Thies CLIMA 5.4103.20.041, 5.4103.20.641, 5.4103.20.741 Instructions For Use Manual

Precipition Sensor
Instruction for Use
with analogue Intensity Output
5.4103.20.041, 5.4103.20.641, 5.4103.20.741
Dok. No. 021335/05/18
T H E W O R L D O F W E A T H E R D A T A
Safety Instructions
Environment
Make sure you retain packaging for storage or transport of products. Should packaging required, please arrange for recycling as the packaging materials
Documentation
Before operating with or at the device/product, read through the operating instructions. This manual contains instructions which should be followed on mounting, start-up, and operation. A non-observance might cause:
- failure of important functions
- endangerment of persons by electrical or mechanical effect
- damage to objects
Mounting, electrical connection and wiring of the device/product must be carried out only by a qualified technician who is familiar with and observes the engineering regulations, provisions and standards applicable in each case.
Repairs and maintenance may only be carried out by trained staff or Adolf Thies GmbH & Co. KG. Only components and spare parts supplied and/or recommended by Adolf Thies GmbH & Co. KG should be used for repairs.
Electrical devices/products must be mounted and wired only in a voltage-free state.
Adolf Thies GmbH & Co KG guarantees proper functioning of the device/ pro du cts prov ided that no
modifications have been made to the mechanics, electronics or software, and that the following points are observed:
All information, warnings and instructions for use included in these operating instructions must be taken into account and observed as this is essential to ensure trouble-free operation and a safe condition of the measuring system / device / product.
The device / product is designed for a specific application as described in these operating instructions.
The device / product should be operated with the accessories and consumables supplied and/or
recommended by Adolf Thies GmbH & Co KG.
Recommendation: As it is possible that each measuring system / device / product may, under certain conditions, and in rare cases, may also output erroneous measuring values, it is recommended using redundant systems with plausibility che cks for security-relevant applications.
As a longstanding manufacturer of sensors Adolf Thies GmbH & Co KG is commi tt ed to
© Copyright Adolf Thies GmbH & Co KG, Göttingen / Germany
Although these operating instruction has been drawn up with due care, Adolf Thies GmbH & Co KG
We can accept no liability whatsoever for any losses arising from the information contained in this
Subject to modification in terms of content.
The device / product should not be passed on without the/these operating instructions.
the objectives of environmental protection and is therefore willing to take back all supplied products governed by the provisions of "ElektroG"
Electronic Equipment Act) and to perform environmentally compatible disposal and recycling. We are prepared to take back all Thies products concerned free of charge if returned to Thies by our customers carriage-paid.
however no longer be are designed to be recycled.
can accept no liability whatsoever for any technical and typographical errors or omissions in this document that might remain.
document.
(German Electrical and
© Adolf Thies GmbH & Co. KG · Hauptstraße 76 · 37083 Göttingen · Germany 021335/05/18 Phone +49 551 79001-0 · Fax +49 551 79001-65 · info@thiesclima.com ·www.thiesclima.com Page 2 of 12
Contents
1 Models available ................................................................................................................4
2 Application .........................................................................................................................4
3 Construction and Mode of Operation ..................................................................................4
4 Recommendation Side Selection/Standard Installation ......................................................5
5 Installation ..........................................................................................................................6
5.1 Mechanical Mounting ................................................................................................ 6
5.2 Electrical Mounting for Precipitation Sensor with Cable Gland .................................. 6
5.3 Electrical Mounting for Precipitation Sensor with Plug Connection ............................ 7
5.3.1 Plug Mounting .......................................................................................................7
6 Taking into Operation .........................................................................................................7
7 Maintenance ......................................................................................................................7
8 Technical Data ...................................................................................................................8
9 Dimension diagram ............................................................................................................9
10 Accessories (Optional) .................................................................................................. 10
11 EC-Declaration of Conformity ........................................................................................ 11
© Adolf Thies GmbH & Co. KG · Hauptstraße 76 · 37083 Göttingen · Germany 021335/05/18 Phone +49 551 79001-0 · Fax +49 551 79001-65 · info@thiesclima.com ·www.thiesclima.com Page 3 of 12

1 Models available

value
Output
Voltage
Order - No. Measuring
5.4103.20.041
5.4103.20.741
5.4103.20.641
Precipitation
intensity
Precipitation
intensity
Precipitation
intensity
Electrical
4…20mA 180 seconds 24 V AC/DC Cable gland
4…20mA 180 seconds 24 V AC/DC
4...20mA 50 seconds 24 V AC/DC Cable gland
Decay time Operating
Connection
7- pole plug
connection

2 Application

The precipitation sensor serves as measuring instrument for the determination of the instantaneous precipitation intensities (mm precipitation / min.). By integration of the precipitation intensities, the precipitation quantity can be calculated, as well. Control- and warning signals can be derived from the precipitation intensity.
The measuring signal is output as intensity-dependent analogue output value. The dimension of the measurement value output is divided into 4 linear characteristic segments, which show a tenth of the slope of the more sensitive segment.
Thus, it is possible to represent an intensity range from approx. 0,001mm/min. (light drizzle) up to 10 mm/min. (extremely heavy rain) with reasonable resolution (quasi-logarithmic output).

3 Construction and Mode of Operation

Precipitation in the form of drizzle, rain, snow, or hail falls through a light band, induced by light diodes, and lead to shadowing effects on the receiving side. The sent light is pulse-modulated so that outside light effects cannot falsify the measurement results. From the extent of shading along with the duration of falling through a factor is calculated that gives the information if there is a precipitation incident or not.
The instrument is equipped with a heating system for extreme weather condition. This avoids ice and snow forming on the housing surface. In addition, the surface retains a temperature of >0° by means of a regulated heating.
At the end of the precipitation the analogue output goes in one time of 180 sec. (5.4103.20.041, 5.4103.20.741) or 50 sec. (5.4103.20.641) to 0 mA. If the precipitation starts, the analog output rises again to the measured value.
© Adolf Thies GmbH & Co. KG · Hauptstraße 76 · 37083 Göttingen · Germany 021335/05/18 Phone +49 551 79001-0 · Fax +49 551 79001-65 · info@thiesclima.com ·www.thiesclima.com Page 4 of 12