ThermoWorks ThermaData Lite Operating Instructions Manual

ThermaData Lite – White 293-010 ThermaData Lite – Black 293-070 ThermaData Lite – Yellow 293-020 ThermaData Lite – Blue 293-050
SPECIFICATIONS
Range
–40 to 185°F (–40 to 85°C)
Accuracy
Resolution
Sample Rate
Sensor Type
Dimensions
±0.9°F (±0.5°C)
0.1°F or °C
Memory
16,000 readings, stops when full. Settable from 6 seconds to 4 hours 15 minutes (255 min) Thermistor
Battery
3.6V 1/2 AA Lithium / 3 year life typical
2.16 dia. X 1 D inches (55 dia. x 25 D mm), plus eyelet tab
Weight
1.4oz (40g)
ThermaData® Lite
ACCESSORIES/RELATED ITEMS
USB-C Interface Cable
(812-510)
Waterproof Pouch
(830-360)
3.6V 1/2AA
Lithium Battery
(T HS- 839 -102 )
GUARANTEE - This instrument carries a two-year warranty against defects in either components or workmanship. During this period, products that prove to be defective will, at the discretion of ThermoWorks, be either repaired or replaced without charge. This warranty does not apply to probes, where a six-month period is offered. Full details of liability are available within ThermoWorks’ Terms & Conditions of Sale at www.thermoworks.com/product-warranty.
For warranty, service, and technical assistance, please contact ThermoWorks’ Technical Support at (801) 756-7705 or email at techsupport@thermoworks.com.
Temperature Loggers
P-12-013-02-b
ThermoWorks, Inc.
Utah, U.S.A.
Ph: 801-756-7705 Fax: 801-756-8948
www.thermoworks.com
Operating Instructions
547-310 / 20.02.19
Your new ThermaData Lite logger uses advanced measurement technology for precision temperature recording. Data is collected within the unit’s onboard memory and a PDF report can be downloaded via USB. Settings and more in-depth data analysis can be adjusted via ThermaData Studio for Windows.
INSTALL SOFTWARE - THERMADATA STUDIO
This data logger utilizes a USB-C to USB-A connection cable, which is purchased separately, to interface with your PC. Along with the hardware, you’ll need to download and install the free software, ThermaData Studio, from www.ThermoWorks.com/software. Once you have the software installed, open and running, and the USB cable connected, you’re ready to plug in your data logger and start working.
NOTE: ThermaData Studio software is updated regularly. Please visit www.ThermoWorks.com/ software to check for any recent updates.
CONNECT AND START YOUR LOGGER*
Once connected to the computer with the USB cable, you can begin setting up your logger. To begin, run ThermaData Studio software and open the ThermaData tab and click on Instrument Settings. Select your ThermaData Lite from the drop-down menu, if prompted. From this screen you can name your logger, set your temperature alarm limits, change your recording interval, add instructions to be included in PDF report, setup a star t delay or timed star t, as well as instructing what to do when the memory is full. If you’re ready to start collecting data, click Send Settings. A message will inform you that proceeding will clear all data currently on the device. Click OK to proceed and wait until the software tells you to its safe to unplug the logger. When your ready to start logging simply press the button on the front of the logger. It will immediately start recording or the logger will star t the delay you setup in the software.
RETRIEVE YOUR DATA*
When you’re ready to retrieve your data, connect the logger to the PC using the USB-C cable. The logger will stop recording and show up in your file explorer as a Mass Storage Device. Saved on the logger will be a PDF report of the logging activity. For a more in-depth data analysis, leave the logger connected to the computer and open ThermaData Studio. On the Home Tab click Download to retrieve all data points from the logger. Be sure to to use the Save option on the Home Tab to save the data in ThermaData Studio.
STOP AND RESTART YOUR LOGGER
You can manually stop ThermaData Lite by pressing and holding the button on the front of the logger for 5 seconds. Both the Red and Green LEDs will light up for 5 seconds, indicating the logger is stopped. To restart your logger after it has been stopped or you have downloaded the data, follow the steps in the section above titled CONNECT AND START YOUR LOGGER.
BATTERIES
This data logger comes with a non-rechargeable, preinstalled 1/2 AA 3.6V lithium battery. When the battery needs to be changed, open the snap-fit cover using a coin in the slot provided. Replace the battery with an 1/2 AA 3.6V lithium battery (THS-839-102) available at www.ThermoWorks.com. Failure to use the same type of battery may result in decreased battery life.
LED LIGHT SEQUENCES
Red & Green together 4 times: Logger is starting.
Red & Green alternating 2 times: Delay started.
Red 3 times: Temperature has exceeded limits.
Green 3 times: Temperature has stayed within limits.
Red & Green on for 5 seconds 1 time: Stop Logging.
Red & Green on for 3 seconds 2 times: Logger is stopped.
Red flashing continuous: Connected to USB.
*Additional details and full operating instructions can be found in the ThermaData Studio software help file.