold Time.
Use the MODE keypad to access the Mode screens.
While a menu is displayed, use the SELECT keypad to move the flashing cursor to the various selection
options.
Pressing the ENTER keypad while in the Main Display screen and in NORMAL mode has no effect.
Pressing the ENTER keypad while in the Main Display screen and CYCLE Mode will PAUSE the cycle.
See CYCLE Mode on page 13 for more details.
Pressing the ENTER keypad while in a Mode menu screen will accept and save the option that corresponds
to the flashing cursor and advance you to the next menu screen or back to the MAIN screen, depending on
the situation.
Use the UP and DOWN keypads to change the set-point temperature while in the Main screen.
Use the UP and DOWN keypads to change numerical options while in any of the Mode screens.
Pressing the MODE keypad at any time while in a mode menu will do one of the following:
o If you are in the PID Mode screen, pressing the MODE keypad will advance to the Sensor Select
menu without changing any of the current PID values.
o If you are in the Sensor Select screen, pressing the MODE keypad will advance to the TMAX/PMAX
Mode screen without changing the sensor selection.
o If you are in the TMAX/PMAX Mode screen, pressing the MODE key will advance to the Display
Units Selection screen without changing the units.
Page 8 TTN051989-D02
Temperature Controller Chapter 5: Stand-Alone Operation
o If you are in the Display Units Selection screen, pressing the MODE keypad will advance to the
NORMAL vs. CYCLE Mode screen without changing the units.
o If you are in the NORMAL vs. CYCLE Mode screen, pressing the MODE keypad will advance to
the Main screen without changing the Operating Mode selection.
Pressing the ENABLE keypad at any time, in any screen* will either enable or disable the heater, depending
on its current status.
* While in the Cycle Parameters screen, you can only disable the heater.
Special Functions – While in the Main Display screen, the SELECT keypad will perform two special
functions:
o If you are using Thorlabs’ gas cell heaters in conjunction with the auxiliary output, pressing the
SELECT key will momentarily display the auxiliary heater’s temperature. This function only works
when the PTC100 sensor is selected.
o If a TMAX error occurs, pressing the SELECT keypad will clear the error.
5.6.2. Selecting and Adjusting the Gains: P, I, D, and Tune
Changing the P, I, or D Gains
1. While in the Main Screen press the MODE keypad once.
2. The screen will now display the P, I, and D values as well as the text “TUNE” or “DETUNE”.
3. The cursor will be flashing over the “P” for the proportional gain value.
4. Adjust the P gain using the UP or DOWN arrow keypads. The P gain is adjustable from 1 to 250.
5. Press the SELECT keypad to move the cursor to the “I” for integral gain.
6. Adjust the I gain using the UP or DOWN arrow keypads. The I gain is adjustable from 0 to 250.
7. Press the SELECT keypad to move the cursor to the “D” for derivative gain.
8. Adjust the D gain using the UP or DOWN arrow keypads. The D gain is adjustable from 0 to 250.
9. Press the MODE keypad to escape to the next Mode screen (Sensor Select) without making any changes
to the current parameters.
10. Alternatively, press the ENTER keypad to accept any changes made to the P, I, or D values and return to
the Main Display screen.
Note: The PID screen is the only Mode screen that will return directly to the Main Display screen upon
pressing the ENTER keypad. All other Mode screens will save changes and advance to the next Mode
screen.
Rev N, July 31, 2017Page 9
Temperature Controller Chapter 5: Stand-Alone Operation
Activating or Deactivating the TUNE Function
Note: The TUNE function is only accessible in NORMAL mode when the heater is enabled. It cannot be accessed
at any time in CYCLE mode or when the heater is disabled.
1. While in the Main Display screen, press the MODE keypad once.
2. The screen will now display the PID values and “TUNE” or “DETUNE” depending on the status of the
TUNE function.
3. Press the SELECT key three times to advance to the TUNE/DETUNE text.
4. Press the ENTER keypad once.
5. The MAIN screen will now be displayed.
6. Please refer to TUNE Function on page 14 for more information on tuning the TC200.
5.6.3. Selecting the Temperature Sensor: PTC100 vs. PTC1000 vs. TH10K
Programming the Appropriate Temperature Sensor
1. While in the Main Display screen press the MODE keypad twice.
2. The screen will now display PTC100, PTC1000, TH10K, and the Beta value.
3. The cursor will be flashing over the “P” for PTC100.
4. Pressing the ENTER keypad at this time will select the PTC100 and advance you to the next Mode screen
(TMAX/PMAX).
5. Otherwise, press the SELECTkeypad to move the cursor to the “P” for PTC1000.
6. Pressing the ENTER keypad at this time will select the PTC1000 and advance you to the next Mode screen
(TMAX/PMAX).
7. Otherwise, press the SELECTkeypad to move the cursor to the “T” for TH10K.
8. Pressing the ENTER keypad at this time will select the TH10K and advance you to the “B” for the Beta
value within the same Mode screen.
9. Use the UP or DOWN keypads to adjust the Beta value in increments of 10 from 2000 to 6000. For entering
exact Beta values, see The Command Line Interface on page 16 or The TC200 Application Software
on page 16.
10. Pressing the ENTER keypad at this time will accept the Beta value and advance you to the next Mode
screen (TMAX/PMAX).
11. Press the MODE keypad to escape to the next Mode screen (TMAX/PMAX) without making any changes
to the current parameters.
Note: Any attempt to change the sensor while the heater is enabled will immediately disable the heater.
