Thies CLIMA 4.3519.00.140, 4.3519.00.167, 4.3519.00.161, 4.3519.00.141, 4.3519.00.173 Instructions For Use Manual

...
Instruction for Use
021075/08/11
Wind Transmitter compact
4.3519.xx.140 ... 961
Hauptstraße 76 37083 Göttingen Germany Box 3536 + 3541 37025 Göttingen Phone ++551 79001-0 Fax ++551 79001-65 www.thiesclima.com info@thiesclima.com
THE WORLD OF WEATHER DATA - THE WORLD OF WEATHER DATA - THE WORLD OF WEATHER DATA
Safety Instructions
Before operating with or at the device/product, read through the operating instructions.
- failure of important functions
- Endangering of persons by electrical or mechanic effect
- Damages at objects
This manual contains instructions which should be followed on mounting, start-up, and operation. A non-observance might cause:
Mounting, electrical connection and wiring of the device/product must be carried out only by a qualified technician who is familiar with and observes the engineering regulations, provisions and standards applicable in each case.
Repairs and maintenance may only be carried out by trained staff or Adolf Thies GmbH & Co. KG. Only components and spare parts supplied and/or recommended by Adolf Thies GmbH & Co. KG should be used for repairs.
Electrical devices/products must be mounted and wired only in voltage-free state.
Adolf Thies GmbH & Co KG guarantees proper functioning of the device/products provided that no
modifications have been made to the mechanics, electronics or software, and that the following points are observed:
All information, warnings and instructions for use included in these operating instructions must be taken into account and observed as this is essential to ensure trouble-free operation and a safe condition of the measuring system / device / product.
The device / product is designed for a specific application as described in these operating instructions.
The device / product should be operated with the accessories and consumables supplied and/or recommended
by Adolf Thies GmbH & Co KG .
Recommendation: As it is possible that each measuring system / device / product under certain conditions, and in rare cases, may also output erroneous measuring values, it is recommended to use redundant systems with plausibility checks with security-relevant applications.
Environment
As a longstanding manufacturer of sensors Adolf Thies GmbH & Co KG is committed to the objectives of environmental protection and is therefore willing to take back all supplied products
governed by the provisions of "ElektroG" and to perform environmentally compatible disposal and recycling. We are prepared to take back all Thies products concerned free of charge if returned to Thies by our customers carriage-paid.
Make sure you retain packaging for storage or transport of products. Should packaging however no longer be required, arrange for recycling as the packaging materials are designed to be recycled.
(German Electrical and Electronic Equipment Act)
Documentation
© Copyright Adolf Thies GmbH & Co KG, Göttingen / Germany
Although this operating instruction has been drawn up with due care, Adolf Thies GmbH & Co KG can accept
no liability whatsoever for any technical and typographical errors or omissions in this document that might remain.
We can accept no liability whatsoever for any losses arising from the information contained in this document.
Subject to modification in terms of content.
The device / product should not be passed on without the/these operating instructions.
2 - 14 021075/08/11
Contents
1 Models .......................................................................................................................................4
2 Application .................................................................................................................................5
3 Mode of Operation.....................................................................................................................5
4 Recommendation Site Selection / Standard Installation............................................................5
5 Installation.................................................................................................................................. 5
5.1 Mechanical Mounting .......................................................................................................... 6
5.2 Electrical Mounting..............................................................................................................6
5.3 Plug mounting ..................................................................................................................... 6
6 Connecting Diagram.................................................................................................................. 7
7 Maintenance ..............................................................................................................................8
8 Technical Data...........................................................................................................................9
9 Dimension diagram..................................................................................................................10
