ThermoWorks Therma Waterproof Operating Instructions Manual

INSTRUMENT RECORD CARD
Model Order Code
Therma Waterproof 232-107
CALIBRATION
)
32°F (0°C
Thermocouple
Calibrated by __________________ Date ____________
This instrument has been checked or calibrated against reference instrument(s) calibrated by a UKAS Accredited Calibration Laboratory.
212°F (100°C
)
752°F (400°C
)
Therma Waterproof
Type T Thermometer
741 E. Utah Valley Dr., American Fork, UT 84003
P-13-006-02-a Ref: 547-721 / 12.04.16
Ph: 801-756-7705 Fax: 801-756-8948
thermoworks.com
Operating Instructions
INSTRUMENT OPERATION - Connect probes via the socket located at the top of the instrument. (The Therma Waterproof uses a miniature type T thermocouple connector). Switch the instrument on and apply the tip of the probe to the substance or medium to be measured. The instrument reading may take a few seconds to stabilize, depending on the nature of the measurement and sensitivity of the probe.
HOLD - Press the HOLD button to stop measurements and freeze the display temporarily, press again to continue measuring.
AUTO-OFF - The unit will switch off automatically after 10 minutes. To disable auto-off press the HOLD button while switching on the instrument (this is reset when the instrument is switched off).
OPEN CIRCUIT/FAULTY PROBE - If the probe has developed an 'Open Circuit' fault the instrument will display 'Err'.
SENSORS - Therma Waterproof should only be used with a type T (copper/constantan) thermocouple probe (BS EN 60584-2) fitted with a miniature type T thermocouple plug (to BS EN 50212). This instrument is designed to be compatible with ThermoWorks manufactured probes.
BATTERY REPLACEMENT - The instrument will display a low battery symbol to warn that the batteries will need changing soon. The instrument continues to measure accurately but after further usage the instrument will display 'flat bat' and shutdown. Replace the three AAA batteries located under the rear case half that is held in place by six screws. Check the seal for damage and carefully replace the rear case half, re-fit and tighten the screws.
MAX/MIN - Press the MAX/MIN button to display the values. The instrument retains the max/min values when switched off. To reset, press the MAX/MIN button for three seconds, 'rst' will be displayed.
BACKLIGHT - Press the HOLD button for two seconds. Please Note: the backlight will not operate if the low battery symbol is being displayed. (Backlight timeout = 20 seconds).
CALIBRATION TRIM - Press the MAX/MIN and ON/OFF buttons at start-up for five seconds to enter ('Offset' is displayed). MAX/MIN then cycles (continuous loop) the trim value in 0.1 steps up to ±2.0ºC (3.6ºF). Press ON/OFF to save the value and exit Trim mode. (Instrument should not require trim
adjustment. Should only be performed by qualified person with appropriate calibration standards.)
CHANGE FROM FAHRENHEIT TO CELSIUS - Press and hold ON/OFF button for five seconds at start-up.
GUARANTEE - This instrument carries a two-year warranty against defects in either components or workmanship. During this period, products that prove to be defective will, at the discretion of ThermoWorks, be either repaired or replaced without charge. This warranty does not apply to probes, where a six-month period is offered. The product warranty does not cover damage caused by fair wear and tear, abnormal storage conditions, incorrect use, accidental misuse, abuse, neglect, misapplication or modification. Full details of liability are available within ThermoWorks' Terms & Conditions of Sale at www.thermoworks.com. In line with our policy of continuous development, we reserve the right to amend our product specification without prior notice.
SPECIFICATION - Full specification details of all products are available at www.thermoworks.com.
SPECIFICATIONS
Range 0.1°: Range 1°: Resolution: Accuracy: Auto-off: Battery: Sensor Type: Display: Water Resistance: Dimensions:
Weight: Calibration:
–148 to 572°F (–99.9 to 299.9°C) 572 to 2,502°F (300 to 1372°C)
0.1° to 572°C (299.9°C) thereafter 1° ±0.36°F (±0.2°C), ±1 digit after 10 minutes 3x 1.5V AAA, 10,000 hours T Thermocouple
0.6 inches (15 mm) LCD IP66/67 rated
5.6 H x 2.8 W x 1.26 D inches (141 H x 71 W x 32 D mm)
0.49lb. (220g) NIST-Traceable certificate included