Page 10 TTN051989-D02
Temperature Controller Chapter 5: Stand-Alone Operation
5.6.4. Selecting and Adjusting the Maximum Temperature and Power: TMAX and PMAX
Changing the TMAX or PMAX Values
1. While in the Main Screen display, press the MODE keypad three times.
2. The screen will now display the TMAX and PMAX values.
3. The cursor will be flashing over the “T” for the TMAX value.
4. Adjust the TMAX value using the UP or DOWN arrow keypads. TMAX is adjustable from 20.0 C to
205.0 C.
5. Press the SELECT Key to move the cursor to the “P” for the PMAX value.
6. Adjust the PMAX value using the UP or DOWN arrow keypads. PMAX is adjustable from 0.1 Watts to
18.0 Watts.
7. Press the MODE keypad to escape to the next Mode screen (Units Select) without making any changes to
the current parameters.
8. Alternatively, press the ENTER keypad to accept any changes made to the TMAX or PMAX values, and
then advance to the next mode screen (Units Select).
5.6.5. Selecting the Units Displayed: Celsius vs. Fahrenheit vs. Kelvin
The Main screen and Mode screens display various information regarding the output temperature. These values
can be displayed as °C, °F, or K, depending on the user’s preferences.
Changing the Format of the Temperature Display
1. While in the Main Display screen, press the MODE keypad four times.
2. The screen will now display the Display Units Selection menu with the text “CELSIUS”, “KELVIN”, and “FAHRENHEIT”.
3. The cursor will be flashing over the “C” in CELSIUS.
4. Pressing the ENTER keypad at this time will select Celsius and advance you to the next Mode screen
(NORMAL vs. CYCLE Mode).
5. Otherwise press the SELECT keypad to move the cursor to the “K” for KELVIN.
6. Pressing the ENTER keypad at this time will select Kelvin and advance you to the next Mode screen
(NORMAL vs. CYCLE Mode).
7. Otherwise press the SELECT keypad to move the cursor to the “F” for Fahrenheit.
8. Pressing the ENTER keypad at this time will select Fahrenheit and advance you to the next Mode screen
(NORMAL vs. CYCLE Mode).
9. Pressing the MODE keypad at any time will advance you to the next Mode screen without changing the
display format.
Rev N, July 31, 2017Page 11
Temperature Controller Chapter 5: Stand-Alone Operation
5.6.6. Selecting the Operating Mode: NORMAL vs. CYCLE
Changing the Operating Mode
1. While in the Main screen press the MODE keypad five times.
2. The screen will now display “NORMAL MODE” and “CYCLE MODE”.
3. The cursor will be flashing over the “N” for NORMAL mode.
4. Pressing the ENTER keypad at this time will select NORMAL mode and advance you to the MAIN screen.
5. Press the MODE keypad to escape to the Main screen without making any changes to the current mode.
6. Otherwise, press the SELECT keypad to move the cursor to the “C” for CYCLE mode.
7. Press the MODE keypad to escape to the Main screen without changing to CYCLE mode.
8. Pressing the ENTER keypad at this time will select CYCLE Mode and advance you to the Cycle Parameters
screen described in the following section.
Changing the Cycle Parameters
The Cycle Parameters menu is a sub-menu of the NORMAL vs. CYCLE Mode menu that can only be accessed if
CYCLE Mode has been selected.
The Cycle Parameters can only be adjusted if the heater is disabled. You can navigate and review the parameter
settings while a cycle is in progress, however.
Once you have accessed the Cycle Parameters menu you can make adjustments as follows:
1. Use the SELECT keypad to move the cursor to the parameter that you wish to change:
The STOP Temperature is located in the upper left, “T=nnn.n”.
The RAMP time is located in the upper right, “Rt=nnn m”
The HOLD time is in the lower left, “Ht=nnn m”
The STEP # is in the lower right.
2. Use the UP and DOWN arrow keypads to increase or decrease the parameters.
3. Use the SELECT keypad to advance to the various parameters.
The valid range for the STOP temperature is 20.0 C to TMAX or 200.0 °C, whichever is lower.
The valid range for the RAMP and HOLD times is 1 to 999 minutes.
4. Pressing the ENTER keypad at any time will accept the changes and advance you to the next step’s
parameter screen. After step 5, you will be returned to the Main screen.
5. Pressing the MODE keypad at any time will return you to the Main screen without changes.
Note: The cycles will initiate the next time the heater is enabled.
Page 12 TTN051989-D02
Temperature Controller Chapter 5: Stand-Alone Operation
5.7. CYCLE Mode
The STOP temperature determines the final temperature that the system will try to go to in the amount of
time determined by the Ramp time value.
The RAMP time is how long it will take to go from a starting temperature to a stopping temperature.
The HOLD time is how long the cycle will stay at the STOP temperature before advancing to the next step,
also known as “dwell time”.
The initial starting temperature is determined to be whatever the TEMP SET value is at the start of a cycle.
After that, the starting temperature of the subsequent step is the stopping temperature from the previous
step. That is, the stopping temperature for step 1 will be the starting temperature for step 2, and so forth.
The number of steps can be reduced by programming in the same STOP Temperature for any steps that
you don’t need. That is, if you only want one cycle that ramps from TEMP SET to the STOP temperature
over a certain amount of time, simply program in the same STOP temperature for steps 2 – 5. Once the
step 1 HOLD time expires, the cycle will “jump” over steps 2 – 5 and resume operation in NORMAL mode.
NOTE: By the same token, if your starting value for TEMP SET is the same as your stopping value for step
1, then step 1 will appear to be jumped over. If ALL off your stopping values are the same as your TEMP
SET value then it will appear as though the cycle never happened.
During the course of a cycle, the following actions can be taken:
1. While in the Main screen, pressing the ENTER keypad at any time will cause the scan to pause. Once
paused, the cycle will hold at the current ramp or hold time count and remain there with the heater still
enabled. Pressing the ENTER keypad again will cause the scan to resume at the current channel.
2. While in any screen, pressing the ENABLE keypad at any time will cause the cycle to terminate
completely, turning off the heater, disabling the output, and changing the operating mode to NORMAL.