10 Accessories.......................................................................................................................... 11
11 EC-Declaration of Conformity...............................................................................................12
Figure
Figure 1: plug mounting.................................................................................................................... 6
Figure 2: Connecting Diagram for Models with fixed Connecting Cable.......................................... 7
Figure 3: Connecting Diagram for Models with Connector............................................................... 8
Figure 4: Dimensional Drawing Model cable gland........................................................................ 10
Figure 5: Dimensional Drawing Model plug.................................................................................... 10
3 - 14 021075/08/11
1 Models
Order - No. Electrical
Output
4.3519.00.140 0...20 mA 0...50 m/s 20 W 12 m Cable LiYCY 6 x 0,25 mm²
4.3519.00.141 4...20 mA 0...50 m/s 20 W 12 m Cable LiYCY 6 x 0,25 mm²
4.3519.00.161 0...10 V 0…50 m/s 20 W 12 m Cable LiYCY 6 x 0,25 mm²
4.3519.00.167 0...2 V 0…50 m/s 20 W 12 m Cable LiYCY 6 x 0,25 mm²
4.3519.00.173 0...5 V 0…50 m/s 20 W 12 m Cable LiYCY 6 x 0,25 mm²
Measuring
range
Heating
power
Connection
4.3519.00.361 0...10 V 0…3 m/s
max. 13,8 V @ >3m/s
4.3519.00.441 4...20 mA 0…40 m/s 20 W 3 m PUR -Cable 6 x 0,25 mm²
4.3519.00.641 4...20 mA 0…60 m/s 20 W 12 m Cable LiYCY 6 x 0,25 mm²
4.3519.00.740 0…20 mA 0…50 m/s 20 W 7 pol. Plug
4.3519.00.741 4...20 mA 0…50 m/s 20 W 7 pol. Plug
4.3519.00.761 0...10 V 0...50 m/s 20 W 7 pol. Plug
4.3519.00.773 0...5 V 0...50 m/s 20 W 7 pol. Plug
4.3519.00.961 0...10 V 0…15 m/s 20 W 12 m Cable LiYCY 6 x 0,25 mm²
4.3519.01.140 0...20 mA 0…50 m/s 20 W 1,5 -3 m Spiral Cable LiYY 6x0,14 mm²
4.3519.02.141 4...20 mA 0…50 m/s 10 W 2 m Cable 6 x 0,56 mm²
4.3519.04.441 4...20 mA 0…40 m/s 20 W 0,95 m PUR- Cable 6 x 0,25 mm²
4.3519.05.141 4...20 mA 0…50 m/s 20 W 15 m Cable LiYCY 6 x 0,25 mm²
4.3519.05.161 0…10 V 0…50 m/s 20 W 15 m Cable LiYCY 6 x 0,25 mm²
4.3519.05.641 4...20 mA 0…60 m/s 20 W 15 m Cable LiYCY 6 x 0,25 mm²
4.3519.10.441 4...20 mA 0…40 m/s
20 W 12 m Cable LiYCY 6 x 0,25 mm²
Without heating
12 m Cable LiYCY 6 x 0,25 mm²
4.3519.20.141 4...20 mA 0...50 m/s 10 W 12 m Cable LiYCY 6 x 0,25 mm²
4.3519.39.141 4...20 mA 0…50 m/s 20 W 12 m Cable LiYCY 6 x 0,25 mm²
with cable lug at the shield
4.3519.40.140 0...20 mA 0...50 m/s 60 W 12 m Cable LiYCY 6 x 0,5 mm²
4.3519.40.141 4...20 mA 0...50 m/s 60 W 12 m Cable LiYCY 6 x 0,5 mm²
4.3519.40.161 0...10 V 0…50 m/s 60 W 12 m Cable LiYCY 6 x 0,5 mm²
4.3519.40.167 0...2 V 0…50 m/s 60 W 12 m Cable LiYCY 6 x 0,5 mm²
4.3519.40.173 0...5 V 0…50 m/s 60 W 12 m Cable LiYCY 6 x 0,5 mm²
4.3519.40.740 0…20 mA 0…50 m/s 60 W 7 pol. Plug
4.3519.40.741 4...20 mA 0…50 m/s 60 W 7 pol. Plug
4.3519.40.761 0...10 V 0...50 m/s 60 W 7 pol. Plug
4 - 14 021075/08/11
2 Application
The wind transmitter detects the horizontal wind speed. The measured values are available at the output as analogue voltage or current signal to control for instance wind power plant..
An electronically-regulated heating system has been installed in some models (see chapter 1) for winter time use, in order to prevent the ball-bearing and the external rotation parts from freezing.
Thanks to the 60-Watt-heating as well as to the optimized regulating characteristic, model no.
4.3519.40.xxx is especially suited for the extremely difficult application in high mountains or at other critical sites, where icing is to be expected.
3 Mode of Operation
The cup star (in ball bearing) is set into rotation by the wind. An opto-electronic speed scanning produces a frequency which is transformed into an analogue signal by an integrated measuring transformer.
The outer parts of the instrument are made of corrosion-resistant materials. Labyrinth gaskets protect the parts inside the instrument against precipitations.
4 Recommendation Site Selection / Standard Installation
In general wind measurement instruments should be able to detect the wind conditions of a large area. In order to obtain comparable values when determining the surface wind, measurements should be taken at a height of 10 meters over an even area with no obstacles. An area with no obstacles means that the distance between the wind direction transmitter and an obstacle should be at least 10 times the height of the obstacle (s. VDI 3786 ). If it is not possible to fulfil this condition then the wind direction transmitter should be set up a height where local obstacles do not influence the measured values to any significant extent (approx. 6-10 m above the obstacle). The wind direction transmitter should be set up in the centre of flat roofs and not on the edge in order to avoid any preferential directions.
5 Installation
Attention:
Storing, mounting and operation under weather conditions is permissible only in vertical position, as otherwise water can get into the instrument.
5 - 14 021075/08/11