6625-D02 Rev B, February 15, 2011 Page 3 www.thorlabs.com
Page 4
High Intensity Fiber Light Source
!
!
Part 1. Warnings
Caution
RISK OF ELECTRIC SHOCK. DO NOT OPEN
INPUT: 180 W, Max
Line Fuse, 5 x 20 mm
99 – 121 VACX, 60 Hz 3 A Quick Acting
200 – 244 VAC, 50 Hz 2 A Quick Acting
This symbol is intended to alert the user to the presence of important operating
and maintenance (servicing) instructions to the literature accom¬panying the
appliance.
This symbol is intended to alert the user to the presence of uninsulated
“dangerous voltage” within the product’s enclosure that may be of sufficient
magnitude to constitute a risk of electric shock to persons.
Important
This equipment has been tested and found to comply with IEC 61010-1 and UL3101-1 as
an Overvoltage Category II: (local level, small appliances, portable equipment, etc. with
smaller transient over voltages than Overvoltage Category III.) Pollution II.
Caution
You are cautioned that any change or modifications not expressly approved in this
manual could void your authority to operate this equipment.
TUV
1.1. Precautions
On Power Sources
•Before operating the product, check that the operating voltage is identical to
your local power supply. See specific descriptions for operating voltages.
•This unit is not disconnected from the AC power source as long as it is
connected to the wall outlet, even if the unit itself has been turned off.
• To disconnect the AC power cord, grasp the plug itself; never pull the cord.
• One blade of the plug is wider than the other for the purpose of safety and will
fit into the wall outlet only one way. If you are unable to fully insert the plug
into the outlet, please contact your local Thorlabs tech support.
•AC power cord must be replaced with IEC compatible cord.
6625-D02 Rev B, February 15, 2011 Page 4 www.thorlabs.com
Page 5
High Intensity Fiber Light Source
On Operation
Before connecting to fiber optic components, be sure that all power connections are
secure.
On Cleaning
Clean the cabinet, panel, and controls with a soft cloth slightly moistened with a mild
detergent solution. Do not use any type of abrasive pad, scouring powder or solvent such
as alcohol or benzine.
6625-D02 Rev B, February 15, 2011 Page 5 www.thorlabs.com
Page 6
High Intensity Fiber Light Source
Part 2. Getting Started
Unpacking – OSL1 and OSL1-EC
Check that you received all items on the packing list.
Hookup Overview
The OSL1 High Intensity Fiber Light Source allows the user to connect and control a
variety of fiber optic components. To learn the locations and names of important
connections see “Illuminator Descriptions” below.
Before You Get Started
• Turn off the power to the OSL1 before making any connections.
• Do not connect the AC power cord to the local power supply until the
connections are completed to the OSL1.
• Be sure to make connections firmly.
• Be sure that the OSL1 and OSL1-EC operating voltage matches the “local”
power supply.
6625-D02 Rev B, February 15, 2011 Page 6 www.thorlabs.com
Page 7
High Intensity Fiber Light Source
1
2
3
4
5
1
2
3
Part 3. Description
Front Panel Light Source Descriptions
1. Variable intensity controller (0 to 100%).
2. On/Off rocker switch.
3. Front panel fasteners (2)
4. Lamp module.
5. V-port with T-handle retainer
Figure 1: Front Panel Description
Rear Panel Light Source Descriptions
1. T-slot mounting for #1/4 - 20 or M6 hexhead fastener; both sides (#10 - 32 or
M5 barnut optional)
2. IEC inlet with line filtering (power cable input)
3. Operating voltage label
6625-D02 Rev B, February 15, 2011 Page 7 www.thorlabs.com
Figure 2: Rear Panel Description
Page 8
High Intensity Fiber Light Source
Lamp
Socket
Lamp
Socket
Lamp
Access Panel
Front Panel
Fasteners (2x)
Part 4. Lamp Replacement
Lamp Replacement Instructions
1. To replace the lamp, shut off the power and remove the power cord from the
AC outlet. To disconnect the AC power cord, grasp the plug itself; never pull
the cord.
2. Located on the lamp access panel, manually loosen the front panel fasteners.
3. With front panel released, remove the lamp module from the unit. Use caution;
fan blades are accessible while lamp module is removed.
4. Allow sufficient time for the unit to cool. CAUTION: Lamp and surfaces may
be extremely hot!
5. Remove the lamp from the socket and place in a safe area.
6. Insert the replacement lamp into the socket; pressing firmly to ensure that pins
are fully seated.
7. Close the access panel.
8. Secure the lamp access panel by manually tightening the front panel fasteners.
Figure 3: Lamp Access Diagram
On Safety
Should any solid object or liquid fall into the assembly while access panel is open, have it
checked by qualified personnel before operating.