3. At no time during an active cycle can the set-point be adjusted from the Main screen using the UP or
DOWN arrow keypads.
4. At no time during an active cycle can any of the cycle parameters be adjusted.
CYCLE Mode will be terminated and changed back to NORMAL Mode as follows:
o Letting a cycle completely run its course will change the mode back to NORMAL, and the
temperature will remain at the last stopping temperature with the heater still enabled.
o Pressing the ENABLE keypad during an active cycle will terminate the cycle and disable the heater.
o From the main screen, press the MODE keypad five times:
The NORMAL vs. CYCLE Mode menu will come on, with NORMAL mode as the active
selection.
Pressing the ENTER keypad at this point will terminate the cycle at the current temperature
and return you to the Main screen. The heater will remain enabled.
Rev N, July 31, 2017Page 13
Temperature Controller Chapter 5: Stand-Alone Operation
5.8. TUNE Function
The TUNE function is a manually enabled feature that digital calculates the offset between the TEMP SET
value and the TEMP ACTUAL value. The offset is only calculated at one time and is not actively adjusted
after that point.
Once the TUNE function is activated, the TEMP ACTUAL value should slowly adjust to within 0.1 C of the
TEMP SET value. The rate of adjustment and accuracy depends on the PID settings.
The TUNE function should only be set once the output has reached a final value and has settled completely.
Activating a TUNE function while the temperature is changing will result in less than ideal performance.
If the unit is being DETUNED, the output will develop an offset between the set-point and actual
temperatures.
Note: For best results, the I and D values should be set to zero. The output must be completely settled before
starting a TUNE function. If for whatever reason the output needs to be retuned, you must select DETUNE from
the Gains mode screen, wait for the offset to settle out, and then select TUNE again.
The offset value calculated by any tune function will be saved to memory and will be loaded on the next
power up of the unit.
To determine if an offset value exists from a tuning action, simply enter the PID Gains mode screen. If the
text reads DETUNE, then there is an offset preset.
In general, the offset should be consistent for a large range of temperatures for the same heating system.
It may be necessary, however, to detune and then re-tune after changing set points.
Adjustments to the P, I, or D gains will affect the offset value and therefore, if the P, I, or D values are
adjusted after a tuning action has been done, the offset will be detuned.
The TUNE function can only be set while in NORMAL Mode. It is not accessible while operating in CYCLE
Mode. Once an offset is tuned in NORMAL mode, however, the tuning action will still apply to the target
temperatures set in the cycle parameters.
5.9. TMAX Function
The TMAX function allows the user to set the maximum temperature to which the control will operate.
The TEMP SET set-point temperature will be limited to the current TMAX or 200.0 C, whichever is lowest.
If the TMAX value is adjusted to a value lower than the current TEMP SET value, the TEMP SET value will
be changed to match TMAX. Note: The capability to attain Maximum Temperature is load dependent.
The STOP temperature values in the cycle parameters are limited to the current TMAX or 200.0 C,
whichever is lowest. If the TMAX value is adjusted to a value lower than any of the current STOP
temperature values, the STOP temperature values will be changed to match TMAX.
If the TC200 reads back a temperature value that equals or exceeds TMAX, the output relay will be opened
and !TMAX! will be displayed in the MODE field of the Main Display screen. The controller, however, will
remain enabled. Once the system temperature reduces down to the TEMP SET set-point value, the output
relay will re-close, connecting the controller to the heater again. If the temperature increases past TMAX
again, the relay will re-open. If this cycle occurs three times, on the third time, the relay will open and the
controller will automatically disable. The TMAX text will continue to be displayed until it is cleared manually
by the user by either pressing the SELECT keypad while in the Main screen, or pressing the ENABLE
keypad.
Page 14 TTN051989-D02
Temperature Controller Chapter 5: Stand-Alone Operation
If the temperature stabilizes to TEMP SET after having crossed the TMAX threshold less than three times,
the !TMAX! alarm will remain active in the Main Display screen to alert a user to the fact that the system
exceeded TMAX momentarily. The !TMAX! display can be cleared by pressing the SELECT keypad while
in the Main Display screen.
Pressing the SELECT keypad while in the Main Display screen will always clear the TMAX count and resets
the alarm.
Pressing the ENABLE keypad at any time with a TMAX alarm present will disable the heater and clear and
reset the TMAX alarm.
5.10. Sensor Alarm
The TC200 can determine whether a temperature sensing element is short circuited or not installed (open
circuit). If this occurs, the TEMP ACTUAL field in the Main LCD display screen will read !SENSOR! If the
sensor alarm occurs during normal operation, the output will immediately be disabled. The unit cannot
become enabled again until the error is resolved.
ATTENTION
The TC200 can only determine shorted or open sensor connections; it cannot determine if the
installed sensor does not match the programmed sensor selection. Care must be taken to ensure that
the settings match the actual sensor used, or damage will occur to the heating system.
A common sense approach to verifying the proper sensor is selected is to simply connect a heating system
to a unit that has been sitting at room ambient temperature for a while. Observe the TEMP ACTUAL value
displayed on the TC200. If the sensor has been properly programmed, the display will show a value
corresponding to the room ambient temperature.
5.11. Setting the PID Gains
The PID gains determine the overall stability and accuracy of the entire heating system. Incorrect values,
especially for the Integral (I) gain, could result in undesired overshoots and instabilities.
A good approach to setting these values is to start with a mid-range value (125) for the P gain and to set
both I and D gains to zero (0).
Enable the heater and observe the response and settling times. Typically, the system will undershoot the
set-point and settle to a value below TEMP SET. Despite the offset, the temperature should remain fairly
stable.
If the temperature offset is too great (greater than 3 to 4°), an increase in P gain might be required.
