CLEAN ALL COMPONENTS
Do not allow blood to dry on components. Ultrasonic cleaning is recommended; however, hand washing with a brush is acceptable.
LUBRICATE ALL THREADED AND CAM COMPONENTS BEFORE AUTOCLAVING
Secure the Elite II Rail Clamp to the table rail over the sterile drape.
CAUTION: If the patient is obese, avoid compressing the ulnar nerve.
When necessary, use a wider OR table or add 2 1/2” to the width of the table by using our Rail Extender #41917.
4
a
b
e
d
Be certain that all Cam Joints are in FULL OPEN POSITION during autoclaving.
STERILIZE:
GRAVITY STEAM (Wrapped) 30 minutes 121ºC (250ºF) @ 1.1kg/cm
HIGH SPEED GRAVITY (FLASH) 5 minutes 132ºC (270ºF) @ 30psi or 132º @ 2kg/cm
PREVACUUM STEAM (Wrapped) 10 minutes @132ºC (270ºF)
CAUTION: Relax retractors periodically to ensure proper blood flow.
2
2
c
a. Press the plunger and insert blade into the handle.
b. Cam should be in “unlock” position
c. Place blade in site
d. Clip joint onto arm
e. Retract and “lock” joint