6625-D02 Rev B, February 15, 2011 Page 8 www.thorlabs.com
Page 9
High Intensity Fiber Light Source
7.13"
(181.1 mm)
3.53"
(89.7 mm)
Ø1.00"
(Ø25.4 mm)
7.91"
(200.9 mm)
6.55"
(166.4 mm)
0.59"
(15 mm)
Part 5. Specifications
Item # OSL1 OSL1-EC
Environmental
Operating Temperature
Humidity Range
Pollution Degree
Electrical (all units)
Operating Frequency
Power Consumption
Power Requirements
Lamp Adjustment Range
Color Temperature
Lamp Life
Physical
Mass
Lamp Type
Fuse Type
* Standard lamp type provided with illuminator.
Dimensional Drawing
-4 to 104°F (-20° to 40°C)
0 to 80%; non-condensing
2
60 Hz (110 VAC) or 50 Hz (220VAC)
180 W at Full Light Intensity
110 VAC; 60 Hz
220 VAC; 50 Hz
0 to 100%
3200 K at Max Intensity
(assuming standard EKE lamp)
250 to 10,000 hours
4.5 lbs. (2.04 kg) approx.
150W EKE, 3250K (#20094145)*
150W EKE-HC, 4200K (#20094142)
150W EJV, 3400K (20094143)
3.15 A, 250 V, 5 x 20 mm
Figure 4: Dimensional Drawing
6625-D02 Rev B, February 15, 2011 Page 9 www.thorlabs.com
Page 10
High Intensity Fiber Light Source
Part 6. Regulatory
As required by the WEEE (Waste Electrical and Electronic Equipment Directive) of the
European Community and the corresponding national laws, Thorlabs offers all end users
in the EC the possibility to return “end of life” units without incurring disposal charges.
• This offer is valid for Thorlabs electrical and electronic equipment:
• Sold after August 13, 2005
• Marked correspondingly with the crossed out
“wheelie bin” logo (see right)
• Sold to a company or institute within the EC
• Currently owned by a company or institute within
the EC
•Still complete, not disassembled and not
contaminated
As the WEEE directive applies to self contained operational
electrical and electronic products, this end of life take back
service does not refer to other Thorlabs products, such as:
•Pure OEM products, that means assemblies to be built into a unit by the user
(e.g. OEM laser driver cards)
• Components
• Mechanics and optics
• Left over parts of units disassembled by the user (PCB’s, housings etc.).
If you wish to return a Thorlabs unit for waste recovery, please contact Thorlabs or your
nearest dealer for further information.
Wheelie Bin Logo
6.1. Waste Treatment is Your Own Responsibility
If you do not return an “end of life” unit to Thorlabs, you must hand it to a company
specialized in waste recovery. Do not dispose of the unit in a litter bin or at a public
waste disposal site.
6.2. Ecological Background
It is well known that WEEE pollutes the environment by releasing toxic products during
decomposition. The aim of the European RoHS directive is to reduce the content of toxic
substances in electronic products in the future.
The intent of the WEEE directive is to enforce the recycling of WEEE. A controlled
recycling of end of live products will thereby avoid negative impacts on the environment.
6625-D02 Rev B, February 15, 2011 Page 10 www.thorlabs.com
Page 11
High Intensity Fiber Light Source
Part 7. Thorlabs Worldwide Contacts
USA, Canada, and South America
Thorlabs, Inc.
435 Route 206
Newton, NJ 07860
USA
Tel: 973-579-7227
Fax: 973-300-3600
www.thorlabs.com
email: feedback@thorlabs.com
Thorlabs SAS
109, rue des Côtes
78600 Maisons-Laffitte
France
Tel: +33 (0) 970 444 844
Fax: +33 (0) 811 381 748
www.thorlabs.de
email: slaes.fr@thorlabs.com
Japan
Thorlabs Japan Inc.
5-17-1, Ohtsuka
Bunkyo-ku, Tokyo 112-0012
Japan
Tel: +81-3-5979-8889
Fax: +81-3-5979-7285
www.thorlabs.jp
email: sales@thorlabs.jp
UK and Ireland
Thorlabs LTD.
1 Saint Thomas Place, Ely
Cambridgeshire CB7 4EX
Great Britain
Tel: +44 (0)1353-654440
Fax: +44 (0)1353-654444
www.thorlabs.de
email: sales.uk@thorlabs.com
Scandinavia
Thorlabs Sweden AB
Box 141 94
400 20 Göteborg
Sweden
Tel: +46-31-733-30-00
Fax: +46-31-703-40-45
www.thorlabs.de
email: scandinavia@thorlabs.com
China
Thorlabs China
Oasis Middlering Centre
3 Building 712 Room
915 Zhen Bei Road
Shanghai
China
Tel: +86 (0)21-32513486
Fax: +86 (0)21-32513480
www.thorlabs.com
email: chinasales@thorlabs.com
6625-D02 Rev B, February 15, 2011 Page 11 www.thorlabs.com
Page 12
Thorlabs, Inc.
435 Route 206N
Newton, NJ 07860 USA
Phone: (973) 579-7227
www.thorlabs.com
♦ Fax: (973) 300-3600
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.