The offset can now be adjusted out by setting the I gain to a suitable value. Always start with a low value
(typically less than 10). Observe any changes in the system response after each adjustment to I gain. If the
value is too large, you will observe overshoot and oscillations about the set-point. If it is too low, you will
not remove the offset.
After arriving at a value for I gain, apply a Step function to the system by increasing the TEMP SET value
by a few degrees. Observe the response and adjust the gains accordingly.
The D gain is the hardest to observe any influence on, as it generally affects the rate at which the system
responds to disturbances. For most small heater systems being controlled by the TC200, a value of zero
(0) for the D gain is sufficient for good operation.
Rev N, July 31, 2017Page 15
Temperature Controller Chapter 6: Operating from a Computer
Chapter 6 Operating from a Computer
6.1. Overview
The TC200 Temperature Controller is fully operational from a remote computer via a USB interface.
Connection from the computer to the TC200 is made with a standard B, USB cable.
The TC200 can be accessed directly using a command line interface and any Terminal Communications
software, such as HyperTerminal.
The TC200 can also be controlled using the TC200 Application Software provided with the unit.
6.2. Installation
Please note that you will require administrative rights on the Computer on which you plan to install the
TC200 Software.
Prior to connecting the TC200 to your computer, the USB drivers must be installed. Insert the CD packad with the
TC200 into the CD drive of your PC and you should be presented with the following “AutoPlayr” dialog box:
Click the Launch Thorlabs, Inc. CD item to start the CD Starter application, you should then be presented with the
following “CD-Starter” dialog box:
.
Page 16 TTN051989-D02
Temperature Controller Chapter 6: Operating from a Computer
Figure 2 CD Starter Screen
6.2.1. Software and USB Driver Installation
Please click “Install Software and Drivers” to begin installation of the Application and drivers, please follow the
On-Screen instructions and dialog boxes presented, to complete the installation.
Note: The USB interface is performed using an RS232 to USB converter. Your computer will reflect this as a COM
Port in your Device Manager, not a USB port.
Note: If using Windows XP, you may be presented with the following dialog box:
You may choose which user(s) are granted access to the TC200 software. To install the software for all users,
you must have administrative rights, if you do not, please contact your Computer Admin. If you have trouble
installing the software as noted above, you may have to Un-Check“Protect my computer and data from
unauthorized program activity”.
Rev N, July 31, 2017 Page 17
Figure 3 Admin Control – Windows XP
Temperature Controller Chapter 6: Operating from a Computer
Set-Up and Connection
In order for the TC200 to communicate with the computer please make sure the USB Driver are installed. Refer to
section 6.2.1 for further details.
When the TC200 is connected to the Computer, it is assigned a “COM Port Number”. It may be necessary to know
which “COM” port number the Computer has assigned the TC200, since the software looks for particular number in
order to communicate. Use to guide below to help determine which “COM Port” number your Computer has
assigned the TC200.
Windows XP:
1. Right-Click on My Computer and click Properties to open System Properties.
2. In the System Properties window click the Hardware tab.
3. In the Hardware tab, click the Device Manager Button.
Windows 7:
1. Right-Click on My Computer and click Properties to open System Properties.
2. Click on Device Manager.
Page 18 TTN051989-D02
Figure 4 Windows XP
Temperature Controller Chapter 6: Operating from a Computer
Figure 5 Windows 7
3. In Device Manager, double-click Ports (COM & LPT)
4. Find “USB Serial Port” and note the COM Port Number in parenthesis.
Rev N, July 31, 2017Page 19
Figure 6 Windows XP
Temperature Controller Chapter 6: Operating from a Computer
Parameter
Value
Baud Rate
115200 Bits Per Second
Data Bits
8
Parity
None
Stop Bits
1
Flow Control
None
Figure 7 Windows 7
6.3. The Command Line Interface
6.3.1. Setup and Connection
1. Open your terminal interface program and, referring to documentation specific to your program, set up the
terminal for the COM Port to which you have connected. Refer to Section 6.2.3 to determine the COM Port number:
2. Apply AC power to the TC200, turn the TC200 on (see page 7), and wait for the unit to complete its poweron initialization.
3. Press the ENTER keypad on your computer. If the connection is working, you will receive the following
message on your terminal:
oCommand error CMD_NOT_DEFINED
4. Followed immediately by the prompt:
o>
Page 20 TTN051989-D02
Temperature Controller Chapter 6: Operating from a Computer
The basic command structure consists of two types: commands and queries, both of which are sent to the
TC200 by a carriage return (CR) or by pressing the ENTER key on your computer.
Most commands follow a format of:
oCommand = Argument (CR)
Command is a keyword and Argument is a numerical value always followed by a carriage return (CR).
The following section gives a listing of all command keywords.
All queries follow a format of:
oquery? (CR)
Query is a keyword always followed by a question mark (?) and carriage return (CR). The following section
gives a listing of all query keywords.
6.3.2. Terminal Commands and Queries
All commands and queries are in lower case letters.
All temperature inputs and read backs are in °C only, regardless of what the display units are for the front
panel LCD of the TC200.
Rev N, July 31, 2017Page 21
Temperature Controller Chapter 6: Operating from a Computer
Command/Query
Function
Description
ens
Toggle Enable State
If the unit is disabled, ens will enable the unit.
Otherwise, it will disable.
tset
Set Temperature
tset=nnn.n (20.0 to 200.0 to TMAX)
tset?
Get TEMP SET Value
Returns the current value of TEMP SET.
tact?
Get Current Temperature
Returns the current value of TEMP ACTUAL.
temps?
Get TEMP SET and TEMP
ACTUAL values
Returns TEMP SET and TEMP ACTUAL.
stat?
Status Query
Returns the Status Byte (see page 23).
mode
Assign Operating Mode
mode=normal or mode=cycle
pgain
Set P Gain
pgain=nnn (1 to 250)
igain
Set I Gain
igain=nnn (0 to 250)
dgain
Set D Gain
dgain=nnn (0 to 250)
pid?
Get PID Gains
Returns P, I, D values, respectively.
unit
Set Degree Units - °C, °F, or K
unit=c, unit=k, or unit=f
sns
Select Sensor
sns=ptc100 or sns=ptc1000 or sns=th10k
sns?
Get Sensor
Returns current sensor selection.
beta
Set TH10K Beta Value
beta=nnnn (2000 to 6000)
PMAX
Set Max Power
PMAX=nn.n (0.1 to 18.0)
TMAX
Set Max Temperature
TMAX=nnn.n (20.0 to 205.0)
PMAX?
Get Max Power
Returns current PMAX value.
TMAX?
Get Max Temperature
Returns current TMAX value.
beta?
Get TH10K Beta Value
Returns current Beta value.
stop
Set StopTmp
stop=nnn.n (20.0 to 200.0 or TMAX- see page 23)
ramp
Set Ramptime
ramp=nnn (1 to 999- see page 23)
hold
Set Holdtime
hold=nnn (1 to 999- see page 23)
cycle
Set Step Number
cycle=n (1 to 5- see page 23)
cycles?
Cycle Mode Parameters
Returns cycle parameters for all 5 steps.
(see page 23)
pause
Pause Cycle
pause (cr). Initiates or terminates
a pause to cycle mode.
config?
Get Current Configuration
Returns all pertinent set-up parameters
(see page 24).
tune
Tune Temperature
tun(cr). Initiates or terminates
(detunes) a tuning function.
tune?
Is Tune Function Set?
Returns offset value. “O” represents
no tune performed.
commands?
Commands Query
Returns a list of available commands.
id?
Identification Query
Returns the unit identification string.
*idn?
Identification Query
Returns the unit identification string.
taux?
Get Temperature of
Auxiliary Heater
Returns the current value of Aux Temp.
Page 22 TTN051989-D02
Temperature Controller Chapter 6: Operating from a Computer
Status Byte Definitions
Bit 0
0 = Disabled
1 = Enabled
Bit 1
0 = Normal Mode
1 = Cycle Mode
Bit 2
0 = TH10K
1 = PTC100 (sensor select bit one)
Bit 3
0 = See Bit2
1 = PTC1000 (sensor select bit two)
Bit 4
0 = See Bit5
1 = degrees C (unit select bit one)
Bit 5
0 = Degrees K
1 = Degrees F (unit select bit two)
Bit 6
0 = No Sensor Alarm
1 = Sensor Alarm
Bit 7
0 = Cycle Not Paused
1 = Cycle Paused
The Status Byte
The “Status Byte” is an 8-bit hexadecimal value that is returned as part of the stat? query. Each bit corresponds to
the following definitions to provide a snapshot of the status of the unit at any given time:
If a TMAXalarm is also present when the stat? query is sent, the text “TMAXERROR” will also be returned.
Setting the CYCLE Parameters from the Command Line
The CYCLE parameters (STOP temp, RAMP time, and HOLD time) can be programmed from the command line
interface as follows:
1. Set the Step number to which the three parameters will apply by typing cycle=n, where n represents the
Step number (1 to 5).
2. Enter valid values for the STOP temp by typing stop=nnn.n (20.0 to 200.0 or TMAX)
3. Enter valid values for the RAMP time by typing ramp=nnn (1 to 999)
4. Enter valid values for the HOLD time by typing hold=nnn (1 to 999)
Until the CYCLE number is changed to another value, any values entered for STOP, RAMP, or HOLD will apply to
the current value of CYCLE.
Typing the cycles? query will return a string of fifteen comma-delineated numbers: five groups of three values.
Each group of three values represents the STOP temp, RAMP time, and HOLD time for each successive cycle step.
The value of the STOP temp is an integer value ten times greater than the actual temperature value. For example,
a value of 600 represents 60.0 degrees.
Rev N, July 31, 2017Page 23
Temperature Controller Chapter 6: Operating from a Computer
The Configuration Command
Typing config? Into the command line interface will return a list of all of the pertinent set up parameters to allow a
user to view all items at a glance. The following is a typical return:
> config?
Tset = 54.3 C
Pgain = 126, Igain = 0, Dgain = 0
Sensor = PTC100
TMAX = 200.0 C
PMAX = 15.9 Watts
Temperature Display Units are CELSIUS
Unit is Normal Mode
Step1:
StopTemp = 60.0 C, RampTime = 1 m, HoldTime = 5 m
Step2:
StopTemp = 70.0 C, RampTime = 5 m, HoldTime = 5 m
Step3:
StopTemp = 80.0 C, RampTime = 5 m, HoldTime = 5 m
Step4:
StopTemp = 90.0 C, RampTime = 5 m, HoldTime = 5 m
Step5:
StopTemp = 100.0 C, RampTime = 5 m, HoldTime = 5 m
Page 24 TTN051989-D02
Temperature Controller Chapter 7: Output Connector Pin-Outs
Pin Assignments
Pin 1
Heater Output Positive
Pin 2
Heater Output Return (Ground)
Pin 3
Reserved (Do Not Connect to this Pin)
Pin 4
Sensor Input (+)
Pin 5
Sensor Input (Ground)
Pin 6
Reserved (Do Not Connect to this Pin)
1 2 3
4 5 6
Chapter 7 Output Connector Pin-Outs
Figure 8 Detail of Main and Auxiliary Connector (HIROSE HR10A-7R-6S)
As Viewed from the Rear Panel
Use HIROSE connector HR10A-7P-6P (73) to mate with the TC200 output connectors.
Rev N, July 31, 2017Page 25
Temperature Controller Chapter 8: Temperature Sensor Specifications
Chapter 8 Temperature Sensor Specifications
The TC200 is compatible with two types of Platinum Thermistors: PT100 and PT1000. The TC200 is also compatible
with the TH10K thermistor. The following specifications are used for determining the set-point and read back values
for these types of thermistors.
8.1. PT100 and PT1000
For the temperature range of 0 to 850 °C:
)
Where:
= 3.9083 x 10
= -5.775 x 10
-3
-7
°C
°C
-2
-1
is the resistance in Ω at temperature T
is the temperature in °C
= 100 Ω for the PT100
= 1000 Ω for the PT1000
8.2. TH10K
Where:
is the resistance in Ω at temperature T
is the temperature in K
is the constant associated with the particular thermistor
Page 26 TTN051989-D02
Temperature Controller Chapter 9: Troubleshooting
Problem
Solution
Unit does not turn on when switching
the power switch to the ON position.
1. Make sure AC line cord is fully inserted into the AC Input
receptacle and plugged into an outlet providing 100 to 240
VAC.
2. Fuse(s) may be open. Refer to page 3228 for information on
replacing open fuses. If the problem persists, please return the
unit to Thorlabs for evaluation.
Unit does not enable the heater when
pressing the ENABLE keypad.
Make sure that a compatible temperature sensor is installed. If
one is not, you should see a message that says “!SENSOR!” on
the front panel LCD.
I can’t adjust the TEMP SET value
using the UP/DOWN arrow keypads.
If you are operating in CYCLE mode, you will not be able to make
adjustments to the set-point using the UP/DOWN keypads in the
Main screen. If you need to make adjustments, first terminate
CYCLE mode and either re-program the CYCLE parameters (see
page 12) or operate the unit in NORMAL mode.
I can’t adjust the TEMP SET value
above a certain value that is below the
200.0 degree maximum capability.
The TEMP SET value is limited by the TMAX maximum
temperature setting. If you need to operate at a higher
temperature, you must first increase TMAX.
When I reach my TEMP SET value, the
unit displays !TMAX!
The TMAX value is the same as the TEMP SET value. Either
increase the TMAX value or lower the TEMP SET value.
The unit appears to be holding the
TEMP SET value, but there is a !TMAX!
message on the display.
The unit exceeded the TMAX setting at some point, but during the
reset phase it settled to the correct level. This is normal; simply
press the SELECT keypad from the Main screen to clear the
alarm.
The unit is enabled but does not seem
to be driving the heater.
1. The PMAX value is too low. Increase the PMAX value to the
maximum rating of the heater you are driving.
2. The output connector is not fully inserted into the MAIN
OUTPUT connector.
The unit overshoots the set-point by a
large amount.
1. Check your I gain setting. It may be too high and if so, should
be lowered.
2. You have a sensor installed, but it is the wrong type.
I can’t connect to the TC200 over the
USB com port.
1. Make sure that the com port is configured correctly for the unit.
Refer to page 24 for the correct com port settings.
2. The incorrect com port is selected on your terminal program or
TC200 application.
3. USB driver was not installed. See Page 17
Chapter 9 Troubleshooting
The following table describes some typical problems that may be encountered while using the TC200 and possible
solutions to these problems.
Rev N, July 31, 2017 Page 27
Temperature Controller Chapter 10: General Maintenance
Fuse Holders
Fuses are installed
in the covers.
Chapter 10 General Maintenance
10.1. AC Line Voltage
Your TC200 Series Temperature Controller will operate from AC line voltages ranging from 100 VAC to 240 VAC
at 50 – 60 Hz. There is no need to configure the unit for specific line voltages.
ATTENTION
Use only 250 mA, 250 VAC, Type T, 5x20 mm style fuse.
If for some reason you need to replace an open fuse, you must perform the following procedure:
1. Remove any AC Input cords that may be connected to the unit.
2. Remove the cover of the unit by removing the two 4-40 Phillips head screws located on the bottom rear of
the unit and sliding the cover off.
3. Refer to Figure 9 below. Locate the two Fuse Holders near the AC Input Module towards the back of the
unit. Remove the cover to each fuse holder. The fuse is installed in the cover. Remove the existing fuse
and install the appropriate replacement fuse.
4. Reinstall the cover and replace the two 4-40 Phillips head screw.
Figure 9 Location of Fuse Holders
10.3. General Maintenance
Aside from the AC Input fuses, there are no user-serviceable parts in the TC200. If you suspect something has
failed on the unit, and if you have first referred to Troubleshooting on page 27, please contact Thorlabs for advice
on returning the unit for evaluation.
10.4. Cleaning
The unit can be cleaned using a soft, slightly damp cloth. Avoid using any solvents on or near the unit.
Page 28 TTN051989-D02
Temperature Controller Chapter 11: Auxiliary Output
Chapter 11 Auxiliary Output
The TC200 provides a secondary output to enable operation of multiple Thorlabs’ Gas Cell Heaters using only one
control unit. The Auxiliary Output shares the total output current of the unit proportionally with the Main Output. The
Main Output current is regulated via the feedback from the temperature sensor in the main heater. The Auxiliary
Output is not regulated and follows the changes in current made to the Main Output. For best results, both heaters
should have similar heater resistances (within a few ohms).
The Auxiliary heater temperature can be momentarily shown on the LCD display by pressing and holding the
SELECT keypad. The temperature will be shown under the ACTUAL TEMP text on the Front Panel display. The
Auxiliary temperature can also be viewed from the command line interface by typing taux?.
Note: the Auxiliary temperature is only updated once every five seconds and only when the sensor is of the PTC100
type.
Note: The Auxiliary input is only compatible with PT100 sensors.
11.1. Operating the Gas Cell Heater
Due to the unbalanced nature of the Auxiliary Output, it is best to connect the heater with the least resistance to the
Main Output and the higher resistance heater to the Auxiliary Output. This will ensure that the Auxiliary heater does
not exceed the TMAX level (which is only monitored on the Main Output).
Determine the main heater by connecting both heaters to the unit, setting the target temperature, and enabling the
TC200. Once the main heater starts to show an increase in temperature, press the SELECT keypad to view the
Auxiliary temperature. If the Auxiliary temperature is greater than the Main temperature, disable the heater, turn off
the power, and exchange the two heaters by connecting the original main heater into the Auxiliary Output. Repeat
this process once the temperatures have settled to verify proper balance.
11.2. Alternate Use of the Auxiliary Output
The Auxiliary output can also be used as an inputfor a second PT100 temperature sensor placed elsewhere in your
thermal system, without the use of a secondary heater load.
Connect the sensor as shown on page 25. Pressing the SELECT keypad while in the Main Display screen will now
show the secondary sensor temperature.
Rev N, July 31, 2017Page 29
Temperature Controller Chapter 12: Specifications
Item #
TC200
Min
Typical
Max
Input Voltage
a
100 VAC
-
240 VAC
Input Power
-
-
25 VA
Output Power
-
-
18 Watts DC
Output Current
-
-
750 mADC
Output Voltage
-
-
24 VDC
Temperature Control Range
b, f
20.0 °C
-
200.0 °C
Set-Point Resolution
0.1 °C
-
-
Actual Temp. Display Accuracy
c
-
±0.1 °C
±0.5 °C
Temperature Stability, 24 hours
d
-
±0.1 °C
-
TC200
Sensor Types
Platinum 100 Ω, Platinum 1000 Ω, TH10K
Output Connector
Mating Connector
HIROSE HR10A-7R-6S
HIROSE HR10A-7P-6P (73)
USB Interface
USB 2.0, Standard B Plug
Operating / Storage Temperatures
10 – 40 °C / 0 – 50 °C
Dimensions (LxWxH)
11.5" (292 mm) x 5.3" (135 mm) x 3.0" (76 mm)
Weight
5 lbs
Chapter 12 Specifications
a
Universal input 50 – 60 Hz; no line switching required. The TEC200 is supplied with a 115 V parallel blade line
cord for North American use only, while the TC200-EC is shipped with a 230 VAC line cord for European use.
For all other applications, use an IEC 320 compatible line cord fitted with a plug appropriate for your particular
AC wall socket.
b
Heating mode only. Ambient temperature must always be less than set-point temperature.
c
Does not account for errors due to sensor tolerances.
d
Measured in a constant ambient temperature. Accuracy is strongly dependent on the individual heating system
and P, I, and D gain parameters.
e
All technical data valid at 23 ± 5 °C and 45 ± 15% humidity.
f
The capability to attain Maximum Temperature is load dependent.
12.1. Accessories
The following accessories are directly compatible with the TC200 Temperature Controller:
1. PPLN Ovens
o PV10: PPLN Oven for 10 mm Long Crystals
o PV20: PPLN Oven for 20 mm Long Crystals
o PV40: PPLN Oven for 40 mm Long Crystals
2. Gas Cell Heaters
o GCH09-150: Heater for 9 x 150 mm Reference Cell
o GCH09-50: Heater for 9 x 50 mm Reference Cell
o GCH09-75: Heater for 9 x 75 mm Reference Cell
o GCH19-100: Heater for 19 x 100 mm Reference Cell
o GCH19-75: Heater for 19 x 75 mm Reference Cell
o GCH25-75: Heater for 25 x 75 Reference Cell
Page 30 TTN051989-D02
Temperature Controller Chapter 13: Warranty
Chapter 13 Warranty
Thorlabs, Inc. warrants the material and production of the TC200 Temperature Controller for a period of 24 months
from the date of shipment. During this warranty period, Thorlabs, Inc. will repair or exchange any units found to be
defective in material or workmanship.
For warranty repairs or service, the unit must be returned to Thorlabs, Inc. or one of our worldwide offices. All
returns must be accompanied by a Returned Material Authorization (RMA) number issued by Thorlabs. The
customer is responsible for all costs incurred to ship the unit back to Thorlabs; in the case of warranty repairs,
Thorlabs will cover the cost of shipment back to the customer.
In the case of shipments from outside of the United States, duties, taxes etc. shall be the responsibility of the
customer.
Thorlabs, Inc. warrants all firmware and software used for the TC200 to operate fault-free, provided they are used
in accordance with the requirements stated in this manual. However, Thorlabs does not warrant a fault-free
operation for special applications.
Thorlabs, Inc. does not guarantee that this manual is error free and reserves the right to change this manual, the
specifications, or the data of the described unit at any time, without notice.
Thorlabs, Inc. is not liable for any incidental damage caused by failure of this unit.
This warranty does not cover errors in operation, defects, or damage caused by the use of software not supplied
by Thorlabs, modifications or misuse, or damage caused if the unit is used outside the operating parameters
specified in this manual.
This warranty does not cover failure due to incorrectly connected heater and/or sensor elements, or failure of heater
and/or sensor elements.
ATTENTION
The TC200 is not intended for use in medical and/or life support situations.
Rev N, July 31, 2017 Page 31
Temperature Controller Chapter 14: Regulatory
Wheelie Bin Logo
Chapter 14 Regulatory
As required by the WEEE (Waste Electrical and Electronic Equipment Directive) of the European Community and
the corresponding national laws, Thorlabs offers all end users in the EC the possibility to return “end of life” units
without incurring disposal charges.
This offer is valid for Thorlabs electrical and electronic equipment:
Sold after August 13, 2005
Marked correspondingly with the crossed out “wheelie bin” logo (see right)
Sold to a company or institute within the EC
Currently owned by a company or institute within the EC
Still complete, not disassembled and not contaminated
As the WEEE directive applies to self-contained operational electrical and electronic products, this end of life take
back service does not refer to other Thorlabs products, such as:
Pure OEM products, that means assemblies to be built into a unit by the user (e. g. OEM laser driver cards)
Components
Mechanics and optics
Left over parts of units disassembled by the user (PCB’s, housings etc.).
If you wish to return a Thorlabs unit for waste recovery, please contact Thorlabs or your nearest dealer for further
information.
Waste Treatment is Your Own Responsibility
If you do not return an “end oflife” unit to Thorlabs, you must hand it to a company specialized in waste recovery.
Do not dispose of the unit in a litter bin or at a public waste disposal site.
Ecological Background
It is well known that WEEE pollutes the environment by releasing toxic products during decomposition. The aim of
the European RoHS directive is to reduce the content of toxic substances in electronic products in the future.
The intent of the WEEE directive is to enforce the recycling of WEEE. A controlled recycling of end of life products
will thereby avoid negative impacts on the environment.
Page 32 TTN051989-D02
Temperature Controller Chapter 15: Certifications and Compliance
Category
Standards or Description
EC Declaration of Conformity –
EMC
Meets intent of Directive 89/336/EEC for Electromagnetic
Compatibility. Compliance is given to the following specifications as
listed in the Official Journal of the European Communities:
-EN 61326
EMC requirements for Class A electrical equipment for measurement,
control, and laboratory use, including Class A Radiated and Conducted
Emissions
a, b
and Immunity
a, c
-IEC 61000-4-2
Electrostatic Discharge Immunity (Performance Criterion C)
-IEC 61000-4-3
Radiated RF Electromagnetic Field Immunity
(Performance Criterion B)
b
-IEC 61000-4-4
Electrical Fast Transient / Burst Immunity (Performance Criterion B)
-IEC61000-4-5
Power Line Surge Immunity (Performance Criterion C)
-IEC61000-4-6
Conducted RF Immunity (Performance Criterion B)
-IEC 61000-4-11
Voltage Dips and Interruptions Immunity (Performance Criterion C)
-EN 61000-3-2
AC Power Line Harmonic Emissions
Australia/New Zealand Declaration
of Conformity – EMC
Complies with the Radio Communications Act per EMC Emission
Standard
, b
:
-AS/NZS 2064
Industrial, Scientific, and Medical Equipment: 1992
-FCC EMC Compliance
Emissions comply with the Class A Limits of FCC Code of Federal
Regulations 47, Part 15, Subpart B
a, b
EC Declaration of Conformity –
Low Voltage
Compliance is given to the following specification as listed in the
Official Journal of the European Communities:
Low Voltage Directive 73/23/EEC, amended by 93/68/EEC
EN 61010-1/A2: 1995
Safety requirements for electrical equipment for measurement control
and laboratory use.
U.S. Nationally Recognized Testing Laboratory Listing
-UL3111-1
Standard for electrical measuring and test equipment.
-ANSI/ISA S82.01:1994
Safety standard for electrical and electronic test, measuring,
controlling, and related equipment.
Canadian Certification
CAN/CSA C22.2 No. 1010.1
Safety requirements for electrical equipment for measurement, control,
and laboratory use.
Additional Compliance
IEC61010-1/A2:1995
Safety requirements for electrical equipment for measurement ,control
and laboratory use.
Equipment Type
Test and Measuring
Safety Class
Class 1 (As Defined in IEC 61010-1, Annex H) – Grounded Product
Chapter 15 Certifications and Compliance
a
Using high-quality shielded interface cables
b
Emissions, which exceed the levels required by these standards, may occur when this equipment is connected to
a test object.
c
Minimum Immunity Test requirement.
Rev N, July 31, 2017Page 33
Temperature Controller Chapter 16: Thorlabs Worldwide Contacts
USA, Canada, and South America
Thorlabs, Inc.
56 Sparta Avenue
Newton, NJ 07860
USA
Tel: 973-300-3000
Fax: 973-300-3600
www.thorlabs.com
www.thorlabs.us (West Coast)
Email: sales@thorlabs.com
Support: techsupport@thorlabs.com
UK and Ireland
Thorlabs Ltd.
1 Saint Thomas Place, Ely
Cambridgeshire CB7 4EX
Great Britain
Tel: +44 (0)1353-654440
Fax: +44 (0)1353-654444
www.thorlabs.com
Email: sales.uk@thorlabs.com
Support: techsupport.uk@thorlabs.com
Thorlabs Sweden AB
Bergfotsgatan 7
431 35 Mölndal
Sweden
Tel: +46-31-733-30-00
Fax: +46-31-703-40-45
www.thorlabs.com
Email: scandinavia@thorlabs.com
France
Thorlabs SAS
109, rue des Côtes
78600 Maisons-Laffitte
France
Tel: +33 (0) 970 444 844
Fax: +33 (0) 825 744 800
www.thorlabs.com
Email: sales.fr@thorlabs.com
Brazil
Thorlabs Vendas de Fotônicos Ltda.
Rua Riachuelo, 171
São Carlos, SP 13560-110
Brazil
Tel: +55-16-3413 7062
Fax: +55-16-3413 7064
www.thorlabs.com
Email: brasil@thorlabs.com
Japan
Thorlabs Japan, Inc.
Higashi-Ikebukuro Q Building, 2F
2-23-2, Higashi-Ikebukuro,
Toshima-ku, Tokyo 170-0013
Japan
Tel: +81-3-5979-8889
Fax: +81-3-5979-7285
www.thorlabs.jp
Email: sales@thorlabs.jp
China
Thorlabs China
Room A101, No. 100
Lane 2891, South Qilianshan Road
Putuo District
Shanghai
China
Tel: +86 (0)21-60561122
Fax: +86 (0)21-32513480
www.thorlabs.hk
Email: chinasales@thorlabs.com
Chapter 16 Thorlabs Worldwide Contacts
Page 34 TTN051989-D02
www.thorlabs.com